The TOX A/B QUIK CHEK™ test is a rapid immunoassay for detecting Clostridium difficile toxins A and B in fecal specimens from persons suspected of having C. difficile disease. The test is to be used as an aid in the diagnosis of C. difficile disease and results should be considered in conjunction with the patient history.
Device Story
TOX A/B QUIK CHEK™ is a rapid membrane-based immunoassay for qualitative detection of C. difficile toxins A and B in fecal specimens. Device utilizes antibodies specific to toxins A and B immobilized on a reaction membrane. Fecal sample is diluted and mixed with horseradish peroxidase-conjugated anti-toxin antibodies; mixture is added to sample well. During 15-minute incubation, toxin-antibody-conjugate complexes form and migrate to the membrane, where they are captured by immobilized antibodies. After washing and substrate addition, presence of blue line at 'T' position indicates positive result; blue line at 'C' position confirms proper reagent activity and migration. Performed by laboratory personnel. Results interpreted visually by clinician to aid in diagnosis of C. difficile disease in conjunction with patient history. Benefits include rapid identification of toxigenic C. difficile, facilitating timely clinical management.
Clinical Evidence
Clinical performance evaluated across 5 studies (n=842) comparing TOX A/B QUIK CHEK™ to tissue culture assay. Results: 90.2% sensitivity (95% CI: 84.1-94.2%), 99.7% specificity (95% CI: 98.8-99.9%), 98.6% PPV, 97.9% NPV, and 98.0% correlation. Reproducibility confirmed at 3 laboratories over 3 days with 100% agreement on known positive and negative samples.
Technological Characteristics
Rapid membrane immunoassay. Components: immobilized anti-toxin A/B antibodies (test line), anti-IgG antibodies (control line), and horseradish peroxidase-conjugated anti-toxin A/B antibodies. Visual readout via blue line appearance. Analytical sensitivity: 0.63 ng/mL (toxin A), 1.25 ng/mL (toxin B). No software or electronic components.
Indications for Use
Indicated for detection of C. difficile toxins A and B in fecal specimens from patients suspected of having C. difficile disease; aid in diagnosis; results to be used with patient history.
Regulatory Classification
Identification
A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
K050891
B. Purpose for Submission:
Substantial equivalence determination
C. Measurand:
Clostridium difficile toxins A and B
D. Type of Test:
A qualitative rapid immunoassay
E. Applicant:
TechLab Inc.
F. Proprietary and Established Names:
TOX A/B Quik Chek
G. Regulatory Information:
1. Regulation section:
21 CFR Part 866.2660 Microorganism Differentiation and Identification
2. Classification:
I
3. Product code:
LLH – Reagents, Clostridium difficile toxin
4. Panel:
Microbiology (83)
{1}
H. Intended Use:
1. Intended use:
The TOX A/B QUIK CHEK™ test is a rapid immunoassay for detecting Clostridium difficile toxins A and B in fecal specimens from persons suspected of having C. difficile disease. The test is to be used as an aid in the diagnosis of C. difficile disease and results should be considered in conjunction with the patient history.
2. Indication for use:
The TOX A/B QUIK CHEK™ test is a rapid immunoassay for detecting Clostridium difficile toxins A and B in fecal specimens from persons suspected of having C. difficile disease. The test is to be used as an aid in the diagnosis of C. difficile disease and results should be considered in conjunction with the patient history.
