K073411 · Light Instruments · GEX · May 13, 2008 · General, Plastic Surgery
Device Facts
Record ID
K073411
Device Name
LITEDUO
Applicant
Light Instruments
Product Code
GEX · General, Plastic Surgery
Decision Date
May 13, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LiteDuo Dental Laser System is intended to aid during dental procedures performed in oral and maxillofacial surgery and dentistry. The LiteDuo Dental Laser System is indicated for a variety of hard tissue (tooth and bone) applications, and for the incision, excision, cutting, ablation, vaporization, and coagulation of soft tissue in oral and maxillofacial surgery and dentistry. In addition, the system is intended for light activation of bleaching materials for teeth whitening.
Device Story
Microprocessor-controlled laser system utilizing Er:YAG and Diode laser applicators. Inputs include user-selected parameters via control panel; system transforms electrical energy via high-voltage power supply, capacitor bank, and switching module to generate laser energy. Er:YAG laser energy verification module ensures output accuracy. Used in dental clinics by dentists/oral surgeons for hard/soft tissue procedures and teeth whitening. Output is laser energy delivered to target tissue; healthcare providers use visual feedback to perform surgical tasks (incision, ablation, coagulation). Benefits include precise tissue management and reduced invasiveness compared to traditional methods.
Clinical Evidence
No clinical data provided. Safety and efficacy are established through scientific research and literature regarding Er:YAG (2.94 micron) and Diode (0.98 micron) laser applications in dentistry. The device parameters (power, wavelength, pulse duration, frequency) are within ranges of previously cleared devices.
Technological Characteristics
Microprocessor-controlled laser system. Components: control panel, Er:YAG laser applicator, Diode laser applicator, high-voltage power supply, capacitor bank, switching module, diode driver, diode laser array, cooling system, Er:YAG energy verification module. Standards: 21 CFR 1040.10/11, EN 60601-1, EN 60825-1, EN 60601-2-22, IEC 60601-1-2.
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## MAY 1 3 2008
Image /page/0/Picture/1 description: The image shows a logo for "LIGHT INSTRUMENTS". The logo consists of a stylized star-like shape above the text. The text is in a simple, sans-serif font, with "LIGHT" on the top line and "INSTRUMENTS" on the bottom line in a smaller font size.
## 510(K) SUMMARY
# LITEDUO Dental Laser System 510(k) Number K O7 341
Applicant's Name: Light instruments Ltd Tavor Building 1 Yokneam Ind. Zone P.O.Box 223 Yokneam Elite 20692 Tel: (972)4-959-7531 Fax: (972)73-244-2610
| Contact Person: | Yoram Levy, Qsite<br>31 Haavoda St.<br>Binyamina, Israel 30500<br>Tel (972)4-638-8837; Fax (972)4-638-0510<br>Yoram@qsitemed.com |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------|
|-----------------|----------------------------------------------------------------------------------------------------------------------------------|
LITEDUO Dental Laser System
Classification: Name: Laser Instrument, Surgical, Powered Product Code: GEX Regulation No: 21 CFR 878.4810 Class: II Panel: General & Plastic Surgery
### Device Description:
Trade Name:
The LITEDUO Dental Laser is an advanced microprocessor-controlled laser system, composed of the following units:
- . The control panel;
- . The Er: Y AG and Diode laser applicators;
- High-voltage power supply, capacitor bank and switching module; �
- . Diode driver and Diode laser array;
- The cooling system;
- An Er: YAG laser energy verification module.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for a company called "LIGHT DIGITAL IMAGERY". The logo consists of a stylized light or star symbol above the word "LIGHT" in a sans-serif font. Below "LIGHT" are the words "DIGITAL IMAGERY" in a smaller font size. The logo appears to be in black and white.
#### Intended Use Statement:
The LiteDuo Dental Laser System is intended to aid during dental procedures performed in oral and maxillofacial surgery and dentistry.
