K081355 · Meditech International, Inc. · ILY · Dec 18, 2008 · Physical Medicine
Device Facts
Record ID
K081355
Device Name
BIOFLEX LD-R100 TREATMENT HEAD
Applicant
Meditech International, Inc.
Product Code
ILY · Physical Medicine
Decision Date
Dec 18, 2008
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BIOFLEX LD-R100 Treatment head device is intended to emit energy in the infrared spectrum to provide topical heating for use when heat is indicated in the temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting relaxation of the muscle tissue.
Device Story
BIOFLEX LD-R100 is a Class II infrared laser treatment head; used as an accessory with existing BioFlex Professional and Prescription systems (K023621, K051875). Device emits infrared energy to penetrate surface tissues; triggers cellular functions via topical heating. Hand-held form factor; operated by healthcare professionals in clinical settings. Output provides therapeutic heat; aids in pain relief and muscle relaxation. Benefits include temporary relief of minor muscle/joint pain, stiffness, and spasms.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Infrared laser treatment head; LED-based; hand-held form factor. Operates via infrared spectrum energy emission for topical heating. Power density, wavelength, and optical power are consistent with predicate devices.
Indications for Use
Indicated for patients requiring topical heating for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and for promoting relaxation of muscle tissue.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
Predicate Devices
BioFlex Prescription Unit and related Treatment Heads (K051875)
BioFlex LD-175 and LD-1200 Treatment Heads (K041885)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Meditech International Incorporated. The word "Meditech" is written in a large, bold, black font. Below the word "Meditech" is the phrase "INTERNATIONAL INCORPORATED" written in a smaller, bold, black font. The logo is simple and professional.
Image /page/0/Picture/2 description: The image shows a logo with the word "BIOFLEX" written vertically on the left side. To the right of the text, there are three curved lines stacked on top of each other, with a circle above them. The word "BIO" is stacked on top of the word "FLEX".
411 HORNER AVE., UNIT #1, TORONTO, ONTARIO, CANADA M8W 4W3 • TELEPHONE: (416) 251-1055 • FACSIMILE: (416) 251-2446
# 510(k) Summary of Safety and Effectiveness
## Submitter:
Dr. Fred Kahn President Meditech International Inc 411 Horner Ave. Unit 1, Toronto, Ontario, Canada M8W 4W3 Tel: 416-251-1055 Fax: 416-251-2446
### Contact:
Dr. Fred Kahn President Tel: 416-251-1055 Fax: 416-251-2446
Date: April 30, 2008
Classification Name: Lamp, infrared
Common/Usual Name: Infrared laser
Product Code/Regulation #: ILY/21 CFR part 890.5500
Proprietary Name: BIOFLEX LD-R100 Treatment head
### Predicate Devices:
We are making the claim that the BioFlex LD-R100 is substantial equivalent to the predicated devices listed in the chart below.
| LEGALLY<br>MARKETED<br>PREDICATE DEVICE | MANUFACTURE<br>NAME | REGULATORY<br>CLASS AND<br>PRODUCT<br>CODE | 510(K)<br>REGISTRATION<br>NUMBER |
|-------------------------------------------------------------|-------------------------------|--------------------------------------------|----------------------------------|
| BioFlex Prescription<br>Unit and related<br>Treatment Heads | Meditech<br>International Inc | Class II/ ILY | K051875 |
| BioFlex LD-175 and<br>LD-1200 Treatment<br>Heads | Meditech<br>International Inc | Class II/ ILY | K041885 |
Image /page/0/Picture/17 description: The image shows the text "MEDITECH INTERNATIONAL INC" at the bottom. Above this text, there is a circular logo with the letters "UL" inside. To the left of the logo is the text "A12233", and to the right is the text "ISO 134" and "ISO 900".
Image /page/0/Picture/19 description: The image shows a circular logo with the text "FDA Cleared" inside. The letters "FDA" are stacked on top of the word "Cleared". The logo is black and white and appears to be a stamp or seal of approval.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Meditech International Incorporated. The logo is in black and features a stylized, bold font. The word "Meditech" is prominently displayed in a large font size, with "INTERNATIONAL INCORPORATED" written in a smaller font size below it.
Image /page/1/Picture/1 description: The image shows a logo for a company called "BIOFLEX". The logo has the word "BIO" stacked vertically on the left side of the image, and the word "FLEX" stacked vertically below it. To the right of the text is a graphic that includes three curved lines and a circle above them. A horizontal line is below the word "FLEX".
