K082887 · Advanced Instrumentations, Inc. · MWI · Jan 16, 2009 · Cardiovascular
Device Facts
Record ID
K082887
Device Name
PM-1000F+PATIENT MONITOR
Applicant
Advanced Instrumentations, Inc.
Product Code
MWI · Cardiovascular
Decision Date
Jan 16, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The PM-1000F+ Patient Monitor is a vital signs monitor used on human patients The target populations are adult, pediatric and neonatal patients The PM-1000F+ Patient Monitor has many features and functions Its use is through an integrated keypad. knob, and intuitive menu system The patient parameters that can be monitored by the PM-1000F+ are ECG (3-Lead or 5-Lead selectable), arrhythmia detection, ST segment analysis Heart Rate (HR), Respiration Rate (RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure (IBP), Carbon Dioxide (CO2), and Anaesthetic Gases (AG) Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs The PM-1000F+ Patient Monitor is intended for use in hospital clinical areas such as intensive care units, cardiac care units, operating room, emergency department, to provide additional information to the medical and nursing staff about the physiological condition of the patient The PM-1000F+ Patient Monitor is intended to be used only under regular supervision of clinical personnel and by qualified medical personnel trained in the use of the equipment The intended location of use is clinics The PM-1000F+ Patient Monitor is not intended or recommended for use in a patient's home or residence, or when it has not been ordered by a physician
Device Story
PM-1000F+ is a battery or line-powered bedside patient monitor; acquires physiological signals including ECG, respiration, NIBP, SpO2, temperature, IBP, CO2, and anesthetic gases. Signals converted to digital data; processed for alarm conditions; presented on color TFT display. Operated by medical staff in hospitals (ICU, CCU, OR, ED) via integrated keypad and knob. Optional built-in recorder and CF memory card provide hard copies and trend data storage. Provides real-time monitoring and audible/visual alarms for parameter excursions; assists clinical decision-making by providing physiological status information to medical/nursing staff.
Clinical Evidence
Bench testing only. Compliance verified against ANSI/AAMI EC13-2002 (cardiac monitors), IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-27, IEC60601-2-30, and ISO14971. No clinical data presented.
Indicated for adult, pediatric, and neonatal patients requiring vital signs monitoring, including ECG, arrhythmia, ST segment, HR, RESP, NIBP, SpO2, TEMP, IBP, CO2, and anesthetic gases. Not for home use; requires physician order and qualified clinical personnel supervision.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
# 510(K) Summary
JAN 1 6 2009
This summary of 510(k) safety and effectiveness information is being submitted ın accordance with the requirements of SMDA 1990 and 21 CFR 807 92
The assıgned 510(k) number ıs K082887
| Submitter / Distributor | Advanced Instrumentations, Inc<br>6800 N W 77th Court<br>Miami, Fl 33166 | | |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|
| | Registration # 1066270 | | |
| | Telephone<br>Fax | 305-477-6331<br>305-477-5351 | |
| Contact Person: | Jorge Millan, PhD<br>Email jmillan@hiatec org | | |
| Manufacturer | Shenzhen Mindray Bio-medical Electronics Co , LTD<br>Mindray Building, Keji 12th Road South, Hi-tech<br>Industrial Park, Nanshan, Shenzhen, 518057, P R<br>China | | |
| | Registration # | 3004365718 | |
| | Tel<br>Fax | + 86-755-2658-2888<br>+ 86-755-2658-2680 | |
- January 15, 2009 Date Prepared .
# Name of the device:
- Trade/Proprietary Name PM-1000F+ Patient Monitor .
- Patient Monitor Common Name: .
#### Classification. .
| 21 CFR 870 2300 | Cardiac monitor (including cardiotachometer and rate alarm) | Class I |
|-----------------|-------------------------------------------------------------|---------|
| 21 CFR 870 1025 | Arrhythmia detector and alarm | Class I |
| 21 CFR 870 1025 | Detector and Alarm, Arrhythmia | Class I |
| 21 CFR 870 1025 | Monitor, ST Segment with Alarm | Class I |
| 21 CFR 870 1130 | Non-Invasive blood pressure measurement system | Class I |
| 21 CFR 870 1110 | Blood pressure computer | Class I |
{1}------------------------------------------------
| 21 CFR 870 2910 | Clinical Electronic Thermometers-<br>Temperature Monitor with Probe | Class II |
|-----------------|---------------------------------------------------------------------|----------|
