K072034 · Quanta System Spa · GEX · Sep 20, 2007 · General, Plastic Surgery
Device Facts
Record ID
K072034
Device Name
DIODE MEDICAL LASER FAMILY, MODELS 4, 6, 30
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 20, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Quanta System Diode Medical Lasers Family, (and the fiber delivery systems and accessories used to deliver laser energy), are indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: gastroenterology, general surgery, genitourinary surgery (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT), orthopedics, ophthalmology, pulmonology, and thoracic surgery.
Device Story
Quanta System Diode Medical Laser Family (Models 4, 6, 30) consists of a laser surgical instrument for soft tissue applications. Input: electrical power; laser source generates 808 nm, 940 nm, or 980 nm wavelengths. Operation: laser energy delivered via optical fibers and handpieces to target tissue for ablation, vaporization, excision, incision, or coagulation. Used in OR, clinic, or dental settings by physicians or dentists. System includes external structure, power electronics, display/control electronics, and opto-mechanical block (laser source, Peltier cooling, fiber launching, aiming beam, power calibration). Safety features include remote interlock and key switch. Output: laser energy for surgical effect. Healthcare providers use the device to perform procedures ranging from endovascular vein treatment and BPH surgery to dental soft tissue management. Benefits include precise tissue interaction and hemostasis.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Diode laser system with 808 nm, 940 nm, and 980 nm wavelengths. Subsystems: external structure, power electronics, display/control electronics, and opto-mechanical block (laser source, Peltier cooling, fiber launching, red diode aiming beam, power calibration). Connectivity: SMA 905 fiber socket. Safety: remote interlock, key switch. Form factor: console-based. Sterilization: accessories (fibers/handpieces) provided sterile or compatible with standard sterilization.
Indications for Use
Indicated for soft tissue ablation, vaporization, excision, incision, and coagulation across multiple surgical specialties including gastroenterology, general surgery, urology, gynecology, neurosurgery, ENT, orthopedics, ophthalmology, pulmonology, and thoracic surgery. Specific populations include men >50 with BPH (28-85cc prostate volume), patients with superficial vein reflux, and dental patients requiring soft tissue procedures or periodontal treatment. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
PhotoMedex, Inc.'s LaserPro Surgical Diode Laser System (K042211)
BioLase Technology, Inc EZLase Dental Diode Laser system (K061898)
INTERmedic Arfran, S.A. INTERmedic Diode Laser Family (K053540)
Dornier MedTech America, Inc Dornier Diode Laser Family (K020339)
Submission Summary (Full Text)
{0}------------------------------------------------
K072034
SEP 20 2007
# 510(k) SUMMARY Quanta System's Diode Medical Laser Family
Title:
| Submitter: | Quanta System SpA<br>via IV Novembre, 116<br>21058 Solbiate<br>Olona VA / Italy |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Dr. Isabella Carrer<br>Medical Division Manager |
| Date Prepared: | May 8, 2007 |
| Device Trade Name: | Quanta System Diode Medical Laser Family |
| Common Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Classification<br>Name: | Instrument, surgical, powered, laser<br>GEX<br>21 C.F.R. 878.4810 |
| Predicate Devices: | PhotoMedex, Inc.'s LaserPro Surgical Diode Laser<br>System (K042211)<br>Diomed, Inc. EVLT Kit and DPlus Lasers<br>(K041957)<br>BioLase Technology, Inc EZLase Dental Diode<br>Laser system (K061898)<br>INTERmedic Arfran, S.A. INTERmedic Diode<br>Laser Family (K053540)<br>Dornier MedTech America, Inc Dornier Diode<br>Laser Family (K020339) |
| Intended Use /<br>Indications for Use: | 808 nm, 940 nm, 980nm Wavelengths<br>The Quanta System Diode Medical Lasers Family, (and<br>the fiber delivery systems and accessories used to<br>deliver laser energy), are indicated for use in surgical<br>applications requiring the ablation, vaporization,<br>excision, incision, and coagulation of soft tissue in<br>medical specialties including: gastroenterology, general<br>surgery, genitourinary surgery (urology), gynecology<br>(GYN), neurosurgery, otolaryngology (ENT),<br>orthopedics, ophthalmology, pulmonology, and thoracic |
{1}------------------------------------------------
### surgery.
The Quanta System Diode Medical Lasers Family are indicated for use in the performance of specific surgical applications in gastroenterology, general surgery, genitourinary surgery (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT), orthopedics. ophthalmology, pulmonology, and thoracic surgery as follows:
## Gastroenterology (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in gastroenterology procedures. Applications include: hemostasis of esophageal varices; palliation of malignant dysphagia; palliative ablation of obstructive neoplasms; hemostasis of colonoscopy.
