COPROSET SALMONELLA AND SHIGELLA, URISET PRESERVATIVE TUBES AND SWAB SET GENERAL USE
Applicant
Diesse Diagnostica Senese S.P.A.
Product Code
LIO · Microbiology
Decision Date
Jul 30, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2900
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
Coproset Salmonella: In vitro Disposable diagnostic device with Selenite Broth intended for the collection and transport, from the collection site to the testing laboratory, of stool samples and for the enrichment of Salmonella spp. Coproset Shigella: In vitro Disposable diagnostic device with GN Broth intended for the collection and transport from the collection site to the testing laboratory of stool samples and for the enrichment of Shigella spp. Uriset preservative tubes: Disposable in vitro diagnostic device for the collection and transport of urine samples from the collection site to the testing laboratory. Swab set General Use: Disposable in vitro diagnostic device with Eugon Broth, intended for the collection and transport of pathogenic agents collected by swabs from the collection site to the testing laboratory and for enrichment of Streptococcus pyogenes, Staphylococcus aureus, Candida spp., Streptococcus agalactiae.
Device Story
Disposable in vitro diagnostic collection and transport systems; includes Coproset (stool), Uriset (urine), and Swab Set (general). Coproset devices utilize liquid media (Selenite or GN Broth) for sample maintenance and enrichment of enteric pathogens. Uriset uses a calibrated vacuum-filled tube with preservative to stabilize bacterial load in urine. Swab Set uses Eugon Broth for transport and enrichment of various pathogens. Used in clinical settings by healthcare professionals to ensure specimen integrity during transit to laboratories. Output is a preserved clinical sample ready for downstream microbiological processing (streaking on agar plates, incubation). Benefits include standardized sample handling, operator safety, and maintenance of pathogen viability.
Clinical Evidence
Bench testing only. Performance evaluation compared pathogen viability and sample preservation between subject devices and predicate devices. Results confirmed that Diesse devices maintain pathogen viability and provide performance comparable to the cited predicate devices.
Technological Characteristics
Disposable specimen collection and transport devices. Components include liquid transport media (Selenite Broth, GN Broth, Eugon Broth) and containers (tubes, swabs, spoons). Uriset features a calibrated vacuum-filled tube mechanism. Non-electronic, non-powered, manual operation.
Indications for Use
Indicated for collection, transport, and enrichment of clinical specimens (stool, urine, or swab-collected pathogens) for laboratory analysis. Applicable to patients requiring diagnostic testing for Salmonella spp., Shigella spp., Streptococcus pyogenes, Staphylococcus aureus, Candida spp., or Streptococcus agalactiae.
Regulatory Classification
Identification
A microbiological specimen collection and transport device is a specimen collecting chamber intended for medical purposes to preserve the viability or integrity of microorganisms in specimens during storage of specimens after their collection and during their transport from the collecting area to the laboratory. The device may be labeled or otherwise represented as sterile. The device aids in the diagnosis of disease caused by pathogenic microorganisms.
{0}------------------------------------------------
Image /page/0/Picture/0 description: This image shows the word "Page" in bold font above the fraction "81/102". The text is centered on a white background. A horizontal line is located below the fraction.
K070062
# Appendix C
JUL 3 0 2007
# 510(k) Summary of Safety and Effectiveness
| Name | DIESSE Diagnostica Senese SpA |
|----------------|----------------------------------------------------------------------------------------|
| Address | Via delle Rose 10, 53035 Monteriggioni SI<br>Tel. 39-0577-587111<br>Fax 39-0577-318690 |
| Contact Person | Dr. Francesco Cocola |
| Phone Number | 39-0577-587143 |
| Fax Number | 39-0577-318379 |
The Following section is included as required by the Safe Medical Device Act (SMDA) 1990.
