KRELL WEIGHT & BODY FAT MONITORING SCALE, MODEL BFA-8530
K081587 · Krell Precision (Yang Zhou) Co., Ltd. · MNW · Oct 24, 2008 · Cardiovascular
Device Facts
Record ID
K081587
Device Name
KRELL WEIGHT & BODY FAT MONITORING SCALE, MODEL BFA-8530
Applicant
Krell Precision (Yang Zhou) Co., Ltd.
Product Code
MNW · Cardiovascular
Decision Date
Oct 24, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The KRELL Weight & Body Fat Monitoring Scale is indicated to measure body weight and impedance and estimate the percentage of Body Fat and body water, Bones mass using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults and children (from age 10 and up) who are not ill, feverish, have a chronic or acute disease, or a condition that affects the level of hydration such as pregnancy.
Device Story
Device is a body fat scale (Model BFA-8530) for home use by consumers. User inputs age, height, and gender; stands barefoot on two electrodes. Device applies 500 µA, 50 kHz current through body to measure bioelectrical impedance. Internal software calculates body fat percentage, body water, and bone mass using impedance, weight, and user-provided variables. Output displayed on scale screen. Helps users monitor body composition. Accuracy validated against DXA.
Clinical Evidence
Bench testing performed per EN/IEC 60601-1 for electrical, mechanical, and environmental safety. Effectiveness and accuracy validated via clinical testing comparing BIA measurements against DXA (Dual Energy X-Ray Absorptiometry) reference standard. All results satisfactory.
Technological Characteristics
Scale with two integrated electrodes. Bioelectrical impedance sensing (500 µA, 50 kHz). Measures weight and impedance. Internal software-based calculation using user-input variables (age, height, gender). Complies with EN/IEC 60601-1 safety standards.
Indications for Use
Indicated for generally healthy adults and children (age 10+) to measure weight and estimate body fat, body water, and bone mass via BIA. Contraindicated for individuals who are ill, feverish, have chronic/acute disease, or conditions affecting hydration (e.g., pregnancy).
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
# 4081587 PAGE 1 OF 3
#### 510(k) Summary
# OCT 2 4 2008
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Date: March 11, 2007
#### 1. Company and Correspondent making the submission:
- Name Krell Precision (Yang Zhou) Co., LTD
- Address Number 28, Xing Yang Road, Yang Zhou, Jiang Su Province,
- China
Telephone -- +86(0)514-87961819
Fax - +86(0)514-87961918
- Contact Mike Zheng
- Email mike zheng@krellprecision.com
- 2. Device :
Trade/proprietary name: KRELL WEIGHT AND BODY FAT SCALE/BFA-
8530
| Common Name | : Analyzer, Body Fat |
|-------------|----------------------|
|-------------|----------------------|
Classification Name : Impedance plethysmograph
- 3. Predicate Devices :
Tanita BC-533, (K040778); Bonso Health o meter BMF583
- 4. Classifications Names & Citations :
21CFR 870.2770, MNW, Body Fat Analyzer, Class2
{1}------------------------------------------------
#### 5. Description :
- 5.1 General
The patient steps on the scale device, where two electrodes are located. The patient must step on the electrodes with bare feet, with normal moisture. Through harmless current stimulation of 500 µA, at 50 kHz, the Krell Body Fat Analyzer calculates the body fat percentage. This calculation is done via the Bioelectrical Impedance Method. The current is passed through the body and the impedance of the body determines the body fat. The calculation is based upon electrical impedance, height, age. and gender. The calculation is performed via internal software, which uses the variables programmed in by the user. There are elements of this process that can produce erroneous readings, such as dry feet or improperprogrammed data. The User's Manual defines items which could cause erroneous readings.
This bioelectrical impedance method has been validated via DXA (Dual Energy X-Ray Absorptiometry).
### 5.2 Direction
As discussed in the General description, the Krell Body Fat Analyzer is relatively simple to use. The user inputs the variable data of age, height, and gender. The user steps onto the scale and the devices measures the user weight and body impedance (via the Bioelectric Impedance through the two electrodes on the scale). The scale displays the user's body fat composition. Upon the user's selection, the device can also display the user's weight.
# 6. Indication for use :
The KRELL Weight & Body Fat Monitoring Scale is indicated to measure body weight and impedance and estimate the percentage of Body Fat and body water, Bones mass using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults and children (from age 10 and up) who are not ill, feverish, have a chronic or acute disease, or a condition that affects the level of hydration such as pregnancy.
{2}------------------------------------------------
- 7. Comparison with predicate device :
Krell Precision (Yang Zhou) Co., LTD, believes that the Krell Weight and Body Fat Scale is substantially equivalent to the Tanita BC-533 (K040778), Bonso Health o meter Model BFM583 (K030349)
- 8. Safety and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1. Clinical testing was used to validate the effectiveness and accuracy of the device. All test results were satisfactory.
CF 2
- 9. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Krell. Inc. concludes that The Krell Weight and Body Fat Scale is safe and effective and substantially equivalent to predicate devices as described herein.
- 10. Krell Precision (Yang Zhou) Co., LTD will update and include in a summary any other information deemed seasonably necessary by the FDA.
END
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 2 4 2008
Krell Precision (Yang Zhou) Co., Ltd. c/o Mr. Charles Mack Principal Engineer International Regulatory Consultants, LLC 340 Shady Grove Road FLINTVILLE TN 37335
Re: K081587
> Trade/Device Name: KRELL Weight & Body Fat Monitoring Scale, Model BFA-8530 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: October 12, 2008 Received: October 20, 2008
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements conceming your device in the Federal Register.
{4}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jorgu M. Whang
Joyce M. Whang, Ph.D Acting Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
Indications for Use
510(k) Number (if known):
K081587
Device Name: Weight & Body Fat Monitoring Scale (Model BFA-8530)
Indications For Use:
The KRELL Weight & Body Fat Monitoring Scale is indicated to measure body weight and impedance and estimate the percentage of Body Fat and body water, Bones mass using BIA (bioelectrical impedance analysis). This product is for use by generally healthy adults and children (from age 10 and up) who are not ill, feverish, have a chronic or acute disease, or a condition that affects the level of hydration such as pregnancy.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
DIVISION Sign-Off)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K081587
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.