### Neurosurgery (Medical 30, 6)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in neurosurgery procedures. Applications include: tumors adjacent to the spinal cord; tumors adjacent to the cortex.
## General Surgery (Medical 30)
Treatment of varicose veins and varicosities associated with superficial reflux of the greater saphenous vein. The ablation, vaporization, excision, incision, and coagulation of soft tissue in general surgery including endoscopic and open procedures. Applications include: Laparoscopic: appendectomy; cholecystectomy; bowel resection. Open: mastectomy; reduction mammoplasty; breast biopsy; rectal and anal hemorrhoidectomy; bowel resection; colectomy; cholecystectomy; liver resection; condyloma; thyroidectomy; thoracotomy; cavernous hemangioma.
# Genitourinary (Urology) (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in genitourinary (urology) procedures. Applications include: Transurethral: transurethral incision of the prostate (TUIP); bladder tumors; bladder neck incisions; urethral strictures; exterior sphincterotomy. Laparoscopic lymphadenectomy. Open: condyloma; circumcision; benign and
{2}------------------------------------------------
## malignant lesions of external genitalia
#### Thoracic Surgery (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in thoracic surgery including endoscopic and open procedures. Applications include: pulmonary resection; coagulation of blebs and bullae; adhesiolysis; pericardiectomy; mediastinal and thoracic lesions and abnormalities; mediastinal lymph node dissection: hemostasis; thoracotomy.
#### Gvnecology (GYN) (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in gynecology (GYN) procedures. Applications include: Laparoscopic excision/lysis of adhesions; endometrial lesions, including ablation of endometriosis; laparoscopic assisted hysterectomy (LAVH); laser uterosacral nerve ablation (LUNA); myomectomy; ovarian cystectomy; ovarian drilling; tubal fimbrioplasty; appendectomy. Open: conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia VIN, VAIN; condyloma acuminata, including cervical, genital, vulvar, perineal, and Bowen's disease, (Erythroplasia of Queyrat) and Bowenoid papulosa (BP) lesions. Intrauterine: Fibroids/polyps/adhesions; Resection of septum.
### Pulmonology (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in pulmonology procedures. Applications include: tracheal bronchial lesions.
## Ophthalmology (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in ophthalmology procedures. Applications include: Oculoplastics; open DCR; endo-nasal DCR; tumor excision and biopsy; eyelid reconstruction; blepharoplasty.
## Orthopedics (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in orthopedic surgery procedures. Applications include: Open: Dissect and coagulate.
{3}------------------------------------------------
## Otolaryngology_(ENT) (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in otolaryngology procedures. Applications include: Nasal/Sinus: turbinectomy and turbinate reduction/ablation; polypectomy of nose and nasal passages; ethmoidectomy; meatal antrostomy; Laryngotracheal: removal of vocal cord/fold nodules, polyps and cysts;arytenoidectomy; tracheal stenosis; Oropharyngeal: uvulopalatoplasty (LAUP, laser UTPP); tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil; hemi glossectomy; Head & Neck: tumor resection on oral, subfacial and neck tissues; parathyroidectomy; thyroidectomy.
In addition, at 808 nm and 940 nm wavelengths, the Quanta system Diode Medical Laser Family is intended for the following use:
## Urology (Medical 30):
Lesions of external genitalia; Circumcision; Condyloma; Bladder tumors; Bladder neck incisions; Vaporization of the prostate
### Neurosurgery (Medical 30):
Percutaneous Disc Decompression; Discectomy; Hemostasis in conjunction with meningiomas
Arthroscopy (Medical 30): Chondromalacia; Synovectomy; Menisectomy
## Pulmonary Surgery (Medical 30);
Endoscopic pulmonary applications; Tracheal bronchial lesions; Benign and malignant pulmonary or stricture
Cardiac Surgery (Medical 30): Coagulation and haemostasis of cardiac tissue
## Dermatology/Aesthetics (Medical 30):
Photocoagulation of vascular & dermatological lesions of the face and extremities; Photocoagulation of telangiectasia, veinulectasias of the legs and face; Treatment of reticular veins and branch varicosities; Pyogenic granuloma, lymphangioma and lymphangiomatosis disease, angiofibromas;
{4}------------------------------------------------
Superficial benign vascular lesions including Telangiectasias, Rosacea, Angioma, venous lakes Couperosis, Chery angioma, Hemangioma, Port wine stains, angiokeratoma, and benign ipidermal pigment lesions as lentigines. Epidermal nevi, spider nevi.; Dermatology surgery: Condyloma acuminate, warts, small non malignant skin tumors, small semi-malignant tumors as basalomas, Bowe, Kaposi sarcom. Warty leucoplasy and ulcers debridment; Seborrheic keratosis; Mixoid cyst; Papillary varix; Acne treatment; Hair removal of unwanted hair from skin type I-V;
#### Plastic Surgery (Medical 30):
Cut, coagulation & vaporization; Blepharoplasy
#### Vascular Surgery (Medical 30):
Endoluminal or endovenus laser surgery for saphenous incompetent veins.
