K073682 · Embla Systems, Inc. · MNR · Aug 8, 2008 · Anesthesiology
Device Facts
Record ID
K073682
Device Name
EMBLETTA GOLD
Applicant
Embla Systems, Inc.
Product Code
MNR · Anesthesiology
Decision Date
Aug 8, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Embletta Gold is a digital recording device designed to be used under the direction of a physician or trained technician but applied by a layperson. The Embletta Gold records multiple physiological parameters for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep-related medical disorders by trained personnel. The Embletta Gold is intended to be used for adult and pediatric (excluding neonatal and infant) studies. Note the recorder is not equipped with an alarm device and is not intended to be used as a life monitor. The intended environments include any clean, dry, dust free environment suitable for a patient's relative comfort.
Device Story
Embletta GOLD is a portable, palm-sized digital polysomnographic recorder; captures 9 channels of physiological data including respiratory, EEG, ECG/EKG, positional, user-triggered events, and oximetry. Device records data to internal memory or connected computer for at least 24 hours. Used in hospitals, clinics, or patient homes; applied by layperson under direction of physician or trained technician. Data downloaded to PC for review/analysis by sleep technologist or physician using separate application software. Device provides no alarms; not a life monitor. Assists clinicians in identifying sleep-related disorders by providing recorded physiological data for review.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1/UL60601-1 safety standards and IEC 60601-1-2 electromagnetic compatibility/immunity standards verified. Performance specifications validated through internal testing and design phase verification.
Technological Characteristics
Portable, palm-sized digital recorder; 9-channel input capability. Connectivity via data interface to PC for download. Powered by internal battery. Safety standards: IEC 60601-1/UL60601-1; EMC standards: IEC 60601-1-2. No sterilization or biocompatibility requirements.
Indications for Use
Indicated for adult and pediatric (excluding neonatal and infant) patients for the recording of physiological parameters to assist in the identification of sleep-related medical disorders. Not for use as a life monitor or alarm device.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
# KO73682
# Traditional 510(k) Notification for Embla Embletta GOLD™ Recorder
#### 510(K) SUMMARY
### SUBMITTER INFORMATION
AUG - 8 2008
| A. Company Name: | Embla Systems Inc. |
|-----------------------|-----------------------------------------------------------------------------------------|
| B. Company Address: | 11001 W. 120th Ave., Suite 200<br>Broomfield, Colorado 80021 |
| C. Company Phone: | (303) 790-1801 |
| D. Company Facsimile: | (303) 790-1810 |
| E. Company Contact: | Robert G. Schueppert<br>Manager, Regulatory Affairs<br>E-mail: bob.schueppert@embla.com |
#### PREPARATION DATE
December 21, 2007
# DEVICE IDENTIFICATION
| A. Device Trade Name: | Embletta GOLD |
|--------------------------|-----------------------------------------|
| B. Device Common Name: | Embletta GOLD polysomnographic recorder |
| C. Classification Name: | Ventilatory Effort Recorder |
| D. Regulation Number: | 21 CFR 868.2375 |
| E. Product Code: | MNR |
| F. Device Class: | Class II |
| G. Classification Panel: | Anesthesiology |
#### PREDICATE DEVICES
- A. Trade Name: Compass F10 System, 510(k) Number: K041904
- B. Trade Name: Somté System, 510(k) Number: K021176
- C. Trade Name: Medipalm 20, 510(k) Number: K031202
- D. Trade Name: Crystal 20 Monitors, 510(k) Number: K042039
{1}------------------------------------------------
HO73688
#### DEVICE DESCRIPTION
The Embletta GOLD polysomnographic recorder is intended for both clinical and research sleep applications, with 9 input channel capability of several types of physiological signals used by a physician or trained technician for the simultaneous acquisition of respiratory, electroencephalogram (EEG), electrocardiogram (ECG, EKG), positional, user triggered event and Oximetry parameters from a patient during sleep related studies.
The device will be capable of a minimum of 24 hours of recording and saving all 9 input channels either to an internal memory or to a connected computer.
The general intended environments are hospitals, institutions, sleep centers, sleep clinics, patient homes but the device should be capable of functioning in any environment where patients can sleep reasonably comfortably.
