The PM-7000 Patient Monitor is intended to be used for monitoring, recording, and alarming of multiple physiological parameters in health care facilities by qualified health care professionals trained in the use of the equipment. The physiological parameters that can be monitored by the PM-7000 Patient Monitor are: ECG(3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Pulse Rate(PR), Carbon Dioxide (CO2), and Anaesthetic Gases(AG). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs. The target populations are human patients ranging from adult to neonate with the exception of the arrhythmia detection and ST Segment analysis, for which the target populations are adult and pediatric only. The Monitor is not intended for use in a patient's home or residence, or when it has not been ordered by a physician.
Device Story
Battery or line-powered patient monitor; acquires physiological signals (ECG, RESP, NIBP, SpO2, TEMP, IBP, CO2, AG). Signals converted to digital data; processed for alarm conditions; presented on color TFT display. Used in healthcare facilities by qualified professionals. Includes optional built-in recorder and CF memory card for data storage/waveforms. Provides audible/visual alarms for parameter excursions. Enables clinical decision-making through real-time monitoring and trend analysis.
Clinical Evidence
Bench testing only. Verification and validation conducted based on differences from predicate device (K070791). No clinical data presented.
Technological Characteristics
Battery or line-powered patient monitor. Color TFT display. Modules for ECG, RESP, NIBP, SpO2, TEMP, IBP, CO2, and AG. Includes optional recorder and CF memory card. Software-based signal processing and alarm management. Identical parameter subsystems to predicate PM-9000 Express.
Indications for Use
Indicated for monitoring, recording, and alarming physiological parameters (ECG, HR, RESP, NIBP, SpO2, TEMP, IBP, PR, CO2, AG) in healthcare facilities. Patient population: neonate to adult; arrhythmia/ST-segment analysis limited to adult and pediatric patients. Contraindicated for home use or without physician order.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
PM Series Patient Monitors, Model PM-9000 Express (K070791)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:
### Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
# NOV 2 9 2007
- Contact Person: .
Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Date Prepared: .
August 8, 2007
Name of the devices:
- Trade/Proprietary Name: PM-7000 Patient Monitor ●
- Common Name: Patient Monitor .
- Classification ●
| 21 CFR 870.1025 Arrhythmia detector and alarm<br>(including ST-segment measurement and alarm) | Class II |
|-----------------------------------------------------------------------------------------------|----------|
| 21 CFR 870.1025 Detector and Alarm, Arrhythmia | Class II |
| 21 CFR 870.1025 Monitor, ST Segment with Alarm | Class II |
| 21 CFR 870.2300 Cardiac monitor (including cardiotachometer and rate alarm) | Class II |
| 21 CFR 870.1130 Non-Invasive blood pressure measurement System | Class II |
| 21 CFR 870.1110 Blood pressure computer | Class II |
| 21 CFR 870.2900 Cable, transducer and electrode, patient, (including connector) | Class II |
| 21 CFR 880.2910 Clinical Electronic Thermometers - | Class II |
Page 1 of 3
{1}------------------------------------------------
| | Temperature Monitor with Probe | Class II |
|-----------------|-------------------------------------------------------------------------|----------|
| 21 CFR 870.2700 | Oximeter, Pulse | Class II |
| 21 CFR 870.2710 | Ear Oximeter, Pulse | Class II |
| 21 CFR 868.1400 | Carbon Dioxide Gas Analyzer | Class II |
| 21 CFR 868.1500 | Enflurane gas analyzer | Class II |
| 21 CFR 868.1500 | analyzer, gas, desflurane, gaseous-phase<br>(anesthetic concentration) | Class II |
| 21 CFR 868.1500 | analyzer, gas, sevoflurane, gaseous-phase<br>(anesthetic concentration) | Class II |
| 21 CFR 868.1500 | analyzer, gas, isoflurane, gaseous-phase<br>(anesthetic concentration) | Class II |
| 21 CFR 868.1620 | Halothane gas analyzer | Class II |
| 21 CFR 868.1700 | Nitrous Oxide gas analyzer | Class II |
| 21 CFR 868.1720 | Oxygen gas analyzer | Class II |
# Legally Marketed Predicate Devices:
PM Series Patient Monitors, Model PM-9000 Express, SHENZHEN K#070791, MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
### Description:
The PM-7000 Patient Monitor is battery or line-powered patient monitor. The patient monitor acquire the physiological signals such as ECG respiration (RESP), non-invasive blood pressure (NIBP), saturation of pulse oxygen (SpO2), temperature (TEMP), invasive pressure (IBP), carbon dioxide (CO2) and anaesthetic gases (AG). These physiological signals are converted into digital data and processed. The monitor examines the data for alarm conditions and presents them on the color TFT display. The monitor also provides advantageous operating control for the user. The optional built-in recorder, the optional CF memory card provides hard copies of all digital data and waveforms as well as tabular and graphic trend information, and storage the previous monitoring data information when power off accidentally.
