K020664 · Sterilogic Waste Systems, Inc. · MMK · May 14, 2002 · General Hospital
Device Facts
Record ID
K020664
Device Name
STERISHARP 3-GALLON RSDC, MODEL RSDC3G
Applicant
Sterilogic Waste Systems, Inc.
Product Code
MMK · General Hospital
Decision Date
May 14, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The SteriSharp™ 3-Gallon RSDC reusable containers are intended for use by healthcare providers (hospitals, laboratories, medical clinics, veterinary clinics, and other such areas and facilities where needles, sharps waste and other infectious waste is generated). The containers are designed to safely contain sharps waste prior to removal from generating facility and until ultimate treatment and disposal of waste. Containers are of such a design and material as to withstand emptying, unloading, washing and disinfecting for reuse.
Device Story
SteriSharp 3-Gallon RSDC is a reusable sharps disposal container; features mail-drop, torturous path tumbler lid for automated sharps deposition. Used in healthcare facilities (hospitals, clinics, labs) by healthcare staff. Device collects sharps waste; when full, tumbler indicates replacement; locking tabs secure container for transport. SteriLogic employees empty, sanitize, and return units for reuse. Durable HDPE construction ensures leak-proof, puncture-resistant containment. Reusable design offers environmental and cost benefits over single-use alternatives; meets OSHA Bloodborne Pathogens standards.
Clinical Evidence
No clinical data. Bench testing only. Simulated life-use testing performed over two days (100 cycles of filling, emptying, washing, and sanitizing) showed no visible failure. Performance testing included puncture resistance (needle penetration force), leak resistance (24-hour water test), vibration (1-hour repetitive bounce), free fall drop (5 drops at 1.2 meters), and stacking (24 hours under 30 kg). All tests passed.
Technological Characteristics
Reusable sharps container; injection-molded HDPE base (0.125" wall thickness) and polypropylene mail-drop tumbler lid. Mechanical design features torturous path for sharps entry. Meets 49 CFR 178.603, 178.606, 178.608 and Health Devices 22 standards. Sanitized via proprietary SteriLogic Sharps Consolidation Unit.
Indications for Use
Indicated for healthcare providers, including hospitals, laboratories, medical clinics, and veterinary clinics, for the containment of needles, sharps waste, and infectious waste generated in these facilities.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
# MAY 1 4 2002
# FDA 510(k) Notification
SteriLogic Waste Systems, Inc. February 15, 2002
KQX064
## 510(k) SUMMARY
807.92(a) (1)
COMPANY NAME:
COMPANY ADDRESS:
STERILOGIC WASTE SYSTEMS, INC.
6691 PICKARD DRIVE SYRACUSE, NY 13211
TELEPHONE: FAX:
(215) 914-1110 (215) 914-0110
CONTACT PERSON:
ERIK R. SYNNESTVEDT FEBRUARY 15, 2002
SUMMARY PREPARATION DATE:
807.92(a)(2)
STERISHARP™ 3-GALLON RSDC TRADE OR PROPRIETARY NAME:
REUSABLE SHARPS DISPOSAL CONTAINER COMMON NAME:
CLASSIFICATION:
CLASS II -- ACCESSORY DEVICE
807.92(a)(3)
## EQUIVALENT DEVICE #1: BEMIS SHARPS DISPOSAL CONTAINER (510[k] #K931664)
CONTAINER LID: The Bemis Sharps Disposal Container lid ('lid'') is used on the top of the SteriSharp™ 3 container. This Bemis Sharps Disposal Container has been approved under 510(k) #K931664.
## EQUIVALENT DEVICE #2: STERISHARP™ 2.5-GALLON RSDC (510]k] #K991662)
CONTAINER BASE: The bottom portion of the unit (the "container") is a reusable polyethy lene cavity with a wall thickness of 0.125" ± 0.002"similar to that of the SteriSharp™ 2.5. The SteriSharp™ 2.5 container has been approved under 510(k) #K991662.
{1}------------------------------------------------
# FDA 510(k) Notification
### 807.92(a)(4)
#### DESCRIPTION:
The SteriSharp™ 3 is a reusable plastic sharps disposal container. It features a mail drop, torturous path tumbler lid for safe and easy disposal of sharps. Simply deposit syringes horizontally into the opening of the lid and they are deposited automatically into the container. When the container is full the tumbler will indicate that it is time to replace the unit. Just insert the locking tabs to secure the container in a closed position and replace it with a clean sanitized unit. Using our proprietary equipment, SteriLogic employees will empty and sanitize the unit and return it to your facility to use for the next switch out. The durable high-density polyethylene construction of the SteriSharp™ 3 makes it leak-proof, puncture resistant and stable. It meets or exceeds all OSHA recommendations for sharps containers, and because it is reusable, it is more environmentally friendly and less expensive than disposable alternatives.
#### 807.92(a)(5)
#### INTENDED USE:
The SteriSharp™ 3 reusable container is intended for use by healthcare providers such as hospitals, laboratories, medical clinics, veterinary clinics, and other facilities where needles, sharps waste and other infectious waste is generated. The containers are designed to safely contain sharps waste prior to removal from generating facility and until ultimate treatment and disposal of waste. Containers are of such a design and material as to withstand emptying, unloading, washing and disinfecting for reuse according to 49 CFR Sections 178.603, 173.4465(d), 173.465(e) and 178.608.
