VERATHON INC. BLADDERSCAN ULTRASOUND SYSTEM, MODEL BVI 9600
Applicant
Verathon Incorporated
Product Code
IYO · Radiology
Decision Date
Dec 1, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The Verathon Inc. BladderScan® BVI 9600 with AortaScan® Mode is a user selectable, dual-function ultrasound device that projects ultrasound energy either into the lower abdomen to obtain an image of the bladder for measuring bladder volume, or into the mid-abdomen to obtain an image of the abdominal aorta for aortic diameter measurement.
Device Story
Portable, battery-operated B-mode ultrasound system; utilizes 3.0 MHz mechanical sector transducer. Operates in two user-selectable modes: BladderScan (bladder volume) and AortaScan (aortic diameter). Transducer collects cross-sectional images from 12 scan planes; Vmode technology constructs 3D finite element model of target organ. System automatically computes bladder volume via volumetric integration or aortic diameter. Output displayed on integrated LCD; thermal printer available for exam reports. Used in clinical settings by healthcare providers. Non-adjustable ultrasonic power. Benefits include accurate, noninvasive measurements compared to 2D ultrasound.
Clinical Evidence
Bench testing only. Accuracy validated using tissue-equivalent phantoms with known dimensions (NIST-traceable). Performance compared against expected phantom values and standard ultrasound systems (Sonosite 180). No clinical patient data presented.
Technological Characteristics
Portable B-mode ultrasound; 3.0 MHz mechanical sector transducer (driven at 1.74 MHz for second harmonic bladder wall detection). Battery-operated. Integrated LCD and thermal printer. Software-based volumetric integration and finite element modeling. Non-adjustable power output. Uses Sontac Ultrasound Coupling Gel.
Indications for Use
Indicated for noninvasive bladder volume measurement and abdominal aortic diameter measurement in adult patients. Contraindicated for fetal use, pregnant patients, and acute clinical events including aortic dissection, ulcer, or rupture.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Verathon Inc. BladderScan® BVI 9400 Ultrasound System (K071217)
SonoSite MicroMaxx High-Resolution Ultrasound System (C3 Series) (K053069)
Reference Devices
Sonosite 180
Submission Summary (Full Text)
{0}------------------------------------------------
K082456
pg. 1 of 4
### 510(K) SUMMARY (As required by 21 CFR 807.92(a))
# DEC 0 1 2008
| Submitter Information | Company: Verathon Inc. |
|-----------------------|---------------------------------------------------------|
| | 20001 North Creek Parkway |
| | Bothell, WA 98011 |
| Phone: | 425-867-1348 ext.5640 |
| Fax: | 425-883-2896 |
| Email: | rzeine@verathon.com |
| Contact: | Richard A. Zeine |
| | Director of Quality Assurance and<br>Regulatory Affairs |
| Date: | August 15, 20008 |
| Device Information | |
Verathon Inc. BladderScan® BVI 9600 Ultrasound System with AortaScan®
Diagnostic Ultrasound System with Accessories
Common Name:
Classification Name(s):
Trade/Proprietary Name:
Regulatory Class: II Review Category: Tier II Classification Panel: Radiology Ultrasonic Pulsed Echo Imaging System
- FR Classification 892.1560 .
- . Product Code 90-1YO
- Diagnostic Ultrasound Transducer
- FR Classification 892.1570 .
- Product Code 90-ITX .
Verathon BladderScan® BVI 9400 Ultrasound System (K071217)
And SonoSite MicroMaxx High-Resolution Ultrasound System (C3 Series) (K053069)
The BVI9600, with AortaScan®, is a B-mode ultrasonic instrument that is portable and battery operated. The BVI 9600 has two user controlled modes of operation; (1) - Bladder Volume Measurement and (2)- Abdominal Aortic Measurement, both noninvasive. Either bladder volume or aortic diameter measurement functions can be chosen by a button press. When the scan button is pressed the BVI9600 uses patented Vmode® technology to measure ultrasonic
Predicate Device:
Device Description:
A.
