BD VACUTAINER PLUS PLASTIC URINE C&S PRESERVAYIVE TUBES AND KITS

K024240 · Bd · JSM · Mar 26, 2003 · Microbiology

Device Facts

Record IDK024240
Device NameBD VACUTAINER PLUS PLASTIC URINE C&S PRESERVAYIVE TUBES AND KITS
ApplicantBd
Product CodeJSM · Microbiology
Decision DateMar 26, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2390
Device ClassClass 1

Indications for Use

The BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tubes and kits are intended for the collection and transport of urine samples for culture and sensitivity testing of bacteria. The BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tube is filled with lyophilized urine maintenance formula and evacuated to draw approximately 4 - 10 mL (depending on the tube size) of urine. The lyophilized urine maintenance formula can maintain the bacterial population in the urine specimen for a period of 48 hours at room temperature levels comparable to those urine specimens without additive, held under refrigeration for the same period of time.

Device Story

Sterile, evacuated plastic tube for urine collection/transport; contains lyophilized urine maintenance formula. Maintains bacterial population for 48 hours at room temperature, equivalent to refrigerated non-additive specimens. Used in clinical laboratory settings for culture and sensitivity testing. Plastic material reduces breakage risk and improves safety compared to glass predecessors. Healthcare providers collect urine into the tube; tube preserves sample integrity for subsequent laboratory analysis. Benefits include reduced specimen breakage and decreased need for repeat collections.

Clinical Evidence

Extensive functional and clinical evaluations performed. Compared subject device to BD Vacutainer™ Brand non-additive tubes and glass C&S preservative tubes. Results demonstrated functional and clinical equivalence in maintaining bacterial populations for 48 hours at room temperature.

Technological Characteristics

Sterile, evacuated plastic tube; contains lyophilized urine maintenance formula. Dimensions/volume: 4-10 mL draw capacity. Principle: Chemical preservation of urine specimens for microbiological analysis. Material: Plastic (vs. glass predicate).

Indications for Use

Indicated for collection and transport of urine samples for bacterial culture and sensitivity testing in patients requiring such diagnostic evaluation.

Regulatory Classification

Identification

A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## K. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS K024240 ### 510(k) Summary Of Safety and Effectiveness #### 1.General Information This Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 § 807.92 #### Establishment: | • Address: | BD Vacutainer Systems, Preanalytical<br>Solutions<br>1 Becton Drive, MC 300<br>Franklin Lakes, NJ 07417-1885 | |------------------------|--------------------------------------------------------------------------------------------------------------| | • Registration Number: | 2243072 | | • Contact Person: | Jing Zhang<br>Manager, Regulatory Affairs<br>Telephone No.:(201) 847-4717<br>Fax No. (201) 847-4858 | | • Date of Summary: | Dec. 20, 2002 | ### Device | • Trade Name: | BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tubes and Kits | |---------------|-------------------------------------------------------------------| |---------------|-------------------------------------------------------------------| - Classification Name: Aerobic Transport System . (Microbiological specimen collection and transport device) - . Classification: Class I - . Performance Standards: None Established under 514 of the Food, Drug and Cosmetic Act {1}------------------------------------------------ #### Safety and Effectiveness Information Supporting the Substantial Equivalence 2. Determination - > Device Description: The BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tube is a sterile, plastic, evacuated urine collection and transport tube for culture and sensitivity testing of bacteria. The benefits of a plastic tube decrease the occurrence of accidental breakage increases the safety of laboratory personnel and reduces the necessity of repeat specimens. - >> Intended Use/Indications for Use: The BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tubes and kits are intended for the collection and transport of urine samples for culture and sensitivity testing of bacteria. The BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tube is filled with lyophilized urine maintenance formula and evacuated to draw approximately 4 - 10 mL (depending on the tube size) of urine. The lyophilized urine maintenance formula can maintain the bacterial population in the urine specimen for a period of 48 hours at room temperature levels comparable to those urine specimens without additive, held under refrigeration for the same period of time. - > Synopsis of Test Methods and Results Extensive functional and clinical evaluations were performed to determine the safety and efficacy of the BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tubes and kits. They were compared to the currently marketed BD Vacutainer™ Brand non-additive tube and Glass Urine C&S Preservative Tubes and kits. The results of the tests demonstrated that the BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tubes provide functionally and clinically equivalent results when compared to the BD Vacutainer™ Brand non-additive tubes and Glass Urine C&S Preservative Tubes. - A Substantial Equivalence Based on comparison of the device features, materials, and intended use, the BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tubes can be shown to be substantially equivalent to the commercially available predicate device identified below: | Manufacturer | Predicate Device | K-Number | Clearance Date | |---------------------------------------------------|---------------------------------------------------------------------|----------|----------------| | BD Vacutainer Systems,<br>Preanalytical Solutions | BD Vacutainer™ Brand Glass Urine<br>C&S Preservative Tubes and Kits | K790336 | 4/4/79 | \$ Jing Zhang Manager, Regulatory Affairs 12/20/02 Date {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three wavy lines that resemble a bird or a symbol representing health and human services. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 # MAR 2 6 2003 Mr. Jing Zhang Manager, Regulatory Affairs Becton, Dickson and company BD Vacutainer Systems, Preanalytical Solutions 1 Becton Drive Franklin Lakes, NJ 07417 - Re: k024240 Trade/Device Name: BD Vacutainer 1M PLUS Plastic Urine C&S Tubes and Kits Regulation Number: 21 CFR 866.2390 Regulation Name: Transport Culture Medium Regulatory Class: Class I Product Code: JSM Dated: December 20, 2002 Received: December 23, 2002 Dear Mr. Zhang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ #### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Putman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # C. INDICATIONS FOR USE 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: BD Vacutainer™ PLUS Plastic Urine C&S Tubes and Kits Indications for Use: The BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tubes and Kits are intended for the collection and transport of urine samples for culture and sensitivity testing of bacteria. The BD Vacutainer™ PLUS Plastic Urine C&S Preservative Tube is filled with lyophilized urine maintenance formula and evacuated to draw approximately 4 - 10 mL (depending on the tube size) of urine. The lyophilized urine maintenance formula can maintain the bacterial population in the urine specimen for a period of 48 hours at room temperature levels comparable to those urine specimens without additive, held under refrigeration for the same period of time. (Please do not write below this line-continue on another page if needed) Or Concurrence of CDRH, Office of Device Evaluation (ODE) V Prescription Use (Per 21 CFR § 801.109) Over-the-Counter Use (Optional format 1-2-96) Freddie Lu Poole (Division Sim-()ff) .. Imeal opy Number _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...