### Dental applications (Medical 30, 6, 4):
Frenectomy; Frenotomy; Biopsy; Pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening
## 808 nm Wavelength
The Quanta System Diode Medical Laser Family (Medical 30) is intended for use in the treatment of superficial vein reflux of the greater saphenous vein associated with varicosities. The Quanta System Diode Medical Laser Family are indicated for treatment of incompetence and reflux of superficial veins in the lower extremity.
#### 940 nm Wavelength
The Quanta System Diode Medical Laser Family (Medical 30) is intended for use in endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux.
The Quanta System Diode Medical Laser Family (Medical 30) is intended for use in the treatment of
{5}------------------------------------------------
symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostates with median and/or lateral lobes ranging in total volume from 28-85cc and for cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes and colonoscopes), or in incision/excision, vaporization, ablation and coagulation of soft tissue in contact or non contact open surgery (with or without a handpiece).
It (Medical 30) is also intended for use in the treatment and/or removal of vascular lesions (tumors) and for the removal of unwanted hair.
It (Medical 30) is also indicated for use in medicine and surgery, in the following medical specialties: Urology, Plastic Surgery, General Surgery, Dermatology, Gynecology, Pulmonary Surgery, Gastroenterology, ENT and Radiology.
The Quanta System Diode Laser Family (Medical 30, 6, 4) is also indicated for incision, excision, vaporization, ablation and coagulation of oral soft tissue including marginal and inter-dental gingival and epithelial lining of free gingival and the following specific indications:Excisional and incisional biopsiesExposure of unerupted teeth; Fibroma removal; Frenectomy; Frenotomy; Gingival troughing for crown impressions; Gingivectomy; Gingivoplasty; Gingival incision and excision; Haemostasis and coagulation; Implant recovery; Incision and drainage of abscess; Leukoplakia; Operculectomy; Oral papillectomies; Pulpotomy; Pulpotomy as an adjunct to root canal therapy; Reduction of gingival hypertrophy; Soft tissue crown lengthening; Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa; Vestibuloplasty
The Quanta System Diode Laser Family (Medical 30, 6, 4) is also indicated for Laser Periodontal procedures, including: Laser soft tissue curettage; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased, infected, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility.)
{6}------------------------------------------------
Model 30 is indicated for Surgical, Dermatological, Dental and Endovascular applications.
Model 6 and Model 4 are indicated for Dental and Surgical applications.
The Diode Laser Family is designed with 4 major subsystems: (1) an external structure; (2) power electronics; (3) display with control electronics, which controls the power electronics, the user interface and the laser source temperature via a thermostat board; and (4) the laser system with an opto-mechanical block composed of the laser source, the Peltier cooling system with dissipater and fans, the fiber launching system, the red diode aiming beam, and the power calibration system. The external accessories include separate optical fibers and hand pieces for dental, dermatological and surgical applications, or for endovascular applications. The fiber is connected to the system through an SMA 905 socket on the front panel. In addition to the four subsystems, the Diode Laser Family incorporates several safety features, including a remote interlock and a key switch..
The Quanta System Diode Medical Laser Family is as safe and effective as the predicate devices. The Diode Medical Laser Family has the same intended uses, and similar indications for use, technological characteristics, and principles of operation as the predicates. The minor technological differences between the Diode Medical Laser Family and its predicates raise no new issues of safety or effectiveness. Thus, the Diode Medical Laser Family is substantially equivalent.
