K072957 · Olympus Medical Systems Corporation · NWB · Mar 18, 2008 · Gastroenterology, Urology
Device Facts
Record ID
K072957
Device Name
EVIS EXERA II 180 SYSTEM
Applicant
Olympus Medical Systems Corporation
Product Code
NWB · Gastroenterology, Urology
Decision Date
Mar 18, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180: This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation. EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180: This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation. VISERA URETERO-RENO VIDEOSCOPE OLYMPUS XURF TYPE V: This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, endo-therapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, ureter, and kidney.
Device Story
EVIS EXERA II 180 System comprises a Xenon light source (CLV-180), video system center (CV-180), and uretero-reno videoscope (XURF-V). System captures endoscopic images via CCD sensor; processes signals for display on monitors. CLV-180 provides white light or Narrow Band Imaging (NBI) via optical filter; NBI enhances visualization of surface structures and vasculature. CV-180 performs image enhancement, recording, and transmission. Clinicians operate system in clinical settings for diagnosis and treatment. Scope ID function enables automatic compatibility recognition. Output allows real-time visualization, aiding clinical decision-making during urological procedures. Benefits include improved contrast of mucous membranes and fine capillary patterns via NBI, facilitating diagnostic accuracy.
Clinical Evidence
Bench testing only. No clinical data presented.
Technological Characteristics
System components: Xenon short-arc lamp (300W), CCD-based imaging, NBI optical filter. CV-180 processes RGB/YIC/VBS/HDTV signals. Connectivity: Scope ID function, foot switch, peripheral device integration. Protection: Class I, Type BF/CF applied parts. Materials: Standard medical-grade materials consistent with predicate endoscopes. Sterilization: Compatible with standard endoscopic reprocessing protocols.
Indications for Use
Indicated for endoscopic diagnosis and treatment within the bladder, urethra, ureter, and kidney in patients requiring urological examination. Used by clinicians in a professional healthcare setting.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
Predicate Devices
OES URETERORENOFIBERSCOPE/CHOLEDOCHOFIBERSCOPE OLYMPUS URF TYPE P2 (K912120)
VISERA CYSTO-NEPHRO VIDEOSCOPES OLYMPUS CYF TYPE V2 (K062049)
EVIS EXERA VIDEO SYSTEM CENTER OLYMPUS CV-160A (K051645)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and the letters are closely spaced together. The word is centered and takes up most of the frame.
# 510(k) SUMMARY
K072957
# EVIS EXERA II 180 SYSTEM
# 1. General Information
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
MAR 1 8 2008
| Applicant: | OLYMPUS MEDICAL SYSTEMS CORP.<br>2951-Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan<br>Establishment Registration No: 8010047 |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Laura Storms-Tyler<br>Vice President<br>Regulatory Affairs & Quality Assurance<br>Olympus Medical Equipment Services America<br>3500 Corporate Parkway, PO Box 610<br>Center Valley PA 18034-0610<br>Phone: (484) 896-5688<br>Facsimile: (484) 896-7128<br>Email:Laura.storms-tyler@olympus.com<br>Establishment Registration No: 2429304 |
| Manufacturer: | |
| Light source/Video system center: | SHIRAKAWA OLYMPUS CO., LTD.<br>3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura,<br>Nishishirakawa-gun, Fukushima, Japan 961-8061<br>Establishment Registration No: 3002808148 |
| Ureteroscope: | OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant<br>34-3 Hirai Hinode-machi, Nishitama-gun,<br>Tokyo, Japan 190-0182<br>Establishment Registration No: 3003637092 |
| Date Prepared: | June 30, 2007 |
| 2. Device Identification | |
| Device Name: | EVIS EXERA II 180 System |
| Common Name: | Endoscopic Video Imaging System |
| Class: | II |
| Regulation Number/Name: | 876.1500 Endoscope and accessories |
| Product Code: | NWB – Endoscope, accessories, narrow band spectrum<br>FGB – Ureteroscope |
| Classification Panel: | Gastroenterology/Urology |
OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-426-42-2891, TELEFAX +81-426-42-3174
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and the letters are tightly spaced. The word is centered and takes up most of the frame.
### 3. Legally Marketed Device to which Substantial Equivalence is Claimed
The following table shows the primary components of the EVIS-EXERATII 180 System and each device to which we claim substantial equivalence (predicate device).
