K973936 · Computerized Medical Systems, Inc. · IYE · Jun 1, 1998 · Radiology
Device Facts
Record ID
K973936
Device Name
FOCUS RTP SYSTEM
Applicant
Computerized Medical Systems, Inc.
Product Code
IYE · Radiology
Decision Date
Jun 1, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The Stereotactic Radiosurgery system is to be used to create a treatment plan for any patient with single or multiple lesions of the brain and for whom radiation therapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
Device Story
FOCUS Stereotactic Radiosurgery Treatment Planning System is an add-on to existing radiation treatment planning software; accepts patient diagnostic imaging (CT/MR) and linear accelerator dosimetry data; allows dosimetrists/physicists to define target volumes and treatment scenarios (beam number, position, energy); uses BRW headframe for patient coordinate system; performs coordinate transformation between headframe and imaging data; calculates and displays 2D/3D radiation dose distributions; output used by clinicians to select optimal treatment plan; hard-copy output for patient files; used in hospital/clinic office settings; benefits patients by enabling precise planning for single-session radiosurgery.
Clinical Evidence
No clinical data. Bench testing only. Seven categories of non-clinical tests performed: CT/MR image transfer, coordinate transformation algorithm verification, fiducial marker localization, ruler/pixel size accuracy, target point verification, isodose alignment, and dose calculation algorithm accuracy. Results met all pre-defined criteria.
Technological Characteristics
Software add-on for existing FOCUS system; runs on UNIX-based RISC workstations; supports 2D/3D planning; utilizes Ratio Tissue Air Ratio (RTAR) dose calculation algorithm; supports BRW headframe for coordinate transformation; no image fusion capability; no physical connection to radiation source.
Indications for Use
Indicated for patients with single or multiple brain lesions requiring radiation therapy. No specific age or gender restrictions.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
ADAC Pinnacle3 APEX Stereotactic Radiosurgery Planning System (K951581)
Nucletron PLATO SRS Stereotactic Radiosurgery System (K940001)
RSA XKNIFE3 Stereotactic Radiosurgery System (K953482)
Submission Summary (Full Text)
{0}------------------------------------------------
K 973936
#### JUN 1 1998
## 510(k) Summary of Safety and Effectiveness Data FOCUS Stereotactic Radiosurgery Treatment Planning System
| Submitter Name: | Computerized Medical Systems, Inc. |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 1195 Corporate Lake Drive<br>St. Louis, MO 63132-1716 |
| Submitter Phone: | 314 993 0003 |
| Submitter Fax: | 314 993 0075 |
| Contact Person: | Michael A. Parsons - Director - Quality Assurance and<br>Regulatory Affairs |
| Date Summary Prepared: | September, 1997 |
| Device Trade Name: | FOCUS Stereotactic Radiosurgery Treatment Planning<br>System |
| Device Common Name: | Radiation Treatment Planning System |
| Device Classification: | System, Simulator, Radiation Therapy per 21 CFR<br>892.5840 (Class II) |
| Substantial Equivalence: | ADAC Pinnacle3 APEX Stereotactic Radiosurgery<br>Planning System - K951581<br>Nucletron PLATO SRS Stereotactic Radiosurgery<br>System - K940001<br>RSA XKNIFE3 Stereotactic Radiosurgery System -<br>K953482 |
Device Description: The FOCUS Radiation Treatment Planning System accepts a) patient diagnostic imaging data from CT or MR scans or from films and b) "source" dosimetry data, typically from a linear accelerator. The system then permits the user to
Computerized Medical Systems Inc. - Premarket Notification - FOCUS Stereotactic Radiosurgery Page 3-1
{1}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness Data FOCUS Stereotactic Radiosurgery Treatment Planning System Page 2 of 3
define a target volume to be treated based on these diagnostic images. Based on the prescribed dose, the user, typically a Dosimetrist or Medical Physicist, can then create multiple treatment scenarios involving the number, position(s) and energy of radiation beams and the use of treatment aids between the source of radiation and the patient (wedges, blocks, ports, etc.). The FOCUS System then produces a display of radiation dose distribution within the patient, indicating not only dose to the target volume but to surrounding tissue and structures. The "best" plan satisfying the prescription is then selected. one which maximizes dose to the target volume while minimizing dose to surrounding healthy volume. The parameters of the plan are output in hard-copy format for later reference and for placement in the patient file.