Technological Characteristics:
Substantial Equivalence:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a seal for the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 1 2008
Quanta System, S.P.A. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, Northwest Buffalo, MN 55313
Re: K072034
Trade/Device Name: Quanta System Diode Medical Laser Family: Models 4, 6, and 30 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II (two) Product Code: OCL, GEX Dated: August 28, 2007 Received: August 29, 2007
Dear Mr. Job:
This letter corrects our substantially equivalent letter of September 20, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
O'Mallein
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known) :
Device Name:
Quanta System Diode Medical Laser Family: Models 4, 6, and 30
Indications For Use:
## 808 nm, 940 nm, 980 nm Wavelengths
The Quanta System Diode Medical Lasers Family, (and the fiber delivery systems and accessories used to deliver laser energy), are indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: gastroenterology, general surgery, genitourinary surgery (urology), gynecology (GYN), neurosurgery, otolaryneology (ENT), orthopedics, ophthalmology, pulmonology, and thoracic surgery,
The Quanta System Diode Medical Lasers Family are indicated for use in the performance of specific surgical applications in gastroenterology, general surgery, genitourinary surgery (urology), gynecology (GYN), neurosurgery, otolaryngology (ENT), orthopedics, ophthalmology, pulmonology, and thoracic surgery as follows:
## Gastroenterology (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in gastroenterology procedures. Applications include: hemostasis of esophageal varices; palliation of malignant dysphagia; palliative ablation of obstructive neoplasms; hemostasis of colonoscopy.
#### Neurosurgery (Medical 30,6)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in neurosurgery procedures. Applications include: tumors adjacent to the spinal cord; tumors adjacent to the cortex.
## General Surgery (Medical 30)
Treatment of varicose veins and varicosities associated with superficial reflux of the greater saphenous vein. The ablation, vaporization, excision, incision, and coagulation of soft tissue in general surgery including endoscopic and open procedures. Applications include: Laparoscopic: appendectomy; cholecystectomy; bowel resection. Open: mastectomy; reduction mammoplasty; breast biopsy; rectal and anal hemorrhoidectomy; bowel resection; colectomy; cholecystectomy; liver resection; condyloma;
thyrgidectorny; thoracotomy; cavernous hemangioma.
Division Sign-Off
Page 1 of 5
Division of General, Restorative, and Neurological Devices
510(k) Number
{10}------------------------------------------------
## Genitourinary (Urology) (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in genitourinary (urology) procedures. Applications include: Transurethral: transurethral incision of the prostate (TUIP); bladder tumors; bladder neck incisions; urethral strictures; exterior sphincterotomy. Laparoscopic lymphadenectomy. Open: condyloma; circumcision; benign and malignant lesions of external genitalia
### Thoracic Surgery (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in thoracic surgery including endoscopic and open procedures. Applications include: pulmonary resection; coagulation of blebs and bullae; adhesiolysis; pericardiectomy; mediastinal and thoracic lesions and abnormalities; mediastinal lymph node dissection; hemostasis; thoracotomy.
### Gvnecology (GYN) (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in gynecology (GYN)
procedures. Applications include: Laparoscopic excision/lysis of adhesions; endometrial lesions, including ablation of endometriosis; laparoscopic assisted hysterectomy (LAVH); laser uterosacral nerve ablation (LUNA); myomectomy; ovarian cystectomy; ovarian drilling; tubal fimbrioplasty; appendectomy. Open: conization of the cervix, including cervical intraepithelial neoplasia (CIN), vulvar and vaginal intraepithelial neoplasia VIN, VAIN; condyloma acuminata, including cervical, genital, vulvar, perineal, and Bowen's disease, (Erythroplasia of Queyrat) and Bowenoid papulosa (BP) lesions. Intrauterine: Fibroids/polyps/adhesions; Resection of septum.
#### Pulmonology (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in pulmonology procedures. Applications include: tracheal bronchial lesions.
#### Ophthalmology (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in ophthalmology procedures. Applications include: Oculoplastics; open DCR; endo-nasal DCR; tumor excision and biopsy; eyelid reconstruction; blepharoplasty.
### Orthopedics (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in orthopedic surgery procedures, Applications include: Open: Dissect and
coagulate.
Mark A. Miller
Page 2 of 5
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number
{11}------------------------------------------------
# Otolaryngology_(ENT) (Medical 30)
The ablation, vaporization, excision, incision, and coagulation of soft tissue in otolaryngology procedures. Applications include: Nasal/Sinus: turbinectorny and turbinate reduction/ablation; polypectomy of nose and nasal passages; ethmoidectomy; meatal antrostomy; Laryngo-tracheal: removal of vocal cord/fold nodules, polyps and cysts;arytenoidectomy; tracheal stenosis; Oropharyngeal: uvulopalatoplasty (LAUP, laser UTPP); tonsillectomy (including tonsillar cryptolysis, neoplasma) and tonsil; hemi glossectomy; Head & Neck: tumor resection on oral, subfacial and neck tissues; parathyroidectomy; thyroidectomy.
