K072449 · Welch Allyn, Inc. · DPS · Nov 29, 2007 · Cardiovascular
Device Facts
Record ID
K072449
Device Name
CP100 & 200 12-LEAD RESTING ELECTROCARDIOGRAPHS
Applicant
Welch Allyn, Inc.
Product Code
DPS · Cardiovascular
Decision Date
Nov 29, 2007
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2340
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Welch Allyn Electrocardiography and Spirometry Products (Subject Devices) are intended for use by trained operators in health facilities. The subject devices provide the following diagnostic functions: Acquiring, viewing, storing and printing ECG waveforms using ECG Front End Modules and associated accessories that provide signal acquisition for up to twelve (12) leads of patient ECG waveforms through surface electrodes adhered to the body. Using optional algorithms to generate measurements, data presentations, graphical presentations and interpretative statements on an advisory basis. These are presented for review and interpretation by the clinician based upon knowledge of the patient, the result of physical examination, the ECG tracings and other clinical findings. Using the optional Spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function including, but not limited to, maximal volume and flow of air that can be moved in and out of the patient's lungs. These measures are used in the diagnosis and monitoring of lung diseases and interventions for the treatment of certain lung diseases. The spirometer should only be used with patients able to understand the instructions for performing the test.
Device Story
Welch Allyn CP 100 and CP 200 are electrocardiographs for clinical use by trained healthcare providers. Devices acquire 12-lead ECG signals via surface electrodes; CP 200 also supports optional spirometry via external transducer/tubing. Inputs are processed to generate ECG waveforms, measurements, and interpretive statements using the MEANS algorithm. Devices feature alphanumeric keyboards, LCD displays (color on CP 200), and internal thermal printers. Data can be stored locally or transmitted to a CardioPerfect workstation for integration with electronic patient records. Interpretive results serve as advisory guidance for physicians to confirm diagnoses. Spirometry functionality uses custom software to replicate Medikro D9 performance on an embedded platform. Benefits include standardized cardiac monitoring and pulmonary function assessment in clinical settings.
Clinical Evidence
Bench testing only. MEANS algorithm validated by Erasmus University Rotterdam (1996) and verified against IEC 60601-2-51:2003. Spirometry functionality validated by LDS Hospital (Dr. Robert O. Crapo) per ATS 1994/2005 standards, confirming accuracy and precision requirements. Compliance with AAMI/ANSI EC11, UL 60601-1, and IEC 60601 series standards demonstrated.
Indicated for adult and pediatric patients to evaluate, diagnose, and monitor cardiac function via 12-lead ECG. Optional spirometry module indicated for assessing pulmonary health status and evaluating symptoms, signs, or abnormal laboratory results in patients capable of following test instructions.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
# WelchAllyn
Abbreviated 510(k)
Section 5 – 510(k) Summary
### 510(k) Summary
[As described in 21 CFR 807.92]
| Submitted by: | Welch Allyn Inc.<br>4341 State Street Road<br>Skaneateles Falls, NY 13153-0220 | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact Person: | John Sawyer<br>Vice President, QA/RA<br>Phone: (315) 685-4571<br>Fax: (315) 685-2532<br>E-mail: sawyerj@welchallyn.com | NOV 2 9 2007 |
| Date Prepared: | August 24, 2007 | |
| Trade Name: | Welch Allyn CP 100TM and CP 200TM Electrocardiographs | |
| Common Name: | Electrocardiograph | |
| Classification Reference: | Class II, Electrocardiograph (21 CFR 870.2340, Product<br>Code DPS) and Diagnostic Spirometer (21 CFR 868.1840,<br>Product Code BZG) | |
| Predicate Device: | PageWriter Trim Series Cardiograph<br>Philips Medical Systems<br>510(k) Number: K001265<br>Electrocardiograph, 21 CFR 870.2340<br>Class II, DPS<br><br>Medikro D9 Spirometer<br>Caird Technology Inc<br>510(k) Number: K031422<br>Diagnostic Spirometer, 21 CFR 868.1840<br>Class II, BZG | |
{1}------------------------------------------------
#### Description of the Device:
#### The Welch Allyn CP 100TM Electrocardiograph:
The Welch Allyn CP 100™ electrocardiograph features a full alphanumeric key board, an LCD display, full-size user-programmable reports, and the ability to operate on either battery or AC power.
ECG tests sent to a memory card are compatible with the Welch Allyn CardioPerfect™ workstation, which in turn can connect with other electronic patient-information systems, such as billing and medical records.
The CP 100™ electrocardiograph is specifically intended for acquiring and printing ECG signals from adult and pediatric patients. It will be used in clinical settings by trained healthcare providers.
