FOTONA FIDELIS III ER:YAG/ND:YAG LASER SYSTEM FAMILY
K070355 · Fotona D.D. · GEX · Mar 4, 2008 · General, Plastic Surgery
Device Facts
Record ID
K070355
Device Name
FOTONA FIDELIS III ER:YAG/ND:YAG LASER SYSTEM FAMILY
Applicant
Fotona D.D.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 4, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fidelis III Er: YAG laser, and its accessories, are in dentistry, dermatology and other surgical areas in the following procedures: In dentistry, for: Caries removal, Cavity Preparation, Enamel Roughening - Intra-oral soft tissue surgery (incision, excision, ablation, coagulation) - -Leukoplakia - Removal of fibromae - Removal granulated tissue - -Sulcular debridement - Apicectomy surgery - Cutting, shaving, contouring and resection of oral osseous tissue (bone) - Osteotomy,osseous crown lengthening, osteoplasty - Tooth preparation to obtain access to root canal debridement and cleaning, root canal preparation including enlargement - -Pulp extirpation - Pulpotomy as adjunct to root canal retreatment - In dermatology and other surgical areas, for: - Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, skin tags, keratoses, verrucae, and skin resurfacing - ENT Surgery Indications:ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia - Oral/Maxillofacial Indications: Oral and glossal lesions, gingivectomy - General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation - Podiatry Indications: warts, plantar verrucae, large mosiac verrucae, matrixectomy - The Fidelis III Nd: YAG laser, and its accessories, are intended for use in dentistry, dermatology and other surgical areas in the following procedures: In dentistry, for: - Excisional and incisional biopsies - Excision and vaporization of herpes simplex I and II - Exposure of unerupted teeth - Fibroma removal - Frenectomy and frenotomy - -Gingival troughing for crown impressions - Gingivectomy - Gingivoplasty - Gingival incision and excision - Hemostasis - Implant recovery - -Incision and drainage of abscess - Laser assisted uvulopaletoplasty (LAUP) - -Operculectomy - -Oral papillectomies - -Reduction of denture hyperplasia - Reduction of gingival hypertrophy - Removal of post-surgical granulations - Soft tissue crown lengthening - Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility) - Tissue retraction for impression - Treatment of aphtous ulcers - Vestibuloplasty - Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium) - Pulpotomy and pulpotomy as an adjunct to root canal therapy - Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy - In dermatology and other surgical areas, for: - Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of -PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin - Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, - port wine stains, hemaongiomae, warts, telengiectasiae, rosacea, venus lake, leg veins and spider veins - Treatment of wrinkles - Treatment of mild to moderate inflammatory acne vulgaris
Device Story
Dual-wavelength laser system (Er:YAG 2940 nm and Nd:YAG 1064 nm) housed in single unit; utilizes flashlamp-pumped crystals; red diode aiming beam (650 nm). Er:YAG energy delivered via articulated arm/optical handpiece; Nd:YAG energy delivered via optical fiber/handpiece. Features Variable Square Pulse (VSP) technology for pulse duration/energy control. Used by clinicians in dental, dermatology, and surgical settings. Provides precise tissue ablation, incision, coagulation, and vaporization. Benefits include improved treatment precision, patient comfort, and safety. Output visualized by clinician via direct observation during procedure; clinical decision-making based on real-time tissue response.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance specifications compared to predicate devices.
Technological Characteristics
Flashlamp-pumped Er:YAG (2940 nm) and Nd:YAG (1064 nm) laser sources. Er:YAG: up to 1.5 J, 1000 us pulse, 50 Hz, 20 W max. Nd:YAG: up to 300 J/cm² (aesthetic) or 15 W (dental), 320 us pulse, 100 Hz. VSP (Variable Square Pulse) technology for pulse control. Articulated arm and optical fiber delivery. Red diode aiming beam (650 nm).
