The PM Series Patient Monitors including models PM-9000 Express and PM-8000 Express are intended to be used for monitoring, recording, and alarming of multiple physiological parameters in health care facilities by qualified health care professionals trained in the use of the equipment. The physiological parameters that can be monitored by PM-9000 Express Patient Monitor are: ECG(3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Pulse Rate(PR), Carbon Dioxide (CO2), and Anaesthetic Gases(AG). The physiological parameters that can be monitored by PM-8000 Express Patient Monitor are: ECG(3-lead or 5-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP) and Pulse Rate(PR) , and Carbon Dioxide (CO2). Their design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs. The target populations are human patients ranged from adult to neonate with the exception of the arrhythmia detection and ST Segment analysis, for which the farget populations are adult and pediatric only. The PM Series Patient Monitors are not intended for use in a patient's home or residence, or when it has not been ordered by a physician.
Device Story
Battery or line-powered patient monitors; acquire physiological signals including ECG, RESP, NIBP, SpO2, TEMP, IBP, CO2, and anesthetic gases. Signals converted to digital data; processed for alarm conditions; presented on color TFT display. Used in healthcare facilities by trained professionals. Optional built-in recorder and CF memory card provide hard copies and data storage. Operators adjust alarm settings for audible/visual notification of excursions. Output assists clinical decision-making by providing real-time monitoring and trend information.
Clinical Evidence
Bench testing only. Validated against FDA DCRND November 1993 guidance. ECG performance verified per ANSI/AAMI EC13-2002 and AAMI EC57-98. CO2 monitoring verified per ISO 21647-2005. System safety and operation validated against ISO 9919, IEC 60601-1 series, and ISO 14971.
Technological Characteristics
Battery or line-powered; color TFT display; CF memory card storage. Parameters: ECG, RESP, NIBP, SpO2, TEMP, IBP, CO2, AG. Standards: ANSI/AAMI EC13-2002, AAMI EC57-98, ISO 21647-2005, ISO 9919, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-27, IEC 60601-2-30, ISO 14971.
Indications for Use
Indicated for monitoring, recording, and alarming physiological parameters in healthcare facilities for patients ranging from neonate to adult. Arrhythmia detection and ST-segment analysis indicated for adult and pediatric patients only. Prescription use only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
K070791
P 1/4
## 510(K) SUMMARY JUN 2 6 2007
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: _______________________________________________________________________________________________________________________________________________
Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
- . Contact Person:
- Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
- � Date Prepared:
March 7, 2007
### Name of the devices:
・
r
- Trade/Proprietary Name: PM Series Patient Monitors (Including Models PM-9000 � Express and PM-8000 Express)
- Common Name: Patient Monitor �
#### ● Classification
| 21 CFR 870.1025 | Arrhythmia detector and alarm<br>(including ST-segment measurement and alarm) | Class II |
|-----------------|-------------------------------------------------------------------------------|----------|
| 21 CFR 870.1025 | Detector and Alarm, Arrhythmia | Class II |
| 21 CFR 870.1025 | Monitor, ST Segment with Alarm | Class II |
| 21 CFR 870.2300 | Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
| 21 CFR 870.1130 | Non-Invasive blood pressure measurement System | Class II |
| 21 CFR 870.1110 | Blood pressure computer | Class II |
| 21 CFR 880.2910 | Clinical Electronic Thermometers - | 0035 |
{1}------------------------------------------------
K070791
P2/4
| | Temperature Monitor with Probe | Class II |
|-----------------|--------------------------------|----------|
| 21 CFR 870.2700 | Oximeter, Pulse | Class II |
| 21 CFR 870.2710 | Ear Oximeter, Pulse | Class II |
| 21 CFR 868.1400 | Carbon Dioxide Gas Analyzer | Class II |
| 21 CFR 868.1500 | Enflura ne gas analyzer | Class II |
| 21 CFR 868.1620 | Halothane gas analyzer | Class II |
| 21 CFR 868.1700 | Nitrous Oxide gas analyzer | Class II |
| 21 CFR 868.1720 | Oxygen gas analyzer | Class II |
#### Legally Marketed Predicate Devices:
- 1) K#053234, PM-9000 Express Patient Monitor, SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
- 2) K#032858, IntelliVue MP60 and MP70 Patient Monitors, Philips Co., Ltd.
- 8000M SYSTEM, GENERAL ELECTRIC MEDICAL 3) K#012467, SOLAR SYSTEMS INFORMATION TECH.
- 4) K#020550. Passport 2 Vital Signs Monitor with View 12 ECG Analysis Module, Datascope Corp.
- PM-8000 EXPRESS PATIENT MONITOR, SHENZHEN MINDRAY 5) K#053193. BIO-MEDICAL ELECTRONICS CO., LTD.
