MIRS ™ is used to create treatment plans for patients for whom radiation therapy has been prescribed. The system will calculate and display three dimensional radiation dose distributions within the patient for a given treatment plan set up. MIRS can be used for different treatment modalities of radiation therapy, as conventional therapy. Conformal Radiation Therapy, Intensity Modulated Radiation Therapy, and Radiosurgery with Conic Collimators.
Device Story
MIRS (Module Integrated Radiotherapy System) is a modular treatment planning system for external beam radiation therapy. It accepts multi-modality medical images, including 3D MRI, CT, functional MRI, PET, and angiography, with or without fiducials. The system performs automatic image registration and calculates 3D radiation dose distributions based on a defined treatment plan. It supports various modalities, including conventional, conformal, IMRT, and radiosurgery with conic collimators, and is compatible with any stereotactic frame. Used in clinical settings by radiation oncology professionals to design treatment plans; the output (dose distribution maps) assists clinicians in optimizing radiation delivery to target volumes while sparing healthy tissue, thereby improving treatment precision and patient outcomes.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and performance comparisons against predicate devices.
Technological Characteristics
Modular software-based radiation therapy planning system. Supports multi-modality image registration (CT, MRI, PET, angiography). Features include 3D dose distribution calculation, support for stereotactic frames, and compatibility with various radiation modalities (conventional, conformal, IMRT, radiosurgery).
Indications for Use
Indicated for patients prescribed radiation therapy requiring 3D dose distribution planning. Supports conventional, conformal, intensity-modulated radiation therapy (IMRT), and radiosurgery with conic collimators.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Ke90682
V. Arredondo 2684-2ª Piso ♦ 1426 Buenos Aires ♦ Argentina Tel: (54) 114896-0989
# 510(K) SUMMARY
MAY 4 2009
MIRS ™
| Submitter: | Nuclemed S.A.<br>V. Arredondo 2684 - 2ª Piso<br>1426 Buenos Aires - Argentina<br>Phone: (54) 11 4896-0989 |
|-----------------|-----------------------------------------------------------------------------------------------------------|
| Contact Person: | Raymond J Kelly IV<br>Arazy Group<br>56B Shadowbrook, Dr Hudson, NH 0305<br>Phone: (603) 595-0095 |
Fax: (603) 595-0095
E-mail: ray@arazygroup.com
December 8th, 2008
*
#### Device Names:
Date:
MIRS ™ Module Integrated Radiotherapy System Trade Name: Module Integrated Radiotherapy System Common Name: System, Planning, Radiation Therapy Treatment Classification Name:
### Legally Marketed Device to Which Substantial Equivalence is Claimed:
The MIRS ™ unit is substantially equivalent (SE) to Rad Calc (K010464) cleared in 2001, CDMS (K082606) cleared in 2008, PLATO SRS (K010784) cleared in 2001, FOCUS SRS (K973936 + K002147) cleared in 1998, IMSure (K031975) cleared in 2000, ERGO++ (K080601) cleared in 2008, ADAC Pinnacle3 (K951581) cleared in 1996, Pinnacle3 (K993923) cleared in 2000, and the FOCUS RTP (K915691) cleared in 1995. .'
Page #:
5.1
Traditional Premarket Notification for MIRS, Nuclemed S.A.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for NUCLEMED. The logo consists of a stylized letter N above the word "NUCLEMED". The letter N is black and white and appears to be made up of several smaller squares. The word "NUCLEMED" is written in a simple, sans-serif font.
Nuclem V. Arredondo 2684-2ª Piso ♦ 1426 Buenos Aires ♦ Argentina Tel: (54) 114896-0989
# 510(K) SUMMARY (CONTINUED)
MIRS ™
#### Device Description:
MIRS ™ is used to create treatment plans for patients for whom external beam radiation therapy has been MIRS ** Is used to create trouncell pails to pay three dimensional radiation dose distributions within the patient for a given treatment plan set up. MIRS (Module Integrated Radiotherapy System) is a modular patient for a given reatment planning system for different treatment modalities of radiation therapy, from three univelsional radiotherapy to Conformal Radiation Therapy, Intensity Modulated Radiation Therapy and Conventional radioneraly with Conic Collimators. MIRS supports any stereotactic frame. MIRS has automatic multimodality image registration, with and without fiducials. Besides 3D RMI and CT images studies with any orientation (axial, coronal or sagital), MIRS supports registration of functional RMI, PET and angiography.
#### Intended Use:
MIRS ™ is used to create treatment plans for patients for whom radiation therapy has been prescribed. The system will calculate and display three dimensional radiation dose distributions within the patient for a given treatment plan set up. MIRS can be used for different treatment modalities of radiation therapy, as conventional therapy. Conformal Radiation Therapy, Intensity Modulated Radiation Therapy, and Radiosurgery with Conic Collimators.
## Technological Characteristics Comparison to Predicate Device:
The proposed device was compared to the predicate devices as described in section 9 of this submission. All of the items contained in the tables of section 9 have been found to be substantially equivalent.
#### Performance Data:
The proposed device's performance was compared to the predicate device's performance as described in section 10 of this submission. All of the items contained in the tables of section 10 have been found to be substantially equivalent.
Traditional Premarket Notification for MIRS, Nuclemed S.A.
Page #: 5.2
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol. The logo is in black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
MAY 4 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nuclemed S.A. % Mr. Raymond Kelly Senior Consultant Arazy Group, Medes Argentina 56B Shadowbrook Dr. HUDSON NH 03051
Re: K090182 -
Trade/Device Name: MIRSTM Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: January 15, 2009 Received: February 3, 2009
## Dear Mr. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse, events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
21 CFR 876.xxx (Gastroenterology/Renal/Urology) (240) 276-0115 (240) 276-0115 (Obstetrics/Gynecology) 21 CFR 884.xxx (Radiology) (240) 276-0120 21 CFR 892.xxx Other (240) 276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Lois McMurtry
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
7
Nuclemed S.A. V. Arredondo 2684-2ª Piso ♦ 1426 Buenos Aires ♦ Argentina Tel: (54) 114896-0989
# INDICATIONS FOR USE
2090182 510(k) Number (if known):
MIRS™M Device Name:
Indications For Use:
MIRS ™ is used to create treatment plans for whom radiation therapy has been prescribed.
The batient the count of the last direct signed redistion dose distributions within t MIRS ™ is used to create treatment plans to patients to whoirs within the patient for The system will calculate and display international treatment modelities of radiation therapy, as
a given treatment plan set up. MIRS during interest the Modulation Therany, a given treatment plan set up. MIRS can be used for unterests and the same of the appy, and Radiosurgery with Conic Collimators.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Traditional Premarket Notification for MIRS, Nuclemed S.A.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.