K051645 · Olympus Medical Systems Corporation · NWB · Oct 13, 2005 · Gastroenterology, Urology
Device Facts
Record ID
K051645
Device Name
EVIS EXERA 160A SYSTEM
Applicant
Olympus Medical Systems Corporation
Product Code
NWB · Gastroenterology, Urology
Decision Date
Oct 13, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EVIS EXERA 160A system consists of Olympus camera heads, endoscopes, video system center, light source, monitors, endo-therapy accessories and other ancillary equipment. This system is intended for endoscopic diagnosis, treatment and video observation of the upper and lower digestive tract.
Device Story
EVIS EXERA 160A System is an endoscopic video imaging platform comprising a Xenon light source (CLV-160A), video system center (CV-160A), and various flexible videoscopes. The system captures endoscopic images via a CCD sensor in the scope; the video center processes these signals for display, recording, and transmission. A key feature is Narrow Band Imaging (NBI), which uses an optical filter in the light source to enhance visualization of mucosal surface structures and capillary patterns. Used in clinical settings by physicians for diagnostic and therapeutic endoscopic procedures. The system provides real-time visual feedback, allowing clinicians to identify lesions or pathology and perform interventions using endo-therapy accessories. Benefits include improved contrast of tissue vasculature via NBI, aiding in clinical decision-making during gastrointestinal examinations.
Clinical Evidence
No clinical data generated by the applicant was provided. The submission relies on clinical literature to support the safety and efficacy of the Narrow Band Imaging (NBI) function.
Technological Characteristics
System includes 300W ozone-free Xenon short-arc lamp, CCD-based videoscopes, and video processing unit. Features NBI optical filtering for enhanced vascular visualization. Connectivity includes RGB, Y/C, VBS, and HDTV outputs. Complies with UL60601-1. Software includes Scope ID recognition and image enhancement algorithms (edge and structure).
Indications for Use
Indicated for endoscopic diagnosis, treatment, and video observation of the upper digestive tract (esophagus, stomach, duodenum) and lower digestive tract (anus, rectum, sigmoid colon, colon, ileocecal valve) in patients requiring endoscopic procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
K05/645
# OCT 1 3 2005
### 510(k) SUMMARY EVIS EXERA 160A SYSTEM
#### 1. General Information
| Applicant | Olympus Medical Systems Corp. |
|-----------|-----------------------------------------|
| | 2951 Ishikawa-cho, Hachioji-shi, |
| | Tokyo, 192-8507, Japan |
| | Establishment Registration No.: 8010047 |
Laura Storms-Tyler I Official Correspondent Executive Director Regulatory Affairs & Quality Assurance Olympus America Inc. Two Corporate Center Drive, Melville, NY 11747-9058, USA Phone: 631-844-5688 FAX: 631-844-5554 Email: Laura.storms-tyler @olympus.com Establishment Registration No.: 2429304
#### ■ Manufacturer
| Light source/Video system center: | Shirakawa Olympus Co., Ltd. |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | 3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura,<br>Nishishirakawa-gun, Fukushima, Japan 961-8061<br>Establishment Registration No.: 3002808148 |
| Gastrointestinal scope/<br>Colonoscope: | Aizu Olympus Co., Ltd. |
| | 500 Aza-Muranishi, Ooaza-Iidera, Monden-cho,<br>Aizuwakamatsu-shi, Fukushima, Japan 965-8520<br>Establishment Registration No.: 9610595 |
#### 2. Device Identification
| Device Name: | EVIS EXERA 160A System |
|--------------|---------------------------------|
| Common Name: | Endoscopic Video Imaging System |
- il ■ Class:
- Classification:
| Table 16-1. Classification of the EVIS EXERA 160A System | | | |
|----------------------------------------------------------|--|--|--|
| | | | |
| Regulation<br>Number | Regulation Name | Product Code | Classification Panel |
|----------------------|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| 876.1500 | Endoscope and accessories | FDS- Gastroscope, Gastro-Urology<br>FDF- Colonoscope, Gastro-Urology<br>KOG- Endoscope and accessories<br>GCT- Light Source, Endoscope,<br>Xenon Arc | Gastroenterology &<br>Urology |
{1}------------------------------------------------
#### 3. Legally Marketed Device to which Substantial Equivalence is Claimed
The following table shows the primary components (part of this submission) of the EVIS EXERA 160A System and each device to which we claim substantial equivalence (predicate device).