3. Special conditions for use statement:
For prescription use only
4. Special instrument requirements:
Not applicable
I. Device Description:
The TOX A/B QUIK CHEK™ uses antibodies specific for toxins A and B of C. difficile. The device contains a Reaction Window with two vertical lines of immobilized antibodies. The test line (“T”) contains antibodies against C. difficile toxins A and B. The control line (“C”) contains anti-IgG antibodies. The Conjugate consists of antibodies to toxins A and B coupled to horseradish peroxidase. The kit contains:
Membrane Devices –25 pouches, each containing 1 device and a desiccant pack
Diluent (14 mL) – Buffered protein solution containing 0.02% thimerosal
Wash Buffer (10 mL) – A buffered solution containing 0.02% thimerosal
Substrate (3.5 mL) – Solution containing tetramethylbenzidine
Conjugate (2 mL) – Mouse monoclonal antibody specific for toxin A coupled to horseradish peroxidase and goat polyclonal antibody specific for toxin B coupled to horseradish peroxidase in a buffered protein solution containing 0.02% thimerosal
Positive Control (1 mL) – Antigen in a buffered protein solution
{2}
3
J. Substantial Equivalence Information:
1. Predicate device names:
Premier toxins A & B, ImmunoCard Toxins A & B, C. difficile TOX A/B II, ProSpecT C. difficile toxin A/B, X/pect C. difficile toxin A/B
2. Predicate 510(k) numbers:
K993914, K041003, K00306 and K030404, K033479 and K041951
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Detection of C.difficile toxins in fecal specimens | Same |
| Specimen type | Human stool | Same |
| Technology | Enzyme immunoassay | Same |
| | | |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Limit of detection | 0.63 ng/mL for toxin A and 1.25 ng/ml for toxin B | ≥ 1.4 ng/ml of toxin A & ≥ 2.4 ng/ml of toxin B |
| Clinical sensitivity | 90.2 % (84.1 - 94.2% C.I.) | 94.7% (88.1 – 98.3% C.I.) |
| Clinical specificity | 99.7 % (98.8 – 99.9% C.I.) | 97.3% (95.4 – 98.5% C.I.) |
K. Standard/Guidance Document Referenced (if applicable):
“Review Criteria for Devices assisting in the diagnosis of C. difficile associated disease” May 1990, ODE/CDRH Guidance Document
L. Test Principle:
The device contains a Reaction Window with two vertical lines of immobilized antibodies (Fig. 1a). The test line (“T”) contains antibodies against C. difficile toxins A and B. The control line (“C”) contains anti-IgG antibodies. The Conjugate consists of antibodies to toxins A and B coupled to horseradish peroxidase. To perform the test, the
{3}
fecal specimen is diluted with Diluent and Conjugate is added to the diluted sample. The diluted sample-conjugate mixture is added to the Sample Well and the device is allowed to incubate at room temperature for 15 minutes. During the incubation, any toxin A and toxin B in the sample bind to anti-toxin antibody-peroxidase conjugate. The toxin-antibody complexes migrate through a filter pad to a membrane where they are captured by the immobilized anti-toxin antibodies in the line. The Reaction Window is subsequently washed with Wash Buffer, and the test is developed with the addition of Substrate. After a 10 minute incubation period, the "T" reaction is examined visually for the appearance of a vertical blue line on the "T" side of the Reaction Window. A blue line indicates a positive test. A positive "C" reaction, indicated by a vertical blue line on the "C" side of the Reaction Window, confirms that the test is working properly and the results are valid.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The reproducibility of the TOX A/B QUIK CHEK™ test was determined using known positive (n=6) and negative (n=2) fecal specimens that were coded and masked. Testing was performed at 3 laboratories, which tested the samples on 3 days. The samples produced the expected results 100% of the time.
#### b. Linearity/assay reportable range:
N/A
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
N/A
#### d. Detection limit:
The analytical sensitivity was determined using serial two-fold dilutions of highly purified toxins A and B. Toxins were purified using standard in-house procedures. Six separate tests were conducted for each toxin and the test was consistently positive at a concentration of 0.63 ng/ml for toxin A and 1.25 ng/ml for toxin B.
#### e. Analytical specificity:
CROSS REACTIVITY
{4}
Fecal specimens inoculated with the following microorganisms to a final concentration of approximately $10^{8}$ or higher organisms per mL did not react in the TOX A/B QUIK CHEK™:
**Bacteria**: Aeromonas hydrophila, Bacillus cereus, Bacillus subtilis, Bacteroides fragilis, Campylobacter coli, Campylobacter fetus, Campylobacter jejuni, Candida albicans, Clostridium bifermentans, Clostridium butyricum, Clostridium perfringens, Clostridium septicum, Clostridium sordellii (nontoxigenic), Clostridium sporogenes, Enterococcus faecalis, Escherichia coli EIEC, Escherichia coli, Escherichia coli 0157 H7, Escherichia coli ETEC, Klebsiella pneumoniae, Peptostreptococcus anaerobius, Proteus vulgaris, Pseudomonas aeruginosa, Salmonella typhimurium, Shigella dysenteriae, Shigella flexneri, Shigella sonnei, Staphylococcus aureus, Staphylococcus aureus (Cowans), Staphylococcus epidermidis, Vibrio parahaemolyticus, Yersinia enterocolitica
**Viruses**: Adenovirus types 1,2,3,5,40,41, Human coronavirus, Coxsackievirus B2,B3,B4,B5, Echovirus 9,11,18,22,33, Enterovirus type 68,69,70,71.
