K083511 · Industrial Water Solutions, Inc. · MMK · Mar 3, 2009 · General Hospital
Device Facts
Record ID
K083511
Device Name
ENVIRO SHARP SOLUTIONS 3
Applicant
Industrial Water Solutions, Inc.
Product Code
MMK · General Hospital
Decision Date
Mar 3, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The Enviro Sharp Solutions 3 reusable container is intended for use by healthcare providers, such as hospitals, laboratories, medical clinics, veterinary clinics and other facilities where needles, sharps waste and other infectious waste are generated. The containers are designed to safely contain sharps waste prior to removal from generating facility and until ultimate treatment and disposal of waste. Containers are of such a design and materials as to withstand emptying, unloading washing and disinfecting for reuse according to 49 CFR 178.603, 173.44655(d), 173465(e) and 178.608.
Device Story
Enviro Sharp Solutions 3 is a reusable plastic sharps disposal container; features mail-drop, torturous path tumbler lid for horizontal syringe disposal. Syringes deposit automatically into container; tumbler indicates when full. Once full, locking tabs secure container for transport. Sci-Med Waste employees empty, sanitize, and return unit to facility. Used in healthcare settings (hospitals, clinics, labs) by healthcare providers. Durable high-density polyethylene (HDPE) construction ensures leakproof, puncture-resistant, and stable performance. Reusable design offers environmentally friendly, cost-effective alternative to disposable containers. Meets OSHA Bloodborne Pathogens Standard requirements.
Clinical Evidence
Bench testing only. Simulated life-use testing (emptying, filling, washing, sanitizing) performed over two days showed no visible signs of failure. Device passed puncture resistance (Health Devices 22), leak resistance (24-hour water test), vibration (1-hour repetition bounce), free fall drop (5 drops at 3.9 feet), and stacking (24 hours under 65 lbs) tests per 49 CFR standards.
Technological Characteristics
Reusable sharps container; high-density polyethylene (HDPE) construction; wall thickness 0.125" ± 0.002". Features mail-drop, torturous path tumbler lid. Mechanical operation; no energy source. Complies with 49 CFR 178.603, 173.44655(d), 173.465(e), 178.608, and OSHA Bloodborne Pathogens Standard.
Indications for Use
Indicated for healthcare providers in hospitals, laboratories, medical clinics, and veterinary clinics to safely contain sharps waste, including needles, prior to removal, transport, and disposal.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
## FDA 510(k) Premarket Notification
K083SII
Industrial Water Solutions, Inc. October 28, 2008
### 807.92(a)(1)
:
1.0.00
COMPANY NAME:
COMPANY ADDRESS:
TELEPHONE NUMBER: FACSIMILE NUMBER:
CONTACT PERSON:
SUMMARY PREPARATION DATE:
807.92(a)(2)
TRADE or PROPRIET ARY NAME:
COMMON NAME:
CLASSIFICATION:
807.92(a)(3)
EQUIVALENT DEVICE #1:
CONTAINER LID:
EQUIVALENT IDEVICE #2:
CONTAINER BASE:
#### 807.92(a)(4)
DESCRIPTION:
510(k) SUMMARY
Industriul Willer Solinions, Inc.
2692 Willowlawn Street Rounoke, Virginia 24018
340/772-0599 540/772-7944
Thomas P. Stewart
November 24, 2008
Enviro Sharp Solutions 3
Reusable Sharps Disposal Container
Class II - Accessory Device
Bernis Sharp Schringlos
The Bornis Sharp Sentinal@ container lid ("lid") is used on the top of the Enviro Sharp Solutions 3 container. This Remis Sharp Sentinato container lid has been upproved under 510(k) #K953797.
SteriSharp™ 3 Gallon RSDC (510(k) #K020664
The bottom portion of the unit (the "comminer") is a rensable polychylene cavity with a wall thickness of 0.125" ± 0.002" similar to that of the StoriSharp™ 3. "The SteriSharp™ 3 comainer has been approvod under 510(k) #K020664
The Eaviro Sharp Solutions 3 is a reusable plustic sharps disposal compiner. Il icatures a mail drop, torturous path tumbler lid for Rale and casy disposal of sharps. Simply deposit syringes horizomally into the opening of the lid, and they are deposited automatically into the container. When the container is full, the tumbler will indicate that it is time to replace the unit. Just insert the locking tobs to secure the container in a closed position and replace it with a clean samilized unit. Using our proprietary equipment, Sci-Med Waste comployees will empry and sanitize the unit and relum it to the facility to use for the next switch out. The durable highdensity polychylone construction of the Enviro Sharp Solutions 3 makes it loakproof, puncture resistant and sinble. It meets or oxcocds all OSTIA recommendations for sharps containers, and because it is rewaible, it is more crivironmentully friendly and less expensive than disposable alternatives.