{1}------------------------------------------------
| Page |
|--------|
| 82/102 |
# 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) Number is:________________________________________________________________________________________________________________________________________________
#### GENERAL INFORMATION 1
#### 1.1 Applicant
| Date | September 19, 2006 |
|----------------|-------------------------------------------------------------------------------------------------------|
| Name | DIESSE Diagnostica Senese SpA |
| Address | Via delle Rose 10, 53035 Monteriggioni SI, Italy<br>Tel. 011-39-0577-587111<br>Fax 011-39-0577-318690 |
| Contact Person | Dr. Francesco Cocola |
| Phone Number | 39-0577-587143 |
| Fax Number | 39-0577-318379 |
{2}------------------------------------------------
#### Device information 1.2
| Proprietary<br>Device Name | Coproset Shigella | Coproset Salmonella | Uriset Preservative tube | Swab Set General<br>Use |
|-------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Device model<br>number | 94052 | 94050 | 94026 | 94320 |
| Type of<br>Submission | Third Party | Third Party | Third Party | Third Party |
| Common Name<br>Or Classification<br>Name | Microbiological<br>specimen collection<br>and transport Device,<br>Coproset<br>(866.2900) | Microbiological<br>specimen collection<br>and transport Device,<br>Coproset<br>(866.2900) | Microbiological<br>specimen collection and<br>transport Device, Uriset<br>(866.2900) | Microbiological<br>specimen collection<br>and transport Device,<br>Swabset general use<br>(866.2900) |
| Establishment<br>Registration<br>Number | 807.87 (b) 9615056 | 807.87 (b) 9615056 | 807.87 (b) 9615056 | 807.87 (b) 9615056 |
| Facility Address | Via delle Rose 10<br>53035 Monteriggioni<br>SI, Italy<br>Tel.: 39-0577-587111<br>Fax: 39-0577-318690 | Via delle Rose 10<br>53035 Monteriggioni<br>SI, Italy<br>Tel.: 39-0577-587111<br>Fax: 39-0577-318690 | Via delle Rose 10<br>53035 Monteriggioni<br>SI, Italy<br>Tel.: 39-0577-587111<br>Fax: 39-0577-318690 | Via delle Rose 10<br>53035 Monteriggioni<br>SI, Italy<br>Tel.: 39-0577-587111<br>Fax: 39-0577-318690 |
| Section 513<br>Device<br>Classification | 807.87 (C) Class I | 807.87 (C) Class I | 807.87 (C) Class I | 807.87 (C) Class I |
| Product code | JTW | JTW | JSM | JTW |
| Panel | Microbiology | Microbiology | Microbiology | Microbiology |
| Reason For Pre-<br>market<br>Notification | New device | New device | New device | New device |
#### 2 Predicate Device
| | Coproset<br>Salmonella | Coproset<br>Shigella | Uriset preservative<br>tubes | Swab set<br>General use |
|--------------------------------|-----------------------------|------------------------------|--------------------------------------------------------------------------|--------------------------|
| Predicate<br>Device Name | BD Difco™<br>Selenite Broth | BD Difco™ GN<br>Broth, Hajna | BD Vacutainer™ PLUS Plastic<br>Urine C &S preservative Tubes and<br>Kits | BD Bacto™<br>Eugon Broth |
| Predicate<br>Device 510(k) | K781304 | K781304 | K024240 | K781304 |
| Product Code | JSN | JSN | JSM | JSN |
| Predicate<br>Device<br>Company | BD Diagnostic<br>System | BD Diagnostic<br>System | BD Diagnostic System | BD Diagnostic<br>System |
{3}------------------------------------------------
#### SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE 3 SUBSTANTIAL EQUIVALENCE
#### Device Description 3.1
## Coproset
disposable in vitro diagnostic device for the collection and transportation of stool samples
and the enrichment of pathogenic organisms present in the sample.
Coproset Salmonella: In vitro Disposable diagnostic device with Selenite Broth intended for the collection and transport, from the collection site to the testing laboratory, of stool samples and for the enrichment of Salmonella spp.
Coproset Shigella: In vitro Disposable diagnostic device with GN Broth intended for the collection and transport from the collection site to the testing laboratory of stool samples and for the enrichment of Shigella spp.
## Uriset Preservative tubes with holder
Disposable in vitro diagnostic device for the collection and transport of urine samples from the collection site to the testing laboratory.
## Swab set General use
Disposable in vitro diagnostic device with Eugon Broth, intended for the collection and transport of pathogenic agents collected by swabs from the collection site to the testing laboratory and for enrichment of Streptococcus pyogenes, Staphylococcus aureus, Candida spp., Streptococcus agalactiae.
#### 3.2 Synopsis of Test methods
# Coproset
COPROSET has been designed for the collection and transport of stool samples and enrichment of the enteric pathogen which may be present.
Stool samples are collected with the spoon and maintained in liquid medium in optimal conditions during transport. The same container is used for enrichment of the pathogen to be detected. , after subsequent streaking on agar plates and appropriated incubation.
# Uriset Preservative tubes with holder
URISET is a calibrated vacuum-filled tube designed for the collection and transportation of urine samples for the detection of pathogenic agents.
Using the tube with the holder, the urine sample can be collected in conditions of safety for the operator and for the sample itself.
The tube fills with about 4 ml of urine and the preservative stabilizes the bacterial load.
