K071815 · Nomir Medical Technologies, Inc. · GEX · Nov 28, 2007 · General, Plastic Surgery
Device Facts
Record ID
K071815
Device Name
NOVEON, MODEL LS1100-01-0968
Applicant
Nomir Medical Technologies, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 28, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Noveon™ (Model LS1100-01-0968) delivers continuous wave laser light in the contact or non-contact mode during surgical procedures of the skin, subcutaneous tissues and nasal passages in dermatology, plastic surgery, podiatry and otolaryngology. The device is indicated for use in applications requiring incision, excision, vaporization, hemostasis, or coagulation of soft tissue.
Device Story
Noveon™ is a Class IV dual-wavelength (870 nm and 930 nm) diode laser surgical instrument. Operated by medical professionals in clinical settings, the device delivers continuous-wave laser light via a flexible optical fiber and end piece. The control unit features touch-sensitive screens for power adjustment (1W to 8W) and a foot pedal for activation. Used for incision, excision, vaporization, hemostasis, and coagulation of soft tissue. The device provides precise thermal effects on target tissues, enabling surgical procedures in dermatology, plastic surgery, podiatry, and otolaryngology. Benefits include controlled tissue ablation and coagulation, facilitating surgical precision and patient recovery.
Clinical Evidence
Bench testing and animal study only. A surgical study on a live Yorkshire pig (n=130 laser passes) demonstrated successful incision, coagulation, debridement, curettage, and excision without adverse events. Bench testing on porcine skin, muscle, and liver samples compared 870 nm and 930 nm wavelengths against 940 nm predicates, measuring ablation and coagulation zones to confirm comparable tissue effects.
Technological Characteristics
Class IV diode laser; dual wavelengths (870 nm, 930 nm); continuous-wave output (1W-8W). Components: control unit, foot pedal, flexible optical fiber. Biocompatibility: ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization/irritation). Electrical safety: IEC 60601-1, 60601-1-1, 60601-1-2; 47 CFR Part 15, Subpart B. Laser safety: 21 CFR 1040.10, 1040.11.
Indications for Use
Indicated for patients requiring soft tissue incision, excision, vaporization, hemostasis, or coagulation in dermatology, plastic surgery, podiatry, and otolaryngology. Applicable to skin, subcutaneous tissues, and nasal passages.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Nomir Noveon™ (Model LS1100-01-0968)- 510(k) Summary
K071815
# 510(k) Summary
### Noveon™ (Model LS1100-01-0968) Dual Wavelength Laser Instrument
【日本】【KITC ミニー " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "
November 19, 2007
#### Submittal information:
Post-approval contact: Richard Burtt Chief Executive Officer Nomir Medical Technologies 275 Grove Street, Suite 2-400 Newton, MA 02466 Phone: 617-901-6637 Fax: 617-663-4801
NOV 2 8 2007
#### Device name and classification
510(k) Number: K071815 Proprietary Name: Noveon™ (Model LS1100-01-0968) Common Name: Medical Diode Laser Classification Name: Class IV Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Classification Panel: General and Plastic Surgery Devices CFR Section: 21 CFR 870.4810 Class: II Product Code: GEX
#### Predicate Devices
- Ceralas D 810 Diode Laser System, K032864, by Biolitec, Inc .
- LaserPro 810, 940, and 980 Dioxide Laser Systems, K040294, by Photomedex, Inc .
- . Vectra Laser System, K060114, by Xintec Corporation
- . BWF-5 Medical Laser Series, K062363, by B&W Tek Inc
- PhoTex15 Diode Laser Series: 980, 810, 940, K060304, by BioTex Inc .
- Q YAG 5 Nd: YAG Laser System, K061436, by Palomar Medical Technologies, Inc .
- . Medlite C3 Q Switched Nd: YAG Laser, K011677, by Continuum Electro-Optics, Inc (now Hoya ConBio, Inc)
#### Device Description
The Noveon™ (Model LS1100-01-0968) is a Class IV Laser Surgical Instrument for use in General and Plastic Surgery and Dermatology. The device consists of the following components and accessories:
- 1. Laser Source and Control Unit: Two Class IV laser diodes, each providing continuous-wave laser light, one at the 870 mm and one at 930 nm wavelength. The power output of each laser is selectable from 1W to 8W in 0.1 W increments. There are two touch sensitive screens to control the optical power output of each laser and a separate emergency on/off button.