| Subject Device<br>(Part of this Submission) | Predicate Device | PD's 510(k) No. |
|-------------------------------------------------------|--------------------------------------------------------|-----------------|
| EVIS EXERA Xenon Light Source<br>Olympus CLV-160A | EVIS Universal Light Source CLV-U40 | K954451 |
| EVIS EXERA Video System Center<br>Olympus CV-160A | EVIS Video System Center CV-140 | K954451 |
| EVIS EXERA Gastrointestinal<br>Videoscope XGIF-N160Y2 | XGIF-N200H<br>Gastrointestinal Videoscope | K001766 |
| EVIS EXERA Gastrointestinal<br>Videoscope XGIF-Q160Y5 | EVIS EXERA Gastrointestinal<br>Videoscope<br>GIF-Q160Z | K011151 |
| EVIS EXERA Gastrointestinal<br>Videoscope XGIF-H160Y2 | | |
| EVIS EXERA Colonovideoscope<br>XCF-Q160W6L | EVIS EXERA Colonoscope CF-<br>Q160AL | |
| EVIS EXERA Colonovideoscope<br>XCF-H160AY2L | | K001241 |
| EVIS EXERA Colonovideoscope<br>XPCF-Q160AY2L | EVIS EXERA Colonovideoscope PCF-<br>160AL | |
Table 16-2 Primary Components & Predicate Devices of the EVIS EXERA 160A System
#### 4. Device Description
The EVIS EXERA 160A System consists of Olympus camera heads, endoscopes, video system center, light source, monitors, endo-therapy accessories and other ancillary equipment. This system is intended for endoscopic diagnosis, treatment and video observation of the upper and lower digestive tract.
The primary components of the subject system, which are part of this submission, are:
- EVIS EXERA Xenon Light Source Olympus CLV-160A, •
- EVIS EXERA Video System Center Olympus CV-160A, .
- EVIS EXERA Gastrointestinal Videoscopes XGIF-N160Y2, XGIF-Q160Y5, . XGIF-H160Y2.
- EVIS EXERA Colonovideoscopes XCF-Q160W6L, XCF-H160AY2L, ، XPCF-Q160AY2L
The EVIS EXERA Xenon Light Source Olympus CLV-160A is intended for endoscopic diagnosis, treatment and video observation. The CLV-160A is basically identical to the predicate device, EVIS Universal Light Source CLV-U40, with the exception that the CLV-160A has an optional filter which allows the user to enhance endoscopic white light images by selective processing of green and blue light. This feature, referred to as Narrow Band Imaging (NBI) employs an optical filter to filter the white light spectrum, changing it from a broad band to a narrow band. Both an endoscopic image by standard white light illumination and that by NBI illumination can be obtained. The user can select either the standard observation mode by pressing the scope switch on the scope or the NBI mode switch on the CLV-160A. In comparison to conventional white light observation provides greater visual contrast of the surface structure and fine capillary patterns of the mucous membranes.
{2}------------------------------------------------
The EVIS EXERA Video System Center Olympus CV-160A is a video processing system intended for use with Olympus endoscopes such as the subject endoscopes. The CV-160A Video System Center contains the video signal processing technology which enables the endoscope to illuminate, enhance, view, record and transmit video data of endoscopic images. The CV-160A Video System Center is equivalent to predicate Olympusdevices which have the same basic functionality. Predicate Olympus devices include the EVIS Video System Center CV-140. The Video System Center CV-140 enabled the physician to utilize a variety of Olympus flexible endoscopes, including colonoscopes, gastroscopes, sigmoidscopes, duodenoscopes and bronchoscopes. The EVIS Video System Center CV-140 contains the video signal processor technology which allowed the various endoscopes to illuminate, enhance, view, record, and transmit video data of endoscopic images.
