JOSTRA PEDIATRIC ARTERIAL PREFUSION CANNULAE

K072172 · Maquet Cardiopulmonary, AG · DWF · Aug 31, 2007 · Cardiovascular

Device Facts

Record IDK072172
Device NameJOSTRA PEDIATRIC ARTERIAL PREFUSION CANNULAE
ApplicantMaquet Cardiopulmonary, AG
Product CodeDWF · Cardiovascular
Decision DateAug 31, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4210
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The Jostra Pediatric Arterial Perfusion Cannulae are designed to be used as perfusion cannulae to return arterial blood from the extracorporeal circuit to the patient during cardiopulmonary bypass up to 6 hours or less.

Device Story

Jostra Pediatric Arterial Perfusion Cannulae are vascular catheters used during cardiopulmonary bypass procedures. The device functions as a conduit to return arterial blood from an extracorporeal circuit to the patient. It is intended for short-term use, specifically 6 hours or less. The device is operated by clinical staff in a surgical or clinical setting. The modification described involves a change in the material used for the white depth markings on the cannula. The device is evaluated for biocompatibility and marker adhesiveness to ensure safety and effectiveness. No changes were made to the fundamental scientific technology, design, or method of operation compared to the predicate device.

Clinical Evidence

Bench testing only. Evaluation included biocompatibility testing and assessment of the adhesiveness of the position marker. No clinical data was required or provided.

Technological Characteristics

Cardiopulmonary bypass vascular catheter. Modification involves a change in the material composition of the white depth markings. Device maintains original design, dimensions, and fundamental scientific technology. Biocompatibility and marker adhesiveness verified via bench testing.

Indications for Use

Indicated for pediatric patients requiring cardiopulmonary bypass for up to 6 hours or less to return arterial blood from the extracorporeal circuit to the patient.

Regulatory Classification

Identification

A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a sequence of handwritten characters, which appear to be 'K072172'. The characters are written in a bold, somewhat rough style, with varying stroke thicknesses. The 'K' is distinct with its angled lines, followed by the rounded '0' and the numbers '7', '2', '1', '7', and '2'. ## MAQUET #### Special 510(k): Device Modification: Jostra Pediatric Arterial Cannulae Fax: July 31, 2007 Arterial Cannulae cannula or tubing 510(k) SUMMARY AUG 3 1 2007 SUBMITTER: Maquet Cardiopulmonary AG Hechinger Strasse 38 72145 Hirrlingen, Germany Phone: (011) 49 7478 921- 151 Jostra Pediatric Arterial Cannulae (011) 49 7478 921- 400 Cardiopulmonary bypass vascular catheter, Katrin Schwenkglenks CONTACT PERSON: DATE PREPARED: DEVICE TRADE NAME: COMMON/USUAL NAME CLASSIFICATION NAME PREDICATE DEVICES OR LEGALLY MARKETED DEVICES Jostra Pediatric Arterial Cannulae #### DEVICE DESCRIPTION / INDICATONS FOR USE STATEMENT The Jostra Pediatric Arterial Perfusion Cannulae are designed to be used as perfusion cannulae to return arterial blood from the extracorporeal circuit to the patient during cardiopulmonary bypass up to 6 hours or less. #### STATEMENT OF TECHNICAL CHARACTERISTICS COMPARISON Jostra Pediatric Arterial Cannulae are identical to the originally cleared Jostra Pediatric Arterial Cannulae, with the only exception that the Jostra Pediatric Arterial Cannulae uses a different material for the white depth markings. Besides this difference the devices are the same in design, intended use, method of operation, components, packaging, and fundamental scientific technology. #### RISK ANALYSIS The risk analysis method used to assess the impact of the modification was done acc. to the logic of a Failure Modes and Effects Analysis (FMEA). The design verification tests were performed as a result of this risk analysis assessment. {1}------------------------------------------------ # MAQUET All possible risks for the user and the patient related to the design change (material change) have been assessed by evaluation or testing acc. to the risk analysis for the Pediatric Arterial Perfusion Cannulae. The following hazards have been addressed: General hazards related to - · biocompatibility Hazards related to manufacturing - · adhesiveness of the position marker The evaluation and test results do not show any kind of risk potential for the user and/or the patient. Based on the test results and evaluation the Pediatric Arterial Perfusion Cannulae are safe and effective for its intended use and are substantially equivalent to the named predicate device. The modification does not alter the fundamental scientific technologies of the Pediatric Arterial Perfusion Cannulae. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 3 1 2007 Maquet Cardiopulmonary AG c/o Ms. Katrin Schwenkglenks Regulatory Affairs Manager Hechinger Strasse 38 72145 Hirrlingen, Germany Re: K072172 > Jostra Pediatric Arterial Perfusion Cannulae Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheter, cannula or tubing Regulatory Class: Class II Product Code: DWF Dated: July 31, 2007 Received: August 6, 2007 Dear Ms. Schwenkglenks: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2- Ms. Katrin Schwenkglenks Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Dwna R. Lachner Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known): Ko72172 Device Name: Jostra Pediatric Arterial Perfusion Cannulae Indications for Use: The Jostra Pediatric Arterial Perfusion Cannulae are designed to be used as perfusion cannulae to return arterial blood from the extracorporeal circuit to the patient during cardiopulmonary bypass up to 6 hours or less. . Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (Please do not write BELOW THIS LINE-Continue on another page OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) una R. Vahner Division Sign Off (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number Koral 72 Page l of i (Posted November 13, 2003)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...