The only non-C. difficile organism to react with the TOX A/B QUIK CHEK™ was C. sordellii VPI 9048. This strain produces toxins HT and LT, which are homologous to toxins A and B, respectively.
## INTERFERING SUBSTANCES
The following substances had no effect on test results when present in feces in the concentrations indicated: mucin (3.5% w/v), human blood (40% v/v), barium sulfate (5% w/v), Imodium® (5% w/v), Kaopectate® (5 mg/mL), Pepto-Bismol® (5% w/v), steric/palmitic acid (40% w/v), Metronidazole (0.25% w/v), Vancomycin (0.25% w/v).
f. Assay cut-off:
The assay was determined to detect Toxin A at 0.63 ng/ml and Toxin B at 1.25 ng/ml
2. Comparison studies:
a. Method comparison with predicate device:
N/A
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity:
{5}
The TOX A/B QUIK CHEK™ test was compared with the tissue culture test at three U.S. hospitals and in-house at TECHLAB®, Inc. Specimens included in the evaluation were submitted to the clinical laboratory for routine testing. The tissue culture test was done according to the in-house procedure. The table below shows a summary of the clinical performance of the TOX A/B QUIK CHEK™ test. The test exhibited a sensitivity and specificity of 90.2% and 99.7%, respectively. The predictive positive and negative values were 98.6% and 97.9%, respectively, and the correlation was 98.0%.
Table 1. Correlation of the TOX A/B QUIK CHEK™ test with tissue culture.
| N = 842 | Tiss cult pos | Tiss cult neg |
| --- | --- | --- |
| TOX A/B QUIK CHEK™ pos | 138 | 2 |
| TOX A/B QUIK CHEK™ neg | 15 | 687 |
| | | 95% CI |
| --- | --- | --- |
| Sensitivity | 90.2 | 84.1 - 94.2 |
| Specificity | 99.7 | 98.8 - 99.9 |
| Predictive Positive Value | 98.6 | 94.4 - 99.8 |
| Predictive Negative Value | 97.9 | 96.4 - 98.7 |
| Correlation | 98.0 | 97.8 - 98.2 |
Of the 2 tissue culture-negative/TOX A/B QUIK CHEK™-positive samples, one was negative in a commercial toxin A+B ELISA. Of the 15 specimens that were tissue culture-positive/TOX A/B QUIK CHEK™-negative, 12 were negative in commercial toxin A+B ELISAs. There were nine specimens that were unreadable. Those specimens were negative by PCR analysis for the genes of toxin A (tcdA) and toxin B (tcdB).
A total of 51 fecal specimens diluted in Cary Blair and 32 fecal specimens diluted in C&S Transport Media were tested in the TOX A/B QUIK CHEK™ test and the results were compared to those obtained by routine testing. The test exhibited an agreement of 97.6% for the detection of C. difficile toxins in specimens prepared in Transport Media.
b. Clinical specificity:
See Sec. M.3.a.
c. Other clinical supportive data (when a. and b. are not applicable):
N/A
{6}
4. Clinical cut-off:
See limit of detection above Sec. M.1.d
5. Expected values/Reference range:
The reported incidence of *C. difficile*-associated disease in patients with antibiotic-associated diarrhea is 10 to 20%. In the studies conducted with this device, the incidence ranged from 10% to 22%. The prevalence of a positive TOX A/B QUIK CHEK™ test will vary from location to location and hospitals may experience rates lower or higher than those observed at the sites used in this evaluation. Clostridium difficile disease is primarily a nosocomial disease of elderly patients, and hospitals that have higher numbers of elderly patients may experience higher rates. It is important to consider any test results in conjunction with clinical symptoms because some healthy adults and large numbers of healthy infants (up to 50%) will be positive for *C. difficile* toxin.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.