The optional interpretation algorithm analyzes these ECG signals to generate measurements and interpretive statements. The interpretive results are intended only as guidance for qualified physicians and must not be relied upon as diagnoses.
Image /page/1/Picture/9 description: The image shows a piece of office equipment, possibly a printer or a fax machine. The device has a keyboard on the front and a display screen on the right side. The device appears to be white or light gray in color. The image is a black and white photograph.
Welch Allyn CP 100™
Image /page/1/Picture/11 description: The image shows a central hub with multiple cables extending outwards. The hub is light-colored and has a curved design. Several of the cables are bundled together with a band, while others are individual and coiled. The cables terminate in various connectors, including what appears to be a VGA connector and several audio jacks.
#### The Welch Allyn CP 200TM Electrocardiograph:
The Welch Allyn CP 200™ electrocardiograph can display, print, save, and send ECGs electronically. It features a full alphanumeric keyboard, a color display to preview ECGs and edit settings, storage for up to 50 ECG and 50 spirometry records, full-size user programmable reports, and the ability to operate on either battery or AC power.
For centralized ECG data storage, the CP 200™ electrocardiograph can connect to a Welch Allyn CardioPerfect™ workstation, which in turn can connect with other electronic patient-information systems, such as billing and medical records.
{2}------------------------------------------------
# WelchAllyn
The CP 200 electrocardiograph is specifically intended for acquiring, viewing, storing, and printing ECG signals from adult and pediatric patients. It will be used in clinical settings by trained health care providers.
The optional interpretation algorithm analyzes these ECG signals to generate measurements and interpretive statements. The interpretive results are intended only as guidance for qualified physicians and must not be relied upon as diagnoses.
The electrocardiograph provides and optional interface to a pulmonary function device.
Communication of ECG and spirometry data with a central data-management system is optional.
Image /page/2/Picture/6 description: The image shows a Welch Allyn CP 200 electrocardiograph machine. The machine has a keyboard, a display screen, and a printer. The machine is white and has a black printer on the left side. The machine is used to record the electrical activity of the heart.
Image /page/2/Picture/7 description: The image shows the text "Welch Allyn CP 200". The text is in a simple, sans-serif font and is left-aligned. The letters are black, and the background is white. The number 200 has a superscripted "T 1/2".
Image /page/2/Picture/8 description: The image shows a collection of white cables and connectors against a black background. The cables are arranged in a radial pattern, emanating from a central hub or connector. Some of the cables are coiled, while others are straight, and they all appear to be interconnected through the central hub.
### Intended Use:
The Welch Allyn Electrocardiography and Spirometry Products (Subject Devices) are intended for use by trained operators in health facilities. The subject devices provide the following diagnostic functions:
- Acquiring, viewing, storing and printing ECG waveforms using ECG Front End . Modules and associated accessories that provide signal acquisition for up to twelve (12) leads of patient ECG waveforms through surface electrodes adhered to the body.
- Using optional algorithms to generate measurements, data presentations, graphical . presentations and interpretative statements on an advisory basis. These are presented for review and interpretation by the clinician based upon knowledge of the patient, the result of physical examination, the ECG tracings and other clinical findings.
- Using the optional Spirometry module and associated accessories to acquire, view, . store and print measures and waveforms of pulmonary function including, but not limited to, maximal volume and flow of air that can be moved in and out of the patient's lungs. These measures are used in the diagnosis and monitoring of lung diseases and interventions for the treatment of certain lung diseases. The
{3}------------------------------------------------
## Welch Allyn
spirometer should only be used with patients able to understand the instructions for performing the test.
#### Indications for Use:
The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.
The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
The optional spirometry module is indicated for use in clinical situations to assess a patient's pulmonary health status and evaluate symptoms, signs, or abnormal laboratory test results.
#### Technological Characteristics:
The Welch Allyn Electrocardiography and Spirometry Products are intended for use by trained operators in healthcare facilities.
The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
The optional spirometry module is indicated for use in clinical situation to assess a patient's pulmonary health status and evaluate symptoms, signs, or abnormal laboratory test results.
The following table summarizes the similarities between the subject Welch Allyn CP 100™ and CP 200™ Electrocardiographs and the predicate Philips PageWriter Trim Series Cardiograph (K031422) and Caird Technology Medikro D9 Spirometer (K971336).
