EXPEDIUM 4.5MM SPINE SYSTEM

K081252 · Depuy Spine, Inc. · NKB · May 30, 2008 · Orthopedic

Device Facts

Record IDK081252
Device NameEXPEDIUM 4.5MM SPINE SYSTEM
ApplicantDepuy Spine, Inc.
Product CodeNKB · Orthopedic
Decision DateMay 30, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion.

Device Story

EXPEDIUM Spine System consists of spinal fixation components (rods, screws, etc.) designed for thoracic, lumbar, and sacral stabilization. Used by surgeons in clinical settings to provide rigid immobilization as an adjunct to spinal fusion. Components are implanted to correct deformities or stabilize instabilities; mechanical fixation supports bone healing. System accepts 4.5mm rods; available in various geometries/sizes to accommodate patient anatomy. Benefit: restoration of spinal stability and support for fusion in patients with degenerative, traumatic, or oncological spinal conditions.

Clinical Evidence

Bench testing only. Performance data per ASTM F 1798 submitted to characterize component mechanical properties.

Technological Characteristics

Materials: ASTM F 136 titanium alloy and ASTM F 138 stainless steel. Components designed for 4.5mm rod fixation. Mechanical testing performed per ASTM F 1798.

Indications for Use

Indicated for skeletally mature patients requiring noncervical spinal immobilization and stabilization as an adjunct to fusion. Conditions include degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, scoliosis, kyphosis, lordosis, tumor, pseudarthrosis, and failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 5. 510(K) SUMMARY | Submitter: | DePuy Spine, Inc.<br>325 Paramount Drive<br>Raynham, MA 02767 | |------------|---------------------------------------------------------------| | | MAY 30 2008 | - Christopher Klaczyk Contact Person: Regulatory Project Manager Voice: (508) 828-2852 (508) 828-3797 Fax: cklaczyk@dpyus.jnj.com E-Mail: May 1, 2008 Date Prepared: Class III Device Class: Classification Name: Pedicle screw spinal fixation per 21 CFR §888.3070 > Spinal interlaminar fixation orthosis per 21 CFR §888.3050 Spinal intervertcbral body fixation orthosis per 21 CFR §888.3060 Classification Panel: Orthopedics FDA Panel Number: 87 | Product Code(s): | NKB, MNH, MNI, KWP, KWQ | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Proprietary Name: | EXPEDIUM Spine System | | Predicate Devices: | EXPEDIUM 4.5mm Spine System (K071495, K073364)<br>MONARCH® Spine System (K010576)<br>ISOLA® Spine System (K964007) | | Device Description: | The subject EXPEDIUM™ Spine System components are<br>designed to accept a 4.5mm rod and are available in various<br>geometries and sizes. | | Intended Use: | The EXPEDIUM Spine System is intended to provide<br>immobilization and stabilization of spinal segments in<br>skeletally mature patients as an adjunct to fusion in the<br>treatment of acute and chronic instabilities or deformities o<br>the thoracic lumbar and sacral spine. | {1}------------------------------------------------ | | The EXPEDIUM™ Spine System is intended for<br>noncervical pedicle fixation and nonpedicle fixation for the<br>following indications: degenerative disc disease (defined as<br>back pain of discogenic origin with degeneration of the disc<br>confirmed by history and radiographic studies);<br>spondylolisthesis; trauma (i.e., fracture or dislocation);<br>spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or<br>lordosis); tumor; pseudarthrosis; and failed previous fusion. | |-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Materials: | Manufactured from ASTM F 136 implant grade titanium<br>alloy and ASTM F 138 implant grade stainless steel. | | Performance Data: | Performance data per ASTM F 1798 were submitted to<br>characterize the subject EXPEDIUM™ Spine System<br>components addressed in this notification. | {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black. The seal is simple and recognizable. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 3 0 2UUd DePuy Spine % Mr. Christopher Klaczyk Regulatory Project Manager 325 Paramount Drive Raynham, Massachusetts 02767 Rc: K081252 Trade/Device Name: EXPEDIUM Spine System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, MNH, MNI, KWP, KWQ Dated: May 1, 2008 Received: May 2, 2008 Dear Mr. Klaczyk: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 - Mr. Christopher Klaczyk This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N Millican Mark N. Melkerson Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 4. INDICATIONS FOR USE STATEMENT 510(k) Number (if known): Device Name: EXPEDIUM Spine System Indications For Use: The EXPEDIUM Spine System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The EXPEDIUM Spine System is intended for noncervical pedicle fixation and nonpedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudarthrosis; and failed previous fusion. Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nélke agle for rkm Division Sign Off (Division Sign-C Division of General, Restorative, and Neurological Devices 510(k) Number k081252
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...