K060114 · Xintec Corporation, Dba, · GEX · Apr 7, 2006 · General, Plastic Surgery
Device Facts
Record ID
K060114
Device Name
VECTRA LASER SYSTEM
Applicant
Xintec Corporation, Dba,
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscope. The device is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose, and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, neurosurgery (peripheral nervous system), pulmonary and cardiothoracic surgery, dental applications, and endovenous occlusion of the greater saphenous vein.
Device Story
Vectra Laser System delivers 980nm laser light to soft tissue via contact or non-contact modes, including endoscopic delivery. Operated by physicians in clinical or surgical settings, the system uses a footswitch for activation. Laser energy facilitates incision, excision, vaporization, ablation, hemostasis, or coagulation of soft tissue across multiple surgical specialties. By providing precise thermal tissue interaction, the device enables surgeons to perform procedures ranging from tumor excision to endovenous vein occlusion, potentially reducing bleeding and improving surgical outcomes compared to traditional methods.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use comparisons to legally marketed predicate devices.
Technological Characteristics
Diode laser system; 980nm wavelength (±10%); output power 10W-120W; continuous or pulsed operating modes; air-cooled; footswitch activation; green aiming beam; 110/220VAC power requirement.
Indications for Use
Indicated for soft tissue incision, excision, vaporization, ablation, hemostasis, or coagulation in patients undergoing otolaryngology, arthroscopic, gastrointestinal, general, dermatological, plastic, podiatric, urological, gynecological, neurosurgical, cardiac, pulmonary, or dental procedures, and for endovenous occlusion of the greater saphenous vein in patients with superficial vein reflux.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
#### 7 2006 APR
# 510(k) Summary Vectra™ Laser System and Accessories
Xintec Corporation, dba, Submitter's Name: Convergent Laser Technologies, 1660 S. Loop Road Alameda, CA 94502 510-832-2130 Phone Number: 510-832-1600 Fax Number: Marilyn M. Chou, Ph.D. Contact Person: February 23, 2006 Date Prepared: Vectra™ Laser System and Accessories Name of Device: Xintec Corporation, dba, Sponsor: Convergent Laser Technologies, 1660 S. Loop Road Alameda, CA 94502 Diode laser Classification: Ceralas Diode 980 nm Laser System (Model 100) Predicate Device:
Intended Use/Indications for Use:
The device is intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscope. The device is generally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose, and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, neurosurgery (peripheral nervous system), pulmonary and cardiothoracic surgery, dental applications, and endovenous occlusion of the greater saphenous vein.
The device is specifically indicated for use as follows:
# Ear. Nose and Throat and Oral Surgery (Otolaryngology)
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include:
Removal of benign lesions from the ear, nose and throat
Excision and vaporization of vocal cord nodules and polyps
KO 60114
{1}------------------------------------------------
1060 4
Incision and excision of carcinoma in situ Ablation and vaporization of hyperkeratosis Excision of carcinoma of the larynx Laryngeal papillomectomy Excision and vaporization of herpes simplex I and II Neck dissection
## Arthroscopy
Hemostasis, incision, excision, coagulation, vaporization and ablation of joint tissues during arthroscopic surgery. Examples include:
Menisectomy Synovectomy Chondromalacia
## Gastroenterology
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts an also with endoscopic procedures. Examples include:
Hemostasis of upper and lower GI bleeding
Excision and vaporization of colorectal carcinoma Excision of polyps
## General Surgery, Dermatology, Plastic Surgery and Podiatry
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include:
Matrixectomy Excision of neuromas Excision of periungual and subungual warts Excision of plantar warts Excision of keloids Liver resection Excision of cutaneous lesions Hemorrhoidectomy Appendectomy Debridement of decubitus ulcers Hepatobiliary tumors Mastectomy Dermabrasion Vaporization and hemostasis of capillary hemangioma Excision, vaporization and hemostasis of abdominal tumors Excision, vaporization and hemostasis of rectal pathology Pilonidal cystectomy Herniorapphy Adhesiolysis
{2}------------------------------------------------
406011-
Parathyroidectomy Laparoscopic cholecystectomy Thyroidectomy Resection of organs Debridement of wounds Photocoagulation of teleangectasia of the legs and face Photocoagulation of vascular lesions of the face and extremities Endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux Treatment of reticular veins and branch varicosities
## Urology
Excision, vaporization, incision, coagulation, ablation and homeostasis of urological, including BPH/prostatic, tissues. Examples include:
Vaporization of uretheral tumors Release of urethral stricture Removal of bladder neck obstruction Excision and vaporization of condyloma Lesions of external genitalia
## Gynecology
Ablation, excision, incision, coagulation, hemostasis and vaporization of gynechological tissue. Examples include:
Endometrial ablation Excision or vaporization of condylomata acuminate Vaporization of cervical intraepithelial neoplasia Cervical conization Menorrhagia
## Neurosurgery
Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue.