| Designation | Philips Medical Systems<br>PageWriter Trim Series Cardiograph<br>(Trim II) | Welch Allyn Medical Systems<br>CP Series Electrocardiographs (Model CP200) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating Principle | Electrocardiographs | Electrocardiographs |
| Display | Monochrome display | Color LCD:<br>View and adjust the ECG waveforms.<br>With spirometry option, also view patients' spirometry<br>efforts and results. |
| Keyboard | Full alphanumeric keyboard | Full alphanumeric keypad |
| ECG Storage | Storage for 50 ECGs | Storage for up to 50 ECG and 50 Spirometry records |
| Designation | Philips Medical Systems<br>PageWriter Trim Series Cardiograph<br>(Trim II) | Welch Allyn Medical Systems<br>CP Series Electrocardiographs (Model CP200) |
| Battery Operation | - Typically 40 ECGs on a single charge or 30 minutes of<br>continuous rhythm recording and no fail operation during<br>ECG printing.<br>- Battery recharge – eight hours to full capacity. | - Use the electrocardiograph almost anywhere. On<br>battery power, 4hours of continuous use or continuous<br>printing of 100 ECG pages. The continuous test is<br>based on performing 5 ECG's in a period of 4 hours.<br>- Battery recharge – 12 hours to full capacity. |
| Power | The electrocardiograph can run on AC or battery power. | The electrocardiograph can run on AC or battery<br>power. |
| Printer | Thermal printer (internal) | Thermal printer (internal) |
| Sterility | Wipe the external surfaces of the cardiograph with a soft<br>cloth dampened in any of the approved cleaning solutions<br>listed below:<br>Mild soap and water<br>Isopropyl alcohol<br>Disinfect the patient cable. Use a damp soft cloth with one<br>of the disinfectants or cleaning agents listed below.<br>Cidex Ortho Phthaladehyde<br>Cetylcide<br>Vesphene 2 Aqueous Phenolic Germicidal Agent | Wipe the exterior of the patient cable and<br>electrocardiograph and with a damp cloth using mild<br>detergent diluted in water.<br>Disinfect the patient cable. Use a damp cloth of<br>chemical disinfectants containing one of the following:<br>ethanol (70% - 80%)<br>propanol (70% - 80%)<br>aldehydes (2% - 4%) |
| Safety | Testing to applicable standards:<br>• AAMI EC11<br>• IEC 60601-1<br>• UL 2601.1<br>• IEC 60601-2-25<br>• IEC 60601-1-2 | Recognized and applicable standards:<br>• EC11 (AAMI/ANSI)<br>• UL 60601-1<br>• IEC 60601-1<br>• IEC 60601-1-1<br>• IEC 60601-1-2<br>• IEC 60601-1-4<br>• IEC 60601-2-25<br>• IEC 60601-2-51 |
| Weight | 16.3 lb | 11.6 lb |
| Dimensions | 15.3 x 12.2 x 6.9 in | 16.2 x 15.6 x 6.2 in |
| Filters | > AC noise<br>> Artifact<br>> High and low pass | > 0.5 Hz high-performance baseline filter<br>> 35 Hz muscle-tremor filter<br>> AC interference filter |
| ECG Acquisition | ECG signal acquisition of up to 12 leads | ECG signal acquisition of up to 12 leads |
| Network Connection | 10 Base-T Ethernet LAN card | Wireless |
| Fax | Support for PCMCIA fax modem | None |
| Modem | V.90, K56flex | None |
| Spirometry | None | Support for Spirometry<br>(K971336 - Medikro D9 Spirometer) |
| Indication for Use | Where the clinician decides to evaluate the<br>electrocardiogram of adult and pediatric patients as part of<br>decisions regarding possible diagnosis, potential treatment,<br>effectiveness of treatment or to rule-out causes for<br>symptoms. | The electrocardiograph is one of the tools that<br>clinicians use to evaluate, diagnose, and monitor<br>patient cardiac function. The 12-lead ECG interpretive<br>algorithm provides a computer-generated analysis of<br>potential patient cardiac abnormalities, which must be<br>confirmed by a physician with other relevant clinical<br>information. The optional spirometry module is<br>indicated for use in clinical situations to assess a<br>patient's pulmonary health status and evaluated<br>symptoms, signs, or abnormal laboratory test results. |
| Designation | Philips Medical Systems<br>PageWriter Trim Series Cardiograph<br>(Trim II) | Welch Allyn Medical Systems<br>CP Series Electrocardiographs (Model CP200) |
| Target Population | Adult and pediatric patients. | - Resting ECG: Adult and pediatric patients.<br>- Resting ECG Interpretation: Adult and pediatric<br>patients.<br>- Spirometry: Adult and pediatric patients. |
| Where Used | Healthcare facilities | Healthcare facilities |
{4}------------------------------------------------
### WelchAllyn
长072449
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Welch Allyn. The text is in a bold, sans-serif font. The logo is simple and recognizable.