### Table 15-1: Primary Components & Predicate Devices of the EVIS EXERA II 180 System
| Subject Device | Predicate Device | 510(k) NO. |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------|
| VISERA URETERO-RENO VIDEOSCOPE<br>OLYMPUS XURF TYPE V<br>(Part of this submission) | OES URETERORENOFIBERSCOPE/<br>CHOLEDOCHOFIBERSCOPE<br>OLYMPUS URF TYPE P2 | K912120 |
| | VISERA CYSTO-NEPHRO<br>VIDEOSCOPES<br>OLYMPUS CYF TYPE V2 | K062049 |
| EVIS EXERA II XENON LIGHT SOURCE<br>OLYMPUS CLV-180 (K062049) | EVIS EXERA XENON LIGHT SOURCE<br>OLYMPUS CLV-160A | |
| EVIS EXERA II VIDEO SYSTEM CENTER<br>OLYMPUS CV-180 (K062049) | EVIS EXERA VIDEO SYSTEM CENTER<br>OLYMPUS CV-160A | K051645 |
### 4. Device Description
The EVIS EXERA II 180 System consists of Olympus camera heads, endoscopes, video system center, light source, monitors, endo-therapy accessories and other ancillary equipment. This system is intended for endoscopic diagnosis, treatment and video observation.
The primary components of the subject system, are:
- EVIS EXERA II Xenon Light Source Olympus CLV-180
- EVIS EXERA II Video System Center Olympus CV-180
The endoscope to be added for the system, is:
- VISERA Uretero-Reno Videoscope Olympus XURF type V (hereinafter referred to as XURF-V)
The EVIS EXERA II Xenon Light Source Olympus CLV-180 is intended for endoscopic diagnosis, treatment and video observation. The CLV-180 is substantially identical to the predicate device, EVIS EXERA Xenon Light Source CLV-160A cleared under K051645 except that the device size has been slightly changed. The CLV-180 has an optional filter which allows the user to enhance endoscopic white light images by selective processing of green and blue light. This feature, referred to as Narrow Band lmaging (NBI), employs an optical filter to filter the white light spectrum, changing it from a broad band to a narrow band. Both an endoscopic image by standard white light illumination and that by NBI illumination can be obtained. The user can select either the standard observation mode by pressing the scope switch on the scope or the NBI mode switch on the CLV-180. In comparison to conventional white light observation, NBI observation provides greater visual contrast of the surface structure and fine capillary patterns of the mucous membranes.
#### OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-426-42-2891, TELEFAX +81-426-42-3174
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and the letters are closely spaced together. The word is centered and takes up most of the frame. The background is white.
23ER32P
The EVIS EXERA II Video System Center Olympus CV-180 is a video processing system intended for use with Olympus endoscopes such as the subject endoscopes. The CV-180 Video System Center contains the video signal processing technology which enables the endoscope to illuminate, enhance view, record and transmit video data of endoscopic images. The CV-180 is identical to the predicate device, EVIS EXERA Video System Center CV-160A, cleared under K051645 except that the device size has been slightly changed.
The CV-180 incorporates the following features:
1. The CV-180 is compatible with any specified Olympus flexible, both video and fiberoptic, and rigid endoscopes.
The CV-180 processes both standard white light and NBI images, generated by the CLV-180 light 2. source and captured by the endoscope's Charged Coupled Device (CCD). NBI images create an enhanced image of the tissue's vasculature.
Both the CLV-180 and CV-180 can be used with any specified Olympus flexible and rigid endoscope models, including gastroscopes, ultrasound gastroscopes, duodenoscopes, colonoscopes, sigmoidscopes, choledochoscopes, bronchoscopes, tracheal intubation scopes, transnasal esophago scopes, hysteroscopes, cystoscopes, laparo-thoracoscopes, for conventional white light endoscopy. The flexible endoscopes which are the subject of this premarket notification are cysto-nephroscope models listed in Table 15-1.
Additionally, when they are combined with the new uretero-reno videoscope XURF-V, both an endoscopic image by white light illumination and that by NBI illumination can be obtained. The user can select either the NBI mode or white light mode by pressing the scope or the filter mode switch on the CLV-180; the NBI filter in the CLV-180 is inserted on the light axis when the NBI mode is selected.