{1}------------------------------------------------
reflections on multiple planes inside the body to produce a 3dimensional image. Based on these images and the mode of operation chosen, the BV19600 either calculates and displays the bladder volume (BladderScan® mode) or calculates and displays the abdominal aortic diameter (AortaScan® mode). The measurements made with Vmode ultrasound are more accurate than those from conventional 2-dimensional ultrasound, as they are based on a more complex. 3-D image. Bladder volume (in BladderScan® mode), aortic diameter (in AortaScan® mode), mode selection, directional aiming, battery status, and usage rate indicators are all displayed on the BV19600 LCD display. The instrument contains a thermal printer that allows the user to print exams with a single button click.
New device: The Verathon Inc. BladderScan® BVI 9600 with AortaScan® is a user selectable dual function ultrasound device that projects ultrasound energy either into the lower abdomen to obtain an image of the bladder for measuring bladder volume or the mid abdomen to obtain an image of the abdominal aorta for aortic diameter measurement.
The BladderScan® BVI 9600 is not intended for fetal use or use on pregnant patients.
The BladderScan® BVI 9600 is not intended for acute events such as aortic dissection, ulcer or rupture.
The Verathon Inc. BladderScan® BVI 9600 Ultrasound System with AortaScan® retains the same bladder volume measurement features of the Verathon inc. BladderScan® BVI 9400 System (K071217). The BVI 9600 is identical to the BVI 9400, with regards to hardware, software, transducer and calibration method.
The Verathon Inc. BladderScan® BVI 9600, with AortaScan®, and its integrated 3.0 MHz mechanical sector transducer operate only in B-mode to locate and automatically measure bladder volume (in BladderScan® Mode) or abdominal aortic diameter (in AortaScan® Mode). The same transducer is driven at 1.74 MHz to obtain a second harmonic for enhanced bladder wall detection. Bladder volume (in BladderScan® Mode), aortic diameter (in AortaScan® Mode), patient gender, non optimal directional aiming, battery status, and usage rate indicators are all displayed on the Verathon Inc. BladderScan® BVI 9600 Ultrasound System scanner. The ultrasonic power transmitted from the system is not user adjustable.
The portable Verathon Inc. BladderScan® BVI 9600 Ultrasound System with AortaScan® is applied to the
Intended Use:
Contraindications (U.S. Only):
Comparison of Required Technological Characteristics:
C.
{2}------------------------------------------------
K082456
Pg. 3 of 4
patient's abdomen along with Sontac Ultrasound Coupling Gel to obtain aortic measurement. The BladderScan® BVI 9600 Ultrasound System transducer collects cross- sectional images of the bladder or aorta from twelve (12) scan planes. From this information, the Verathon Inc. BladderScan® BVI 9600 Ultrasound System constructs a finite element model of the bladder (in BladderScan® Mode) or aorta (in AortaScan® Mode) and automatically computes the volume of urine via volumetric integration (for the bladder measurement) or the diameter of the aorta (for aortic measurement), depending on the mode of operation chosen (AortaScan® or BladderScan®).
In order to demonstrate the BladderScan® accuracy claimed in 0270-0422 -xx-55, two third party vendors were contracted to build tissue equivalent phantoms with known dimensions. The supplier of the aortic phantom, Blue Phantom, is known for supplying medical imaging phantoms to the medical marketplace as is the manufacturer of the bladder phantom, Computer Imaging Reference Systems (CIRS).
The AortaScan® phantom is essentially a tube roughly the length of an adult aorta, with a bulge in the length of the tube that represents an aortic aneurism, filled with blood mimicking material that is surrounded by tissue mimicking material. Known diameters of the tube can then be compared to the AortaScan® measurements from the BVI 9600.
The BladderScan® Phantom is essentially a balloon which gets filled with urine mimicking material and then the filled balloon is surrounded with tissue mimicking material. This balloon is a known volume and can be compared to the BladderScan® measurements from the BVI 9600.
