CIVASTRING PD-103

K082159 · Civatech Oncology, Inc. · KXK · Sep 30, 2008 · Radiology

Device Facts

Record IDK082159
Device NameCIVASTRING PD-103
ApplicantCivatech Oncology, Inc.
Product CodeKXK · Radiology
Decision DateSep 30, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

CivaString"d103 is intended for medical purposes to be placed into a body cavity or tissue as a source of nuclear radiation for therapy.

Device Story

CivaStringPd103 is a linear, low dose rate (LDR) brachytherapy source composed of polymeric materials; contains Pd-103 isotope distributed into wells along device length. Wells can be filled or left empty to customize radiation delivery based on therapy plan. Device functions as a sealed source delivering therapeutic radioactive energy (20-23 keV x-rays) to localized tumors. Unlike predicate 'seed' configurations, CivaString provides a linear string form factor. Used by clinicians for permanent interstitial implantation in body cavities or tissue. Benefits include targeted radiation delivery for primary or residual tumor treatment.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, material, function, and intended use comparisons to legally marketed predicate devices.

Technological Characteristics

Linear LDR brachytherapy source; polymeric materials; Pd-103 isotope; 20-23 keV x-ray emission; sealed source configuration; linear form factor.

Indications for Use

Indicated for permanent interstitial brachytherapy for treatment of selected localized tumors. Used as primary treatment or for residual disease post-excision. May be used concurrently or sequentially with other modalities like external beam radiation or chemotherapy.

Regulatory Classification

Identification

A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.

Special Controls

*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Section 5 - 510(k) Summary SEP 3 0 2008 #### 1. Applicant Contact: Lois V. Smart Regulatory Consultant Smart Consulting Services, Inc. 1317 Buchanan Drive Mebane, NC 27302 Phone: (919) 563-0869 Fax: (888) 622-4075 Email: loissmart@yahoo.com Date Prepared: July 30, 2008 - CivaStringPd103 2. Name of Device: Common Name: Radionuclide Brachytherapy Source Classification Name: Brachytherapy Radionuclide Regulation 21 CFR 892.5730, Product Code KXK ### Identification of device(s) to which the submitted claims equivalence: 3. The CivaString 1103 is substantially equivalent to the following predicate devices: - RadioMed™ Source by RadioMed Corporation, 510(k) K001070 � - OptiSource103 by International Brachytherapy, Inc., 510(k) K040766 ● - Readi-Strand Model PSStrand by Worldwide Medical Technologies, Inc., � 510(k) K023179 #### 4. Device Description: CivaString" d103 is a linear LDR (low dose rate) brachytherapy source comprised of polymeric materials. CivaString 1103 contains Pd-103 isotope that is distributed into "wells" along the length of the device. Wells may be filled or left empty (to provide gaps in regions where no radiation is desired) based on the overall therapy plan prescription. #### 5. Intended Use of the Device: CivaString"d103 is intended for medical purposes to be placed into a body cavity or tissue as a source of nuclear radiation for therapy. {1}------------------------------------------------ ## Indications for Use 6. CivaStringPd103 is indicated for use as a permanent interstitial brachytherapy source for the treatment of selected localized turnors. The device may be used either as the primary treatment or for treatment of residual disease after excision of the primary tumor. CivaString 9103 may be indicated for use concurrently with or sequentially with other treatment modalities, such as external beam radiation therapy or chemotherapy. ### 7. Characteristics of the device in comparison to those of the predicate device(s) When compared with the Intended Use and Indications for Use of the three (3) predicates, the Intended Use and Indications for Use does not change. The design of each of the predicates and CivaString 8103 is a sealed source from which a therapeutic dosage of radioactive energy is delivered. The predicate devices use a sealed source in a "seed" configuration. CivaString"d103 is a sealed source in the form of a string and is linear versus current point sources. The energy emitted from CivaStringPd103 is exactly the same as current Pd 103 sources: 20-23 keV x-rays. ### 8. Safety and Performance: The difference between the CivaString" and the above mentioned predicate devices do not raise any questions regarding the safety and effectiveness of the device. The device, as designed, is as safe and effective as its predicate devices. ### 9. Conclusion Based on the design, material, function and intended use discussed herein. CivaTech Oncology, Inc. believes the CivaString 8103 is substantially equivalent to predicate devices currently marketed under the Federal Food, Drug and Cosmetic Act. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized depiction of an eagle, represented by three curved lines that suggest the bird's wings and body. SEP 3 0 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 CivaTech Oncology, Inc. % Ms. Lois Smart Regulatory Consultant Smart Consulting Services, Inc. 1317 Buchanan Drive MEBANE NC 27302 Rc: K082159 Trade/Device Name: CivaStringPd103 Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class; II Product Code: KXK Dated: July 30, 2008 Received: July 31, 2008 ## Dear Ms. Smart: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ # Page 2 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jaque M. Whang Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Section 4 - Indications for Use Statement K082159 510k number if known: Device Name: CivaStringPd103 Indications for Use: CivaStringPd103 is indicated for use as a permanent interstitial brachytherapy source for the treatment of selected localized tumors. The device may be used either as the primary treatment or for treatment of residual disease after excision of the primary tumor. CivaString 4103 may be indicated for use concurrently with or sequentially with other treatment modalities, such as external beam radiation therapy or chemotherapy. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...