The new endoscope is basically identical to each predicate device shown in Table 15-1 in intended use. and similar in specifications, performance and materials. The CV-180 identifies an NBI-compatible scope when it is connected by using the Scope ID function provided with the scopes.
### 5. Indications for Use
#### EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180 (K062049)
This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
#### EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180 (K062049)
This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
#### VISERA URETERO-RENO VIDEOSCOPE OLYMPUS XURF TYPE V
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, endo-therapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, ureter, and kidney.
> OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-426-42-2891, TELEFAX +81-426-42-3174
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in bold, black letters. The font is sans-serif and the letters are tightly spaced. The word is centered and takes up most of the frame. The background is plain white.
### 6. Comparison of Technological Characteristics
Each primary-component of the EVIS EXERA II 180 System is basically identical to its predicate device in intended use, and similar in specifications except for the NBI function. Comparison between the subject and predicate devices is shown in Table 15-2 to 15-5.
#### Table 15-2. Comparison of Specifications Subject Device: EVIS EXERA II Xenon Light Source Olympus CLV-180 Predicate Device: EVIS EXERA Xenon Light Source Olympus CLV-160A (K051645)
| | Specification | Compliance Value |
|-------------------------------------------------------------------|-----------------------------|--------------------------------------------------------|
| Power Supply | 100-120V~±10%, 50/60 Hz±1Hz | 100-240V~±10%,<br>50/60Hz±1Hz |
| Over-current Protection | Same as PD. | Fuse type |
| Input Current | Same as PD. | 500VA (at observation) |
| Size | 383(W)×162(H)×536(D)mm | 381(W) × 162(H)×536(D)mm |
| Weight | Same as PD. | 15.4kg |
| Compatible Endoscopes | Same as PD. | Videoscope, Fiberscope,<br>Rigid scope |
| Examination Lamp | Same as PD. | Xenon short-arc lamp<br>(ozone-free)300W |
| Average Lamp Life | Same as PD. | Approximately 500 hours of<br>continuous use |
| Emergency Lamp | Same as PD. | Halogen lamp 12V 35W |
| Average Emergency Lamp<br>Life | Same as PD. | Approximately 500 hours |
| NBI Filter | Same as PD. | Provided. |
| Brightness Control | Same as PD. | Automatic and Manual |
| Automatic Exposure | Same as PD. | 17 steps |
| Photography Function | Same as PD. | Not provided. |
| Air Feeding | Same as PD. | 4 levels available<br>(off, low, mid, high) |
| Air Feeding Pump | Same as PD. | Diaphragm type pump |
| System Connector | Same as PD. | Provided |
| Foot Switch Connector | Same as PD. | Provided |
| CV Connector | Same as PD. | Provided |
| Cooling Air Direction | Same as PD. | Rear |
| Type of Protection against<br>Electric Shock | Same as PD. | Class I |
| Degree of Protection<br>against Electric Shock of<br>Applied Part | Same as PD. | TYPE BF or CF applied part<br>(Depend on applied part) |
OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-426-42-2891, TELEFAX +81-426-42-3174 4/6
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and the letters are closely spaced together. Below the word, there is a thin line of smaller text that is difficult to read due to its size and resolution.
#### Table 15-3. Comparison of Specifications
#### Subject-Dovice: EVIS EXERA-Il-Video-System-Genter-Olympus-GV-180-Predicate Device: EVIS EXERA Video System Center Olympus CV-160A (K051645) STARS (Spicifican Station) 1. 2014 Regal since at Page . 10:10 11 - 14 - 14 - 14 - 14 - 11 -2017 11:00:00 169,314 13191 Same as PD. 100-240V~±10%、50/60Hz土1Hz Power Supply Over-current Protection Same as PD. Fuse type Same as PD. 15076 Input Current 382(W)×91(H)× 370(W)×91(H)×462 (D)mm Size 490 (D)mm Weight 10 kg -------10.6 kg · Fiber/rigid scope via camera head Same as PD. Compatible Endoscopes · Videoscope Video Signal RGB:3 YIC:4 VBS:4 HDTV:1 Same as PD. Output Automatically adjusted using the white balance switch. At the time of connection with the Auto White Same as PD. scope in which Scope ID is provided, Balance compensation ાંડ performed automatically. Standard Color Color bar image Same as PD. Chart Output R: ±8 steps B: ±8 steps Color Tone Same as PD. CHROMA : ±8steps Adiustment Observation Automatic Gain Same as PD. MAX gain: 18dB Control (AGC) Edge enhancement: [OFF] [Low] [Med] [High] 4 levels available lmage Same as PD. Structure enhancement:[OFF] [Low] Enhancement [Med] [High] 4 levels available Iris Mode AUTO/PEAK Same as PD. EXPOSURE Electrical shutter Selection × 1/× 1.2 /× 1.5: 3-Mode Optical Zoom Same as PD. NBI function NB1 Observation Same as PD. The image of an external device instrument connected to this is Picture in Same as PD. displayed on the main monitor together Picture with the endoscopic image. Same as PD. Provided Communication with Scope Provided Foot Switch Connector Same as PD. Provided Same as PD. Record to Memory Card of Protection Type against Same as PD. Class I Electric Shock TYPE BF or CF applied part Degree of Protection against Same as PD. (Depend on applied part) Electric Shock of Applied Part UL60601-1 Same as PD. Applicable Standard
OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-426-42-2891, TELEFAX +81-426-42-3174 કાદ
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, black letters. The font is sans-serif and the letters are closely spaced. Below the word, there is a thin, horizontal line that spans the width of the word. The background is white.