# Examples include:
Hemeostasis in conjunction with menigiomas
## Cardiac Surgery
Hemostasis and coagulation of soft tissue, including cardiac tissue
## Pulmonary Surgery
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include:
Tracheobronchial malignancy or stricture
Benign and malignant pulmonary obstruction
Endoscopic pulmonary applications
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a series of the number 1 and the number 2. There are many instances of the number 1, and only one instance of the number 2. The numbers are arranged in a single line.
## Dental Applications
Indicated for the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingival and epithelial lining of free gingival): frenectomy, frenotomy, biopsy, operculectorny, implant recovery, gingivoplasty, gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery, debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectorny, vestibulopasty, excision of lesions, exposure of unerupted/partially erupted teeth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
# Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux
Indicated for use in the endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
## Technical Characteristics
The Vectra™ Laser System and Accessories is substantially equivalent to the Optica and Odyssey Laser Systems (Xintec Corporation, dba, Convergent Laser Technologies, Alameda, CA) and Ceralas D 980 Diode Laser Systems (East Longmeadow, MA)) previously cleared for marketing under applicable 510(k) pre-market notification regulations.
Table I summarizes device specifications of the Vectra Laser Systems compared to the Xintec Corporation Optica and Odyssey Laser Systems (Xintec Corporation, dba, Convergent Laser Technologies, Alameda, CA) and Ceralas D 980 Diode Laser System (East Longmeadow, MA)) which have been previously cleared for marketing under applicable 510(k) pre-market notification regulations.
## Table I: DEVICE DESCRIPTION AND EQUIVALENCE INFORMATION
# Specifications
| | Vectra™ | Optica ™<br>510K# K901710 | Odyssey™<br>510K# 951910 | Ceralas<br>510L#K050824 |
|-------------------------|------------------------------------|---------------------------|----------------------------|-------------------------|
| Wavelength | 980nm± 10% | 1064nm± 10% | 2100nm± 10% | 980nm |
| Maximum<br>Output Power | 10/20/30W; 60W;<br>80W; 100W; 120W | 60W; 80W;<br>100W; 120W | 15W; 30W; 45W;<br>60W; 80W | 15W; 25W;<br>50W; 100W |
{4}------------------------------------------------
| Operating Modes | Continuous or<br>pulsed | Quasi-Pulsed | Pulse | Continuous or<br>pulsed |
|----------------------|-------------------------|--------------|------------|-------------------------|
| Aiming Beam | Green | Red | Green | Red |
| Laser Activation | Footswitch | Footswitch | Footswitch | Footswitch |
| Cooling | Air-cooled | Air-cooled | Air-cooled | Air-cooled |
| Weight | ~200 lbs | 250 lbs | 175 lbs | 50 lbs |
| Power<br>Requirement | 110/220VAC | 110/220VAC | 110/220VAC | 110/220VAC |
Note: These prototype specifications are subject to modifications for the final production models.
.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/5/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with its wings spread, with three lines representing the feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" are arranged in a circular pattern around the eagle.