Indications for Use
Indicated for dental, dermatological, and surgical procedures including soft/hard tissue surgery, caries removal, cavity preparation, periodontal treatments, hair reduction (Fitzpatrick I-VI), vascular/pigmented lesion treatment, acne vulgaris, and wrinkle treatment. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of a human figure in profile, with three faces overlapping to symbolize the department's focus on health and well-being. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 26, 2017
Fotona d.d. Stojan Trost Quality Assurance and Regulatory Affairs Manager Stegne 7 1210 Ljubljana Slovenia
Re: K070355
Trade/Device Name: Fotona Fidelis III Er:YAG/Nd:YAG Laser System Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: February 22, 2008 Received: February 27, 2008
Dear Stojan Trost:
This letter corrects our substantially equivalent letter of December 16, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
K070355 pg 1 of
# 4. Indications for Use Statement
510(k) Number (if known):
Device Name: Fotona Fidelis III Er: YAG/ Nd: YAG Laser System Family
Indications For Use:
Er: Y AG laser (2940 nm wavelength) in dentistry:
- Intra-oral soft tissue surgery (incision, excision, ablation coagulation) -
- -Leukoplakia
- = Pulpotomy as adjunct to root canal retreatment
- Pulp extirpation -
- Removal of fibromae -
- Removal of granulated tissue ﺳﯿﺴﺖ
- Caries removal, cavity preparation, enamel roughening -
- Sulcular debridement -
- Tooth preparation to obtain access to root canal, root canal debridement and cleaning, root canal preparation including enlargement
- ー Cutting, shaving, contouring and resection of oral osseous tissue (bone)
- Osteotomy, osseous crown lengthening, osteoplasty -
- Apicectomy surgery -
Er: Y AG laser (2940 nm wavelength) in dermatology and other surgical areas:
- Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and skin resurfacing;
- ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia; l
- Oral/Maxillofacial Indications: Oral and glossal lesions, gingivectomy; -
- General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation;
- Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy; -
Nd:YAG laser (1064 nm wavelength) in dentistry:
- -Excisional and incisional biopsies
- Excision and vaporization of herpes simplex I and II -
- Exposure of unerupted teeth -
- Fibroma removal —
- Frenectomy and frenotomy -
- Gingival troughing for crown impressions -
- Gingivectomy
Nil R.P. Ogden for mxm
and Neurological De
**510(k) Number** K070355
{3}------------------------------------------------
K070355
P g R = (R
- Gingivoplasty
- Gingival incision and excision =
- Hemostasis -
- Implant recovery J
- Incision and drainage of abscess -
- Laser assisted uvulopaletoplasty (LAUP) -
- Operculectomy =
- Oral papillectomies =
- Pulpotomy and pulpotomy as an adjunct to root canal therapy -
- Reduction of denture hyperplasia -
- -Reduction of gingival hypertrophy
- Removal of filling material such as gutta percha or resin as adjunct treatment during root e canal therapy
- Removal of post-surgical granulations ।
- Soft tissue crown lengthening =
- Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- Tissue retraction for impression -
- Treatment of aphtous ulcers -
- Vestibuloplasty -
- Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium)
Nd: YAG laser (1064 nm wavelength) in dermatology and other surgical areas:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited । to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins
- Treatment of wrinkles -
- Treatment of mild to moderate inflammatory acne vulgaris
Prescription Use: X (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use: (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
ViRBE
P. Oslen for
and Neurologics
510(k) Submission: Fotona Fidelis III Laser System Family
Concurrence of CDRH, Office of Device Evaluation (C
510(k) Numbe
{4}------------------------------------------------
K070355
Pg 1 of 3
## 5. 510(k) Summary
| Submitter's Name: | Fotona d.d.<br>Stegne 7<br>1210 Ljubljana, Slovenia<br>Phone: +386 15009100<br>Fax: + 386 5009 200 |
|-------------------|----------------------------------------------------------------------------------------------------|
| Contact Person: | Stojan Trošt, QA&RA Manager<br>Phone: + 386 1 5009 299<br>E-mail: stojan.trost@fotona.si |
MAR - 4 2008
Date: January 31, 2007
| Device Name: | |
|----------------------|-----------------------------------------------------|
| Trade name: | Fotona Fidelis III Er:YAG/Nd:YAG Laser System Famil |
| Common name: | Er:YAG/Nd:YAG Surgical Laser |
| Classification name: | Instruments, Surgical, Powered, Laser<br>79-GEX |
## DEVICE DESCRIPTION
The Fotona Fidelis III laser system family is based on Er: Y AG (2940 nm) and Nd: YAG (1064 nm) laser technology. It combines two flashlamp-pumped laser sources in one housing, with optical cavities containing the Er.Y AG and Nd: Y AG crystals. A red diode aiming beam (650 mm) is combined with both therapeutic laser beams. The combined therapeutic and aiming beams are guided through an articulated arm to an optical hand piece (in the case of the Er: YAG laser), or through an optical fiber delivery system to an optical hand piece (in the case of the Nd: YAG laser).
