K082499 · Techlab, Inc. · LLH · Mar 26, 2009 · Microbiology
Device Facts
Record ID
K082499
Device Name
C. DIFF QUIK CHEK COMPLETE
Applicant
Techlab, Inc.
Product Code
LLH · Microbiology
Decision Date
Mar 26, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2660
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The C. DIFF QUIK CHEK COMPLETE™ test is a rapid membrane enzyme immunoassay for the simultaneous detection of C. difficile glutamate dehydrogenase antigen and toxins A and B in a single reaction well. The test detects C. difficile antigen, glutamate dehydrogenase, as a screen for the presence of C. difficile and confirms the presence the toxigenic C. difficile by detecting toxins A and B in fecal specimens from persons suspected of having C. difficile disease. The test is to be used as an aid in the diagnosis of C. difficile disease. As with other C. difficile tests, results should be considered in conjunction with the patient history.
Device Story
Rapid membrane enzyme immunoassay for simultaneous detection of C. difficile glutamate dehydrogenase (GDH) antigen and toxins A and B in fecal specimens. Device utilizes antibodies specific for GDH and toxins A/B immobilized on a reaction window membrane. Fecal specimen is diluted and mixed with antibody-horseradish peroxidase conjugate; mixture is added to sample well. During 15-minute incubation, GDH/toxin in sample binds to conjugate; complexes migrate to membrane and are captured by immobilized antibodies. After wash, substrate is added; 10-minute incubation produces blue lines indicating positive results. Visual examination by healthcare provider. Positive antigen line screens for C. difficile; positive toxin line confirms toxigenic C. difficile. Results aid clinical diagnosis in conjunction with patient history.
Clinical Evidence
Clinical study included 1126 fecal samples from patients aged 5 months to 100 years. Antigen detection compared to bacterial culture showed 90.5% sensitivity and 93.1% specificity. Toxin A/B detection compared to tissue culture assay showed 87.8% sensitivity and 99.4% specificity. Reproducibility studies across 3 sites over 5 days showed consistent performance.
Technological Characteristics
Rapid membrane enzyme immunoassay; utilizes immobilized antibodies against C. difficile GDH and toxins A/B; horseradish peroxidase-conjugated antibodies; visual readout via blue line formation; manual test format; no instrumentation required.
Indications for Use
Indicated for use in persons suspected of having C. difficile disease, including patients of all ages (5 months to 100 years), to detect C. difficile glutamate dehydrogenase antigen and toxins A and B in fecal specimens as an aid in diagnosis.
Regulatory Classification
Identification
A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K082499
B. Purpose for Submission:
Substantial equivalence determination for a new toxin A and B and antigen combined as one test system
C. Measurand:
Clostridium difficile (C. diff) toxins A and B and the antigen glutamate dehydrogenase (GDH)
D. Type of Test:
A qualitative rapid immunoassay
E. Applicant:
Techlab, Inc.
F. Proprietary and Established Names:
C. DIFF QUIK CHEK COMPLETE™
G. Regulatory Information:
1. Regulation section:
866.2660
2. Classification:
Class I
3. Product code:
LLH – Reagents, Clostridium difficile toxin
{1}
MCB- Antigen, Clostridium difficile
4. Panel:
83- Microbiology
H. Intended Use:
1. Intended use(s):
The C. DIFF QUIK CHEK COMPLETE™ test is a rapid membrane enzyme immunoassay for the simultaneous detection of C. difficile glutamate dehydrogenase antigen and toxins A and B in a single reaction well. The test detects C. difficile antigen, glutamate dehydrogenase, as a screen for the presence of C. difficile and confirms the presence of toxigenic C. difficile by detecting toxins A and B in fecal specimens from persons suspected of having C. difficile disease. The test is to be used as an aid in the diagnosis of C. difficile disease. As with other C. difficile tests, results should be considered in conjunction with the patient history.