During the manufacturing processes, both Blue Phantom and CIRS measure key parameters of the physical parts as the parts are constructed. These measurements are all NIST traceable and each phantom comes with a certification sheet listing the measurements.
These results obtained through the Verathon Inc. BladderScan® BVI 9600 Ultrasound System are compared to the expected results derived from the NIST traceable measurements. In addition to these measurements, additional measurements are taken by standard ultrasound systems, such as the Sonosite 180, to compare bladder volume measurements and aortic measurements.
AortaScan® accuracy has been demonstrated by comparing the AortaScan® Phantom measurements to the measurements of the Verathon Inc. BladderScan® BVI 9600 Ultrasound System.
{3}------------------------------------------------
K082454
pg. Hof 4
The calibration system for the Verathon Inc. BladderScan® BVI 9600 Ultrasound System with AortaScan® is exactly the same as the Verathon Inc. BladderScan® BVI 9600 Ultrasound System (K071217).
Summary and Conclusion of Non Clinical and Clinical Testing
Design and testing of the Verathon Inc. BladderScan® BVI 9600 Ultrasound System with AortaScan® indicates that the BVI 9600 is substantially equivalent to the Verathon Inc. BladderScan® BVI 9400 Ultrasound System (K071217) for bladder volume measurements and equivalent to the Sonosite MicroMaxx Ultrasound System (K053069)for aorta scan measurements.
All acoustic output measurements for the BladderScan® BVI 9600 Ultrasound System with AortaScan® are not changed from the BladderScan® BVI 9400 Ultrasound System and remain within Pre-amendment limits.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing the world, representing the department's mission to protect the health of all Americans.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Richard A. Zeine Director of Quality and Regulatory Affairs Verathon Incorporated 20001 North Creek Parkway BOTHELL WA 98011
DEC 0 1 2008
Re: K082456
Trade/Device Name: Verathon Inc. BladderScan® BVI 9600 with AortaScan Ultrasound System Regulation Number: 21 CFR 892.1560
Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: October 7, 2008 Received: October 8, 2008
Dear Mr. Zeine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Verathon Inc. BladderScan® BVI 9600 with AortaScan Ultrasound System, as described in your premarket notification:
Transducer Model Number3.0/1.74 MHz Second Harmonic Transducer
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Lauren Hefner at (240) 276-3666.
Sincerely yours.
Jorju M. Whay
Joycé M. Whang, Ph.D. Acting Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): K082456
## Device Name: Verathon Inc. BladderScan® BVI 9600 with AortaScan® Ultrasound System
Indications for Use:
- Abdomen, B-Mode, per Indications for Use Ultrasound Form
- The Verathon Inc. BladderScan® BVI 9600 with AortaScan® Mode is a user selectable, dual-function ultrasound device that projects ultrasound energy either into the lower abdomen to obtain an image of the bladder for measuring bladder volume, or into the mid-abdomen to obtain an image of the abdominal aorfa for aorlic diameter measurement.
Contraindications:
- The BladderScan® BVI9600 is contraindicated for fetal use and for use on pregnant patients.
The BladderScan® BVI 9600 is not intended for acute events such as aortic dissection, ulcer or rupture.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hondh Whly
(Division S Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
{7}------------------------------------------------
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM System: BladderScan® BVI 9600 Ultrasound System 3.0 / 1.74 MHz Second Harmonic Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|------------------------------------------------------------------|-------------------------------------------------------|-------------------|---|-----|-----|---------------|---------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | P | | | | | | |
| | Intra-operative<br>(Abdominal organs and<br>vascular) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (breast,<br>thyroid, testicles) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel.<br>(Convent.) | | | | | | | |
| | Musculo-skel.<br>(Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (Abdominal Aorta<br>Measurement) | N | | | | | | |
| | Peripheral vessel | | | | | | | |
| | Other (Bladder) | P | | | | | | |
| N= new indication; P= previously cleared by FDA (K071217) | | | | | | | | |
| (Division Sign-Off) | | | | | | | | |
| Division of Reproductive, Abdominal, and<br>Radiological Devices | | | | | | | | |
108245
510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.