K071130 · Universal Medical, Inc. · DXH · May 14, 2008 · Cardiovascular
Device Facts
Record ID
K071130
Device Name
HEARTRAK AF
Applicant
Universal Medical, Inc.
Product Code
DXH · Cardiovascular
Decision Date
May 14, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2920
Device Class
Class 2
Indications for Use
Heartrak Smart AF is a hand-held, portable, externally applied, cardiac event recorder that is intended for transtelephonic use. Patient calls a receiving center at the hospital or physician's office from the patient's home to play back the recording. Heartrak Smart AF converts electrocardiogram (ECG) signals into audio tones which are transmitted over the telephone lines. Heartrak Smart AF does not deliver any energy, administer any drugs, or control a patient's life. Heartrak Smart AF is not a diagnostic tool and performs no diagnostic functions.
Device Story
Hand-held, portable, externally applied cardiac event recorder; used by patients at home. Device captures ECG signals; converts signals into audio tones for transmission over standard telephone lines to a receiving center at a hospital or physician's office. Physician/staff at receiving center listen to/process audio output to review cardiac rhythm. Device includes an added firmware module for atrial fibrillation detection. Does not deliver energy, administer drugs, or provide automated diagnostic conclusions. Benefits include remote cardiac rhythm monitoring without requiring patient presence in clinic.
Clinical Evidence
Bench testing only. Atrial fibrillation detection module validated using annotated MIT-BIH AFIB database, MIT-BIH Arrhythmia database (Series 100 and 200), and MIT Wave Form Data Base (WFDB) software. Testing performed in compliance with ANSI/AAMI EC38:1998, IEC 60601-1, IEC 61000-4, ANSI/AAMI EC57:1998/R2003, and AAMI ECAR (1987) protocols.
Technological Characteristics
Hand-held, portable, externally applied cardiac event recorder. Identical materials and design to predicate. Energy source: battery (implied). Connectivity: transtelephonic audio transmission. Firmware: includes arrhythmia and atrial fibrillation detection module. Standards: ANSI/AAMI EC38:1998, IEC 60601-1, IEC 61000-4, ANSI/AAMI EC57:1998/R2003.
Indications for Use
Indicated for patients requiring transtelephonic cardiac event recording. Device is a hand-held, portable, externally applied recorder used by patients at home to transmit ECG signals via telephone to a receiving center at a hospital or physician's office. Not a diagnostic tool; does not deliver energy or drugs.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
K071130 R /3
MAY 1 4 2008
## "510(k) Summary" As Required By 21 CFR Section 807.92(c)
#### Applicant's Name: Universal Medical Incorporated 275 Phillips Blvd, Ewing, New Jersey 08618 Phone (609) 671-1790 Fax (609) 671-1765
Establishment Registration Number: 2248680
Contact Person:
Steven Adamsky
Date:
09-19-2007
{1}------------------------------------------------
K071130
p.2/3
### Proprietary Name: Heartrak Smart AF (K071130)
Classification Name: Telephone Hlectrocardiograph Transmitter and Receiver 21 CFR 870.2920.
Predicate Devices: Heartrak Smart AT (K033451)
#### Device Description:
Heartrak Smart AF is a hand-held, portable, externally applied transtelephonic cardiac event recorder device, which is intended for transtelephonic use.
#### Intended Use:
Heartrak Smart AF is a hand-held, portable, externally applied transtelephonic cardiac event recorder, which is intended for transtelephonic use.
#### Indication for Use:
Heartrak Smart AF is a hand-held, portable, externally applied transtelephonic cardiac event recorder device, which is intended for transtelephonic use. Patient calls a receiving center at the hospital or physicians office from the patient's home to play back the recording. Heartrak Smart AF converts ECG signals into audio tones which are transmitted over the telephone lines. Heartrak Smart AF does not deliver any energy. administer any drugs, or controls a patient's life. Heartrak Smart AF is not a diagnostic tool and performs no diagnostic functions.
#### Technological Characteristics:
All of the technological characteristics are identical to the predicate device, including design, material, and energy source. A module to better detect Atrial Fibrillation was added to the Heartrak Smart AT (predicate device) firmware to develop the Heartrak Smart AF (subject device). The Heartrak Smart AT device and Heartrak Smart AF device are otherwise identical in materials and technology.
