K082775 · Siemens Medical Solutions USA, Inc. · IYE · Feb 19, 2009 · Radiology
Device Facts
Record ID
K082775
Device Name
PRESCISION OPTION
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYE · Radiology
Decision Date
Feb 19, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer. The LINAC is a high-dose and high-dose rate medical linear accelerator optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT) and when used in conjunction with the PreScision™ option, supports precision stereotactic applications. The stereotactic applications include single-session radiosurgery, fractionated stereotactic radiation therapy, fractionated stereotactic intensity modulated radiation therapy for lesions, tumors and conditions anywhere in the body where radiation therapy is indicated. The syngo® Suite for Oncology Workspaces: The syngo®workspaces includes a number of syngo®based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording.
Device Story
PreScision™ is an optional feature for Siemens ARTISTE™, ONCOR™, and PRIMUS™ linear accelerators (LINACs). It enables high-dose rate (up to 2,000 MU/min) unflattened X-Ray photon beam delivery (7UF energy) for precision stereotactic applications (SRS/SRT). Input includes patient treatment plans managed via syngo®-based workspaces (COHERENCE™/PRIMEVIEW™). The system uses MLCs (58, 82, or 160-leaf) and optional third-party gating/positioning hardware to deliver radiation. Operated by radiation therapists in clinical settings, the device provides high-precision dose delivery for tumors. Output is therapeutic radiation; clinicians use the system to verify patient positioning via EPID (OPTIVUE™) and record treatment delivery. Benefits include reduced treatment times and enhanced precision for stereotactic procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness are supported by risk analysis, software verification, and validation testing of system requirements.
Technological Characteristics
Optional feature for existing LINACs. Delivers 6.67 MV unflattened X-Ray photon energy. Supports 5cm x 5cm to 40cm x 40cm field sizes. Integrates with MLCs (58, 82, 160-leaf), aSi EPID, and third-party gating/fixation devices. Connectivity via syngo®-based RTT/COHERENCE workspaces and PRIMEVIEW™ record and verify systems. Software architecture is syngo®-based.
Indications for Use
Indicated for patients requiring radiation therapy for lesions, tumors, and conditions anywhere in the body. Supports 3D conformal radiation therapy, IMRT, stereotactic radiosurgery (SRS), and stereotactic radiation therapy (SRT).
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
#### Siemens Medical Solutions USA, Inc. Oncology Care Systems
KOS2225
FEB 1 9 2009
# Section 5 510(k) Summary
Submitter: Siemens Medical Solutions USA, Inc. Oncology Care Systems 4040 Nelson Avenue Concord, CA 94520 Contact: Christine Dunbar Senior Regulatory Affairs Specialist Phone: (925)602-8139 Fax: (925)602-8008 Email: christine.dunbar.ext@siemens.com Proprietary Name: PreScision™ Option Common Name: An accessory to: Accelerator, Linear, Medical Classification: 892.5050 Product Code: IYE
Substantial Equivalence Claimed To:
| PRODUCT | Clearance | Claim of Equivalence For: |
|------------------------------|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SIEMENS ARTISTE™ | K072485 | The Stereo mode with the 160 MLC |
| SIEMENS ONCOR™<br>Expression | K060226 | ONCOR™ and PRIMUS™ linear accelerator families<br>with Stereo Mode and COHERENCE Workspaces (or the<br>re-branded syngo® Suite for Oncology Workspaces)<br>and the 82 leaf MLC marketed as OPTIFOCUS. |
| SIEMENS PRIMUSTM | K993425 | PRIMUS™ family of linear accelerators with the Stereo<br>Mode, PRIMEVIEW 3i system, the 58 leaf MLC |
| Varian Trilogy™ | K061140 | Stereotactic (SRT) and Stereotactic Radio-Surgery<br>(SRS) Radiation Therapy with high dose beam rate. |
| TomoTherapy Hi-Art™ | K060912 | Stereotactic (SRT) and Stereotactic Radio-Surgery<br>(SRS) with unflattened beam delivery and extended<br>treatment field for Stereotactic Body Radiation Therapy. |
| Accuray Cyberknife™ | K052325 | Stereotactic (SRT) and Stereotactic Radio-Surgery<br>(SRS) robotic delivery system and increased dose rates<br>for reduced treatment times. |
The ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerators with the PreScision™ Option as described in this premarket notification has similar intended use and fundamental scientific technical characteristics as the devices listed above.
