K030290 · Millennium Dental Technologies, Inc. · GEX · Jul 26, 2004 · General, Plastic Surgery
Device Facts
Record ID
K030290
Device Name
PERIOLASE ND:YAG DENTAL LASER SYSTEM
Applicant
Millennium Dental Technologies, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 26, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PerioLase Nd: YAG Dental Laser System is to provide the ability to perform intraoral soft tissue dental, general, oral maxillo-facial and cosmetic surgery. The PerioLase is intended for ablating, incising, excising, vaporization and coagulation of soft tissues using a contact fiber optic delivery system. The device will be used in the following areas: general and cosmetic dentistry, otolaryngology, arthroscopy, gastroenterology, general surgery, dermatology & plastic surgery, neurosurgery, gynecology, urology, ophthalmology and pulmonary surgery. The following is the additional oral-pharngeal indication for use for which the device will be marketed: Abscess Incision and Drainage Apthous Ulcers Treatment Biopsies Excision and Incision Crown lengthening Hemostatic assistance Fibroma Removal Frenectomy Frenotomy Gingival Incision and Excision Gingivectomy Gingivoplasty Operculectomy Oral Papillectomy Tissue retraction for Impression Vestibuloplasty. Selective ablation of enamel (first degree) caries Exposure of unerupted / partially erupted teeth Implant recovery Lesion (tumor) removal Leukoplakia Pulpotomy Pulpotomy as adjunct to root canal therapy Removal of filling material such as gutta percha or resin as adjunct treatment during root canal re- treatment Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) to improve clinical indices including gingival index, gingivaltooth mobility tooth mobility. Laser assisted new attachment procedure (cementum-mediated periodontal ligament new- attachment to the root surface in the absence of long junctional epithelium.)
Device Story
PerioLase is a pulsed Nd:YAG laser (1064nm) for soft tissue surgery. System comprises a cabinet housing laser head, power supply, cooling system, and microprocessor control panel, connected to a fiber optic delivery system. Operated by clinicians in dental/surgical settings. Laser energy is delivered via contact fiber optic to perform ablation, incision, excision, vaporization, and coagulation of soft tissues. Clinical benefit includes precise tissue management and periodontal pocket debridement. Output is controlled by the clinician via the control panel to adjust pulse durations (100-650µsec).
Clinical Evidence
Clinical evidence consists of human histological and radiographic data from a controlled prospective university-based clinical study supporting the laser-assisted new attachment procedure (LANAP) claim.
Technological Characteristics
Nd:YAG laser; 1064nm wavelength; variable pulse durations (100-650µsec). System includes laser head, power supply, cooling system, and microprocessor control panel. Fiber optic delivery system. Class II device.
Indications for Use
Indicated for patients requiring intraoral soft tissue dental, oral maxillo-facial, and cosmetic surgery, including procedures like gingivectomy, frenectomy, lesion removal, and sulcular debridement. Also indicated for laser-assisted new attachment procedures (LANAP) and various general surgical applications across multiple medical specialties.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Ko 30 290
| | | C. Status and Children and Children and Children and Children and Children and Children and<br>The for the state the count of may and support of the collections and contract and contributed and controlled on the fight to the fight of the fight of the fight of the fight |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| THE WATER THE LEAD LE LE LE LE SELL SELL COLLECT LAND STATE LA LA LA LA LA LA LA LES SELLER LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE LE L | STATUS STATE COLLECTION COLLEGION COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLLECT COLL<br>MARKET LE COLLECT COLLEGE LE LE SECTION LE STATUS COLLECTION CONSULT AND CONSULTER COLLECTION CONSULTERS CONTRACTOR COLLECTION CONSULTION CONSULTION CONSULTION CONSULTION CON | A F . No. 1978 minutes while to MC<br> |
| Carles Children Children Children Children Childer Comments and A. | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | |
| | | Parks . Naturi<br>1 11 1 1 1 1 1 1 1 1 1 6 - 2 - 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 |
| | | SP |
| = CASSE = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = " = | | *** **************** |
| EAR and Career parties a<br>.<br>. | | |
| Concerner of the concession in the contribution in<br>Callery of Children Children Children<br>1575-5575-55-58-11-11-11 | | |
| the first of the seen the same a<br>Annual Partic Partic Partic Company of Concession Come | | |
| arman and state and section and the program<br>the first to the many of the state and the county of | | |
| and the street to seen the security of the support of the support of<br><br>.<br>Carrelation Concession, Canada Andrew<br>Call College of Children Children and Children Children | | . |
ד / / / >
This 510(K) Summary of safety and effectiveness for the Millennium Dental Technologies PerioLase Dental Laser This 310(K) Submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.