| 21 CFR 870 2700 | Oximeter, Pulse | Class II |
| 21 CFR 870 2710 | Ear Oximeter, Pulse | Class II |
| 21 CFR 868 1400 | Carbon Dioxide Gas Analyzer | Class II |
| 21 CFR 868 1500 | Enflurane gas analyzer | Class II |
| 21 CFR 868 1620 | Halothane gas analyzer | Class II |
| 21 CFR 868 1700 | Nitrous Oxide gas analyzer | Class II |
| 21 CFR 868 1720 | Oxygen gas analyzer | Class II |
### Legally Marketed Predicate Device.
- PM-8000 Patient Monitor, Shenzhen Mindray Bro-medical K043348 Electronics Co . Ltd
- PM-9000 Express Patient Monitor, Shenzhen Mindray Bio-medical K053234 Electronics Co . Ltd
- PM Series Patient Monitors, Shenzhen Mindray Bıo-medical K070791 Electronics Co . Ltd
#### Device Description.
The PM-1000 F+ Patient Monitor is a battery or line-powered patient monitor The Patient Monitor acquires the physiological signals such as ECG, respiration (RES), non-invasive blood pressure (NIBP), saturation of pulse oxygen (SpO2), temperature (TEMP), invasive pressure (IBP), carbon dioxide (CO2) and anesthetic gases (AG) These physiological signals are converted into digital data and processed The PM 1000F+ Patient Monitor examines the data for alarm conditions and presents them on the color TFT display The Patient Monitor also provides advantageous operating control for the user The optional built-in recorder, the optional CF memory card provides hard copies of all digital data and waveforms as well as tabular and graphic trend information, and storage the previous monitoring data information when power off accidentally
#### Statement of Intended Use.
The PM-1000F+ Patient Monitor is a vital sıgns monıtor used on human patıents The target populations are adult, pediatric and neonatal patients The PM-1000F+ Patient Monitor has many features and functions Its use is through an ıntegrated keypad. knob, and intuitive menu system
The patient parameters that can be monitored by the PM-1000F+ are ECG (3-Lead or 5-Lead selectable), arrhythmia detection, ST segment analysis Heart Rate (HR), Respiration Rate (RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure (IBP), Carbon Dioxide (CO2), and Anaesthetic Gases (AG) Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs
{2}------------------------------------------------
The PM-1000F+ Patient Monitor is intended for use in hospital clinical areas such as intensive care units, cardiac care units, operating room, emergency department, to provide additional information to the medical and nursing staff about the physiological condition of the patient The PM-1000F+ Patient Monitor ıs intended to be used only under regular supervision of clinical personnel and by qualified medical personnel trained in the use of the equipment The intended location of use is clinics The PM-1000F+ Patient Monitor is not intended or recommended for use in a patient's home or residence, or when it has not been ordered by a physician
## Comparison of Technological Characteristics:
The PM-1000F+ Patient Monitor is substantially equivalent to systems currently marketed predicate devices The design, components, storage technology and energy source the PM-1000F+ Patient Monitor are similar to the predicate devices The PM-1000F+ provides a means for interfacing with a patient, collecting parameter specific physiological data, and processing the data for alarm generation and display of numeric values and waveforms on a bedside or central monitoring system The parameter's specification of the PM-1000F+ ıncludıng ECG, RESP, HR, PR, RESP, NIBP, SpO2, CO2, TEMP, IBP, IS ıdentıcal to the predicate devices
## Testing
Testing was conducted to validate and verify that the PM-1000F+ Patient Monitor met all design specifications and was substantially equivalent to the predicate devices Testing was performed to demonstrate compliance with the ANSI/AAMI standards EC13-2002, "Cardiac monitors, heart rate meters, and alarms " The PM-1000F+ has also been tested to assure compliance with the requirements of various published standards, including IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-27, IEC60601-2-30 and ISO14971
## Conclusion:
The conclusions drawn form clinical and bench testing of the PM-1000F+ Patient Monitor demonstrates that the device is as safe, as effective, and performs as well to the legally marketed predicate device
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Advanced Instrumentations, Inc c/o Jorge Millan, PhD Hialeah Technology Center 601 W 20th Street Hialeah, FL 33010
JAN 1 6 2009
K082887 Re
PM-1000F+ Patient Monitor Regulation Number 21 CFR 870 2300 Regulation Name Cardiac monitor (including cardiotachometer and alarm) Regulatory Class Class II (two) Product Code DSIMWI Dated December 22, 2008 Received December 22, 2008
Dear Dr Mıllan
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set
1151-1600 - 115 11 25 25 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11
{4}------------------------------------------------
Page 2- Dr Jorge Millan
forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Mısbrandıng by reference
to premarket notification" (21CFR Part 807 97) You may obtain other general information on your responsibilities under the Dryision of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry/support/index html
Sincerely yours,
and the may be the state of the minister
Bram D Žuckerman, M D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
IKE LA 18 12 2017
{5}------------------------------------------------
#### Indications for Use
#### 510(k) Number (if known) K082887
#### Device Name
PM-1000F+ Patient Monitor
#### Indications For Use
The PM-1000F+ Patient Monitor is a vital signs monitor used on human patients The target populations are adult, pediatric and neonatal patients
The patient parameters that can be monitored with the PM-1000F+ Patient Monitor are ECG(3-lead or 5-lead selectable), Heart rate (HR), Pulse Rate (PR), Respiration rate (RESP), Non-invasive blood pressure (NIBP), Saturation of Pulse Oxygen (SpO2), Temperature (TEMP), Invasive Blood pressure (IBP), Carbon dioxide (CO2) and Anaesthetic gases (AG) Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs
This monitor is intended for use in a health care facility setting It is intended for use by qualified medical personnel trained in the use of the equipment
The PM-1000F+ Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**510(k) Number** K082587
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.