In addition, at 808nm and 980nm wavelengths, the Quanta system Diode Medical Laser family is intended for the following use:
## Urology (Medical 30):
Lesions of external genitalia; Circumcision; Condyloma; Bladder turnors; Bladder neck incisions; Vaporization of the prostate
### Neurosurgery (Medical 30):
Percutaneous Disc Decompression; Discectomy; Hemostasis in conjunction with meningiomas
#### Arthroscopy (Medical 30):
Chondromalacia; Synovectomy; Menisectomy
## Pulmonary Surgery (Medical 30):
Endoscopic pulmonary applications; Tracheal bronchial lesions; Benign and malignant pulmonary or stricture
### Cardiac Surgery (Medical 30):
Coagulation and haemostasis of cardiac tissue
## Dermatology/Aesthetics (Medical 30):
Photocoagulation of vascular & dermatological lesions of the face and extremities; Photocoagulation of telangiectasia, veinulectasias of the legs and face; Treatment of reticular veins and branch varicosities; Pyogenic granuloma, lymphangioma and lymphangiomatosis disease, angiofibromas; Superficial benign vascular lesions including Telangiectasias, Rosacea, Angioma, venous lakes Couperosis, Chery angioma, Hemangioma, Port wine stains, angiokeratoma, and benign ipidermal pigment lesions as lentigines. Epidermal nevi, spider nevi ; Dermatology surgery: Condyloma acuminate, warts, small non malignant
Mark A. Muller
Page 3 of 5
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
**510(k) Number**.
{12}------------------------------------------------
skin tumors, small semi-malignant tumors as basalomas, Bowe, Kaposi sarcom. Warty leucoplasy and ulcers debridment; Seborrheic keratosis; Mixoid cyst; Papillary varix; Acne treatment; Hair removal of unwanted hair from skin type I-V;
#### Plastic Surgery (Medical 30):
Cut, coagulation & vaporization; Blepharoplasy
#### Vascular Surgery (Medical 30):
Endoluminal or endovenus laser surgery for saphenous incompetent veins.
### Dental applications (Medical 30, 6, 4):
Frenectomy; Frenotomy; Biopsy; Pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening
#### 808 nm Wavelength
The Quanta System Diode Medical Laser Family (Medical 30) is intended for use in the treatment of superficial vein reflux of the greater saphenous vein associated with varicosities. The Quanta System Diode Medical Laser Family are indicated for treatment of incompetence and reflux of superficial veins in the lower extremity.
#### 940nm Wavelength
The Quanta System Diode Medical Laser Family (Medical 30) is intended for use in endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux.
The Quanta System Diode Medical Laser Family (Medical 30), is intended for use in the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostates with median and/or lateral lobes ranging in total volume from 28-85cc and for cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, gastroscopes, cystoscopes and colonoscopes), or in incision/excision, vaporization, ablation and coagulation of soft tissue in contact or non contact open surgery (with or without a handpiece).
It (Medical 30) is also intended for use in the treatment and/or removal of vascular lesions (tumors) and for the removal of unwanted hair.
Mark A. Milliken
Page 4 of 5
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
K07203
**510(k) Number**
{13}------------------------------------------------
It (Medical 30) is also indicated for use in medicine and surgery, in the following medical specialties: Urology, Plastic Surgery, General Surgery, Dermatology, Gynecology, Pulmonary Surgery, Gastroenterology, ENT and Radiology.
The Quanta System Diode Laser Family (Medical 30, 6, 4) is also indicated for incision, excision, vaporization, ablation and coagulation of oral soft tissue including marginal and inter-dental gingival and epithelial lining of free gingival and the following specific indications:Excisional and incisional biopsiesExposure of unerupted teeth; Fibroma removal; Frenectomy; Gingival troughing for crown impressions; Gingivectorny; Gingivoplasty; Gingival incision and excision; Haemostasis and coagulation; Implant recovery; Incision and drainage of abscess; Leukoplakia; Operculectomy; Oral papillectomies; Pulpotomy; Pulpotomy as an adjunct to root canal therapy; Reduction of gingival hypertrophy; Soft tissue crown lengthening; Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa; Vestibuloplasty
The Quanta System Diode Laser Family (Medical 30, 6, 4) is also indicated for Laser Periodontal procedures, including: Laser soft tissue curettage; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility.)
Prescription Use: > AND/OR (Part 21 C.F.R. 801 Subpart D)
Over-The-Counter Use: (Part 21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office off Device Evaluation (ODE)
Mash A. Millan
Page 5 of 5
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
72039
**510(k) Number**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.