Description Summary:
# 000026
510(k) for PreScision™ Option
CONFIDENTIAL Page Sect. 5/1
{1}------------------------------------------------
## Siemens Medical Solutions USA, Inc. Oncology Care Systems
| Market Name | Internal naming convention |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ARTISTE™ | ARTISTE™ linear accelerator and ACCEL release 4+ |
| ONCOR™ Expression | ONCOR™ linear accelerator and ACCEL release 2+ |
| ONCOR™ Avant-Garde | ONCOR™ linear accelerator and ACCEL release 2+ |
| PRIMUS™ | PRIMUS™ linear accelerator and ACCEL release 2+ |
| PRIMEVIEW™ | Siemens proprietary verify and record system. The syngo® based<br>PRIMEVIEW is hosted on the COHERENCE™ Therapist<br>Workspace.<br>The syngo base version is marketed as PRIMEVIEW3i and is used<br>on the PRIMUS™ linear accelerator systems. |
| COHERENCE™ Therapist Workspace | RTT Workspace contains the SIEMENS proprietary verify and<br>record system as well as access to the Oncology Information<br>System and directly connects to the LINAC control console. |
| MVCB | Mega-Voltage Cone Beam - a method of obtaining 3 dimensional<br>data for portal imaging. |
| 160 MLC™ | 160-leaf multi-leaf collimator |
| OPTIFOCUS™ | 82-leaf multi-leaf collimator |
| OPTIVUE™ | aSi flat panel electronic portal imaging device (EPID) AL7 model |
| OPTIVUE 1000ST | aSi flat panel electronic portal imaging device (EPID) AG9 model |
| SRT / SRS | Stereotactic Radiation Therapy/ Stereotactic radiosurgery ~ a<br>method of using high energy, X-Ray photon radiation for precision<br>treatment. |
| syngo® Therapist Workspace, RTT<br>Express™ | RTT Workspace contains the SIEMENS proprietary verify and<br>record system as well as access to the Oncology Information<br>System and directly connects to the LINAC control console on the<br>ARTISTE™ linear accelerator system. |
| syngo® Suite for Oncology<br>Workspaces | Syngo based workstation, re-branded COHERENCE workspaces. |
| syngo® | Siemens proprietary software architecture and hosting SIEMENS<br>software applications organized by task cards on a dedicated<br>workstation. |
Within the submission the following internal naming conventions are used:
For further definitions of the terms used in this submission, refer to the Glossary in Section 24.
# Technological Characteristics:
# The PreScision™ Option:
The PreScision™ Option package is an optional feature to the existing SIEMENS ARTISTE™,
ONCOR™ and PRIMUS™ family of medical linear accelerator devices [LINAC]. The basic design, safety features and function of the LINAC remain unchanged. The PreScision features upports Stereotation Therapy (SRT) and Stereotactic Radio-surgery (SRT) using the conventional linear accelerator and subsystems for the delivery of precision high dose X-Ray photon energy for treatment of lesions, turnors and conditions anywhere in the body where radiation therapy is indicated.
The PreScision option supports the delivery of up to 2,000 (± 2%) MU / Minute X-Ray photon beam for either single-session radiosurgery or hypo-fractionated stereotactic treatment using a calibrated 6.67 MV photon energy spectrum consisting of an unflattened beam geometry, a selectable field size
{2}------------------------------------------------
of 5cm x 5 cm up to 40cm x 40 cm, collimated by any validated third party stereotactic accessory such as a stereotactic cone or a Multi-leaf collimator, with or without the use of patient motion detection or physiologic gating.
The control console will support the PreScision feature with one unflattened beam energy labeled as "7UF" to differentiate from the previous 6 MV unflattened beam energy used by the predicate Stereo feature. The 7UF option replaces the existing "Stereo" option previously cleared under the PRIMUS™ (K993425) and ONCOR™ (K031764, K060226) and ARTISTE (K072585) devices.
The unflattened X-Ray photon energy spectrum is similar to the 6 MV energy spectrum as used by the PRIMUS™, ONCOR™, ARTISTE Linear Accelerator systems for the predicate Stereo option and is similar to the energy spectrum of the TomoTherapy Hi-Art system as described in Section 11, Design Description.
The SIEMENS linear accelerator product requirements for the PreScision™ option consists of the following:
A SIEMENS ARTISTE™, ONCOR™ or PRIMUS™ Family of Linear Accelerators with the minimum configuration of:
- COHERENCE™ Therapist Workspace or the syngo® Radiation Therapist (RTT) with . embedded PRIMEVIEW™ Record and Verify system or the RT Express™ system for the ONCOR™or ARTISTE™ linear accelerator systems.