Compared to these predicate devices, the subject CV-160A incorporates the following changes: 1. The CV-160A is compatible with any specified Olympus flexible, both video and fiberoptic, and rigid endoscope.
- 2. The CV-160A processes the NBI image, generated by the CLV-160A light source and captured by the endoscope's Charged Coupled Device (CCD), creating an enhanced image of the tissue's vasculature.
Both the CLV-160A and CV-160A can be used with any specified Olympus flexible and rigid endoscope models, including gastroscopes, ultrasound gastroscopes, duodenoscopes, colonoscopes, sigmoidscopes, choledochoscopes, bronchoscopes, rhino-laryngoscopes, tracheal intubation scopes, transnasal esophago scopes, hysteroscopes, cystoscopes, ureterorenoscopes, laparo-thoracoscopes for conventional white light endoscopy. The flexible endoscopes which are the subject of this premarket notification are gastroscope and colonoscope models listed in Table 16-2.
Additionally, when they are combined with the new Gastrointestinal Videoscopes (XGIF-N160Y2, XGIF-Q160Y5, XGIF-H160Y2), and Colonovideoscopes (XCF-Q160W6L, XCF-H160AY2L, XPCF-Q160AY2L), both an endoscopic image by white light illumination and that by NBI illumination can be obtained. The user can select either the NBI mode or normal mode by pressing the scope switch on the scope or the NBI mode switch on the CLV-160A; the NBI filter in the CLV-160A is inserted on the light axis when the NBI mode is selected.
The new endoscopes are basically identical to each predicate device shown in Table 16-2 in intended use, and similar in specifications, performance and materials. The CV-160A identifies an NB+compatible scope when it is connected by using the Scope ID function provided with the scopes.
#### 5. Indications for Use
The EVIS EXERA 160A system consists of Olympus camera heads, endoscopes, video system center, light source, monitors, endo-therapy accessories and other ancillary equipment. This system is intended for endoscopic diagnosis, treatment and video observation of the upper and lower digestive tract. The indications for use of each primary component are shown below.
#### EVIS EXERA Xenon Light Source Olympus CLV-160A
This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
#### EVIS EXERA Video System Center Olympus CV-160A
{3}------------------------------------------------
This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
### EVIS EXERA Gastrointestinal Videoscope XGIF-N160Y2
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories (such as a biopsy forceps) and other ancillary equipment for transoral or transnasal observation and surgery within the upper digestive tract (including the esophagus, stomach and duodenum).
### EVIS EXERA Gastrointestinal Videoscope XGIF-Q160Y5, XGIF-H160Y2
These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories (such as a biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach and duodenum).
### EVIS EXERA Colonovideoscope XCF-Q160W6L, XCF-H160AY2L, XPCF-Q160AY2L
These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories (such as a biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve).
#### 6. Comparison of Technological Characteristics
Each primary component of the EVIS EXERA160A System is basically identical to its predicate device in intended use, and similar in specifications except for the addition of the NBI function. Comparison between the subject and predicate devices is shown in Table 16-3 to 16-10. The clinical literatures provided in this submission supports the safety and efficacy of NBI imaging in endoscopy.
