The CPS Courier Guidewires are intended for use in the coronary and peripheral vasculature.
Device Story
Disposable, single-use guidewire; 0.014" diameter stainless steel core; distal end reduced via centreless grinding for variable stiffness; polymer-coated body; distal tip features 0.010" platinum/tungsten alloy coil for x-ray visibility; attached via solder joints. Used by physicians in clinical settings (coronary/peripheral vasculature) to facilitate catheter placement. Provides tactile feedback and radiopaque visualization to assist navigation through vasculature.
Clinical Evidence
Bench testing only. Performance testing included tensile strength, torque strength, torque response, coating performance, radiopacity, tip flexibility, catheter compatibility, accelerated aging, and biocompatibility (ISO 10993-1). Results demonstrated device performs within design parameters and is equivalent to predicate.
Technological Characteristics
0.014" stainless steel core; platinum/tungsten alloy distal coil; polymer coating. Dimensions vary by model (stiffness/tip shape). Sterilization and packaging consistent with predicate. No software or electronic components.
Indications for Use
Indicated for use in coronary and peripheral vasculature for patients requiring guidewire navigation during interventional procedures.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Brivant Medical Engineering. The word "BRIVANT" is in large, bold, sans-serif font. Below "BRIVANT" is the phrase "MEDICAL ENGINEERING" in a smaller, sans-serif font. There is a line separating the two phrases.
CPS Courier Guidewire Special 510(k)
# K073082
1/2
### 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
(1) Submitter's Name / Contact Person: Brivant Limited,
Parkmore West Business Park Galway. Ireland
JAN - 9 2008
Contact Person:
Tomas Furey, Vice President Regulatory Affairs Tel: +353 91 385037 Fax: +353 91 766598
(2) Summary Preparation Date:
October 25, 2007
(3) Device Name and Classification:
Trade Name:
Common Name:
Classification Name:
Device Classification:
CPS Courier Guidewire, Extra Firm, Straight Tip. Catalogue Number: 901008-01
CPS Courier Guidewire, Extra Firm, J Tip. Catalogue Number: 901008-02
CPS Courier Guidewire, Firm, Straight Tip. Catalogue Number: 901008-03
CPS Courier Guidewire, Firm, J Tip. Catalogue Number: 901008-04
CPS Courier Guidewire, Medium, Straight Tip. Catalogue Number: 901008-05
CPS Courier Guidewire, Medium, J Tip. Catalogue Number: 901008-06
CPS Courier Guidewire, Soft, Straight Tip. Catalogue Number: 901008-07
CPS Courier Guidewire, Soft, J Tip. Catalogue Number: 901008-08
CPS Courier Guidewire, Extra Soft, Straight Tip. Catalogue Number: 901008-09
CPS Courier Guidewire, Extra Soft, J Tip. Catalogue Number: 901008-10
Guidewire
Catheter, Guidewire
Class II, 21 CFR §870.1330
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "BRIVANT" in large, bold letters. Below "BRIVANT" is the text "MEDICAL ENGINEERING" in a smaller font. There is a line separating the two texts.
## CPS Courier Guidewire Special 510(k)
K073082
Pg 2/2
#### (4) Summary of Substantial Equivalence
The CPS Courier quidewires have minor design differences with the predicate Brivant guidewire. The Brivant Guidewire has the same intended use and fundamental technology as the predicate Brivant Guidewire. The technical features are closely similar and where there are differences these differences are minor such that they do not affect safety or effectiveness and there is also supporting data to demonstrate that the new technical features have not diminished safety or effectiveness - as a result the CPS Courier Guidewire is substantially equivalent to the Brivant Guidewire predicate.
#### (5) Description of the Device:
The Courier Guide wire is a disposable medical device designed for single use only. It consists of a 0.014" diameter stainless steel core wire, the distal end of which is reduced in diameter in a progressive fashion through a centreless grinding operation. The profile of this reduced section affords the product a reduced area of stiffness. The quide wire comes in 5 unique base models distinguished by their distal flexibility portions, the profile varied to provide a range of products of mixed stiffness. This reduced section is covered with a polymer coating of equivalent diameter to the main core body. Underneath the polymer at the distal tip there is a 0.010" platinum/tungsten alloy platinum coil attached to the distal tip of the core by means of solder joints. This provides greater visibility on x-ray equipment.
#### (6) Statement of Intended Use:
The CPS Courier Guidewires are intended for use in the coronary and peripheral vasculature.
#### (7) Technological Characteristics.
Comparisons of the proposed and predicate device show that the technological characteristics such as materials, performance characteristics, sterilization and packaging are substantially equivalent to the currently marketed device.
#### (8) Summary of Testing:
Performance testing involving the following testing has been completed tensile strength, torque strength, torque response, coating performance, radiopacity, tip flexibility, catheter compatibility testing, accelerated aging and biocompatibility testing in compliance with ISO 10993-1 has been successfully completed. The successful tests demonstrated the CPS Courier Guidewires consistently performed within their design parameters, are safe and effective and perform as well as the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird-like figure.
Public Health Service
JAN - 9 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Brivant, LTD c/o Mr. Thomas Furey Vice President Regulatory Affairs Parkmore West Business Park Galway, Ireland
Re: K073082
> Trade Name: CPS Courier Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guidewire Regulatory Class: Class II Product Code: DQX Dated: December 18, 2007 Received: December 20, 2007
Dear Mr. Furey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. Thomas Furey
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Dina R. Vuchner
ABram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
દ્વાર)
Image /page/4/Picture/1 description: The image shows the logo for Brivant Medical Engineering. The word "BRIVANT" is in large, bold, sans-serif font. Below that, in a smaller font, are the words "MEDICAL ENGINEERING". A horizontal line is above the words "MEDICAL ENGINEERING" and below the word "BRIVANT".
## CPS Courier Guidewire Special 510(k)
#### Indications for Use
510(k) Number (if known): K073082
Device Name: CPS Courier Guidewire
Indications for Use:
The CPS Courier Guidewires are intended for use in the coronary and peripheral vasculature.
Prescription Use X (Per 21 CFR 801 Subpart D)
AND/OR
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dune R. Richter
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K073082
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.