- 2. Foot Pedal Switch: turns the lasers on/off when the lasers are enabled.
### Confidential and Proprietary to Nomir Medical Technologies
{1}------------------------------------------------
- 3. Flexible Optical Fiber and End Piece: designed to be lightweight for holding the distal end of the fiber comfortably in the hand to perform procedures.
#### Intended Use
The Noveon™ (Model LS1100-01-0968) delivers continuous wave laser light in the contact or non-contact mode during surgical procedures of the skin, subcutaneous tissues and nasal passages in dermatology, plastic surgery, podiatry and otolaryngology. The device is indicated for use in applications requiring incision, excision, vaporization, hemostasis, or coagulation of soft tissue.
#### Performance Data
#### Performance Standards
- The Noveon™ (Model LS1100-01-0968) complies with applicable performance standards . for light emitting products as outlined in 21 CFR 1040.10 and 21 CFR 1040.11.
- . The device has been tested and verified by Intertek ETI Semko as conforming to 47 CFR Part 15, Subpart B Unintentional Radiator Class A. Prior to introduction into the US market, the device will be tested and verified to conform to the applicable voluntary electrical equipment standards; IEC 60601-1, IEC 60601-1-1, and IEC 60601-1-2.
- The device also conforms to the standards for biocompatibility as given by Cytoxicity: ISO ● 10993-5, Sensitization: ISO 10993-10, and Irritation: ISO 10993-10.
#### Performance Testing
A surgical study with a live Yorkshire pig was conducted to test the capability of the Noveon™ (Model LS1100-01-0968) to perform incision, contact coagulation, non-contact coagulation, debridement, curettage and excision. Results from this study demonstrate the device satisfactorily performed these procedures when used by a medical professional. Importantly, there were no adverse events that were noted during the performance of any of these demonstrations. Altogether there were 130 passes of the laser made during this demonstration, of which 127 were made in the contact mode: incision (5), contact coagulation (4), debridement (19), curettage (31), and excision (68).
Using tissue samples of porcine skin, muscle, and liver we tested individual samples with exposures to 940 nm, 930 nm, and 870 nm. In addition, we tested samples with 870 nm and 930 nm in three different optical power combinations. The testing included both contact and noncontact (free beam) mode. The testing was performed in triplicate. For each test the width and depth of the zone of ablation, the zone of coagulation (necrosis) and the zone of total destruction were recorded. These data substantiate that the difference in tissue effect was comparable to the 940 nm predicate device and any differences were not of clinical significance.
#### Substantial Equivalence
The Noveon™ (Model LS1100-01-0968) share the same indications for use, similar laser and electronics design features, safety and functional features, and therefore are substantially equivalent to the predicate devices listed above.
#### Safety and Effectiveness
The Noveon™ (Model LS1100-01-0968) is designed in accordance with both mandatory and voluntary Standards ensuring it is both safe and effective for the medical procedures indicated above. No new clinical indications are to be provided by the introduction of the device as compared to the predicate devices, identified above, which have previously demonstrated clinical effectiveness.
## Confidential and Proprietary to Nomir Medical Technologies
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" are arranged in a circular pattern around the bird.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 8 2007
Nomir Medical Technologies, Inc % Richard Burtt Chief Executive Officer 275 Grove Street, Suite 2-400 Newton, Massachusetts 02466
Re: K071815
Trade/Device Name: Noveon™ Model LS1100-01-0968 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 19, 2007 Received: November 20, 2007
Dear Mr.Burtt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2 -- Mr. Richard Burtt
This letter will allow you to begin marketing your device as described in your Section 510(k) r ms leter witification. The FDA finding of substantial equivalence of your device to a legally premaince noticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to ( premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours,
Mark N Mulhearn
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use Statement L1
510(k) Number (if known): K071815
Device Name: Noveon™ (Model LS1100-01-0968)
Indications For Use:
The Noveon™ (Model LS1100-01-0968) delivers continuous wave laser light in the contact or non-contact mode during surgical procedures of the skin, subcutaneous tissues and nasal passages in dermatology, plastic surgery, podiatry and otolaryngology. The device is indicated for use in applications requiring incision, excision, vaporization, hemostasis, or coagulation of soft tissue.
Rih
(Division Sig-Off)
Division of General, Restorative, and Neurological Devices
**510(k) Numbe**r: 2601818
Prescription Use X AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.