{4}------------------------------------------------
## Swab set General use
SWAB SET has been designed for the collection and transportation of samples collected on bacteriological swabs, and enrichment of the pathogen which may be present. Samples are collected using the swab accessory and maintained in liquid medium in optimal conditions during transportation.
#### Substantial Equivalence 3.3
As reported in M40-P NCCLS a Transport System is a device that ensures the integrity of the specimen and allows for a safe handling during the interval between specimen collection and processing the specimen in the laboratory. It consists of the liquid or semiliquid transport medium, the container of the transport system and in some cases, also sampling device.
Transport medium: Liquid or semisolid medium designed to preserve and maintain the integrity of the specimen for the time period between specimen collection and laboratory processing of the sample.
Following this definition and considering that Diesse Coproset and Swabset devices ad their respective predicate devices have similar Intended use, indication for use, similar features and that the media are the same , we can consider DIESSE Device Substantially equivalent to their respective predicate device.
Performance evaluation, further demonstrate the substantial equivalence and the safety and effectiveness of Diesse Devices when compared with the respective Predicate Device: in the case of Coproset devices and Swab set device viability of pathogen is maintained and the results obtained are comparable with those obtained for predicate device;
Uriset preservative tubes with holder have the same intended use, Indication for use, material and design than its predicate device and thus it possible to conclude that they are substantially equivalent.
The results obtained in the comparison between Uriset preservative tube and Vacutainer demonstrate that DIESSE Uriset give a good performance in terms of sample preservation and data are comparable with those obtained for the predicate device.
Signature
Date:
Dr. Francesco Cocola Director of Quality Unit/ Regulatory Affairs Director, Research and Development
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Coproset Salmonella, Coproset Shigella, Uriset preservative tubes, Swab set General Use
Indications For Use: Coproset Salmonella
In vitro Disposable diagnostic device with Selenite Broth intended for the collection and transport, from the collection site to the testing laboratory, of stool samples and for the enrichment of Salmonella spp.
### Coproset Shigella
In vitro Disposable diagnostic device with GN Broth intended for the collection and transport from the collection site to the testing laboratory of stool samples and for the enrichment of Shigella spp
### Uriset preservative tubes
Disposable in vitro diagnostic device for the collection and transport of urine samples from the collection site to the testing laboratory.
### Swab set General Use
Disposable in vitro diagnostic device with Eugon Broth, intended for the collection and transport of pathogenic agents collected by swabs from the collection site to the testing laboratory and for enrichment of Streptococcus pyogenes, Staphylococcus aureus, Candida spp., Streptococcus agalactiae.
| Prescription Use <span style="text-decoration: underline;"><b>✓</b></span> | AND/OR | Over-The-Counter |
|----------------------------------------------------------------------------|--------|------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 8) |
ounter Use 01 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Lede
Division Sign-Off
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD).
Office of In Vitro Diagnostic Device Evaluation and Safety
sin(k) KD7
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, facing to the left. The eagle is black and has a sleek, modern design. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - U.S.A.", arranged in a circular fashion around the eagle. The text is in all capital letters and is also black.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Diesse Diagnostica Senese S.P.A. c/o Casey Conroy Senior Project Engineer Underwriters Laboratories, Inc. 1285 Walt Whitman Road Melville, New York 11747
JUL 3 0 2007
Re: k070062
Trade/Device Name: Coproset Salmonella, Coproset Shigella, Uriset Preservative Tubes, Swab Set General Use Regulation Number: 21 CFR 866.2900 Regulation Name: Microbiological Specimen Collection and Transport Device Regulatory Class: Class I Product Code: LIO, JTW Dated: July 16, 2007 Received: July 18, 2007
Dear Mr. Conroy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{7}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If vou desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sally, anton
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Coproset Salmonella, Coproset Shigella, Uriset preservative tubes, Swab set General Use
Indications For Use: Coproset Salmonella
In vitro Disposable diagnostic device with Selenite Broth intended for the collection and transport, from the collection site to the testing laboratory, of stool samples and for the enrichment of Salmonella spp.
### Coproset Shigella
In vitro Disposable diagnostic device with GN Broth intended for the collection and transport from the collection site to the testing laboratory of stool samples and for the enrichment of Shigella spp
#### Uriset preservative tubes
Disposable in vitro diagnostic device for the collection and transport of urine samples from the collection site to the testing laboratory.
### Swab set General Use
Disposable in vitro diagnostic device with Eugon Broth, intended for the collection and transport of pathogenic agents collected by swabs from the collection site to the testing laboratory and for enrichment of Streptococcus pyogenes, Staphylococcus aureus, Candida spp., Streptococcus agalactiae.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
KD7006. 11/16/2)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.