The RadioMed™ Source with activities from 0.1 to 5.0 mCi per centimeter length is indicated for temporary or permanent interstitial implantation or surface application to treat selected localized turnors. They can be used as either primary treatment or as treatment for residual disease after excision of primary or recurrent tumors. The RadioMed M Source may be used concurrently with or following treatment with other interventions. such as external beam therapy, hyperthermia, or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
Device Story
RadioMed™ Source is a brachytherapy device utilizing Palladium-103 (Pd-103) as a radionuclide. The device consists of a coiled rhodium wire, which serves as both the matrix and a radiopaque marker. It is a sealed source designed for interstitial implantation or surface application. The device is delivered to the tumor site using a 17 or 18-gauge needle and stylet, similar to existing brachytherapy seeds. It is intended for use by clinicians to treat localized tumors, either as primary therapy or for residual disease, and may be used in conjunction with external beam therapy, hyperthermia, or chemotherapy. The device is supplied non-sterile and requires steam autoclave sterilization by the end user prior to clinical application. By delivering therapeutic radioactive energy directly to the tumor site, the device aims to destroy malignant tissue while sparing surrounding healthy structures.
Clinical Evidence
Bench testing only. Performance was validated through compliance with ANSI N43.6-1997 (sealed source classification), ANSI N44.1-1973 (integrity/test specifications), ANSI N44.2-1973 (leak testing), and ISO 9978:1992 (leakage test methods). Sterilization validation was performed per AAMI 11134-1994 for steam autoclave. No clinical trial data was provided.
Technological Characteristics
Sealed radioactive source; radionuclide: Pd-103; form factor: coiled rhodium wire; radiopacity: provided by rhodium matrix; energy: 20-23 keV x-rays; delivery: 17 or 18-gauge needle and stylet; sterilization: steam autoclave; standards: ISO 10993-1, ANSI N43.6-1997, ANSI N44.1-1973, ANSI N44.2-1973, ISO 9978, AAMI 11134-1994.
Indications for Use
Indicated for temporary or permanent interstitial implantation or surface application to treat localized tumors (e.g., head, neck, lung, pancreas, prostate) in patients requiring primary treatment or treatment for residual/recurrent disease.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Ir-192 Wires from Amersham International (Preammendments device)
Submission Summary (Full Text)
{0}------------------------------------------------
RadioMed Corporation March 31, 2000
# OCT 2 4 2000
K001070
Lone
RadioMed™ Source Premarket Notification
RadioMed Corporation 9 Linnell Circle Billerica. MA 01821-3902
### 510k Summary
1. Sponsor Name
RadioMed Corporation 9 Linnell Circle Billerica, Massachusetts 01821-3902
Telephone:
(978) 663-7400 voice (978) 663-7771 fax
Contact Individual: John Schwamb
- 2. Device Name
Proprietary Name: Common/Usual Name: Classification Name:
RadioMed™ Source Brachytherapy Radionuclide Source Brachytherapy Radionuclide
- 3. Identification of Predicate or Legally Marketed Device
The predicate devices for the RadioMed™ Source are: Theragenics Pd-103 Seed K874787 Medi-Physics Rapid Strand TM K940632 Ir-192 Wires from Amersham International - Preammendments device
## 4. Device Description
The RadioMed™ source utilizes Pd-103 as the radionuclide for brachytherapy. The RadioMed The Source exists in the form of a coiled rhodium wire. The rhodium matrix also provides the radiopacity of the device. The RadioMed™ Source is a naturally sealed source in the form of a wire.
The RadioMed The Source is packaged non-sterile, single use, and is intended to be sterilized by the end user in accordance with a validated sterilization process. Sterilization is accomplished by exposure to steam autoclave.
The RadioMed" Source will be manufactured. labeled and packaged under GMP controls. Upon completion of the manufacturing and assembly process the device will be inspected to assure compliance to
{1}------------------------------------------------
RadioMed Corporation March 31, 2000
RadioMed™ Source
Premarket Notification
10081620
Page 2 of 3
#### 7 Performance Testing
Summary of standards achieved:
FDA QSR 21 CFR Part 820 Good Manufacturing Practices ISO 10993-1 1992 (E) Biological Evaluation of Medical Devices ANSI N43.6-1997: Classification of sealed radioactive sources ANSI N44.1-1973 Integrity and Test Specifications for selected Brachytherapy Sources ANSI N44.2-1973 Leak testing radioactive brachytherapy sources ISO 9978: 1992(E) "Radiation protection - Sealed radioactive sources - Leakage test methods".
AAMI Standard 11134-1994 Recommended practice for Steam Autoclave
The "performance" of the RadioMed™ source is subdivided into four categories
- 1. Radiation Energy
- 2. Source Strength Measurement: Calibration and Calibration Accuracy
- 3. Dose Comparison
- 4. Sealed Source Testing
{2}------------------------------------------------
specifications. The devices will be tested in accordance with Standard Operating Procedures.
The sources are delivered in the same manner as brachytherapy seeds currently on the market, i.e. a 17 or 18-gauge needle and stylet.
### 5. Intended Use
The RadioMed" Source with activities from 0.1 to 5.0 mCi per centimeter length is indicated for temporary or permanent interstitial implantation or surface application to treat selected localized turnors. They can be used as either primary treatment or as treatment for residual disease after excision of primary or recurrent tumors. The RadioMed M Source may be used concurrently with or following treatment with other interventions. such as external beam therapy, hyperthermia, or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
- 6. Comparison of Technological Characteristics
The design of each of the predicates and the RadioMed™ Source is a sealed source from which to deliver a therapeutic dosage of radioactive energy. The predicate devices use a sealed source in a "can" configuration, often referred to as a seed. The RadioMed™ Source is a sealed source in the form of a wire. The RadioMed source, like Ir-192 wires, is linear. In contrast, the Theragenics Seed and the I-125 Seeds that are inside of the Rapid Strand 10 are designed to emulate point sources.
The energy emitted for the RadioMed™ Source and the Theragenics 103Pd Seed is exactly the same: 20-23 keV x-rays.
The materials that make up the components of the predicates and the RadioMed The Source are both similar and different. The shells for the predicate devices and the RadioMed "" Source are different. Both predicate seeds use titanium. The RadioMed Source uses rhodium. The radionuclide for the predicate Theragenics 103Pd seed is Palladium - 103 which is the same as the RadioMedith Source. The marker in each of the predicate devices is lead, while that in the RadioMed Source is rhodium.
The delivery of the predicates and the RadioMed" Source is through a 17 or 18-gauge needle and stylet.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 4 2000
John Schwamb Quality Manager RadioMed Corporation 9 Linnell Circle Billerica, MA 01821
Re:
K001070 RadioMed™ Source Dated: August 4, 2000 Received: August 7, 2000 Regulatory class: II 21 CFR 892.5730/Procode: 90 KXK
Dear Mr. Schwamb:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. A substantialion assumes compliance with the Current Good Mannfacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations,
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
The text in the image is:
Baniel G. Schultz, M.D. Captain, USPHS Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure (s)
{4}------------------------------------------------
RadioMed Corporation March 31, 2000
RadioMed™ Source Premarket Notification
510(k) Number (if known): K001070
Device Name:
RadioMed™ Source
Indications For Use:
The RadioMed"10 Source with activities from 0.1 to 5.0mCi per centimeter is indicated for temporary or permanent interstitial implantation or surface application to treat selected localized turnors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual disease after excision of primary or recurrent tumors. The Pd-103 sources may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia, or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Simil Ch. Segura
(Division Sign Off) Division of Reproductive, Abdominal, ENT. and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.