| Specifications | Suject Deve | Predicate Device CLV-U40 |
|-------------------------|-----------------------------------------|-----------------------------------------|
| Power Supply | 100-240V~ ±10%,<br>50/60Hz ± 1Hz | 120V~ ±10%,<br>50/60Hz ± 1Hz |
| Over-current Protection | Fuse type | Circuit breaker type |
| Input Current | 500VA (at observation) | 6A ( at observation)<br>8A ( at flash) |
| Size | 381(W) × 162(H) × 536(D)mm | 587(W) × 165(H) × 537(D)mm |
| Weight | 15.4kg | 20kg |
| Compatible Endoscopes | Videoscope<br>Fiberscope<br>Rigid scope | Videoscope<br>Fiberscope<br>Rigid scope |
### Table 16-3. Comparison of Specifications Subject Device: EVIS EXERA Xenon Light Source Olympus CLV-160A Predicate Device: EVIS Universal Light Source CLV-U40 (K954451)
{4}------------------------------------------------
| Examination Lamp | Xenon short-arc lamp<br>(ozone-free)300W | Xenon short-arc lamp<br>(ozone-free)300W | |
|-------------------------------------------------------------------|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Average Lamp Life | Approximately 500 hours of<br>continuous use | Approximately 500 hours of<br>continuous use | |
| Emergency Lamp | Halogen lamp 12V 35W | Halogen lamp(without mirror)<br>24V 150W | |
| Average Emergency Lamp<br>Life | Approximately 500 hours | Approximately 500 hours | |
| NBI Filter | NBI filter | NA | |
| Brightness Control | Automatic and Manual | Automatic and Manual | |
| Automatic Exposure | 17 steps | 17 steps + F | |
| Photography Function | - | Provided | |
| Air Feeding | 4 levels available<br>(off, low, mid, high) | 4 levels available<br>(off, low, mid, high) | |
| Air Feeding Pump | Diaphragm type pump | Linear type pump | |
| System Connector | Provided | Provided | |
| Foot Switch Connector | Provided | Provided | |
| CV Connector | Provided | Provided | |
| Cooling Air Direction | Rear | Left side | |
| Type of Protection against<br>Electric Shock | Class? | Class? | |
| Degree of Protection<br>against Electric Shock of<br>Applied Part | TYPE BF or CF applied part<br>(Depend on applied part) | TYPE BF applied part | |
| Applicable Standard | UL60601-1 | UL544 | |
| Specifications | Subject device CV 160A | Predicate Device CV-140 | |
| Power Supply | | 100-240V~ ±10%? 50/60Hz ±1Hz | 120V~ ±10%? 50/60Hz ± 1Hz |
| Over-current Protection | | Fuse type | Circuit breaker type |
| Input Current | | 150VA | 0.8A |
| Size | | 370(W) $\times$ 91(H) $\times$ 462 (D)mm | 450(W) $\times$ 72(H) $\times$ 465(D)mm |
| Weight | | 10.6 kg | 10 kg |
| Compatible Endoscopes | | · Fiber/rigid scope via camera<br>head<br>· Videoscope | · Gastro/colono/sigmoid/duodeno/<br>broncho videoscope<br>· Fiberscope via video converter |
| | Video Signal<br>Output | RGB:3 Y/C:4 VBS:4 HDTV:1 | RGB:3 Y/C:3 VBS:4 |
| | Auto White<br>balance | Automatically adjusted using the<br>white balance switch.<br>At the time of connection with the<br>scope in which Scope ID is<br>provided, compensation is<br>performed automatically. | Automatically adjusted using the<br>white balance switch |
| Observation | Standard Color<br>Chart Output | Color bar image | Color bar image<br>50% white |
| | Color Tone<br>Adjustment | R: ±8 steps<br>B: ±8 steps<br>CHROMA: ±8steps | R control: ±7 steps<br>B control: ±7steps<br>HSR control: ±7steps |
| | Automatic<br>Gain Control<br>(AGC) | Provided | Provided |
| | Image<br>Enhancement | Edge enhancement: [OFF] [Low]<br>[Med] [High] 4 levels available<br>Structure enhancement:[OFF] [Low]<br>[Med] [High] 4 levels available | Image enhancement: [Low] [Med]<br>[High] 3 levels available |
| | Iris Mode<br>Selection | AUTO/ PEAK EXPOSURE<br>Electrical shutter | AVE/ PEAK |
| | Optical Zoom | $\times$ 1/ $\times$ 1.2 / $\times$ 1.5: 3-Mode | |
| | NBI<br>Observation | NBI function | |
| | Picture in<br>Picture | The image of an external device<br>connected to this instrument is<br>displayed on the main monitor<br>together with the endoscopic image. | The frozen endoscopic image is<br>displayed on the sub screen when<br>the freeze mode is selected. |
| Communication with Scope | | Provided | |
| Foot Switch Connector | | Provided | |
| Record to Memory Card | | Provided | |
| Type of Protection against<br>Electric Shock | | Class? | Class? |
| Degree of Protection against | | TYPE BF or CF applied part<br>(Depend on applied part) | |
| | Electric Shock of Applied Part | | TYPE BF applied part |
| Applicable Standard | | UL60601-1 | UL544 |
:
・
:
.
