BIOFLEX PRESCRIPTION SYSTEM AND RELATED TREATMENT HEADS

K051875 · Meditech International, Inc. · ILY · Mar 13, 2006 · Physical Medicine

Device Facts

Record IDK051875
Device NameBIOFLEX PRESCRIPTION SYSTEM AND RELATED TREATMENT HEADS
ApplicantMeditech International, Inc.
Product CodeILY · Physical Medicine
Decision DateMar 13, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The BioFlex™ Prescription Unit and related Treatment Heads is used by trained health care professionals and is indicated for the use to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, arthritis, muscle spasm, relieving stiffness and promoting relaxation of muscle tissue.

Device Story

BioFlex Prescription Unit and Treatment Heads are low-level light therapy (LLLT) devices. System consists of electronic control unit and multiple treatment heads (LEDs/laser diodes). Device emits infrared energy to penetrate skin and underlying tissues; triggers cellular functions to relieve pain. Used in clinical settings by trained healthcare professionals. Practitioner selects treatment protocols via controller. Output is topical heating; intended to provide temporary relief of minor muscle/joint pain, muscle spasm, and arthritis-related stiffness. Benefits include surgery-free, drug-free pain management. Device does not store patient data.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and efficacy supported by factory calibration tests to verify accuracy and maintenance of calibration over device life, and adherence to Quality Systems regulations.

Technological Characteristics

Infrared lamp (21 CFR 890.5500). Multi-mode low-level light therapy system. Components: electronic control unit, multiple treatment heads (LEDs and laser diodes). Energy source: electrical power. Connectivity: standalone unit. Software-controlled protocol delivery. Sterilization: not specified.

Indications for Use

Indicated for trained health care professionals to provide topical heating via infrared energy for temporary relief of minor muscle and joint pain, arthritis, muscle spasm, stiffness, and to promote muscle relaxation.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAR 1 3 2006 ## Exhibit 19 Summary of Safety & Effectiveness 051875 2 May 2005 & L The BioFlex™ Prescription Unit and related Treatment Heads are designed for clinical applications to provide low level light therapy and and display system parameters as part of the treatment session. This device does NOT store patient data. As such, this device is a Class II device, having Regulation Number: 21 CFR part 890.5500. Product Code ILY This summary is submitted in behalf of: Meditech International Inc. 411 Horner Ave., Unit #1, Etobicoke, Ontario, Canada M8W 4W3 Voice phone number 416 251 1055 Fax phone number 416 251 2446 This summary is submitted by: ﺔ ﺍ Richard Keen Compliance Consultants 1151 Hope Street Stamford, Connecticut, 06907 Voice phone number (203) 329 2700) Fax phone number (203) 329 2345. The BioFlex™ Prescription Unit and related Treatment Heads are described as a Class II Low Level Laser treatment heads that apply energy, which penetrates the skin surface to the underlying tissues, and triggers normal cellular functions that lead to a surgery-free, drug-free, and low cost benefit to the patient, the practitioner and the health care system. The scientific concept on which this device is based is the principle that by stimulating a local area with low level laser to relieve pain. The intended use of this device is for a trained health care professional to diagnose that specific patients would benefit from this therapy and treat patients for specific ailments using specific protocols. The "Indications for Use" for this device is intended to emit energy in the infrared spectrum to provide topical heating for use when heat is indicated in the temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis, and promoting relaxation of the muscle tissue. This is a prescription only device. The labeling, instructions and user operations are designed for health care professionals. Meditech International Inc. has determined that the BioFlex™ Prescription Unit and related Treatment Heads are substantially equivalent to the performance of these predicate devices: {1}------------------------------------------------ ## Exhibit 19 Summary of Safety & Effectiveness - BioFlex"" Professional System consisted of four low level light therapy (LLT) treatment . heads (LEDs) powered by an electronic control unit that was driven by a P.C. using software, K023621 (completed in 10 April 2003). - BioFlex ™ LD-175 & LD-1200 Treatment Heads consists of two laser diode treatment heads . that were designed to work with the BioFlex" Professional System. These laser diode treatment heads were determined substantially equivalent in K041885 (completed on February 24th, 2005). A series of factory calibration tests are conducted to verify the device is accurate and calibrated (and can maintain calibration over its useful life). The BioFlex™ Prescription Unit and related Treatment Heads has benefited from design, development, testing and production procedures that conform to Quality Systems. This device is safe and effective for the application for which it is intended and has been tested to confirm safety and efficacy. Meditech International Inc. continues to search all appropriate sources for information relating to safety and effectiveness and maintains an in-house reporting system to identify adverse safety and effectiveness information and as such, applicable data is recorded for this product. #### CERTIFICATION: I hereby certify this Summary of Safety and Effectiveness applies for the above indicated device. Fred Kohn Dr. Fred Kahn President Meditech International Inc. 411 Horner Ave., Unit #1, Etobicoke, Ontario, Canada M8W 4W3 Voice 416 251 1055 416 251 2446 Fax {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and head. The eagle is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 3 2006 Meditech International, Inc. c/o Compliance Consultants Mr. Richard Keen 1151 Hope Street Stamford, Connecticut 06907 Re: K051875 Trade/Device Name: BioFlex Prescription Unit and Related Treatment Heads Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY, GEX Dated: February 10, 2006 Received: February 14, 2006 Dear Mr. Keen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Fedcral statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 – Mr. Richard Keen This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark N. Melkerson Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Exhibit 2 Indications for Use 510(K) Number (If known): K051875 ... Device Name: BioFlex™ Prescription Unit and related Treatment heads_ #### Indication For Use: The BioFlex™ Prescription Unit and related Treatment Heads is used by trained health care professionals and is indicated for the use to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, arthritis, muscle spasm, relieving stiffness and promoting relaxation of muscle tissue. The BioFlex™ Prescription Unit and related Treatment Heads is a" multi-mode" low level light treatment system configured for multiple treatment heads and powered by a flexible protocol controller that delivers various treatment protocols. | Prescription Use _XXX_ (Part 21 CFR 801 Subpart D) | AND/OR | Over - The - Counter Use (21 CFR 807 Subpart C) | |----------------------------------------------------|--------|-------------------------------------------------| |----------------------------------------------------|--------|-------------------------------------------------| (PLEASE DO NOT WRITE BELOW THIS LINE - DIVISION SIGN-OFF) CONTINUE ON ANOTHER PAGE IF NEEDED Division of General, Restorative, Concurrence of CDRH, Office of Device Evaluation (ODE) and Neurological Devices Sheet 1 of 1 KOGIATS 510(k) Number_ **Sheet 1 of 1**
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...