WELCH ALLYN CARDIOPERFECT WORKSTATION SOFTWARE (CPWS) V 1.62
K082478 · Welch Allyn, Inc. · BZG · Mar 9, 2009 · Anesthesiology
Device Facts
Record ID
K082478
Device Name
WELCH ALLYN CARDIOPERFECT WORKSTATION SOFTWARE (CPWS) V 1.62
Applicant
Welch Allyn, Inc.
Product Code
BZG · Anesthesiology
Decision Date
Mar 9, 2009
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems. The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below. Stress Electrocardiograph - Intended Use Using the optional ECG module and associated accessories the user can acquire, view, store and print ECG waveforms. Resting Electrocardiograph - Intended Use The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis. These are presented for review and interpretation by the clinician. Spirometry - Intended Use Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. The spirometer should only be used with patients able to understand the instructions for performing the test. Ambulatory Blood Pressure - Intended Use Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history.
Device Story
CardioPerfect Workstation Software (CPWS) operates on standard Windows-based PCs to manage ECG, ambulatory blood pressure (ABP), and spirometry data. Input signals are acquired via optional hardware modules; software processes these signals to generate waveforms, measurements, and interpretive statements. Clinicians in hospitals or clinics use the software to review advisory interpretations, manage patient records, and export data to EMRs. The system assists in diagnosing and monitoring cardiac and pulmonary conditions. Version 1.6.2 introduces the PEDMEANS algorithm for pediatric ECG interpretation. Output is displayed on-screen or printed for physician review to support clinical decision-making.
Clinical Evidence
Bench testing only. Verification of MEANS and PEDMEANS algorithms against IEC 60601-2-51:2003 requirements. Risk management conducted per ISO 14971. No clinical trials performed; equivalence established via technical comparison and software validation.
Technological Characteristics
Software-based diagnostic workstation for ECG, ABP, and spirometry. Operates on off-the-shelf Windows PCs. Includes MEANS and PEDMEANS interpretive algorithms. Complies with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-25, and AAMI/ANSI EC11 standards. Software Level of Concern: Moderate.
Indications for Use
Indicated for adult and pediatric patients requiring cardiac screening or diagnostic differentiation (ECG), pulmonary function assessment (spirometry) for symptoms like shortness of breath, cough, or wheezing, and ambulatory blood pressure monitoring for conditions including white coat hypertension, nocturnal hypertension, resistant hypertension, diabetes, or pregnancy-related hypertension. Spirometry contraindication: patients unable to understand test instructions.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
Predicate Devices
CardioPerfect Workstation Software Version 1.5.0 (K052158)
{0}------------------------------------------------
# 510(k) Summary
Kof 2478
[As described in 21 CFR 807.92]
9 2009 MAR
Submitted by:
Welch Allyn Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220
Contact Person:
John Sawyer Vice President, QA/RA Phone: (315) 685-4571 Fax: (315) 685-2532 E-mail: sawyerj@welchallyn.com
August 22, 2008 Date Prepared:
Trade Name:
CardioPerfect Workstation Software (CPWS) Version 1.6.2
Common Name:
Medical Device Software for interpretive and stress testing electrocardiographs and diagnostic spirometry.
Classification Reference: Class II, Diagnostic Spirometer (21 CFR 868.1840, Product Code BZG) and Cardiovascular (Product Code LOS).
Predicate Device:
CardioPerfect Workstation Software Version 1.5.0 (K052158)
{1}------------------------------------------------
### Description of the Device:
CardioPerfect Workstation software is used to create a computer platform on which ECG, ABP, & Spirometry applications can operate within inherent capabilities of an off-theshelf desktop or laptop personal computer utilizing a Windows operating system.
### Intended Use:
The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems.
The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below.
# Stress Electrocardiograph - Intended Use
Using the optional ECG module and associated accessories the user can acquire, view, store and print ECG waveforms.
# Resting Electrocardiograph - Intended Use
The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis. These are presented for review and interpretation by the clinician.
# Spirometry - Intended Use
Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. The spirometer should only be used with patients able to understand the instructions for performing the test.
# Ambulatory Blood Pressure - Intended Use
Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history.
{2}------------------------------------------------
### Indications for Use:
# Electrocardiograph - Indications for Use
Indications for electrocardiography range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department.
# Spirometry - Indications for Use
Indications for spirometry include, but are not limited to, the following:
- Shortness of breath .
- . Chronic cough
- Occupational exposure to dust or chemicals .
- Assist in the diagnosis of Bronchitis .
- Assist in the diagnosis of Asthma .
