K031202 · Braebon Medical Corp. · OLV · May 1, 2003 · Neurology
Device Facts
Record ID
K031202
Device Name
MEDIPALM-20
Applicant
Braebon Medical Corp.
Product Code
OLV · Neurology
Decision Date
May 1, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The BRAEBON Medical Corporation MEDIPalm-20™ is intended for use in collecting, recording and disclaying physiological data from FDA-cleared sensors during studies in both a cinical and home environment and for transmitting this data to an electrically isolated FDA-cleared polysomnographic analysis station for analysis. The target population of the MEDIPalm-20TM is all children and adult patients who are screened during sleep disorder studies. The majority of the screenings occur at a sleep laboratory but may occur at the patient's house through the use of the MEDIPalm-20TM. The MEDIPalm-20™ is intended to be used only by or on the order of a physician.
Device Story
Hand-held portable recorder; acquires/stores physiological signals from FDA-cleared sensors; records up to 16 channels including EEG, EOG, EMG, EKG, chest/abdominal effort, airflow pressure, snoring, body position, SaO2, and pulse rate. Inputs processed and stored; transmitted via USB to external computer/analysis station. Used in sleep labs, hospitals, or patient homes; operated by clinicians or patients under physician order. Real-time viewing on built-in LCD or remote computer. Data output in European Data Format (EDF) for analysis on external polysomnographic stations. Facilitates sleep disorder diagnosis by enabling ambulatory or clinical monitoring.
Clinical Evidence
Bench testing only. No clinical studies required. Functional testing verified performance specifications and signal readability. Simulation tests compared subject device to Sandman Digital predicate, demonstrating identical performance. Software verification and validation performed per FDA guidance.
Technological Characteristics
Hand-held recording device; 16-channel acquisition; USB connectivity. Powered by AA batteries or IEC 60601-1 compliant AC/DC supply. Integrates headbox, respiratory effort, airflow, and pulse oximetry into one case. No patient-contacting materials. Data output in European Data Format (EDF).
Indications for Use
Indicated for children and adult patients undergoing sleep disorder screening. Used for collection and recording of physiological signals for transfer to an FDA-cleared polysomnographic analysis station. No specific contraindications listed.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for BRAEBON MEDICAL CORPORATION. The word "BRAEBON" is in large, bold, sans-serif font on the top line. Below that, "MEDICAL CORPORATION" is in a smaller, sans-serif font. The text is black on a white background.
800 Proctor Avenue
Ogdensburg, NY
13669
**MAY - 1 2003**
120 Walgreen Drive Suite 1, RR#3 Carp, Ontario Canada, KOA 1LO Tel: 613.831.6690 Fax: 613.831.6699 www.braebon.com
031202
# 510(K) Summary
## BRAEBON Medical Corporation MEDIPalm-20™
March 20, 2003
The following safely and effectiveness summary has been prepared according to the requirement for 510(k) summaries specified in 21 CFR 807.92c.
#### 1.0 Manufacturer Name
BRAEBON Medical Corporation 120 Walgreen Drive Suite 1, RR#3 Carp. Ontario Canada, KOA 1LO
### 2.0 Proprietary Name of Device
MEDIPalm-20TM
### 3.0 Common Name of Device
Electroencephiograh/polysomnograph
### 4.0 Device Classification
Devices of this type have been class II by the Neurology Devices Panel. Devices of this classification have a classification code of OLV. Electroencephalograph (21 CFR 882.1400).
#### 5.0 Intended Use
The BRAEBON Medical Corporation MEDIPalm-20™ is intended for use in collecting, recording and disclaying physiological data from FDA-cleared sensors during studies in both a cinical and home environment and for transmitting this data to an electrically isolated FDA-cleared polysomnographic analysis station for analysis.
The target population of the MEDIPalm-20TM is all children and adult patients who are screened during sleep disorder studies. The majority of the screenings occur at a sleep laboratory but may occur at the patient's house through the use of the MEDIPalm-20TM.
The MEDIPalm-20™ is intended to be used only by or on the order of a physician.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for BRAEBON MEDICAL CORPORATION. The text is in all caps and is black on a white background. The word "BRAEBON" is in a larger font than the words "MEDICAL CORPORATION."
800 Proctor Avenue
Ogdensburg, NY
13669
120 Walgreen Drive Suite 1, RR#3 Carp, Ontario Canada, KOA 1LO Tel: 613.831.6690 Fax: 613.831.6699 www.braebon.com
# 6.0 Device Description
The MEDIPalm-20™ is a hand-held recording device capable of acquiring and storing physiological signals from FDA-cleared sensors and transmitting the physiological data to a computer through the Universal Serial Bus (USB) port.
The MEDIPalm-20™ records up to 16 channels of physiological signals: any combination of eight differential analog channels of electroencephalogram (EEG), electronyogram (EOG), electrocardiogram (EMG), and electrocardiogram (EKG) signals; as well as chest effort; abdominal effort; airflow pressure; snoring; body position; arterial oxygen saturation (SaO2); pulse rate; and SaO2 probe status. The signals can be viewed in real time on a built-in LCD or on a remote computer. The data can be downloaded after collection for analysis on any FDA-deared polysomnographic analysis station capable of reading European Data Formatted files.