2. Indication(s) for use:
The C. DIFF QUIK CHEK COMPLETE™ test is a rapid membrane enzyme immunoassay for the simultaneous detection of C. difficile glutamate dehydrogenase antigen and toxins A and B in a single reaction well. The test detects C. difficile antigen, glutamate dehydrogenase, as a screen for the presence of C. difficile and confirms the presence of toxigenic C. difficile by detecting toxins A and B in fecal specimens from persons suspected of having C. difficile disease. The test is to be used as an aid in the diagnosis of C. difficile disease. As with other C. difficile tests, results should be considered in conjunction with the patient history.
3. Special condition for use statement(s):
For prescription use only
4. Special instrument requirements:
Not applicable.
I. Device Description:
The C. DIFF QUIK CHEK COMPLETE™ test uses antibodies specific for glutamate dehydrogenase and toxins A and B of C. difficile. The device contains a Reaction Window with three vertical lines of immobilized antibodies. The antigen test line ("Ag") contains antibodies against C. difficile glutamate dehydrogenase. The control line ("C") is a dotted line that contains anti-horseradish peroxidase (HRP) antibodies. The toxin A and B test line ("Tox") contains antibodies against C. difficile toxins A and B. The
{2}
Conjugate consists of antibodies to glutamate dehydrogenase and antibodies to toxins A and B coupled to horseradish peroxidase.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
There are four predicate device referenced to be substantially equivalent to the C. DIFF QUIK CHEK COMPLETE™.
- C. DIFF QUIK CHEK
- TOX A/B QUIK CHEK
- C. DIFF CHEK-60
- C. DIFFICILE TOX A/B II
2. Predicate 510(k) number(s):
- K053572
- K050891
- K030992
- K003306
3. Comparison with predicate:
| Similarities | | | | | |
| --- | --- | --- | --- | --- | --- |
| Item | C. DIFF QUIK CHEK COMPLETE™ | C. DIFF QUIK CHEK® | TOX A/B QUIK CHEK® | C. DIFF CHEK™ - 60 | C. DIFFICILE TOX A/B II™ |
| Intended Use | Detection of C. difficile toxins A and B and GDH | Detection of C. difficile GDH | Detection of C. difficile toxins A and B | Detection of C. difficile GDH | Detection of C. difficile toxins A and B |
| Specimen Types | Human Fecal Specimens | Human Fecal Specimens | Human Fecal Specimens | Human Fecal Specimens | Human Fecal Specimens |
| Technologies | Enzyme Immunoassay - Rapid Membrane | Enzyme Immunoassay - Rapid Membrane | Enzyme Immunoassay - Rapid Membrane | Enzyme Immunoassay - Microassay Plate | Enzyme Immunoassay - Microassay Plate |
{3}
| Differences | | | | | |
| --- | --- | --- | --- | --- | --- |
| Item | C. DIFF QUIK CHEK COMPLETE™ | C. DIFF QUIK CHEK® | TOX A/B QUIK CHEK® | C. DIFF CHEK™ - 60 | C. DIFFICILE TOX A/B II™ |
| Limit of Detection (Analytical Sensitivity) | Toxin A ≥ 0.63 ng/mL
Toxin B ≥ 0.16 ng/mL
GDH ≥ 0.8 ng/mL | GDH ≥ 0.8 ng/mL | Toxin A ≥ 0.63 ng/mL
Toxin B ≥ 1.25 ng/mL | GDH ≥ 0.8 ng/mL | Toxin A ≥ 0.8 ng/mL
Toxin B ≥ 2.5 ng/mL |
| Clinical Sensitivity | Toxins A&B: 87.8%
GDH: 90.5% | GDH: 92.8% | Toxins A&B: 90.2% | GDH: 70.5% | Toxins A&B: 92.2% |
| Clinical Specificity | Toxins A&B: 99.4%
GDH: 93.1% | GDH: 92.6% | Toxins A&B: 99.7% | GDH: 91.2% | Toxins A&B: 100% |
## K. Standard/Guidance Document Referenced (if applicable):
This 510(k) Premarket Notification was prepared and referenced the following guidance document and recognized standard:
- “Review Criteria for Devices assisting in the diagnosis of C. difficile associated disease” May 1990, ODE/CDRH Guidance Document.