- . Wheezing
- Assist in the monitoring of bronchodilator ●
# Ambulatory Blood Pressure - Indications for Use
Indications for ambulatory blood pressure measurement (as listed in Journal of Hypertension 2003, 21 :821-848, E. O'Brien et. at.) include, but are not limited to, the following:
- · suspected "white coat" hypertension
- · Suspected nocturnal hypertension
- To establish dipper status
- · Resistant hypertension
- · Elderly patient
- · To monitor antihypertensive drug treatment
- · Type 1 diabetes
- · Hypertension of pregnancy
- · Evaluations of hypotension
- · Autonomic failure
- · Masked hypertension
### Technological Characteristics:
The Welch Allyn Electrocardiography and Spirometry Products are intended for use by trained operators in healthcare facilities.
The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
{3}------------------------------------------------
The optional spirometry module is indicated for use in clinical situation to assess a patient's pulmonary health status and evaluate symptoms, signs, or abnormal laboratory test results.
The CPWS version 1.6.2 is substantially equivalent to the previously released CPWS version 1.5.0 (K.052158). The major difference being the change to intended use. This submission intends to change the indications for use to include interpretive ECG software for the pediation. The previously marketed CPWS version 1.5.0 (K052158) did not provide interpretive software support for patients less than 16 years of age. The Pediatric Modular ECG PEDMEANS Interpretive ECG algorithm developed by P. Rijnbeek, 2007, in Analysis System thesis entitled 'Automatic Interpretation of Pediatric Electrocardiograms', Erasmus University Rotterdam. This PEDMEANS algorithm, is also used in our CP 100™ and CP 200™ ECG device cleared under K072449.
## Summary of Effectiveness:
The Welch Allyn CPWS team has determined that the software "Level of Concern" is Moderate. (See section 10 for CPWS software Level of Concern)
The implementation of the PEDMEANS interpretive software in the CPWS 1.6.2 electrocardiograph device, and its performance, is equivalent in every respect to the implementation of this software in the CP 100™ and CP 200™ submitted to the agency by Welch Allyn and cleared per K072449.
The MEANS algorithms were verified and found to be consistent with the requirements of IEC 60601-2-51:2003, particular requirements for safety, including essential performance, of recording and analyzing single channel and multi-channel electrocardiographs.
All requirements of the recognized and applicable standards are in compliance: EC11 (AAMI/ANSI), UL 60601-1. IEC 60601-1. IEC 60601-1-1. IEC 60601-1-2. IEC 60601-1-4 and IEC 60601-2-25.
Additionally, risk management (risk, SFMEA and safety analysis) activities have been conducted in accordance with ISO 14971 Medical Devices - Application of risk management to medical devices and comply with IEC 60601-1-4 Medical Electrical Equipment Part 1: General Requirements for Safety, Part 4: Programmable Electrical Medical Systems.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The logo is black and white.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
8 2005
Welch Allyn, Inc. c/o Mr. John Sawyer Vice President QA/RA 4341 State Street Road Skaneateles Falls, NY 13153-0220
K082478 Re:
> Trade/Device Name: Welch Allyn CardioPerfect Workstation Software (CPWS), Version 1.6.2 Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II (Two) Product Code: BZG Dated: February 4, 2009 Received: March 5, 2009
### Dear Mr. Sawyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. John Sawyer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
J. Atkin
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use Statement
KO82478 510(k) Number (if known):
Device Name: Welch Allyn CardioPerfect Workstation Software (CPWS) V 1.6.2
### Intended Use:
The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems.
The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below.
# Stress Electrocardiograph - Intended Use
Using the optional ECG module and associated accessories the user can acquire, view, store and print ECG waveforms.
## Resting Electrocardiograph - Intended Use
The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis. These are presented for review and interpretation by the clinician.
# Spirometry - Intended Use
Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. The spirometer should only be used with patients able to understand the instructions for performing the test.
# Ambulatory Blood Pressure - Intended Use
Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history.
Atill Stein for B. Zuckerman
(Division Sign-Off) Division of Cardiovascular Device 510(k) Number KOS2478
{7}------------------------------------------------
### Indications for Use:
### Electrocardiograph - Indications for Use
Indications for electrocardiography range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department.
### Spirometry - Indications for Use
Indications for spirometry include, but are not limited to, the following:
- Shortness of breath .
- Chronic cough .
- Occupational exposure to dust or chemicals ●
- Assist in the diagnosis of Bronchitis .
- Assist in the diagnosis of Asthma .
- Wheezing ♥
- Assist in the monitoring of bronchodilator .
## Ambulatory Blood Pressure - Indications for Use
Indications for ambulatory blood pressure measurement (as listed in Journal of Hypertension 2003, 21 :821-848, E. O'Brien et. at.) include, but are not limited to, the following:
- · suspected "white coat" hypertension
- · Suspected nocturnal hypertension
- To establish dipper status
- · Resistant hypertension
- · Elderly patient
- · To monitor antihypertensive drug treatment
- · Type 1 diabetes
- · Hypertension of pregnancy
- · Evaluations of hypotension
- · Autonomic failure
- · Masked hypertension
X AND/OR Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Stittser for B Zuckerman
Mision Sign-Off) Division of Cardlovascular Device 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.