ACROMED CABLE/WIRE PEDICLE SCREW

K964007 · Acromed Corp. · MNH · Dec 16, 1996 · Orthopedic

Device Facts

Record IDK964007
Device NameACROMED CABLE/WIRE PEDICLE SCREW
ApplicantAcromed Corp.
Product CodeMNH · Orthopedic
Decision DateDec 16, 1996
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AcroMed Cable/Wire Pedicle Screw is designed to be utilized with the ISOLA Spine System (K951657). When used with pedicle screws, the ISOLA implants are intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained. Levels of attachment for the indication range from L3 to the sacrum. The ISOLA Spinal System, when not used with pedicle screws, is intended for hook, wire, and/or sacral/iliac screw fixation from T1 to the ilium/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, lordosis and kyphosis), tumor, fracture and previous failed surgery. As a whole, the ISOLA Spinal System is intended for T1-sacral fixation. Screw fixation is from L3-S1.

Device Story

Pedicle screw with modified grooved head and base hole; used with ISOLA Spine System. Input: 4.75mm or 6.35mm diameter rods; 18-gauge wire or Songer cable. Operation: wire/cable threaded through screw head base and secured over rod to provide spinal fixation. Used by surgeons in clinical settings for spinal stabilization. Output: mechanical fixation of spinal segments. Benefit: facilitates solid fusion in spondylolisthesis and other spinal pathologies; intended for removal post-fusion.

Clinical Evidence

Bench testing only. Static and fatigue testing demonstrate performance consistent with previously cleared components.

Technological Characteristics

Implant grade stainless steel (ASTM F1314). Wire (ASTM F1350) and Songer cables (ASTM F138). Pedicle screw with grooved head and base hole. Cancellous lengths: 25.0mm, 30.0mm, 35.0mm.

Indications for Use

Indicated for grade 3 or 4 spondylolisthesis at L5-S1 (with pedicle screws) or T1-sacrum fixation for spondylolisthesis, degenerative disc disease, deformities (scoliosis, lordosis, kyphosis), tumor, fracture, and failed surgery (non-pedicle screw use).

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} K964007 1 of 2 # ISOLA SYSTEM ## AcroMed Cable/Wire Pedicle Screw ### 510(k) SUMMARY **COMPANY:** AcroMed Corporation 3303 Carnegie Avenue Cleveland, OH 44115 DEC 16 1996 **TRADENAME:** AcroMed Cable/Wire Pedicle Screw **CLASSIFICATION:** Spondylolisthesis Spinal Fixation Device Unclassified, preamendment device system **DESCRIPTION:** The AcroMed Cable/Wire Pedicle Screw is a pedicle screw with a modified grooved head and hole through the base of the head. The ISOLA rod is placed on top of the screw so that the rod fits in the groove. The groove is sized to accommodate both the 4.75 mm and 6.35 mm diameter rods. The AcroMed Cable/Wire Pedicle Screw is intended for use with 18 gauge wire (304.8 mm in length) (K895439), which adjoins the screw to the rod. Alternatively, stainless steel Songer cable (K935481) may be used in place of 18 gauge wire. The wire or cable is threaded through the hole at the base of the screw head and the wire is twisted or the cable is crimped over the top of the rod. The screw is available in cancellous lengths of 25.0 mm, 30.0 mm, and 35.0 mm. **MATERIAL:** The AcroMed Cable/Wire Pedicle Screw is manufactured from implant grade stainless steel conforming to ASTM F1314 specifications. The wire is made from stainless steel conforming to ASTM F1350 specifications. Stainless steel Songer cables are manufactured from stainless steel conforming to ASTM F138 specifications. 1 {1} K964007 2 of 2 ## INDICATIONS: The AcroMed Cable/Wire Pedicle Screw is designed to be utilized with the ISOLA Spine System (K951657). When used with pedicle screws, the ISOLA implants are intended for use in grade 3 or 4 spondylolisthesis at L5-S1 utilizing autologous bone graft and intended to be removed after solid fusion is attained. Levels of attachment for the indication range from L3 to the sacrum. The ISOLA Spinal System, when not used with pedicle screws, is intended for hook, wire, and/or sacral/iliac screw fixation from T1 to the ilium/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, lordosis and kyphosis), tumor, fracture and previous failed surgery. As a whole, the ISOLA Spinal System is intended for T1-sacral fixation. Screw fixation is from L3-S1. ## PERFORMANCE DATA: Static and fatigue testing show the AcroMed Cable/Wire Pedicle Screw to perform consistent with previously cleared components. ## SUBSTANTIAL EQUIVALENCE: The AcroMed Cable/Wire Pedicle Screw is equivalent to other AcroMed pedicle screws as cleared under K951657 and to the Harrington Lag Screw, manufactured by Zimmer beginning in the 1960’s.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...