- User Protocol for Evaluation of Qualitative Test Performance, Clinical Laboratory Standards Institute (CLSI) Approved Guideline, EP12-A (August 2002)
## L. Test Principle:
The device contains a Reaction Window with two vertical lines of immobilized antibodies and center dots of immobilized antibody. One line consists of antibodies against C. difficile glutamate dehydrogenase and the second line contains antibodies against C. difficile toxins A and B. After the incubation period, "Ag" reaction is examined visually for the appearance of a vertical blue line on the "Ag" side of the Reaction Window. A blue line indicates a positive test. If the "Ag" is positive, then the "Tox" reaction should be examined visually for the appearance of a blue line on the "Tox" side of the Reaction Window. A blue line indicates a positive test. A positive "C" reaction, indicated by a vertical dotted blue line under the "C" portion of the Reaction Window, confirms that the test is working properly and the results are valid.
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
{4}
# a. Precision/Reproducibility:
The reproducibility of the C. DIFF QUIK CHEK COMPLETE™ test was determined using 12 fecal specimens that were coded to prevent their identification during testing. Testing was performed at 3 independent laboratories, which tested the samples for 3 days. The samples produced the expected results $100\%$ of the time.
An additional 5-day study was performed at 3 sites by running a set of low positive, moderate positive and high negative fecal samples that were spiked with glutamate dehydrogenase, Toxin A and Toxin B. The samples were run in triplicate, twice a day over a 5-day period by multiple technicians at each site. The combined antigen and toxin data from the 5-day reproducibility study is shown in Table 7. The antigen data for Sample 1 was below $90\%$ negative at one site and was consistently negative for antigen at the other two sites. The toxin data for Sample A was below $90\%$ positive at one site, and was consistently positive for toxin at the other two sites. No single site reported results below expectations for both antigen and toxin.
Table 7 Summary of 5-day reproducibility study
| Sample ID | Antigen Positive | Antigen Negative | Toxin Positive | Toxin Negative |
| --- | --- | --- | --- | --- |
| Sample 1 (95% negative results expected) | 15 (16.7%) | 75 (83.3%) | 0 (0%) | 90 (100%) |
| Sample A (95% positive results expected) | 89 (98.9%) | 1 (1.1%) | 65 (72.2%) | 25 (27.8%) |
| Sample B (mod positive) | 87 (96.7%) | 3 (3.3%) | 86 (95.5%) | 4 (4.5%) |
The Reproducibility study was satisfactory for this type of assay.
# b. Linearity/assay reportable range:
Not applicable.
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable.
# d. Detection limit:
The cutoff for GDH was determined using highly purified recombinant C. difficile glutamate dehydrogenase (rGDH). The cutoff was determined by testing dilutions of purified rGDH in diluent, in replicates of 20. The Limit of Detection (LOD) was established at $0.20\mathrm{ng / mL}$ . The LOD for Toxin A and B GDH was determined using highly purified C. difficile Toxin A and B. The LOD for each toxin was determined by testing dilutions of purified Toxin A and B in diluent, in replicates of 20. The LOD was established at $0.45\mathrm{ng / mL}$ for Toxin A and $0.06\mathrm{ng / mL}$ for Toxin B.
{5}
The methodologies above supports the product claims for the cutoff for Toxin A at 0.63 ng/mL, 0.16 ng/mL for Toxin B and 0.8 ng/mL for GDH.
