POWERPORT(TM)IMPLANTED TITANIUM PORT W/8 FR. CHRONOFLEX CATHETER-MULTIPLE
K060812 · C.R. Bard, Inc. · LJT · Jul 14, 2006 · General Hospital
Device Facts
Record ID
K060812
Device Name
POWERPORT(TM)IMPLANTED TITANIUM PORT W/8 FR. CHRONOFLEX CATHETER-MULTIPLE
Applicant
C.R. Bard, Inc.
Product Code
LJT · General Hospital
Decision Date
Jul 14, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PowerPort™ Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with the PowerLoc™ Safety Infusion Set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s. The PowerLoc™ Safety Infusion Set is an intravascular administration set with a non-coring right angle needle and manually activated needle-stick safety mechanism. The device is used to access surgically implanted vascular ports. The PowerLoc™ Safety Infusion Set is indicated for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. When used with the PowerPort™ device, the PowerLoc™ Safety Infusion Set is also indicated for power injection of contrast media into the central venous system. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s for 19 Ga. and 20 Ga. needles, and 2 ml/s for 22 Ga. needles.
Device Story
PowerPort System comprises titanium implanted port and PowerLoc Safety Infusion Set (SIS). System enables repeated vascular access for drug/fluid delivery and blood sampling; specifically designed to support high-pressure power injection of contrast media. Port features purple body and unique three-sided shape with septum bumps for identification. PowerLoc SIS includes non-coring Huber needle and manual safety mechanism to prevent needle-stick injuries. Used in clinical settings by healthcare professionals. System allows clinicians to perform contrast-enhanced imaging procedures via implanted port. Benefits include reliable vascular access and capability for power injection without requiring additional access points.
Clinical Evidence
Bench testing only. No clinical data. Performance verified through design verification testing per FDA guidance for implanted infusion ports and intravascular administration sets. Biocompatibility met ISO-10993; sterilization validated per ISO 11135.
Technological Characteristics
Titanium port body; 8 Fr ChronoFlex (polyurethane) catheter. PowerLoc SIS uses non-coring Huber needle with manual safety guard. Biocompatible materials per ISO-10993. Sterilization via ethylene oxide (ISO 11135). System designed for power injection up to 5 ml/s.
Indications for Use
Indicated for patients requiring repeated vascular access for infusion of medications, fluids, parenteral nutrition, blood products, and blood sampling. Also indicated for power injection of contrast media when used as a system (PowerPort + PowerLoc). Contraindicated for use with non-power injectable ports.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
{0}------------------------------------------------
K060812
i+4
PowerPort™ System
Traditional 510(k)
# Section 11 510(K) Summary of Safety & Effectiveness Information
The following summary of safety and effectiveness takes account of both the primary device (PowerPort™ implanted titanium port) and accessory (PowerLoc™ of bon the planting dower injecting, are required to work together as a system.
#### 11.1 Submitter Information
## Primary Device & Accessory Device
| Submitter Name: | Bard Access Systems, Inc. (BAS)<br>[Subsidiary of C. R. Bard, Inc.] |
|----------------------|---------------------------------------------------------------------|
| Address: | 5425 W. Amelia Earhart Drive<br>Salt Lake City, UT 84116 |
| Telephone Number: | (801) 595-0700, Ext. 5484 |
| Fax Number: | (801) 595-5425 |
| Contact Person: | Susan Scott |
| Date of Preparation: | March 23. 2006 |
#### 11.2 Device Name
# PowerPort™ Port/Catheter Device
| Trade/Device Name: | PowerPort™ Implanted Titanium Port with 8 Fr. ChronoFlex® Catheter |
|-----------------------|-------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Implanted Infusion Port |
| Classification Name: | Class II, 80 LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter (21 CFR §880.5965) |
| Classification Panel: | General Hospital |
## PowerLoc™ SIS Device
| Proprietary Name: | PowerLoc™ Safety Infusion Set |
|-----------------------|------------------------------------------------------------------------|
| Common/Usual Name: | Huber Needle Intravascular Administration Set |
| Classification Name: | Class II, FPA – Intravascular Administration Set<br>(21 CFR §880.5441) |
| Classification Panel: | General Hospital and Personal Use |
#### 11.3 Predicate Device(s):
PowerPort™ System Indications Predicate:
| Trade/Device Name: | 5 Fr Dual Lumen PowerPICC® Catheter |
|-------------------------|----------------------------------------------------------------------------|
| Common/Usual Name: | Peripherally Inserted Central Catheter (PICC) |
| Classification Name: | Class II, 80 LJS -- Long Term Intravascular Catheter<br>(21 CFR §880.5970) |
| Classification Panel: | General Hospital |
| Premarket Notification: | K051672, Substantial Equivalence: November 23, 2005 |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows handwritten text and printed text. The handwritten text at the top reads 'K060812' and '2of4'. The printed text below reads 'PowerPort System' and 'Traditional 510(k)'. The text appears to be part of a document or label.