The software addition to add stereotactic radiosurgery treatment planning capability to FOCUS was developed at the Mallinckrodt Institute of Radiology of the Washington University School of Medicine in St. Louis, Missouri. The addition of Stereotactic Radiosurgery Therapy Planning capability to the system is a logical extension of the original intended use of the system. Stereotactic Radiosurgery is actually a sub-set of standard treatment planning in that it accepts the same inputs in the same formats as that for general planning. One significant, simplifying difference is that, in Stereotactic Radiosurgery Therapy, there are no treatment aids between the source and the patient. All therapy is performed using what are referred to as "open fields". A second difference is the introduction of "headframes" which, as the name implies, attach to the patients head and provide a unique patient coordinate system. The significant issue introduced here is one of coordinate transformation between the headframe and the patient diagnostic imaging data from the CT or MR machine. The system has been validated and marketed to support only the BRW headframe for CT or MR scans.
Device Intended Use: The Stereotactic Radiosurgery system is to be used to create a treatment plan for any patient with single or multiple lesions of the brain and for whom radiation therapy has been prescribed.
The system will calculate and display, both on-screen and in hard-copy, either two- or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
Summary of Technological Characteristics Compared To Predicate Devices: The Stereotactic Radiosurgery Treatment Planning capability described in this 510(k) has been designed as an "add-on" to the existing FOCUS Radiation Treatment Planning System previously cleared under K915691 in February, 1995. The FOCUS System with Stereotactic Radiosurgery Planning capability incorporates no technological characteristics not currently existent in the predicate Stereotactic Radiosurgery devices.
{2}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness Data FOCUS Stereotactic Radiosurgery Treatment Planning System Page 3 of 3
The FOCUS System supports all Linear Accelerators, has 3-D visualization capabilities and can perform both 2-D and 3-D treatment planning. Beam's Eye View (BEV) presentation is available as is a Time/MU calculator. All predicate systems, and FOCUS, use some form of the Ratio Tissue Air Ratio (RTAR) Algorithm to calculate dose. Likewise, all systems run under UNIX on a RISC-based processor workstation. The FOCUS System, unlike most of the predicate devices, has been validated only for use with the Brown-Roberts-Wells (BRW) headframe (there are at least four more on the market) and it does not have image fusion capabilities (the ability to map a MR scan onto a CT scan to obtain better imaging data of soft tissue).
Summary of Non-clinical Tests: Seven categories of tests were performed to evaluate the performance of the system: 1) Transfer of CT and MR Images, 2) Verification of Coordinate Transformation Algorithm, 3) Fiducial Marker Localization Verification, 4) Verification of Accuracy of Linear and Spherical Rulers and Checking of Image Pixel Size Detection, 5) Verification of Target Points In Different Software Modules of the System, 6) Verify Alignment of Isodoses with Image Anatomy and 7) Verification of Accuracy of the Dose Calculation Algorithm. This last test was performed by verifying system outputs against calculated and measured data collected especially for this project. The Coordinate Transformation verification was performed against a current marketed device.
Conclusions From Non-Clinical Tests: All testing provided results which met the criteria set. The system has been judged to be substantially equivalent to predicate devices and safe and effective for clinical use in the planning of Stereotactic Radiosurgery therapy situations using the BRW headframe. Normal QA of treatment planning activities, including review of the plans by a Medical Physicist and a Radiation Oncologist/Neurosurgeon prior to their use, is recommended.
Summary of Clinical Tests: Clinical testing was not performed as part of the development of this feature by the Mallinckrodt Institute of Radiology of the Washington University School of Medicine in St. Louis, MO. Clinical testing is not required to demonstrate substantial equivalence or safety and effectiveness.
{3}------------------------------------------------
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard Rockville MD 20850
Image /page/3/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of a human figure, possibly representing a family or individual, with arms outstretched.
JUN 12:08
Michael A. Parsons Director, Quality Assurance & and Regulatory Affairs Computerized Medical Ssytem (CMS) 1195 Corporate Lake Drive St. Louis, MO 63132
Re:
K973936 FOCUS Stereotactic Radiation Treatment Planning System Dated: October 8, 1997 Received: October 15, 1997 Regulatory class: II Procode: IYE 90, CFR 892.5050
Dear Mr. Parsons:
ﻨﺎ
We have reviewed your Section 5100%) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Feceral Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Fecteral Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odch/dsmaldsmamain.html".