The Limit of Detection and Cutoff studies were satisfactory for this type of assay.
e. Analytical specificity: CROSS REACTIVITY
Fecal specimens inoculated with the following microorganisms to a final concentration of approximately 10⁸ or higher organisms per mL did not react in the antigen or toxin portion of the C. DIFF QUIK CHEK COMPLETE™ test:
**Bacterium or Pathogen:** Aeromonas hydrophila, Bacillus cereus, Bacillus subtilis, Bacteroides fragilis, Campylobacter coli, Campylobacter fetus, Campylobacter jejuni, Candida albicans, Clostridium butyricum, Clostridium clostridiforme, Clostridium haemolyticum, Clostridium histolyticum, Clostridium novyi, Clostridium perfringens, Clostridium septicum, Clostridium sordellii (nontoxigenic), Clostridium sporogenes, Enterobacter aerogenes, Enterobacter cloacae, Enterococcus faecalis, Escherichia coli EIEC, Escherichia coli, Escherichia coli O157:H7, Escherichia coli ETEC, Klebsiella pneumoniae, Peptostreptococcus anaerobius, Proteus vulgaris, Pseudomonas aeruginosa, Salmonella typhimurium, Serratia liquifaciens, Shigella dysenteriae, Shigella flexneri, Shigella sonnei, Staphylococcus aureus, Staphylococcus aureus (Cowans), Staphylococcus epidermidis, Vibrio cholerae, Vibrio parahaemolyticus, Yersinia enterocolitica
The non-C. difficile organism, Clostridium bifermentans, reacted in the antigen portion of the C. DIFF QUIK CHEK COMPLETE™ test. The non-C. difficile organism, Clostridium sordellii, reacted in the toxin portion of the C. DIFF QUIK CHEK COMPLETE™ test was VPI 9048. This strain produces toxins HT and LT, which are homologous to toxins A and B, respectively. Appropriate recommendations will be made in the product insert, Limitations of the Procedure section to reflect this cross-reactivity.
The following viruses of 10³.³ to 10⁸.²⁵ TCID units per 0.2 mL did not react in the C. DIFF QUIK CHEK COMPLETE™ test:
**Viruses:** Adenovirus types 1, 2, 3, 5, 40, 41, Human coronavirus, Coxsackievirus B2, B3, B4, B5, Echovirus 9, 11, 18, 22, 33, Enterovirus type 68, 69, 70, 71, Rotavirus.
INTERFERING SUBSTANCES
The following substances had no effect on test results when present in feces in the concentrations indicated: mucin (3.5% w/v), human blood (40% v/v), barium sulfate (5% w/v), Imodium® (5% v/v), Kaopectate® (5% v/v), Pepto-Bismol® (5% v/v), steric/palmitic acid (40% w/v), Metronidazole (0.25% w/v), Vancomycin (0.25% w/v).
{6}
f. Assay cut-off:
The assay cutoff was determined to be at a concentration of $0.63\mathrm{ng / mL}$ for toxin A, $0.16\mathrm{ng / mL}$ for toxin B, and $0.8\mathrm{ng / mL}$ for glutamate dehydrogenase. This fulfils the requirements of demonstrating a cutoff for the C. DIFF QUIK CHEK COMPLETE™.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Clinical studies were conducted at three external sites in Pennsylvania, Ohio and Virginia using the C. DIFF QUIK CHEK COMPLETE™. Clinical truth was determined at Techlab, Inc. by testing samples using the tissue culture assay for analysis of the toxins A and B portion of the test and bacterial culture for analysis of the GDH antigen portion of the test. A total of 1126 fecal samples were collected and tested in the study. The study subjects ranged in age from 5 months-100 years, and $60.1\%$ were female.
The antigen portion of the C. DIFF QUIK CHEK COMPLETE™ test was compared to bacterial culture. Specimens included in the evaluation were submitted to the clinical laboratories for routine testing. The bacterial culture test was performed according to in-house procedures. The results are shown in Table 1.