The XURF-V is similar to the predicate device URF-P2 in specifications except for the optical system. As-for-the optical system, the XURF-V-is-similar-to-the-predicate-device-GYF-V2-Gomparison-betweenthe subject and predicate devices is shown below.
| Field of View | 90° | 90° | 120° |
|-----------------------------------------|-----------------------|--------------------------|-------------------------|
| Direction of View | 0° (Forward Viewing) | 0° (Forward Viewing) | 0°(Forward viewing) |
| Depth of Field | 2-50 mm | 1-50 mm | 3-50mm |
| Optical System | Color CCD | Image guide fiber bundle | Color CCD |
| Outer Diameter of<br>Distal End | 8.5 Fr (bullet-shape) | 9.3 Fr (φ3.1 mm) | 12.9 Fr (bullet-shaped) |
| Outer Diameter of<br>Insertion Tube | φ3.3mm | φ3.3mm | φ5.4mm<br>φ6.7mm (Max.) |
| Bending Section<br>Angulation | Up: 180°/ Down: 275° | Up: 180°/ Down: 100° | Up: 210°/ Down: 120° |
| Working Length | 670 mm | 700 mm | 380mm |
| Inner Diameter of<br>Instrument Channel | φ1.2 mm | φ1.2 mm | φ2.2mm |
#### Table 15-4. Comparison of Specifications
### 6. Conclusion
1 - 1
When-compared to the predicate device, the EVIS EXERA II 180 System does not incorporate any ... significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device.
#### OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho Hachioji-shi Tokyo, 192-8507 JAPAN TELEPHONE +81-426-42-2891, TELEFAX +81-426-42-3174 ରାଚ
{6}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Olympus Medical Systems Corporation % Ms. Laura Storms-Tyler Regulatory Affairs & Ouality Assurance Olympus America, Incorporated 3500 Corporate Parkway P.O. Box 610 CENTER VALLEY PA 18034-0610
MAR 1 8 2008
Re: K072957
Trade/Device Name: EVIS EXERA II 180 System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NWB and FGB Dated: February 20, 2008 Received: February 21, 2008
Dear Ms. Storms-Tyler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/6/Picture/10 description: The image is a circular logo with the letters "FDA" in large, bold font in the center. Above the letters, the numbers "1906 - 2006" are displayed. Below the letters, the word "Centennial" is written in a cursive font. The logo is surrounded by a circular border with text that is difficult to read due to the image quality.
Protecting and Promoting Public Health
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally. marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
**Enclosure**
{8}------------------------------------------------
## Indications for Use
K072957
510(k) Number (if known):
K072957
Device Name: EVIS EXERA II 180 S
Indications For Use:
#### EVIS EXERA II XENON LIGHT SOURCE OLYMPUS CLV-180
This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
#### EVIS EXERA II VIDEO SYSTEM CENTER OLYMPUS CV-180
This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
#### VISERA URETERO-RENO VIDEOSCOPE OLYMPUS XURF TYPE V
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, endo-therapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, ureter, and kidney.
Prescription Use V AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
oncurrence of CDRH, Office of Device Evaluation (ODE)
Hub Lema
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices,
510(k) Number K072957
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.