FEB 2 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Xintec Corporation, dba c/o Marilyn M. Chou, Ph.D. Convergent Laser Technologies 1660 South Loop Road Alameda, CA 94502
Re: K060114
Trade/Device Name: Vectra™ Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II (two) Product Code: OCL, GEX Dated: February 23, 2006 Received: February 24, 2006
Dear Dr. Chou:
This letter corrects our substantially equivalent letter of April 7, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Marilyn M. Chou, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
O'Malli
Bram D. Zuekerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K#060114
## Device Name: Vectra™ Laser System
Indications for Use:
The device is intended for delivery of laser light to soft tissue in the contact and non-contact mode during surgical procedures including via endoscope. The device is gencrally indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in ear, nose, and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, neurosurgery (pcripheral nervous system), pulmonary and cardiothoracic surgery, dental applications, and endovenous occlusion of the greater saphenous vein.
The device is specifically indicated for use as follows:
### Ear, Nose and Throat and Oral Surgery (Otolaryngology)
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue from the car, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include:
Removal of benign lesions from the car, nose and throat Excision and vaporization of vocal cord nodules and polyps Incision and excision of carcinoma in situ Ablation and vaporization of hyperkeratosis Excision of carcinoma of the larynx Laryngeal papillomectomy Excision and vaporization of herpes simplex I and II Neck dissection
### Arthroscopy
Hemostasis, incision, excision, coagulation, vaporization of joint tissues during arthroscopic surgery. Examples include:
Menisectomy
Synovectomy
Chondromalacia
### Gastroenterology
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts an also with endoscopic procedures. Examples include:
Hemostasis of upper and lower GI bleeding
Excision and vaporization of colorectal carcinoma Excision of polyps
## General Surgery, Dermatology, Plastic Surgery and Podiatry
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, excision, vaporization, ablation and debulking of soft tissue, abdominal, rectal, skin, fa and dermabrasion. Examples include:
Matrixectomv Excision of neuromas
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
KOGOJI4 510(k) Number
{8}------------------------------------------------
Excision of periungual and subungual warts Excision of plantar warts Excision of keloids Liver resection Excision of cutaneous lesions Hemorrhoidectomy Appendectomy Debridement of decubitus ulcers Hepatobiliary tumors Mastectomy Dermabrasion Vaporization and hemostasis of capillary hemangioma Excision, vaporization and hemostasis of abdominal tumors Excision, vaporization and hemostasis of rectal pathology Pilonidal cystectomy Herniorapphy Adhesiolysis Parathyroidectomy Laparoscopic cholecystectomy Thyroidectomy Resection of organs Debridement of wounds Photocoagulation of teleangectasia of the legs and face Photocoagulation of vascular lesions of the face and extremities Endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial
### vein reflux
Treatment of reticular veins and branch varicosities
### Urology
Excision, vaporization, incision, coagulation, ablation and homeostasis of urological, including BPH/prostatic, tissues. Examples include:
Vaporization of uretheral tumors
Release of urethral stricture
Removal of bladder neck obstruction
Excision and vaporization of condyloma
Lesions of external genitalia
### Gynecology
Ablation, excision, incision, coagulation, hemostasis and vaporization of gynechological tissue. Examples include:
Endometrial ablation
Excision or vaporization of condylomata acuminate Vaporization of cervical intraepithelial neoplasia Cervical conization
Menorrhagia
elgndl
(Division Sign-Off Division of General, Restorative, and Neurological Devices
510(k) Number_ KOSDI14
{9}------------------------------------------------
## Neurosurgery
Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. Examples include:
Hemeostasis in conjunction with menigiomas
### Cardiac Surgery
Hemostasis and coagulation of soft tissue, including cardiac tissuc
### Pulmonary Surgery
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include:
Tracheobronchial malignancy or stricture Benign and malignant pulmonary obstruction
Endoscopic pulmonary applications
### Dental Applications
Indicated for the following applications on intraoral soft tissue (including marginal and interdental gingival and epithelial lining of free gingival): frenectomy, frenotomy, biopsv, operculectomy, implant recovery, gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgerv, debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions. papillectomy, vestibulopasty, excision of lesions, exposure of unerupted/partially erupted tecth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
### Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux
Indicated for use in the endovascular coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
, },
ﺪ
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
|-------------------------------------------------|---|
AND/OR
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
|------------------------------------------------|--|
|------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence Afice of Device Evaluation (ODE)
(Division Sign-Off) Division of General, Restorative. and Neurological Devices
510(k) Number KO6014
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.