- PRIMEVEIW 3i Record and Verify system used with the PRIMUS™ linear accelerator . system.
- 58 leaf multi-leaf collimator (MLC) marketed as 3D-MLC, 82 leaf MLC marketed as . OPTIFOCUS™ or the 160 Leaf MLC for:
- Stereotactic Radiotherapy 0
- Stereotactic Radiosurgery o
- Fixed fields, 0
- o Auto-sequenced,
- 0 Arc (Rotation) and
- Intensity Modulated Radiation Therapy (IMRT) treatment delivery methods, 0
- . MLC interface to support the SIEMENS ModuLeaf MLC (K030609) for precision Stereotactic or Stereotactic Radiosurgery Radiation TherapyTreatments.
- . Gating interface to support Third party, cleared, gating devices for physiologic or patient motion detection.
- . Stereotactic mode to support Third party, cleared, stereotactic hardware and patient fixation positioning devices.
- An amorphous Silicon (aSi) flat panel electronic portal imaging device (EPID) marketed as . OPTIVUE 1000ST
- patient treatment couch; 550 TxT™ (K050422), or the ZXT™ (K910971). �
Refer to Section 11 Design Description, for the Product Specification regarding these specific product features.
#### General Safety and Effectiveness:
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software development, verification of
{3}------------------------------------------------
requirements and validation testing. To minimize electrical and radiation hazards. SIEMENS adheres to recognized and established industry practice and relevant international standards.
#### Intended Use:
The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
The LINAC is a high-dose and high-dose rate medical linear accelerator optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT) and when used in conjunction with the PreScision™ option, supports precision stereotactic applications. The stereotactic applications include single-session radiosurgery, fractionated stereotactic radiation therapy, fractionated stereotactic intensity modulated radiation therapy for lesions, turnors and conditions anywhere in the body where radiation therapy is indicated.
#### The syngo@ Suite for Oncology Workspaces:
The syngo@workspaces includes a number of syngo@based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording.
#### Summary:
In summary, it is SIEMENS' belief that the addition of the PreScision option for enhanced stereotactic radiation therapy and stereotactic radiosurgery (SRT / SRS ) does not introduce any new potential safety risks and is substantially equivalent to, and performs as well as, the predicate devices.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font. The words are arranged horizontally, with "Public" and "Health" on the first line and "Service" on the second line.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 1 9 2009
Ms. Christine Dunbar Senior Regulatory Affairs Specialist Siemens Medical Solutions, USA, Inc. Oncology Care Systems 4040 Nelson Avenue CONCORD CA 94520
Re: K082775 .
> Trade/Device Name: PreScision™ Option Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 19, 2008 Received: September 22, 2008
#### Dear Ms. Dunbar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our laceling regulation (21) . CFR Part 801) ), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.
Sincerely yours,
Lain B. Morris
anine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{6}------------------------------------------------
#### Siemens Medical Solutions USA, Inc. Oncology Care Systems
# Section 4
# Indication For Use Statement
510(k) Number (if known): _ K082775
#### Device Name: PreScision™ Option
#### Indications for Use:
The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.
The LINAC.is a high-dose and high-dose rate medical linear accelerator optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy-{IMRT}, and when used in conjunction with the PreScision™ option, supports precision stereotactic applications. The stereotactic applications include single-session radiosurgery, fractionated stereotactic radiation therapy, fractionated stereotactic intensity modulated radiation therapy for lesions, turnors and conditions anywhere in the body where radiation therapy is indicated.
# The syngo Suite for Oncology Workspaces:
The syngo@workspaces includes a number of syngo@based software applications whose indication for use include the viewing, processing, filming, and archiving of medical images. The workspaces also permit patient data management, patient selection/setup, patient positioning verification, treatment planning, treatment delivery/verification, and treatment recording.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | (Division Sign-Off) |
| | Division of Reproductive, Abdominal and |
| | Radiological Devices |
| 510(k) Number | K082775 |
| Prescription Use<br>(Per 21 CFR 801.109) | <div style="display:inline-block;"><img alt="[X]" src="data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAUAAAAFCAYAAACNbyblAAAAHElEQVQI12P4//8/w+bAwNDg4OB/YGBgYGBgYABWMAhYAwAAAABJRU5ErkJggg=="/></div> OR Over-the-Counter Use ______ |
510(k) for PreScision™ Option
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.