The Er.Y AG laser is capable of delivering up to 1.5 J of laser energy in pulses with durations of up to 1000 us and frequencies (repetition rates) of up to 50 Hz. The maximum average output power is 20 W.
For aesthetic indications, the Nd:YAG laser is capable of delivering laser fluences up to 300 1/cm² in pulses with durations of up to 25 ms. For dental indications, it is capable of delivering laser pulses with durations of up to 320 us, frequencies (repetition rates) up to 100 Hz and a maximum output power of 15 W.
Fotona's power supply Variable Square Pulse (VSP) Technology, integrated into the laser system, allows ultimate control of the laser energy and the laser pulse duration. This ensures treatment precision, patient comfort, safety and ease-of-use in all treatments.
### INTENDED USE
The Fidelis III Er: YAG laser, and its accessories, are in dentistry, dermatology and other surgical areas in the following procedures:
In dentistry, for:
- Caries removal, Cavity Preparation, Enamel Roughening -
- Intra-oral soft tissue surgery (incision, excision, ablation, coagulation) -
- -Leukoplakia
{5}------------------------------------------------
K070355 pg. 2 of 3
- Removal of fibromae -
- Removal granulated tissue -
- -Sulcular debridement
- ﯿﮯ Apicectomy surgery
- Cutting, shaving, contouring and resection of oral osseous tissue (bone) -
- Osteotomy,osseous crown lengthening, osteoplasty -
- Tooth preparation to obtain access to root canal debridement and cleaning, root canal preparation including enlargement
- -Pulp extirpation
- Pulpotomy as adjunct to root canal retreatment -
#### In dermatology and other surgical areas, for:
- Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, skin tags, keratoses, verrucae, and skin resurfacing
- = ENT Surgery Indications:ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia
- Oral/Maxillofacial Indications: Oral and glossal lesions, gingivectomy ー
- General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation
- Podiatry Indications: warts, plantar verrucae, large mosiac verrucae, matrixectomy -
The Fidelis III Nd: YAG laser, and its accessories, are intended for use in dentistry, dermatology and other surgical areas in the following procedures:
In dentistry, for:
- Excisional and incisional biopsies 는
- Excision and vaporization of herpes simplex I and II జ
- Exposure of unerupted teeth -
- Fibroma removal =
- Frenectomy and frenotomy -
- -Gingival troughing for crown impressions
- Gingivectomy =
- Gingivoplasty -
- Gingival incision and excision -
- Hemostasis =
- Implant recovery --
- -Incision and drainage of abscess
- يي Laser assisted uvulopaletoplasty (LAUP)
- -Operculectomy
- -Oral papillectomies
- = Reduction of denture hyperplasia
- Reduction of gingival hypertrophy -
- Removal of post-surgical granulations -
- Soft tissue crown lengthening
{6}------------------------------------------------
K070555 013-2+3
- Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- Tissue retraction for impression -
- Treatment of aphtous ulcers =
- Vestibuloplasty -
- Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-ー attachment to the root surface in the absence of long junctional epithelium)
- Pulpotomy and pulpotomy as an adjunct to root canal therapy -
- Removal of filling material such as gutta percha or resin as adjunct treatment during root canal therapy
#### In dermatology and other surgical areas, for:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of -PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited to, ー port wine stains, hemaongiomae, warts, telengiectasiae, rosacea, venus lake, leg veins and spider veins
- Treatment of wrinkles =
- Treatment of mild to moderate inflammatory acne vulgaris —
### STATEMENT OF SUBSTANTIAL EQUIVALENCE
The Fidelis III laser system family shares the same indications for use, similar design and functional features with, and therefore Fotona d.d. believes that its Fidelis III laser system family is substantially equivalent to, the
a) Fotona Fidelis Er: Y AG Laser System (K001527) previously cleared for incision, excision, vaporization, ablation and coagulation of soft and hard tissue in the mouth;
b) Fotona Fidelis Plus Nd: Y AG Laser System (K024204) previously cleared as an accessory for Fotona Fidelis Er.YAG Laser System (K001527) for incision, excision and coagulation of intra oral soft tissue, including the marginal and inter dental gingiva;
c) Fotona XP Plus Nd: Y AG Family (K050293) previously cleared for surgical and aesthetic applications in soft tissue in the medical specialities of general and plastic surgery and dermatology;
d) Fotona Dualis Nd: YAG/Er: YAG Laser System (K021548) previously cleared for surgical incision/excision, vaporization and coagulation of soft and hard tissue;
e) Hoya ConBio -VersaWave Dental Er:YAG Laser System (K041710) previously cleared for various hard tissue and soft tissue dental indications;
f) Millennium Periolase Nd: Y AG Laser System (K030290) previously cleared for various dental indications;
g) Cooltouch Nd: YAG Laser System previously cleared for the treatment of mild to moderate inflammatory acne vulgaris (K040131).