Table 1. Summary of clinical performance comparing C. DIFF QUIK CHEK COMPLETE ${}^{\mathrm{{TM}}}$ test to bacterial culture
| N = 1126 | Bacterial Culture Positive | Bacterial Culture Negative |
| --- | --- | --- |
| C. DIFF QUIK CHEK COMPLETE™ Antigen Line Positive | 201 | 62 |
| C. DIFF QUIK CHEK COMPLETE™ Antigen Line Negative | 21 | 842 |
| | | 95% Confidence Limits |
| --- | --- | --- |
| Sensitivity | 90.5% | 85.7 – 93.9 |
| Specificity | 93.1% | 91.2 – 94.7 |
| Predictive Positive Value | 76.4% | 70.7 – 81.3 |
| Predictive Negative Value | 97.6% | 96.2 – 98.4 |
| Correlation | 92.6% | 91.8 – 93.4 |
{7}
The antigen portion of the C. DIFF QUIK CHEK COMPLETE™ test was compared to the tissue culture assay for the detection of C. difficile toxin. Specimens included in the evaluation were submitted to the clinical laboratories for routine testing. The results are shown in Table 2. The antigen portion of the C. DIFF QUIK CHEK COMPLETE™ test detected 98.7% of the tissue culture-positive samples.
Table 2. Summary of clinical performance comparing C. DIFF QUIK CHEK COMPLETE™ test to the tissue culture assay
| N = 1126 | Tissue Culture Positive | Tissue Culture Negative |
| --- | --- | --- |
| C. DIFF QUIK CHEK COMPLETE™
Antigen Line Positive | 154 | 109 |
| C. DIFF QUIK CHEK COMPLETE™
Antigen Line Negative | 2 | 861 |
Clinical evaluation of the toxin portion of the C. DIFF QUIK CHEK COMPLETE™ test
The toxin portion of the C. DIFF QUIK CHEK COMPLETE™ test was compared to the tissue culture assay at two clinical laboratories and in-house at TECHLAB®, Inc. Specimens included in the evaluation were submitted to the clinical laboratories for routine testing. The results are shown in Table 3.
Table 3. Summary of clinical performance comparing C. DIFF QUIK CHEK COMPLETE™ test to the tissue culture assay
| n = 1126 | Tissue Culture Positive | Tissue Culture Negative |
| --- | --- | --- |
| C. DIFF QUIK CHEK COMPLETE™
Toxin Line Positive | 137 | 6 |
| C. DIFF QUIK CHEK COMPLETE™
Toxin Line Negative | 19 | 964 |
| | | 95% Confidence Limits |
| --- | --- | --- |
| Sensitivity | 87.8% | 81.4 - 92.3 |
| Specificity | 99.4% | 98.6 - 99.7 |
| Predictive Positive Value | 95.8% | 90.7 - 98.3 |
| Predictive Negative Value | 98.1% | 96.9 - 98.8 |
| Correlation | 97.8% | 97.6 - 98.0 |
b. Clinical specificity:
See section M.3.a
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
{8}
The cutoff for the assay was established at concentrations of 0.63 ng/mL for toxin A, 0.16 ng/mL for toxin B, and 0.8 ng/mL for glutamate dehydrogenase.
5. Expected values/Reference range:
The reported incidence of *C. difficile* disease in patients with antibiotic-associated diarrhea may range from 5 to 20%, and hospitals may experience rates lower or higher than this range. It is important to consider any test results in conjunction with clinical symptoms because some healthy adults and large numbers of healthy infants (up to 50%) will be positive for *C. difficile* toxin. In addition, *C. difficile* carriage rates of 22% to 32% have been reported in cystic fibrosis patients (1,3). In the studies conducted for this device, using symptomatic patients, the incidence of toxins A and B was 12% and GDH was 18%. A positive result in the antigen portion of the *C. DIFF QUIK CHEK COMPLETE™* test confirms the presence of *C. difficile* in a fecal specimen; a negative result indicates the absence of the organism. A positive result in the toxin portion of the *C. DIFF QUIK CHEK COMPLETE™* confirms the presence of *C. difficile* toxin in a fecal specimen; a negative result indicates the absence of toxin or insufficient levels of toxin for detection.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The information submitted in this premarket notification is complete and supports a substantial equivalent decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.