### 807.92(a)(6)
The SteriSharp™ 3 reusable container is substantially equivalent to the Bemis Sharps Disposal Container in that the lid is identical; a semi-transparent natural colored polypropylene mail drop, torturous path tumbler lid. It is injection molded and varies in thickness from 0.05" to 0.07". The base container is similar in design and ergonomic characteristics to the Bemis Sharps Disposal Container. While the base of the Bemis unit is manufactured using polypropylene, the SteriSharp unit is manufactured with a thicker wall (0.125" + 0.002" as compared to 0.06" +0.01") and stronger polymer (HDPE) to help confer reusable qualities.
| COMPARISON TABLE (LID) | Bemis 3-Gallon | SteriSharpTM3 |
|------------------------|-----------------------------|----------------|
| Indications for use | healthcare sharps | same |
| Target population | healthcare professionals | same |
| Design | torturous path, mail-drop | same |
| Materials | polypropylene | same |
| Performance | single use | multiple use |
| Mechanical safety | mail-drop | same |
| Where used | healthcare facilities/labs | same |
| Standards met | 49 CFR / HD 22 (single use) | same/multi-use |
{2}------------------------------------------------
The SteriSharp™ 3 reusable container is also substantially equivalent to the SteriSharp™ 2.5 reusable container. They are both a combination of a polypropylene tumbler lid with a polyethylene base. The SteriSharp™ 2.5 base is a red, rotationally molded linear low density Polyethylene cavity while the SteriSharp™ 3 base is a red, injection molded high density polyethylene cavity. Both units are designed to be opened and sanitized automatically using the SteriLogic Sharps Consolidation Unit which is consistent with all applicable OSHA and DOH standards.
| COMPARISON TABLE (CONTAINER) | SteriSharp™2.5 | SteriSharp™3 |
|------------------------------|-----------------------------|------------------------|
| Indications for use. | healthcare sharps | same |
| Target population. | healthcare professionals | same |
| Design. | rotationally molded | injection mold tapered |
| Materials. | LLDPE | HDPE |
| Performance. | multiple use | same |
| Where used. | healthcare facilities/labs. | same |
| Standards met | 49 CFR / HD 22 | same |
## 807.92(b)(1)
Over a two-day period from July 25, 2001 to July 26, 2001, all components of a random sampling of the SteriSharp™ 3 container were subjected to a simulated life-use of 100 cycles of closing, filling, depositing waste, opening, emptying, washing and sanitizing. A detailed report is included below under section "TAB 3." After simulated use, there were no visible signs of failure. Containers were then performance tested and passed. (see below)
## 807.92(b)(2) and (3)
The SteriSharp™ 3-Gallon RSDC meets and exceeds the primary design characteristics needed to comply with the OSHA Bloodborne Pathogens Standard. Data for the following tests have been provided and are as follows:
| Puncture | Health Devices 22 | Needle penetration force | Pass |
|-----------------|-------------------|--------------------------|------|
| Leak Resistance | Health Devices 22 | 24 hrs filled with water | Pass |
| Vibration | 49 CFR 178.608 | 1 hour repetitive bounce | Pass |
| Free Fall Drop | 49 CFR 178.603 | 5 drops 1.2 meter | Pass |
| Stacking | 49 CFR 178.606 | 24 hrs. under 30 kg | Pass |
"The package, as submitted and tested, visually appears to satisfy the test criteria and is capable of preventing the loss or dispersal of the contents for conditions normal to transport." (Container Testing Laboratory, Inc.)
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Erik R. Synnestvedt SteriLogic Waste Systems, Incorporated 6691 Pickard Drive Syracuse, New York 13211
Re: K020664
Trade/Device Name: Sterisharp 3-Gallon RSDC, Model RSDC3G Regulation Number: 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK and FMI Dated: February 15, 2002 Received: March 1, 2002
Dear Mr. Synnestvedt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
# Page 2 - Mr. Synnestvedt
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Tunney
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) NOTIFICATION
STERILOGIC WASTE SYSTEMS, INC. FEBRUARY 15, 2002
# 510(k) NUMBER (IF KNOWN) :
DEVICE NAME : STERISHARP 3-GALLON RSDC, MODEL RSDC3G
INDICATIONS FOR USE :
The SteriSharp™ 3-Gallon RSDC reusable containers are intended for use by healthcare providers (hospitals, laboratories, medical clinics, veterinary clinics, and other such areas and facilities where needles, sharps waste and other infectious waste is generated). The containers are designed to safely contain sharps waste prior to removal from generating facility and until ultimate treatment and disposal of waste. Containers are of such a design and material as to withstand emptying, unloading, washing and disinfecting for reuse.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (per 21 CFR 801.109) | | OR | | Over-The-Counter-Use (optional format 1-2-96) |
|---------------------------------------|--|----|--|-----------------------------------------------|
| | | | | |
| | | | | |
| | | | | |
| | | | | |
*Sftr Clin*
(Division Sign-Off)
Division of Dental, Infection Control,
and General Hospital Devices
| 510(k) Number | K020664 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.