#### Description:
The PM Series Patient Monitors including models PM-9000 Express are battery or line-powered patient monitors. The patient monitors acquire the physiological signals such as ECG, respiration (RESP), non-invasive blood pressure (NIBP), saturation of pulse oxygen (SpO2), temperature (TEMP), invasive pressure (IBP), carbon dioxide (CO) and anaesthetic gases (AG, PM-9000 Express only). These physiological signals are converted into digital data and processed. The monitors examine the data for alarm conditions and present them on the color TFT display. The monitors also provide advantageous operating control for the user. The optional built-in recorder, the optional CF memory card provides hard copies of all digital data and waveforms as well as tabular and graphic trend information, and storage the previous monitoring data information when power off accidentally.
#### Statement of intended Use:
The PM Series Patient Monitors including models PM-9000 Express are intended to be used for monitoring, recording, and alarming of multiple physiological parameters in health care facilities by qualified health care professionals trained in the use of the equipments.
The physiological parameters that can be monitored by PM-9000 Express Patient Monitor are:
0036
Page 2 of 4
{2}------------------------------------------------
ECG(3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Pulse Rate(PR), Carbon Dioxide (CO2), and Anaesthetic Gases(AG).
The physiological parameters that can be monitored by PM-8000 Express Patient Monitor are: ECG(3-lead or 5-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP) and Pulse Rate(PR), and Carbon Dioxide (CO2).
Their design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The target populations are human patients ranged from adult to neonate with the exception of the arrhythmia detection and ST segment analysis, for which the target populations are adult and pediatric only.
The PM Series Patient Monitors are not intended for use in a patient's home or residence, or when it has not been ordered by a physician.
#### Comparison of Technological Characteristics:
The modified devices are substantially equivalent to systems currently marketed predicate devices. The design, components, storage technology and energy source of the modified devices are similar to the predicate devices. Both the modified devices and the predicate devices provide a means for interfacing with a patient, collecting parameter specific physiological data, and processing the data for alarm generation and display of numeric values and waveforms on a bedside or central monitoring system. The parameters' specifications and technologies of the modified device are similar to the predicate devices.
The technological differences do not affect the safety or efficacy of the devices. Any safety issues that may be raised by software controlled medical devices are the same issues already addressed by the predicate devices and are addressed in the systems hazard analysis and in the system validation.
#### Testing:
Bench testing has been conducted to validate and verify that the PM Series Patient Monitors including models PM-9000 Express and PM-8000 Express meet all design specifications and are substantially equivalent to the predicate devices. The testing consisted of all environmental testing identified in the FDA's DCRND November 1993 "Reviewer Guidance Document for Premarket notification Submissions" Draft Guidance Document. Additional testing of ECG portion of the system has been performed to demonstrate compliance with the
Passion of C
{3}------------------------------------------------
ANSI/AAMI EC13-2002, "Cardiac monitors, heart rate meters, and alarms" and AAMI EC57-98, "Testing and Reporting Performance Results of Cardiac Rhythm and ST-Segment Measurement Algorithms". Additional Testing of the Carbon Dioxide (CO2) portion of the system has been conducted according to the requirements outlined in the Standards ISO 21647-2005, "Medical electrical equipment - Particular requirements for the basic safety and essential performance of respiratory gas monitors".
Finally, a hazard analysis of the systems and their software have been performed and testing has been conducted to validate the systems overall operation. The PM Series Patient Monitors including models PM-9000 Express and PM-8000 Express have also been tested to assure compliance with the requirements of various published standards, including 1SO9919, IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-4, IEC60601-2-27, IEC60601-2-30 and ISO14971, etc..
#### Conclusion:
The conclusions drawn from clinical and bench testing of the PM Series Patient Monitors including models PM-9000 Express and PM-8000 Express demonstrate that the devices are as safe, as effective, and performs as well as or better than the legally marketed predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of a bird-like figure with three lines representing the wings and a stylized head and body. The overall design is simple and clean, with a focus on the department's name and a symbolic representation of health and human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 2 6 2007
Shenzhen Mindray Bio-medical Electronics Co., LTD c/o Mr. Li Dongling Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P.R. China
Re: K070791
PM Series Patient Monitors Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: May 30, 2007 Received: June 1, 2007
Dear Mr. Dongling:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Li Dongling
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: PM Series Patient Monitors
Indications for Use:
The PM Series Patient Monitors including models PM-9000 Express and PM-8000 Express are intended to be used for monitoring, recording, and alarming of multiple physiological parameters in health care facilities by qualified health care professionals trained in the use of the equipment.
The physiological parameters that can be monitored by PM-9000 Express Patient Monitor are: ECG(3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Pulse Rate(PR), Carbon Dioxide (CO2), and Anaesthetic Gases(AG).
The physiological parameters that can be monitored by PM-8000 Express Patient Monitor are: ECG(3-lead or 5-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP) and Pulse Rate(PR) , and Carbon Dioxide (CO2).
Their design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The target populations are human patients ranged from adult to neonate with the exception of the arrhythmia detection and ST Segment analysis, for which the farget populations are adult and pediatric only.
The PM Series Patient Monitors are not intended for use in a patient's home or residence, or when it has not been ordered by a physician.
| Prescription Use | X (Part 21 CFR 801 Subpart D) |
|----------------------|-------------------------------|
| AND/OR | |
| Over-The-Counter Use | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K0 7029 |
| Page | 1 of 1 |
|------|--------|
| | 0033 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.