411 HORNER AVE, UNIT #1, TORONTO, CANADA M8W 4W3 • TELEPHONE: (416) 251-1055 • FACSIMILE: (416) 251-2446
The rationale of declaring the new BioFlex LD-R100 is substantial equivalent to the above 2 predicate devices are based on the following:
- > Same Indications for use: is intended to emit energy in the infrared spectrum to provide topical heating for use when heat is indicated in the temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting relaxation of the muscle tissue.
- > Similar key design technical characteristics- All contain LED lasers that are applied on the surface of the human skin.
- Same/similar power density, wavelength, and optical power
- > Similar size (hand held), weight, power source, and performance
### Description:
The BIOFLEX LD-R100 Treatment head device is to be used with the Low Intensity Laser Therapy System, BioFlex Professional System & the Prescription Unit and Related Treatment Heads. These systems were cleared under 510(k) submission K023621 & K051875 respectively. The treatment head device is a Class II Low Level Laser treatment head that apply energy, which penetrates the surface to the underlying tissues, and triggers normal cellular functions.
### Indications for Use:
Is intended to emit energy in the infrared spectrum to provide topical heating for use when heat is indicated in the temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting relaxation of the muscle tissue.
### Technological Comparison to the Predicate:
Technologically, the BIOFLEX LD-R100 is substantially equivalent to the listed 2 predicated devices above. The risks, safety or effectiveness and benefits for the BIOFLEX LD-R100 are also comparable. The table of Comparison in Section 5.1 will provide additional information illustrating that the new BIOFLEX LD-R100 is substantially equivalent to the BioFlex Prescription Unit and related Treatment Heads and the BioFlex LD-I75 and LD-I200 Treatment Heads.
### Conclusion:
As stated above, Meditech International's conclusion is that the BioFlex LD-R100 is safe and effective and complies with the appropriate medical standards and is substantially equivalent to the above-identified predicate devices.
Image /page/1/Picture/16 description: The image shows the text "MEDITECH INTERNATIONAL INC" at the bottom. Above this text, there is a UL registered firm logo. To the left of the logo, the text "A12233" is visible. To the right of the logo, the text "ISO 134" and "ISO 900" are stacked on top of each other.
Image /page/1/Picture/18 description: The image shows a circular logo with the text "FDA Cleared" inside. The letters "FDA" are stacked on top of the word "Cleared". The logo appears to be a stamp or seal of approval.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular fashion around the eagle. The eagle is depicted in black, and the text is also in black. The logo is simple and recognizable, representing the department's role in protecting the health of all Americans.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 8 2008
Meditech International Inc. % Mr. Mark Slonchka Director of Product Development and Technical Services 411 Horner Avenue, Unit #1 Toronto, Ontario, Canada M8W 4W3
Re: K081355
Trade/Device Name: BIOFLEX LD-R100 Treatment Head Regulation Number: 21 CFR 890.5500 Regulation Name: Infrarcd lamp Regulatory Class: II Product Code: ILY Dated: November 19, 2008 Received: November 20, 2008
Dear Mr. Slonchka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Mark Slonchka
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark H. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Meditech International Incorporated. The word "Meditech" is written in a bold, sans-serif font, with the letters slightly overlapping each other. Below the word "Meditech", the words "INTERNATIONAL INCORPORATED" are written in a smaller, sans-serif font. The logo is black and white.
Image /page/4/Picture/1 description: The image shows a logo for BioFlex. The logo consists of the word "BIO" stacked vertically on the left, followed by a stylized image of three curved lines resembling waves or a fingerprint. To the right of the curved lines is a circle. Below the curved lines and circle, the word "FLEX" is written.
411 HORNER AVE., UNIT #1, TORONTO, ONTARIO, CANADA M8W 4W3 • TELEPHONE: (416) 251-1055 • FACSIMILE: (416) 251-2446
## Indications for Use
510(k) Number (if known): KK81355
Device Name: BIOFLEX LD-R100 Treatment Head
Indications for Use:
The BIOFLEX LD-R100 Treatment head device is intended to emit energy in the infrared spectrum to provide topical heating for use when heat is indicated in the temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting relaxation of the muscle tissue.
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sheila P. Dahl
(Division Sign-Off Division of General, Restorative, and Neurological Devices
510(k) Number K081355
Image /page/4/Picture/16 description: The image shows the text "MEDITECH INTERNATIONAL INC" at the bottom. Above this text, there are two logos. The logo on the left contains the text "A12233". The logo on the right contains the text "ISO 13485 ISO 9001".
Image /page/4/Picture/18 description: The image shows a circular logo with the text "FDA Cleared" inside. The letters "FDA" are stacked on top of the word "Cleared". The logo is enclosed in a thin, black circle.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.