PowerPort™ Port/Catheter Device Predicate:
| Trade/Device Name: | PowerPort™ Implanted Titanium Port with 8 Fr ChronoFlex Catheter |
|-------------------------|-------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Implanted Infusion Port |
| Classification Name: | Class II, 80 LJT - Subcutaneous, Implanted, Intravascular Infusion Port & Catheter (21 CFR §880.5965) |
| Classification Panel: | General Hospital |
| Premarket Notification: | K050310, Substantial Equivalence: April 18, 2005 |
## PowerLoc™ SIS Device Predicate:
| Trade/Device Name: | MiniLoc™ Safety Infusion Set |
|-------------------------|------------------------------------------------------------------------|
| Common/Usual Name: | Huber Needle Intravascular Administration Set |
| Classification Name: | Class II, FPA – Intravascular Administration Set<br>(21 CFR §880.5440) |
| Classification Panel: | General Hospital |
| Premarket Notification: | K050600, Substantial Equivalence: May 12, 2005 |
#### 11.4 Device Description
PowerPort™ Implanted Titanium Port Device
- The PowerPort™ Implanted Titanium Port is a titanium port with attachable open-ended, 45 . cm, 8 Fr ChronoFlex® polyurethane catheter.
- The PowerPort™ is designed for power injection of contrast media when used with the power . injectable PowerLoc™ SIS. Confirmation of PowerLoc™ and PowerPort™ as a requirement for power injection is directed in all associated literature.
- The purple color of the port body differentiates the PowerPort™ from other implantable . vascular access ports, highlighting the power injectable capability of the PowerPort™.
- The unique shape of the PowerPort™ (three sided port with three septum bumps) aids in . identification.
PowerLoc™ SIS Device
- The PowerLoc™ Safety Infusion Set is a standard non-coring Huber type needle and . administration set with an integral safety needle-stick prevention feature.
#### 11.5 Intended Use
The intended use of each of the components of the PowerPort™ System has not changed from that of the predicate.
## PowerPort™ Port/Catheter Device
The PowerPort™ Implanted titanium Port is a totally implantable vascular access device designed to provide long term repeated access to the vascular access tervice intended use as compared to the predicate Implanted titanium Port.
## PowerLoc™ SIS Device
The PowerLoc™ SIS is intended for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. The intended use has not changed. This is the same intended use as the predicate MiniLoc™ Safety Infusion Set [K050600].
{2}------------------------------------------------
#### 11.6 Indications for Use
The Indications for Use were modified to reflect the addition of the power injection of contrast media procedure. The Indications for Use Statements can be found in Section 1.2.
## PowerPort™ Port/Catheter Device
The PowerPort™ Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with the PowerLoc™ Safety Infusion Set, the PowerPort™ device is indicated for power injection of contrast media. The maximum recommended infusion rate is 5 ml/s.
## PowerLoc™ SIS Device
The PowerLoc™ Safety Infusion Set is an intravascular administration set with a non-coring right anqle needle and manually activated needle-stick safety mechanism. The device is used to access surgically implanted vascular ports.
The PowerLoc™ Safety Infusion Set is indicated for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports.
When used with the PowerPort™ device, the PowerLoc™ Safety Infusion Set is also indicated for power injection of contrast media into the central venous system only with an implanted port that is also indicated for power injection. The maximum recommended infusion rate is 5 ml/s for 19 Ga. And 20 Ga. Needles, and 2 ml/s for 22 Ga. Needles.
#### 11.7 Technological Characteristics Summary
The following Summary Information refers to both the primary device (PowerPort™ port) and the accessory device (PowerLoc™ infusion set). Although these devices can be assessed individually for routine infusions, they are also addressed together as a system for power injection.
## New device is compared to Marketed Device?
- > Yes. Both the port and infusion set devices are compared to legally marketed predicates.
#### Does the new device have the same indication statement?
- A No. The Indications For Use were expanded to include power injection of contrast media for both the port and infusion set devices.
## Do the differences alter the intended therapeuticletc. effect (i.e. in deciding, impact on safety and effectiveness may be considered)?
- No, the differences do not alter the intended use of the port or infusion set device. A
## Does the new device have the same technological characteristics, e.g. design, material, etc?
- a No, not in all regards.
- o The PowerPort™ Implanted Port has some minor differences from the predicate Implanted titanium Port, however, the fundamental scientific technology of the catheter has not changed.
- The PowerLoc™ Safety Infusion Set has some minor differences from the predicate o MiniLoc™ Safety Infusion Set. The basic fundamental scientific technology of the device has not changed.