A label will be attached in the outside of the container specifying the intended use of the container. A sample of this lubel is attached at the end of this document. (Sec Attachment 1)
{1}------------------------------------------------
## 807.92(n)(5)
INTENDED USE:
The Haviro Sharp Solutions 3 reusable container is intended for use by healthcare providers, such as hospitals, luboratories, medical clinics, versinary clinics und other facilities where needles, sharps waste und other infectious waste are generated. The containers are designed to suffely contain sharps waste prior to removal from generating facility and until ultimate treatment and disposal of waste. Containers are of such a design and materials as to withstand craptying, unloading washing and disinfecting for reuse according to 49 CFR 55 178,603, 173,44655(d), 173465(e) and 178.608.
## 807.92(a)(6)
The Enviro Sharp Shiple on Inchiner is substantially equivalent to the SteriSharp™ 3, and west the Benting Nectination (510k)/K953797 lid that is a Remi-transpirent nuil drop, 10/voronylene muil drop, 10/0rous path tunbler lid. It is injection molded and varies in thickness from 0.0.5" to 0.07". The base vontainer is similar in design and ergonomic characteriales to the SteriShurp™ 3. While the base of the Starish is manufactured using polypropying, the Enviro Sharines is unit is manufactured viri a thicker wall (0.125" ±0,002" as compared to 0.06" ±0.01") and stronger polymer (HDPE) to help contier reusable qualities.
| COMPARISON TABLE (IIJ) | BEMIS SHARP SENTINAL® | NAME OF YOUR CONTAINER |
|------------------------|----------------------------|------------------------|
| Indications for use | Healthcare Sharps | Same |
| Target population | Healthcare Professionals | Same |
| Design | Torturous path, mail-drop | Same |
| Materials | Polypropylene | Same |
| Performance | Multiple Use | Same |
| Mechanical safety | Mail-drop | Same |
| Where used | Healthcare facilities/labs | Same |
| Standards met | 49 CFR/HD 22 (multi-use) | Same |
The Laviro Shap Solidions 3 reusable contuner is also submantially equivelent to the Steckliner They are Inch a combination of a polypropylane turnhir lid with upolyedrylone hase. The SteriShap™ 3 base is a road continuel motion lower under liner low dersity Polyethylone cavity, while the Enviro Sharp Solutions 3 is a roll, injection nolded high density.
| COMPARISON TABLE (container) | SteriSharp™ 3 | NAME OF YOUR CONTAINER |
|------------------------------|----------------------------|------------------------|
| Indications for use | Healthcare Sharps | Same |
| Target population | Healthcare Professionals | Same |
| Design | Injection mold tapered | Same |
| Materials | HDPE | Same |
| Performance | Multiple Use | Same |
| Where used | Healthcare facilities/labs | Same |
| Standards met | 49 CFR/HD 22 | Same |
#### 807.92(b)(1)
Over a two-day poriod from November 5, 2008 to November 5, 2008, all components of a randon sampling of the Enviro Sharp Solutions 3 container were subjected to a simulated life-use of elming, filling, depositing waste, one in an one in an onlying washing and subtizing. After simulated use, there were no visible signs of Inilure. Containces tested and passed (Scc helow )
# 807.92(b)(2) and (3)
The Enviro Sharp Solutions 3 meets und exceeds the primary design churacteristics neceded to comply with the OSILA Blood borno Pathogens Standard. Date for the following tests have been provided and are as follows:
| Puncture | Health Devices 22 | Needle penetration force | Pass |
|-----------------|-------------------|----------------------------|------|
| Leak Resistance | Health Devices 22 | 24 Hours filled with water | Pass |
| Vibration | 49 CFR 178.608 | 1 hour repetition bounce | Pass |
| Free Fall Drop | 49 CFR 178.603 | 5 drops 3.9 feet | Pass |
| Stacking | 49 CFR 178.606 | 24 hours under 65 lbs | Pass |
"The package, us submitted und tested, visually appents to satisfy the test critchi und is capable of preventing the loss or Ilspersal of the contents for continues normal to transport" (Container Testing Laboratory, Inc.) (See Attischnered 2)
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The eagle is facing to the right, and its wings are angled upwards. The text is in all capital letters and is evenly spaced around the circle.
6 2009 MAR
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Tom Stewart Industrial Water Solutions, Incorporated 2692 Willowlawn Street Roanoke, Virginia 24018
Re: K083511
> Trade/Device Name: Enviro Sharp Solution 3 Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: January 15, 2009 Received: February 11, 2009
Dear Mr. Stewart:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Stewart
Please be advised that FDA's issuance of a substantial equivalence determination does not rnean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding postmarket surveillance, please contact: CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Anthony D. Austin for
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K08351)
Device Name: Enviro Sharp Solutions 3
Indication For Use:
The container is intended for use by healthcare providers, such as hospitals, laboratories, medical clinics, veterinary clinics and other facilities where needles are generated. The container is designed to safely contain sharps prior to removal from the generating facility, during transportation and until ultimate treatment and disposal of the sharps.
Prescription Use (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety
Division Sign-Off Office of In Vitro Diagnostic Device
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K083511
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.