The Portland Cobalt Chrome Femoral Head has the following indications for use: · The patient should be skeletally mature. • The patient's condition should be due to one or more of the following: 1. Osteoarthritis. 2. Rheumatoid arthritis. 3. Tumor conditions involving the upper third of the femur or of the Acetabular. 4. Ankylosing spondylitis. 5. Psoriatic arthritis. 6. Old osteomyelitis - with a long infection-free period and a normal WBC, ESR and C-reactive protein. 7. Non union of femoral neck fracture or avascular necrosis of the femoral head. 8. Post-traumatic fracture/dislocation of the hip. 9. Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where low back pain or knee pain is becoming disabling. 10. Revision of an unsuccessful cemented or un-cemented hip replacement, providing sufficient bone stock is present. 11. Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, cup arthroplasty or hemi arthroplasty.
Device Story
Portland Cobalt Chrome Femoral Head is a single-component metal prosthesis; manufactured from cobalt chrome alloy; designed for use in hip arthroplasty. Device features internal 12/14 taper lock for connection to Portland Orthopaedics femoral hip stems; provides articulation with acetabular modules. Used by orthopedic surgeons in clinical/surgical settings. Device restores hip joint function; alleviates pain; improves mobility in patients with degenerative, traumatic, or revision-related hip conditions.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Cobalt chrome alloy femoral head; 12/14 taper lock interface; various sizes available; mechanical component for hip arthroplasty.
Indications for Use
Indicated for skeletally mature patients requiring hip arthroplasty due to osteoarthritis, rheumatoid arthritis, psoriatic arthritis, ankylosing spondylitis, tumor conditions of the upper femur/acetabulum, old osteomyelitis (infection-free), non-union femoral neck fracture, avascular necrosis, post-traumatic hip fracture/dislocation, or revision of previous unsuccessful hip procedures (arthrodesis, cemented/uncemented replacement, osteotomy, Girdlestone resection, cup/hemi-arthroplasty).
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Submission Summary (Full Text)
{0}------------------------------------------------
\$\qquad 063278\$
# SECTION 1
JAN 1 2 2007
# 510(K) SUMMARY
# SUBMITTED FOR:
| Company Name: | Portland Orthopaedics Limited |
|-------------------------|------------------------------------------------------------------------------------------|
| Address: | Unit 3, 44 McCauley St<br>Matraville, NSW, 2036 Australia |
| Telephone: | + 61-2-9666-8444 |
| Fax: | + 61-2-9666-8544 |
| SUBMITTED BY: | Elaine Duncan, M.S.M.E., RAC<br>President, Paladin Medical, Inc. |
| Address: | Paladin Medical, Inc.<br>PO Box 560<br>Stillwater, MN 55082,<br>United States of America |
| Telephone: | +1-715-549-6035 |
| Fax: | +1-715-549-5380 |
| CONTACT PERSON: | Elaine Duncan |
| DATE PREPARED: | October 25, 2006 |
| TRADE NAME: | Portland Cobalt Chrome Femoral Head |
| COMMON NAME: | Metal Femoral Head Prosthesis |
| DEVICE PROCODE & PANEL: | Orthopaedics 87 JDI, LWJ, LPH |
| REGULATION: | CFR § 888.3358, Class II |
#### DESCRIPTION of the DEVICE:
The Portland Cobalt Chrome Femoral Head is provided as a single component. The device is manufactured from a cobalt chrome alloy and is available in various sizes. The internal bore of the femoral head is designed to taper lock (12/14) with the external male taper on a Portland Orthopaedics femoral hip stem providing articulation with a suitable acetabular module.
#### INDICATIONS FOR USE:
510(k) Portland Orthopaedics, Ltd.
{1}------------------------------------------------
#### Page 2: 510(k) Summary
The Portland Cobalt Chrome Femoral Head has the following indications for use:
· The patient should be skeletally mature.
• The patient's condition should be due to one or more of the following:
1. Osteoarthritis.
2. Rheumatoid arthritis.
3. Tumor conditions involving the upper third of the femur or of the Acetabular.
4. Ankylosing spondylitis.
5. Psoriatic arthritis.
6. Old osteomyelitis - with a long infection-free period and a normal WBC, ESR and C-reactive protein.
7. Non union of femoral neck fracture or avascular necrosis of the femoral head.
8. Post-traumatic fracture/dislocation of the hip.
9. Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where low back pain or knee pain is becoming disabling.
10. Revision of an unsuccessful cemented or un-cemented hip replacement, providing sufficient bone stock is present.
11. Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, cup arthroplasty or hemi arthroplasty.
### SUBSTANTIAL EQUIVALENCE INFORMATION
The Portland Cobalt Chrome Femoral Head described in this submission is substantially equivalent to cobalt chrome femoral heads manufactured by Biomet and Exacted, based on similarities of design, intended use, material and manufacturing meth 173x40As demonstrated by the test results and materials information, the differences in the Portland Cobalt Chrome Femoral Head do not raise any new issues of safety and effective roris.
#### SUMMARY of TESTING:
Portland Orthopaedics, Ltd. has provided analytical and mechanical testing to demonstrate the substantial equivalence of and compliance to standards for the Cobalt Chrome Femoral Head.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Paladin Medical®, Inc. % Elaine Duncan, M.S.M.E., RAC President P.O. Box 560 Stillwater, Minnesota 55082
JAN 1 2 2007
Re: K063278
Trade/Device Name: Portland Cobalt Chrome Femoral Head Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: JDI, LPH, MEH Dated: October 25, 2006 Received: October 31, 2006
Dear Ms. Duncan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 – Ms. Elaine Duncan, M.S.M.E., RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Barbary Muchm
tor
Mark N. M lelkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Portland Cobalt Chrome Femoral Head
The Portland Cobalt Chrome Femoral Head_has the following indications for use:
· The patient should be skeletally mature.
· The patient's condition should be due to one or more of the following:
- 1. Osteoarthritis.
2. Rheumatoid arthritis.
3. Tumor conditions involving the upper third of the femur or of the Acetabular.
4. Ankylosing spondylitis.
5. Psoriatic arthritis.
6. Old osteomyelitis - with a long infection-free period and a normal WBC, ESR and C-reactive protein.
7. Non union of femoral neck fracture or avascular necrosis of the femoral head.
8. Post-traumatic fracture/dislocation of the hip.
9. Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where low back pain or knee pain is becoming disabling.
10. Revision of an unsuccessful cemented or un-cemented hip replacement, providing sufficient bone stock is present.
11. Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, cup arthroplasty or hemi arthroplasty.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X |
|-------------------------------------------------|---|
| AND/OR | |
| Over-The-Counter Use<br>(21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Barbara buclun
Division of General, Restorative, and Neurological Devices
510(k) Number K063278
12 :
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.