# Statement of intended Use:
The PM-7000 Patient Monitor is intended to be used for monitoring, recording, and alarming of multiple physiological parameters in health care facilities by qualified health care professionals trained in the use of the equipment.
The physiological parameters that can be monitored by the PM-7000 Patient Monitor are: ECG(3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse
{2}------------------------------------------------
Oxygen Saturation (SpO2), Temperature (TEMP), Invasive Blood Pressure(IBP), Pulse Rate(PR), Carbon Dioxide (CO2), and Anaesthetic Gases(AG).
Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The target populations are human patients ranging from adult to neonate with the exception of the arrhythmia detection and ST Segment analysis, for which the target populations are adult and pediatric only.
The Monitor is not intended for use in a patient's home or residence, or when it has not been ordered by a physician.
# Comparison of Technological Characteristics:
Both the PM-7000 Patient Monitor and the predicate PM-9000 Express Patient Monitor (K070791) have the same parameters modules, measurement accessories and modules (software/hardware). That is to say, they have identical parameters subsystems. Both the devices have the main board and host software in common, and identical screen layout (user interface and display style). Parameter measurement performance and function specification of PM-7000 could be proved by the verification and validation of PM-9000 Express (K070791) because both are ensured by same modules and same software.
There are differences only in power supply, key board, and display Screen. The technological differences do not affect the safety or efficacy of the devices. And safety and effectives of PM-7000 could be ensured by the same design control procedure as PM-9000 Express.
## Testing:
By risk analysis, verification and validation caused by differences between PM-7000 and PM-9000 Express have been completely conducted according to testing methods that have been cleared in PM-9000 Express (K070791).
#### Conclusion:
The PM-7000 Patient Monitor is as safe, as effective, and performs as well as the legally marketed predicate device PM Series Patient Monitors, Model PM-9000 Express (K070791).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most complete and the bottom profile being the least complete. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular pattern.
NOV 2 9 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shenzhen Mindray Bio-Medical Electronics Co. Ltd. c/o Ms. Susan D. Goldstein-Falk mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021
Re: K072346
> Trade/Device Name: PM-7000 Patient Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: November 1, 2007 Received: November 2, 2007
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Blymmumor for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
072346
510(k) Number (if known):
Device Name: PM-7000 Patient Monitor
Indications for Use:
The PM-7000 Patient Monitor is intended to be used for monitoring, recording, and alarming of multiple physiological parameters in health care facilities by qualified health care professionals trained in the use of the equipment.
The physiological parameters that can be monitored by the PM-7000 Patient Monitor are: ECG(3-lead or 5-lead or 12-lead selectable), arrhythmia detection, ST Segment analysis, Heart Rate(HR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Pulse Oxygen Saturation (SpO₂), Temperature (TEMP), Invasive Blood Pressure(IBP), Pulse Rate(PR), Carbon Dioxide (CO2), and Anaesthetic Gases(AG).
Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The target populations are human patients ranging from adult to neonate with the exception of the arrhythmia detection and ST Segment analysis, for which the target populations are adult and pediatric only.
The Monitor is not intended for use in a patient's home or residence, or when it has not been ordered by a physician.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page ot
Bymmamo
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.