{5}------------------------------------------------
### Table 16-4. Comparison of Specifications Subject Device: EVIS EXERA Video System Center Olympus CV -160A Predicate Device: EVIS Video System Center CV-140 (K954451)
{6}------------------------------------------------
### Table 16-5. Comparison of Specifications Subject Device: EVIS EXERA Gastrointestinal Videoscope XGIF-N160Y2 Predicate Device: XGIF-N200H Gastrointestinal Videoscope (K001766)
| Specifications | Subject Device<br>XGIF-N160Y2 | Predicate Device<br>XGIF-N200H |
|-----------------------------------------|-------------------------------|---------------------------------------------------------|
| Field of View | 120 ° | 120 ° |
| Depth of Field | 3-100mm | 3-100mm |
| Direction of Forward View | 0 ° | 0 ° |
| Type of CCD | Color | Monochrome |
| Outer Diameter of Distal End | 4.9mm | 6mm |
| Outer Diameter of Insertion Tube | 4.9mm | 6mm |
| Bending Section Angulation | UP: 210 °<br>DOWN:120 ° | UP: 180 °<br>DOWN: 180 °<br>RIGHT: 160 °<br>LEFT: 160 ° |
| Working Length | 1100mm | 925mm |
| Inner Diameter of Instrument<br>Channel | 2.0mm | 2.0mm |
#### Table 16-6. Comparison of Specifications Subject Device: EVIS EXERA Gastrointestinal Videoscope XGIF-Q160Y5 Predicate Device: EVIS EXERA Gastrointestinal Videoscope GIF-Q160Z (K011151)
| Specifications | Subject Device<br>XGIF-Q160Y5 | Predicate Device<br>GIF-Q160Z |
|-----------------------------------------|----------------------------------------------------|----------------------------------------------------------|
| Field of View | 140 ° | 140 ° (in WIDE position)<br>75 ° (in TELE position) |
| Depth of Field | 3-100mm | 8-100mm (in WIDE position)<br>1.5-3mm (in TELE position) |
| Direction of Forward View | 0 ° | 0 ° |
| Type of CCD | Color | Color |
| Outer Diameter of Distal End | 8.8mm | 10.8mm |
| Outer Diameter of Insertion Tube | 8.8mm | 10.9mm |
| Bending Section Angulation | UP:210 °<br>DOWN:90 °<br>RIGHT:100 °<br>LEFT:100 ° | UP: 210°<br>DOWN: 90 °<br>RIGHT:100 °<br>LEFT:100 ° |
| Working Length | 1030mm | 1030mm |
| Inner Diameter of Instrument<br>Channel | 2.8mm | 2.8mm |
| Zoom Magnification Function | No | Yes |
{7}------------------------------------------------
### Table 16-7. Comparison of Specifications Subject Device: EVIS EXERA Gastrointestinal Videoscope XGIF-H160Y2 Predicate Device: EVIS EXERA Gastrointestinal Videoscope GIF-Q160Z (K011151)
| Specifications | Subject Device<br>XGIF-H160Y2 | Predicate Device<br>GIF-Q160Z |
|--------------------------------------|------------------------------------------------------|----------------------------------------------------------|
| Field of View | 140 ° | 140 ° (in WIDE position)<br>75 ° (in TELE position) |
| Depth of Field | 2-100mm | 8-100mm (in WIDE position)<br>1.5-3mm (in TELE position) |
| Direction of Forward View | 0 ° | 0 ° |
| Type of CCD | Color | Color |
| Outer Diameter of Distal End | 9.8mm | 10.8mm |
| Outer Diameter of Insertion Tube | 9.8mm | 10.9mm |
| Bending Section Angulation | UP: 210 °<br>DOWN: 90 °<br>RIGHT:100 °<br>LEFT:100 ° | UP: 210 °<br>DOWN: 90 °<br>RIGHT:100 °<br>LEFT:100 ° |
| | Working Length | 1030mm |
| Inner Diameter of Instrument Channel | 2.8mm | 2.8mm |
| Zoom Magnification Function | No | Yes |
Table 16-8. Comparison of Specifications Subject Device: EVIS EXERA Colonovideoscope XCF-Q160W6L Predicate Device: EVIS EXERA Colonovideoscope CF-Q160AL (K001241)