#### Substantial Equivalence:
The Heartrak Smart AF is substantially equivalent to the predicate device, Heartrak Smart AT. The safety and effectiveness of this device is substantially equivalent to the predicate device. There are no known contradictions for use of this type of device. All of the features in this device present a non-significant risk to the user. The Substantial Equivalence Table is presented in the submission, Section 12 Substantial Equivalence Discussion.
A module to better detect Atrial Fibrillation was added to the Heartrak Smart AT (predicate device) firmware to develop the Heartrak Smart AF (subject device). The Heartrak Smart AT device and Heartrak Smart AF device are otherwise identical in materials and technology.
{2}------------------------------------------------
# 17130 3/3
#### Conclusions Based on Tests Submitted:
All testing performed on the Heartrak Smart AF device was derived from the risk assessment that evaluated the effects of the addition of the atrial fibrillation detection module to the existing Heartrak Smart AT arrhythmia detection firmware. Heartrak Smart AF testing included environmental and software validation testing and device testing,
The atrial fibrillation detection module was tested, using available annotated MIT-BIH AFIB databasc and MIT-BIH Arrhythmia database (Series 100 and Series 200), which are annotated ECG databases, and the MIT Wave Form Data Base (WFDB) software to derive performance statistics for Heart Smart AF. Testing was performed in compliance with standard cvaluation protocols. These protocols have been adopted as parts of the ANSI/AAMI EC38:1998; IEC 60601-1; IEC 61000-4, and ANSI/AAMI EC57:1998/R2003 recommendations for testing cardiac rhythm and reporting the results of those tests. They include earlier evaluation protocols developed for an AAMI Recommended Practice, Testing and Reporting Performance Results of Ventricular Arrhythmia Detection Algorithms (AAMI ECAR, 1987).
The subject device, Heartrak Smart AF, has indications for use the same as the predicate device, Heartrak Smart AT. The testing included in our submission in Section 18 Performance Testing - Live Patient/Field Test and Bench Test demonstrates that there are no differences in the technological characteristics of the subject device and the predicate device; therefore there are no new issues of safety or effectiveness. Heartrak Smart AF is substantially equivalent to the predicate device, Heartrak Smart AT.
#### Promotional Material:
Instructional manual, labeling.
#### Address for Manufacturing Site:
Heartrak Smart AF will be manufactured by Universal Medical, Inc. at its offices at 275 Phillips Boulevard, Ewing, New Jersey 08618.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 4 2008
Universal Medical, Inc. c/o Mr. Steven Adamsky Executive Vice President 275 Philips Blvd. Ewing, NJ 08618
Re: K071130
Heartrak Smart AF Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter and Receiver Regulatory Class: Class II (two) Product Code: DXH Dated: April 24, 2008 Received: April 25, 2008
Dear Mr. Adamsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Steven Adamsky
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Prem P. Zukerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Universal Medical Incorporated, a Mednet company. The logo features a stylized letter "U" with a star at the top, enclosed within a circle. The words "UNIVERSAL MEDICAL" are stacked above the word "INCORPORATED", and the phrase "A MEDNET COMPANY" is at the bottom of the logo.
#### UNIVERSAL MEDICAL INCORPORATED
275 Phillips Blvd, Ewing, NJ 08618 (800) 606-5511 Fax (800) 889-5383
# Indications for Use
510(k) Number (if known): K071130
Device Name: Heartrak Smart AF
Indications for Use:
Heartrak Smart AF is a hand-held, portable, externally applied, cardiac event recorder that is intended for transtelephonic use. Patient calls a receiving center at the hospital or physician's office from the patient's home to play back the recording. Heartrak Smart AF converts electrocardiogram (ECG) signals into audio tones which are transmitted over the telephone lines.
Heartrak Smart AF does not deliver any energy, administer any drugs, or control a patient's life. Heartrak Smart AF is not a diagnostic tool and performs no diagnostic functions.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| (Division Sign-Off) |
|---------------------|
|---------------------|
| Division of Cardiovascular Devices |
|------------------------------------|
|------------------------------------|
| 510(k) Number | k071130 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.