The MEDIPalm-20™ is powered by either two standard AA batteries, a Medical Grade IEC 60601-1 compliant AC/DC power supply, or an electrically isolated computer that uses an IEC 60601-1 compliant isolation transformer and connects to the MEDIPalm-20™ via the MEDIPalm USB data and power cable. To protect the patient when the MEDIPalm-20™ is used with a computer, the computer and all its peripherals must be connected to an IEC 60601-1 compliant isolation fransformer. The sensors connect to the patient and the MEDIPalm-20™ does not contain any patient contacting materials.
# 7.0 Predicate Device Equivalence
We are claiming substantial equivalence to the following devices:
| Airsep Corp. | Monet | K001013 |
|-----------------------------|-------------------------------------------------------|---------|
| EB Neuro. S.P.A. | Sandman Digital | K003154 |
| Nellcor Puritan Bennett | Suzanne | K990565 |
| Nonin Medical, Inc. | PalmSAT, Model 2500 Pulse Oximeter with finger probes | K002690 |
| BRAEBON Medical Corporation | Ultima Smartbelt | K001743 |
| BRAEBON Medical Corporation | Ultima Airflow Pressure Sensor | K984431 |
# 8.0 Similarities and Differences Between Subject and Predicate Devices
| Intended Use | No difference |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications Statement | No difference |
| Method of Connection to Patient | No difference |
| Power Source | No significant difference. The MEDIPalm-20TM, like the predicate devices, is powered by<br>batteries or a Medical Grade IEC 60601-1 compliant AC/DC power supply. The<br>MEDIPalm-20TM can also be powered using the MEDIPalm Universal Serial Bus (USB)<br>data cable. When using the MEDIPalm-20TM with the computer, the computer and all its<br>peripherals must be connected to an IEC 60601-1 compliant isolation transformer to<br>protect the patient. |
| Safety Characteristics | No difference. |
| Reuse and Hygiene Characteristics | No difference. |
| Design | No difference. The MEDIPalm-20TM incorporates the headbox, respiratory effort sensor,<br>airflow pressure sensor, and pulse oximeter into one palm-sized case rather than<br>multiple cases. |
| Performance Data Conclusions | No difference. |
{2}------------------------------------------------
### 9.0 Performance Testing
Functional testing was performed to confirm that MEDIPalm-20™ is stated performance specifications and that the device output is readable. MEDIPalm-20™ passed all tests.
All software testing was performed in accordance with the May 29, 1998 "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" published by the Office of Device Evaluation. The MEDIPalm-20™ passed all tests.
All software verification and validation testing was performed in accordance with the January 11, 2002 "General Principles of Software Validation; Final Guidance for Industry and FDA Staff." The MEDIPalm-20TM passed all tests.
All environmental and electrical salety lesting was performed in accordance with the November of the may and graph Devices Guidance for 510(k) Content." The MEDIPalm-20TM passed all tests.
No clinical studies were required to support a substantal equivalence determination. The MEDIPalm-2014 was connected to a healthy person and was run to verify that readable, appropriate signals were being recorded. Simulation lests comparing the MEDIPalm-20™ and the Sandman Digital were also performed to gather comparative performance of the BRAEBON Medical Corporation MEDIPalm-20™ (subject device) was identical to that of Sandman Digital Recording System (predicate device).
#### 10.0 Conclusions
We conclude that MEDIPalm-20™ is equivalent in safely and performance to the legally marketed predicate devices. The MEDIPalm-20™ meets its stated performance specifications and the Reviewers Guidance publication specified above, and the MEDIPalm-20™ will operate safely in the intended environment and fulfill its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Braebon Medical Corporation % Mr. Neil E. Devine, Jr. Entela, Incorporated 3033 Madison Avenue, SE Grand Rapids, Michigan 49548
APR
Re: K031202
Trade/Device Name: MEDIPalm-20TM Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLV Dated (Date on orig SE ltr): April 16, 2003 Received (Date on orig SE ltr): April 16, 2003
Dear Mr. Devine:
This letter corrects our substantially equivalent letter of May 1, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Neil E. Devine, Jr.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melvin R. Fudelman, M.D.
Malvina B. Eydelman Director Division of Ophthalmic. Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510 (k) Number_K 031202-(To be assigned)
| Device Name: | MEDIPalm-20™ |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | The MEDIPalm-20™ is a portable data recorder used to<br>collect physiological signals for transfer to an FDA-cleared<br>polysomnographic analysis station capable of reading<br>European Data Format (.edf) files. |
| Target Population: | Children and adult patients who are screened during sleep<br>disorder studies |
| Environment of Use: | The majority of the screenings occur at a sleep laboratory,<br>hospital or at the patient's home. |
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K031202
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.