| Specifications | Subject Device<br>XCF-Q160W6L | Predicate Device<br>CF-Q160AL |
|-----------------------------------------|-------------------------------|-------------------------------|
| Field of View | 170 ° | 140 ° |
| Depth of Field | 3-100mm | 3 -100mm |
| Direction of Forward View | 0 ° | 0 ° |
| Type of CCD | Color | Color |
| Outer Diameter of Distal End | 13.2mm | 12.8mm |
| Outer Diameter of Insertion Tube | 12.8mm | 12.8mm |
| Bending Section Angulation | UP: 180 ° | UP: 180 ° |
| | DOWN: 180 ° | DOWN: 180 ° |
| | RIGHT:160 ° | RIGHT:160 ° |
| | LEFT:160 ° | LEFT:160 ° |
| Working Length | 1680mm | 1680mm |
| Inner Diameter of Instrument<br>Channel | 3.7mm | 3.7mm |
| Flexibility of Insertion Tube | Adjustable by the user. | Adjustable by the user. |
{8}------------------------------------------------
#### Table 16-9. Comparison of Specifications Subject Device: EVIS EXERA Colonovideoscope XCF-H160AY2L Predicate Device: EVIS EXERA Colonovideoscope CF-Q160AL (K001241)
| Specifications | Subject Device<br>XCF-H160AY2L | Predicate Device<br>CF-Q160AL |
|-----------------------------------------|------------------------------------------------------|------------------------------------------------------|
| Field of View | 170 ° | 140 ° |
| Depth of Field | 2 -100mm | 3 -100mm |
| Direction of Forward View | 0 ° | 0 ° |
| Type of CCD | Color | Color |
| Outer Diameter of Distal End | 13.9mm | 12.8mm |
| Outer Diameter of Insertion Tube | 12.8mm | 12.8mm |
| Bending Section Angulation | UP: 180 °<br>DOWN:180 °<br>RIGHT:160 °<br>LEFT:160 ° | UP: 180 °<br>DOWN:180 °<br>RIGHT:160 °<br>LEFT:160 ° |
| Working Length | 1680mm | 1680mm |
| Inner Diameter of Instrument<br>Channel | 3.7mm | 3.7mm |
| Flexibility of Insertion Tube | Adjustable by the user. | Adjustable by the user. |
Table 16-10. Comparison of Specifications Subject Device: EVIS EXERA Colonovideoscope XPCF-Q160AY2L Predicate Device: EVIS EXERA Colonovideoscope PCF-160AL (K001241)
| Specifications | Subject Device<br>XPCF-Q160Y2L | Predicate Device<br>PCF-160AL |
|-----------------------------------------|--------------------------------|-------------------------------|
| Field of View | 140 ° | 140 ° |
| Depth of Field | 3-100mm | 3-100mm |
| Direction of Forward View | 0 ° | 0 ° |
| Type of CCD | Color | Color |
| Outer Diameter of Distal End | 11.3mm | 11.3mm |
| Outer Diameter of Insertion Tube | 11.5mm | 11.5mm |
| Bending Section Angulation | UP: 180 ° | UP: 180 ° |
| | DOWN: 180 ° | DOWN: 180 ° |
| | RIGHT:160 ° | RIGHT:160 ° |
| | LEFT:160 ° | LEFT:160 ° |
| Working Length | 1680mm | 1680mm |
| Inner Diameter of Instrument<br>Channel | 3.2mm | 3.2mm |
| Flexibility of Insertion Tube | Adjustable by the user | Adjustable by the user |
#### 6. Conclusion
When compared to the predicate devices, the EVIS EXERA 160A System does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the system.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a black and white logo of the Department of Health & Human Services. The logo is circular in shape, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the perimeter of the circle. Inside the circle is a stylized image of a bird with three lines extending from its body.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 3 2005
Ms. Laura Storms-Tyler Executive Director, Regulatory Affairs and Quality Assurance Olympus America, Inc. Two Corporate Center Drive MELVILLE NY 11747-3157