Abbreviated 510(k) Section 5 - 510(k) Summary
K072449 2617
| Subject Area | Caird Technology Inc.<br>Medikro D9 Spirometer | Welch Allyn Medical Systems<br>CP Series Electrocardiographs (CP200)<br>Spirometer |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510 (k) Number | K971336 | OEM Device (Medikro D9 Spirometer - K971336) |
| Target Population | Adult and pediatric patients. | Spirometry: Adult and pediatric patients. |
| Where Used | Hospital and Clinic use only | Healthcare facilities |
| Indication for Use | Some of the conditions for use are as follows:<br>• Shortness of Breath<br>• Chronic Cough<br>• Occupational Exposure to Dust or Chemicals<br>• Assist in the Diagnosis of Bronchitis<br>• Assist in the Diagnosis of Asthma<br>• Wheezing<br>• Assist in the monitoring of Bronodialators | The optional spirometry module is indicated for use in<br>clinical situations to assess a patient's pulmonary<br>health status and evaluate symptoms, signs, or<br>abnormal laboratory test results. |
The Welch Allyn CP 100™ and CP 200™ Series electrocardiograph/spirometry has equivalent electrocardiograph characteristics as the Philips PageWriter Trim Series and Medikro D9 Spirometer.
The technological differences do not affect the safety or effectiveness of the device.
#### Summary of Effectiveness:
The Welch Allyn CP Series team has determined that the software "Level of Concern" is Moderate. (See section 10 for CP 100™ & CP 200™ software Level of Concern)
The implementation of the MEANS interpretive software in the CP 100™ and CP 200™ electrocardiographs device, and its performance, is equivalent in every respect to the implementation of this software in the CardioPerfect™ with MEANS device submitted to the agency by Cardio Control BV (= Welch Allyn acquired Cardio Control BV) and cleared per K962854. Both of these product families have implemented the 1996 version of the MEANS algorithms. The Department of Medical Informatics of Erasmus University Rotterdam independently validated the MEANS algorithms in 1996. A copy
{6}------------------------------------------------
## Welch Allyn
of the specificity and sensitivity from that study, as presented for the Welch Allyn CardioPerfect™ with MEANS submission K962854.
The MEANS algorithms were verified and found to be consistent with the requirements of IEC 60601-2-51:2003, particular requirements for safety, including essential performance, of recording and analyzing single channel and multi-channel electrocardiographs.
All requirements of the recognized and applicable standards are in compliance: EC11 (AAMI/ANSI), UL 60601-1, IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-25 and IEC 60601-2-51.
As a point of clarification, only the CP 200™ product will include spirometry functionality. The CP 200™ uses the same flow transducer, signal-conditioning electronics and tubing as the Medikro D9 Spirometer (K971336); all of the components external to the personal computer of the Medikro D9. The CP 200™ makes use of its embedded computer in place of the personal computer used in the Medikro D9. Because the user interface, display and embedded computer differ from that of the Medikro D9, the CP 200™ requires custom software to provide the functionality of the Spiro2000 software provided with the Medikro D9.
The spirometry functionality of the CP 200™, including the custom software, was validated by Robert O. Crapo, M.D., Medical Director of the pulmonary Laboratory at LDS Hospital in Salt Lake City, Utah. He performed testing using the standard "Standardization of Spirometry: 1994 Update and 2005 version" published by the Americal Thoracic Society (ATS). The test report issued by Dr. Crapo's laboratory -LDS Hospital, dated March 2, 2007, the CP 200™ device meets all stated accuracy and precision requirements defined in the ATS standard.
Additionally, risk management (risk, SFMEA and safety analysis) activities have been conducted in accordance with ISO 14971 Medical Devices - Application of risk management to medical devices and comply with IEC 60601-1-4 Medical Electrical Equipment Part 1: General Requirements for Safety, Part 4: Programmable Electrical Medical Systems.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol of medicine and healing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular pattern around the caduceus.
Public Health Service
NOV 2 9 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Welch Allyn, Inc. c/o Mr. John Sawyer Vice President, QA/RA 4341 State Street Road Skaneateles Falls, NY 13153-0220
Re: K072449
Welch Allyn CP 100™ and CP 200™ Electrocardiographs Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS and BZG Dated: October 22, 2007 Received: October 24, 2007
Dear Mr. Sawyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. John Sawyer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blyminston for
Bram &. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a handwritten string of characters, "K072449". The characters are written in a casual, slightly uneven style, suggesting they were written quickly. The numbers are clearly distinguishable, and the letter 'K' is capitalized.
### Indications for Use Statement
510(k) Number (if known): K072449
Device Name: Welch Allyn CP 100™ and CP 200™ Electrocardiographs
#### Indications for Use:
The electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.
The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information.
The optional Spirometry module is indicated for use in clinical situations to assess a patient's pulmonary health status and evaluate symptoms, signs, or abnormal laboratory test results.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|

| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | K072449 |
Welch Allyn CP100™ & CP200™
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.