The LiteDuo Dental Laser System is indicated for a variety of hard tissue (tooth and bone) applications, and for the incision, excision, cutting, ablation, vaporization, and coagulation of soft tissue in oral and maxillofacial surgery and dentistry. In addition, the system is intended for light activation of bleaching materials for teeth whitening. This includes the following:
#### Er:YAG Laser:
#### Hard Tissue Indications of Erbium Laser Energy:
Caries removal, Cavity preparation, Enamel etching, Enameloplasty, excavation of pits and fissures for placement of sealant,
#### Bone Indications of Erbium Laser Energy:
Contact and non-contact cutting, shaving, contouring, and resection of oral osseous tissue (bone), Apicoectomy - amputation of the root end, Cutting bone to prepare a window access to the apex (apices) of the root(s), Osseoplasty, Osteotomy, Osseous crown lengthening
#### Soft Tissue and Periodontal Indications of Erbium Laser Energy:
Excisional and incisional biopsies, Exposure of unerupted teeth, Incision drainage of abscesses, Gingival incision and excision, and a Gingivoplasties, Gingivectomies, Gingivectomy in case of hyperplasias of the gingival or excision of hyperplasias, Gingival troughing for crown impressions. Hemostasis. Implant recovery. Frenectomies and frenotomies, Fibromatosis (fibroma removal), Benign and malignant lesion removal, Operculectomy, Oral papillectomies, Reduction of gingival hypertrophy, Soft tissue crown lengthening, Preprosthetic surgery: flabby alveolar ridge, vestibuloplasty, exposure of implants, hyperplasia, epulides, papillomas, fibromatoses benign growths, Vestibuloplasty, Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
#### Endodontal Indications of Erbium Laser Energy:
Tooth preparation to obtain access to root canal, Pulpotomy, Pulpotomy as an adjunct to root canal therapy, Pulp extirpation, Root canal debridement and cleaning, Root canal preparation including enlargement.
#### Diode Laser:
Marginal and interdental gingiva and epithelial lining of free gingiva, frenectomy frenotomy, biopsy, operculectomy, implant recovery, gingivectomy, gigngivioplasty, gingival troughing crown lengthening,
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo for a company called "LIGHT". The logo consists of a circular graphic above the word "LIGHT", which is written in a stylized font. Below "LIGHT" is the text "MEASUREMENT", also in a stylized font, but smaller than the word "LIGHT". The graphic above the word "LIGHT" appears to be an abstract design.
hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery, debridement of diseased epithelial lining, incision and draining of tissue retraction, for impressions, papillectomy, abscesses. vestibuloplasty, excision of lesions, exposure of unerupted/partially erupted teeth, removal of hyperplastic tissues, treatment of aphtous ulcers. leukoplakia, sulcular debridement (removal of diseased or inflamed soft tissue, in the periodontal pocket), pulpotomy, pulpotomy as adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
Predicate Devices: Substantial equivalence to the following predicate devices is claimed:
| Device Name | 510k No | Date of approval |
|--------------|---------|------------------|
| LITETOUCH | K061966 | 27 Sep, 2006 |
| SIROLaser | K053161 | 18 Jan, 2006 |
| Opus 10 | K011769 | 30 Aug, 2001 |
| Vectra Laser | K060114 | 7 Apr. 2006 |
#### Performance Standards
LITEDUO Dental Laser System complies with U.S. Federal Performance Standards 21 CFR 1040.10 and 21 CFR 1040.11 for class IV Laser Products.
- . In addition, the device complies with the European Medical Directive 93/42/EEC concerning medical devices (Annex II) and with the following voluntary standards:
- . EN 60601-1 (Medical Electrical Equipment-Part 1: General Requirements for Safety-1. Collateral Standard: Safety Requirements for Medical Electrical Systems).
- . EN 60825-1 (Safety of laser products);
- t EN 60601-2-22 (Medical device equipment, Particular Requirements for the safety and diagnostics and therapeutic laser equipment).
- � IEC 60601-1-2 (Electromagnetic compatibility (EMC)
A detailed description appears in Section 14.
#### Summary of Clinical performance data
The safety and efficacy of Er: YAG Laser devices with wavelength of 2.94 micron and power up to 8.4 Watts, and Diode Laser devices with wavelength of 0.98 micron and power up to 7 Watts are well established
#### LITEDUO Dental Laser System - 510k Notification
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for Light Investments. The logo consists of a stylized star-like symbol above the word "LIGHT" in a sans-serif font. Below "LIGHT" is the word "INVESTMENTS" in a smaller font size.
in scientific research and literature including procedures performed in hard and soft oral tissue, Cosmetic Endodontology and Periodontology.