Re: K051645
Trade/Device Name: EVIS EXERA 160A System Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Codes: NWB, FDS and FDF Dated: September 2, 2005 Received: September 6, 2005
Dear Ms. Storms-Tyler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act actively subjects to the for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your I tpp. o tal), it the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the I or 's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{10}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I ins icter with anow you're ought mains of substantial equivalence of your device to a legally prematication. "The I Drivination for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific actifice for your ac ween the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | | 240-276-0115 |
|-----------------|----------------------------------|--|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | | 240-276-0120 |
| Other | | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, picaso note the regulation entitled, formation on your responsibilities under the Act from the 807.97). Tou may obtain outer general missistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indications for Use
510(k) Number (if known): Device Name: EVIS EXERA 160A System Indications For Use:
### EVIS EXERA Xenon Light Source Olympus CLV-160A
This light source has been designed to be used with Olympus endoscopes, video system center, and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
## EVIS EXERA Video System Center Olympus CV-160A.
This video system center has been designed to be used with Olympus camera heads, endoscopes, light source, monitors, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis, treatment and video observation.
### EVIS EXERA Gastrointestinal Videoscope XGIF-N160Y2
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories (such as a biopsy forceps) and other ancillary equipment for transoral or transnasal observation and surgery within the upper digestive tract (including the esophagus, stomach and duodenum).
Prescription Use_ (Part 21 CFR 801 Subpart D) ANDIOR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
of Reproductive, Abdomin and Radiological Dev 510(k) Number
Page 1 of _2__
{12}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: EVIS EXERA 160A System
Indications For Use:
### EVIS EXERA Gastrointestinal Videoscope XGIF-Q160Y4, XGIF-Q160Y5, XGIF-H160Y1, XGIE-H160Y2
These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories (such as a biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the upper digestive tract (including the esophagus, stomach and duodenum).
# EVIS_EXERA_Colonovideoscope_XCF_Q160W3L,_XCF-Q160W6L,_XCF-H160AYL,_XCE-H160AY2L, XPCF-Q160AYL, XPCF-Q160AY2L
These instruments have been designed to be used with an Olympus video system center, light source, documentation equipment, video monitor, endo-therapy accessories (such as a biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract (including the anus, rectum, sigmoid colon, colon and ileocecal valve).
Prescription Use_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Larson
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Device
Page 2 of __2_________________________________________________________________________________________________________________________________________________________________
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.