The users are the general public, trained physicians, trained sleep technicians (RPGST) or people working under the supervision of one of these professionals. The user may or may not possess knowledge of the physiological signals or test criteria.
The recorder does not provide any alarms and is not intended to be a monitor.
A trained sleep technologist (polysomnographer) and a physician would typically review and analyze the recorded signals when downloaded and presented on a PC using a separate application software program.
#### INTENDED USE
The Embletta Gold is a digital recording device designed to be used under the direction of a physician or trained technician but applied by a layperson. The Embletta Gold records multiple physiological parameters for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep-related medical disorders by trained personnel, The Embletta Gold is intended to be used for adult and pediatric (excluding neonatal and infant) studies. Note the recorder is not equipped with an alarm device and is not intended to be used as a life monitor.
The intended environments include any clean, dry, dust free environment suitable for a patient's relative comfort.
{2}------------------------------------------------
#### Traditional 510(k) Notification for Embla Embletta GOLD™ Recorder
#### COMPARISON TO PREDICATE DEVICES
The Embletta GOLD recorder is substantially equivalent in the following technological ways to the intended use and application in the identified predicate devices:
- Indications for Use O
- Target population O
- Basic design and architecture ੁ
- Where used 0
- Standards met 0
- o Data Interface
- Signals recorded o
The four predicate device data recorders, (1) the Medcare Flaga Compass F10, (2) the Compumedics Somté System, (3) the Braebon Medipalm - 20, and (4) the Cleveland Medical Devices Crystal 20 Monitor are all small palm-size portable devices that connect to one or more probes or sensors on the patient to record a variety of physiological signals. This data is then downloaded into a separate computer where the polysomnographic application software presents the signals in a format that can be read by a polysomnographic technologist or physician.
#### TESTING AND PERFORMANCE DATA
Safety tests have been completed to verify compliance with IEC 60601-1/UL60601-1 and all applicable particular standards in this family of international safety standards to ensure that there are no potential hazards on patients, operators, or the surroundings. The results of these tests demonstrate compliance with these safety standards.
Electromagnetic Compatibility tests according to IEC 60601-1-2 have been completed to ensure that no intolerable electro-magnetic disturbances are introduced. The results of these tests demonstrate compliance with this standard.
Immunity tests to IEC 60601-1-2 have been completed to ensure that the device operates satisfactorily in an electromagnetic environment. The results of these tests demonstrate compliance with this standard.
The internal testing, verification in various design phases, and validation of performance specifications have been completed. The results demonstrate the safety and effectiveness of the device in accordance with the intended use.
{3}------------------------------------------------
K073682
## Traditional 510(k) Notification for Embla Embletta GOLD™ Recorder
No other specific guidance document on performance is required for this type of device.
# BIOCOMPATIBILITY, STERILIZATION, PACKAGING, AND BENCH TESTING
Biocompatibility and Sterilization do not apply.
#### CONCLUSION
The signals and channels recorded by the Embletta GOLD recorder were compared to signals and channels recorded by the predicate device recorders. The result of the comparison is that all signals recorded by the Embletta Gold are currently recorded by one or more of the predicate devices.
It is therefore the conclusion of Embla Systems Inc. that the Embletta GOLD recorder is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Robert G. Schueppert Manager, Regulatory Affairs Embla Systems Incorporated 11001 West 120th Avenue, Suite 200 Broomfield, Colorado 80021
AUG - 8 2008
Re: K073682
Trade/Device Name: Embletta GOLD Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: July 28, 2008 Received: July 29, 2008
Dear Mr. Schueppert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Schueppert
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,
Sincerely yours,
TM Samuels
Judnd
foell
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
July 28, 2008
#### STATEMENT OF INDICATIONS OF USE
510(k) Number (if known): KO7 308 2
Device Name: Embletta GOLD
Indications For Use:
The Embletta Gold is a digital recording device designed to be used under the direction of a physician or trained technician but applied by a layperson. The Embletta Gold records multiple physiological parameters for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep-related medical disorders by trained personnel.
The Embletta Gold is intended to be used for adult and pediatric (excluding neonatal and infant) studies. Note the recorder is not equipped with an alarm device and is not intended to be used as a life monitor.
The intended environments include any clean, dry, dust free environment suitable for a patient's relative comfort.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ..............................................................................................................................................................................
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mr. Hh
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K073682
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.