Due to the comprehensive animal and clinical study performed in scientific research and published in literature, and since the power, wavelength, pulse duration and frequency of the LITEDUO Dental Laser System are well within the previous cleared values, Light Instruments believes that animal and clinical studies are not required to determine the safety and efficacy of the device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 2008
Light Instruments Limited % Mr. Yoram Levy Light Instrument QA/RA Consultant Tavor Building 1 Yokneam Ind. Zone P.O. Box 223 Yokneam Elite, Israel 20692
Re: K073411
Trade/Device Name: LITEDUO Dental Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: April 10, 2008 Received: May 05, 2008
Dear Mr. Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Yoram Levy
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Light Instruments. The logo consists of a stylized geometric shape above the words "LIGHT" and "INSTRUMENTS", which are stacked on top of each other. The geometric shape appears to be a stylized representation of light or a star.
## INDICATIONS FOR USE STATEMENT
LITEDUO Dental Laser System
#### 510(k) Number (if known):
Device Name:
Indications for Use:
The LiteDuo Dental Laser System is intended to aid during dental procedures performed in oral and maxillofacial surgery and dentistry.
The LiteDuo Dental Laser System is indicated for a variety of hard tissue (tooth and bone) applications, and for the incision, excision, cutting, ablation, vaporization, and coagulation of soft tissue, in oral and maxillofacial surgery and dentistry. In addition, the system is intended for light activation of bleaching materials for teeth whitening.
This includes the following:
Er:YAG Laser:
#### Hard Tissue Indications of Erbium Laser Energy:
Caries removal, Cavity preparation, Enamel etching, Enameloplasty, excavation of pits and fissures for placement of sealant.
#### Bone Indications of Erbium Laser Energy:
Contact and non-contact cutting, shaving, contouring, and resection of oral osseous tissue (bone), Apicoectomy - amputation of the root end. Cutting bone to prepare a window access to the apex (apices) of the root(s), Osseoplasty, Osteotomy, Osseous crown lengthening
### Soft Tissue and Periodontal Indications of Erbium Laser Energy:
Excisional and incisional biopsies. Exposure of unerupted teeth, Incision and drainage of abscesses, Gingival incision and excision, Gingivoplasties, Gingivectomies, Gingivectomy in case of hyperplasias of the gingival or excision of hyperplasias, Gingival troughing for crown impressions, Hemostasis, Implant recovery, Frenectomies and frenotomies, Fibromatosis (fibroma removal), malignant lesion removal, Operculectomy, Oral Benign and papillectomies, Reduction of gingival hypertrophy, Soft tissue crown Preprosthetic surgery: flabby alveolar lengthening, ridge. vestibuloplasty, exposure of implants, hyperplasia, epulides, papillomas, fibromatoses benign growths, Vestibuloplasty, Sulcular debridement (removal of diseased or inflamed soft tissue in the
Nilreole fr mkm
(Division Sign-Off)
Division of General, Restorative. and Neurological Devices
**510(k) Number** K073411
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the word "LIGHT" in a stylized font above the word "ENTERTAINMENT". Above the word "LIGHT" is a circular design with a square in the center. The logo appears to be for an entertainment company called "Light Entertainment".
periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
#### Endodontal Indications of Erbium Laser Energy:
Tooth preparation to obtain access to root canal, Pulpotomy, Pulpotomy as an adjunct to root canal therapy, Pulp extirpation, Root canal debridement and cleaning, Root canal preparation including enlargement.
#### Diode Laser:
Marginal and interdental gingiva and epithelial lining of free gingiva, frenectomy frenotomy, biopsy, operculectomy, implant recovery, gigngivioplasty, gingival troughing gingivectomy, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery, debridement of diseased epithelial lining, incision and draining of abscesses, tissue retraction, for impressions, papillectomy, vestibuloplasty, excision of lesions, exposure of unerupted/partially erupted teeth, removal of hyperplastic tissues, treatment of aphtous ulcers, leukoplakia, sulcular debridement (removal of diseased or inflamed soft tissue, in the periodontal pocket), pulpotomy, pulpotomy as adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Office of Device Evaluation (ODE) Concurrence of CDRH. (Division Sign-off Division of General, Restorative and Neurological Devices 510(k) Number Ko?"?
LITEDUO Dental Laser System - 510k Notification
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.