h) Laserscope Gemini Laser System previously cleared for the treatment of mild to moderate inflammatory acne vulgaris (K034011)
Details are provided in the Substantial Equivalence Discussion Section of this submission.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is presented in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Fotona d.d. % Stojan Trošt QA & RA Manager Stegne 7 1210 Ljubljana, Slovenia
MAR - 4 2008
Re: K070355
Trade/Device Name: Fotona Fidelis III Er: YAG/Nd: YAG Laser System Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 22, 2008 Received: February 27, 2008
Dear Stojan Trošt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Stojan Trošt
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Milhusan
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
K070355 pg 1 of
# 4. Indications for Use Statement
510(k) Number (if known):
Device Name: Fotona Fidelis III Er: YAG/ Nd: YAG Laser System Family
Indications For Use:
Er: Y AG laser (2940 nm wavelength) in dentistry:
- Intra-oral soft tissue surgery (incision, excision, ablation coagulation) -
- -Leukoplakia
- = Pulpotomy as adjunct to root canal retreatment
- Pulp extirpation -
- Removal of fibromae -
- Removal of granulated tissue ﺳﯿﺴﺖ
- Caries removal, cavity preparation, enamel roughening -
- Sulcular debridement -
- Tooth preparation to obtain access to root canal, root canal debridement and cleaning, root canal preparation including enlargement
- ー Cutting, shaving, contouring and resection of oral osseous tissue (bone)
- Osteotomy, osseous crown lengthening, osteoplasty -
- Apicectomy surgery -
Er: Y AG laser (2940 nm wavelength) in dermatology and other surgical areas:
- Dermatology and Plastic Surgery Indications: Epidermal nevi, actinic cheilitis, verrucae, skin tags, keratoses and skin resurfacing;
- ENT Surgery Indications: ENT lesions, cysts, polyps, hyperkeratosis, oral leukoplakia; l
- Oral/Maxillofacial Indications: Oral and glossal lesions, gingivectomy; -
- General Surgery Indications: Surgical incision/excision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation;
- Podiatry Indications: Warts, plantar verrucae, large mosaic verrucae, matrixectomy; -
Nd:YAG laser (1064 nm wavelength) in dentistry:
- -Excisional and incisional biopsies
- Excision and vaporization of herpes simplex I and II -
- Exposure of unerupted teeth -
- Fibroma removal —
- Frenectomy and frenotomy -
- Gingival troughing for crown impressions -
- Gingivectomy
Nil R.P. Ogden for mxm
and Neurological De
**510(k) Number** K070355
{10}------------------------------------------------
K070355
P g R = ( R
- Gingivoplasty
- Gingival incision and excision =
- Hemostasis -
- Implant recovery J
- Incision and drainage of abscess -
- Laser assisted uvulopaletoplasty (LAUP) -
- Operculectomy =
- Oral papillectomies =
- Pulpotomy and pulpotomy as an adjunct to root canal therapy -
- Reduction of denture hyperplasia -
- -Reduction of gingival hypertrophy
- Removal of filling material such as gutta percha or resin as adjunct treatment during root e canal therapy
- Removal of post-surgical granulations ।
- Soft tissue crown lengthening =
- Sulcular debridement or soft tissue curettage (removal of diseased or imflamed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility)
- Tissue retraction for impression -
- Treatment of aphtous ulcers -
- Vestibuloplasty -
- Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium)
Nd: YAG laser (1064 nm wavelength) in dermatology and other surgical areas:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. The laser is indicated for all skin types, Fitzpatrick I-VI, including tanned skin
- Photocoagulation and hemostasis of pigmented and vascular lesions, such as, but not limited । to, port wine stains, hemaongiomae, warts, telangiectasiae, rosacea, venus lake, leg veins and spider veins
- Treatment of wrinkles -
- Treatment of mild to moderate inflammatory acne vulgaris
Prescription Use: X (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use: (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
NEEDED)
Vilbe
Oselm for
and Neurologics
510(k) Submission: Fotona Fidelis III Laser System Family
Concurrence of CDRH, Office of Device Evaluation (C
510(k) Numbe
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.