| Applicant: | Millennium Dental Technologies, Inc. |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 10945 South Street, Suite 104-A<br>Cerritos, CA 90703 |
| Contact Person: | David M. Harris, Ph.D. |
| Telephone: | (562) 860-2908 - Phone<br>(562) 860-2429 - FAX |
| Preparation Date: | January 01, 2003 |
| Device Trade Name: | PerioLase Dental Laser |
| Common Name: | Nd:YAG Pulsed Laser |
| Classification Name: | Instrument, Surgical, Powered, Laser<br>79-GEX<br>21 CFR 878-48 |
| Legally Marketed Predicate<br>Device: | PerioLase Dental Laser |
| Description of the Millennium<br>Dental Technologies PerioLase<br>Dental Laser | The PerioLase is a FR Nd:YAG laser producing laser emission at<br>1064nm with variable pulse durations (100-650µsec). The laser<br>consists of two interconnected sections: The cabinet which houses<br>the laser head, the power supply, the cooling system and the<br>microprocessor with control panel; and the fiber optic delivery<br>system. |
| Clinical Performance Data: | The new clinical outcome claim is based on human histological and<br>radiographic data from a controlled prospective University-based<br>clinical study. |
| Summary Basis of Equivalence: | N/A |
{1}------------------------------------------------
Kozc kp
1/2
| Summary of Safety and Effectiveness Information PerioLase<br>Dental Laser System<br>Premarket Notification, Section 510(k) | MILLENNIUM DENTAL<br>TECHNOLOGIES, INC.<br><br>JANUARY 17, 2003 |
|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use: | The following are the oral-pharngeal indications for use<br>for which the device will be marketed:<br>Abscess Incision and Drainage<br>Apthous Ulcers Treatment<br>Biopsies Excision and Incision<br>Crown lengthening<br>Hemostatic assistance<br>Fibroma Removal<br>Frenectomy<br>Frenotomy<br>Gingival Incision and Excision<br>Gingivectomy<br>Gingivoplasty<br>Operculectomy<br>Oral Papillectomy<br>Tissue retraction for Impression<br>Vestibuloplasty.<br>Selective ablation of enamel (first degree) caries<br>Exposure of unerupted / partially erupted teeth<br>Implant recovery<br>Lesion (tumor) removal<br>Leukoplakia<br>Pulpotomy<br>Pulpotomy as adjunct to root canal therapy<br>Removal of filling material such as gutta percha or<br>resin as adjunct treatment during root canal re-<br>treatment<br>Sulcular debridement (removal of diseased or inflamed<br>soft tissue in the periodontal pocket) to improve<br>clinical indices including gingival index, gingivaltooth mobility tooth mobility. |
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and a pair of wings at the top.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 6 2004
Millenium Dental Technologies, Inc. c/o David M. Harris, Ph.D. Bio-Medical Consultants, Inc. 4256 Heyer Avenue Castro Valley, California 94546
Re: K030290
Trade/Device Name: PerioLase Nd:YAG Dental Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: April 21, 2004 Received: April 27, 2004
Dear Dr. Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic rooduct radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - David M. Harris, Ph.D.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will and w you'ls begal finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Melkersen
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
## 510(k) Number : K030290
## PerioLase Nd:YAG Dental Laser System Device Name(s):
## indications For Use:
The PerioLase Nd: YAG Dental Laser System is to provide the ability to perform intraoral soft tissue dental, general, oral maxillo-facial and cosmetic surgery. The PerioLase is intended for ablating, incising, excising, vaporization and coagulation of soft tissues using a contact fiber optic delivery The device will be used in the following areas: general and cosmetic dentistry system. otolaryngology, arthroscopy, gastroenterology, general surgery, dermatology & plastic surgery, neurosurgery, gynecology, urology, ophthalmology and pulmonary surgery. The following is the additional oral-pharngeal indication for use for which the device will be marketed:
- Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-. attachment to the root surface in the absence of long junctional epithelium.)
Prescription Use (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Milkerson
I, Restorative. and Neurological Devices
Page_1 of
510(k) Number K030290
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.