{3}------------------------------------------------
1060817. PowerPort™ System Traditional 510(k)
## Could the new characteristics affect safety or effectiveness?
- > Yes. The expanded indication to include power injection of contrast media through the PowerPort™/PowerLoc™ system could affect safety or effectiveness.
## Do the new characteristics raise new types of safety and effectiveness questions?
- > No. There are no new types of safety and effectiveness questions.
# Do accepted scientific methods exist for assessing effects of the new characteristics?
> Yes.
The FDA's Guidance on 510(k) Submissions for Implanted Infusion Ports, dated October 1990 was used to evaluate the PowerPort™ device's performance.
The following two FDA documents were used to evaluate the PowerLoc™ device's performance: Guidance for Industry and FDA Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)], dated April 15, 2005; and ODE Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Guards: Guidance for Industry and FDA, dated December 31, 2002.
Sterilization requirements of ISO 11135:1994, Medical Devices – Validation and Routine Control of Ethylene Oxide Sterilization, will be met.
Biocompatibility requirements of ISO-10993, Biological Evaluation of Medical Devices Part-1: Evaluation and Testing, were met for both the PowerPort™ implanted port and PowerLoc™ infusion set.
## Are performance data available to assess effects of new characteristics?
- Yes. Verification testing was performed according to protocols based on the above-A referenced guidance document recommendations and additional standards.
#### Do performance data demonstrate equivalence?
- Yes. Performance data gathered in design verification testing demonstrated the following: A one, that the PowerPort™ Implanted Port is substantially equivalent to the predicate Implanted Titanium Port; and two, that the PowerLoc™ Safety Infusion Set is substantially equivalent to the predicate MiniLoc™ Safety Infusion Set.
#### 11.8 Conclusion
The PowerPort™ System met all established acceptance criteria for performance testing and design verification testing. Based on FDA's decision tree, the components of the PowerPort™ System are substantially equivalent to their respective predicates: the PowerPort™ Implanted Titanium Port as compared to the Titanium Implanted Port [K050310, cleared on April 18, 2005]; and the PowerLoc™ Safety Infusion Set as compared to the MiniLoc™ Safety Infusion Set [K050600, cleared on May 12, 2005].
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized caduceus, a symbol often associated with medicine and healthcare, alongside the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged in a circular fashion around the symbol. The caduceus is depicted with three lines forming a stylized shape, and the text is in uppercase letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
C.R. Bard, Incorporated Ms. Susan Scott Regulatory Affairs Specialist Bard Access Systems, Incorporated 5425 West Amelia Earhart Drive Salt Lake City, Utah 84116
JUL 1 4 2006
Re: K060812
Trade/Device Name: PowerPort™ Implanted Titanium Port, PowerLoc™ Safety Infusion Set Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT, FPA Dated: June 23, 2006 Received: June 26, 2006
Dear Ms. Scott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Ms. Scott
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Syngite Y. Michael Davis.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
KO60812
127712 PowerPort™ System Traditional 510(k)
# Section 1.2 A
# Indications for Use
| 510(k) Number (if known): | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | PowerPort™ Implanted Titanium Port |
| Indications for Use: | The PowerPort™ Implanted Port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. |
| | When used with the PowerLoc™ Safety Infusion Set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s. |
| Prescription Use<br>(Part 21 CFR §801 Subpart D) | <div>✓</div> |
|--------------------------------------------------|--------------|
|--------------------------------------------------|--------------|
AND/OR
| Over-The-Counter Use<br>(21 CFR §801 Subpart C) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
. ... n Control, Dental Devices
1660512 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the text "K060812 2 of 2 PowerPort System Traditional 510(k)". The text appears to be part of a document or label. The phrase "PowerPort System" is written in a slightly larger font size compared to the other text. The text is black and the background is white.
# Section 1.2 B
## Indications for Use
510(k) Number (if known): Device Name: PowerLoc™ Safety Infusion Set Indications for Use: The PowerLoc™ Safety Infusion Set is an intravascular administration set with a non-coring right angle needle and manually activated needle-stick safety mechanism. The device is used to access surgically implanted vascular ports. The PowerLoc™ Safety Infusion Set is indicated for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports. When used with the PowerPort™ device, the PowerLoc™ Safety Infusion Set is also indicated for power injection of contrast media into the central venous system. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s for 19 Ga. and 20 Ga. needles, and 2 ml/s for 22 Ga. needles. Prescription Use > Over-The-Counter Use AND/OR (Part 21 CFR §801 Subpart D) (21 CFR §801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
![]()
S.JI) Jign-Cit) n of Anesthesiology, General Hospital, uon Control, Dental Devices
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
page 47
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.