Sincerely yours,
Kilian Yu
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
JUN 1 1998
#### Statement of Indication for Use
# 510(k) Number: K973936
#### Device Name: FOCUS Stereotactic Radiosurgery Treatment Planning System
Purpose: To calculate and display, both on-screen and in hard-copy, either two- or threedimensional radiation dose distributions within a patient diagnosed as having single or multiple brain lesions based on user-defined radiation treatment plan parameters.
Function: The FOCUS Radiation Treatment Planning System accepts a) patient diagnostic imaging data from CT or MR scans or from films and b) "source" dosimetry data for a linear accelerator. The system then permits the user to define a target volume to be treated based on these diagnostic-images. Knowing the prescribed dose, the user, typically a Dosimetrist or Medical Physicist, can then create multiple treatment seenarios involving the number, position(s), shape (achieved by use of treatment aids such as wedges, blocks, and ports between the source of radiation and the patient) and energy of the radiation beams. The FOCUS System then produces a display of radiation dose distribution within the patient, indicating not only dose to the target volume but to surrounding tissue and structures. The "best" plan satisfying the prescription is then selected, one which maximizes dose to the tumor volume while minimizing dose to surrounding healthy volume. The parameters of the plan are output in hard-copy format for later reference and for placement in the patient file.
The addition of Stereotactic Radiosurgery Therapy Planning capability to the system is a logical extension of the original intended use of the system. Stereotactic Radiosurgery is actually a sub-set of standard treatment planning in that it accepts the same inputs in the same formats as that for general planning. One significant difference is that, in Stereotactic Radiosurgery There are no treatment aids between the source and the patient. All therapy is performed using what are referred to as "open fields". One new item is the introduction of "headframes" which, as the name implies, attach to the patients head, enhance target immobilization, and provide a unique patient coordinate system. The significant issue introduced here is one of coordinate transformation between the headframe and the patient diagnostic imaging data from the CT or MR machine. The system has been validated and marketed to support only the Brown .-. Roberts-Wells (BRW) headframe for CT or MR scans.
One difference in delivery of treatment between Stereotactic Radiosurgery and other Treatment Planning is the number of therapy sessions required to deliver the presectbed dose. In stereotactic radiosurgery, the total prescribed dose is typically delivered in a single treatment. For therapy to target volumes in other locations of the anatomy, treatment is typically broken into multiple (20-30) fractions. While this change in therapy delivery introduces no significant change in the treatment planning system, it does increase the amount of patient risk resulting from an incorrect treatment plan. For
{5}------------------------------------------------
### Statement of Indication for Use (continued from Page 5-1)
patients receiving fractionated therapy, checks of dose received by the patient are made weekly and any mistake in the plan or delivery may be corrected by raising or lowering the delivered dose over the remaining fractions of therapy. For patients undergoing only one session of radiation, once a mistake is made involving lesion overdose, it cannot be adjusted.
Target Population: The Stereotactic Radiosurgery system is to be used to create a treatment plan for any patient with single or multiple lesions of the brain and for whom radiation therapy has been prescribed. Patient demographics are not an issue in creating the treatment plan.
Intended Environment for Use: The System is typically used in an office setting within a hospital/clinic. It is not located in the same room as the Linear Accelerator or other source of radiation. There is no need for physical connection between the source of the radiation and the Treatment Planning System.
Claims for the device: The FOCUS Stereotactic Radiosurgery Treatment Planning System will provide displays, both on-screen and in hard-copy format, of the radiation dose distributions for a given beam set-up(s) defined by the treatment planner. No claims of dose calculation accuracy, accuracy for CT localization, or center of radiation versus actual center of the lesion is indicated.
Substantial Equivalence: The system is substantially equivalent to at least three currently legally marketed devices. It is a subset of these devices, having no features not already found on one or more of the systems. See Section 7 of this Premarket Notification for details of Substantial Equivalence determination.
> Concurrence of the Center for Devices and Radiological Health, Office of Device Evaluation (ODE)
Thrid b. Jysom
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devic 510(k) Number .
Prescription Use ... V.. Over the Counter Use OR per 21 CFR 801.109
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.