VORTEX CT PORT ACCESS SYSTEM AND LIFEGUARD SAFETY INFUSION SET
K062414 · Rita Medical Systems, Inc. · LJT · Feb 1, 2007 · General Hospital
Device Facts
Record ID
K062414
Device Name
VORTEX CT PORT ACCESS SYSTEM AND LIFEGUARD SAFETY INFUSION SET
Applicant
Rita Medical Systems, Inc.
Product Code
LJT · General Hospital
Decision Date
Feb 1, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vortex® CT Port Access System is indicated for any patient requiring repeated access of the vascular system or other selected body site, for the delivery of medications, nutritional supplementation, fluids, blood, blood products, and the sampling of blood. When used with non Y site LifeGuard Safety infusion sets in 20 Ga or 19Ga sizes, the Vortex® CT Port Access System is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec.
Device Story
Vortex® CT Port Access System is a totally implanted vascular access port; consists of titanium port body with silicone septum and polyurethane or silicone catheter. Used for repeated vascular access; delivery of medications, fluids, blood products; and blood sampling. Enables power injection of contrast media at 5 ml/sec when accessed with specific 19G/20G LifeGuard® safety infusion sets. Port features toroidal reservoir and tangential outlet stem to facilitate flushing. Catheter attached via silicone locking mechanism. Implanted via percutaneous puncture; distal tip placed at superior vena cava/right atrium junction. Physician-operated; used in clinical settings. Output is fluid delivery or contrast media injection for imaging studies. Benefits patient by providing reliable, repeated access for therapy and diagnostic imaging.
Clinical Evidence
No clinical tests were performed. Substantial equivalence is supported by non-clinical bench testing, including sterilization exposure, physiological conditioning, power injection performance (18 samples), simulated power injection (30 samples), dynamic failure, static burst, life cycle power injection, cyclic testing, port patency, and puncture life testing.
Technological Characteristics
Titanium Grade 2 port body; silicone (NuSil MED 4850) septum; 7.5F to 9.6F polyurethane or silicone (NuSil MED 4765) catheters with 15-25% BaSO4 radiopacifiers. Toroidal reservoir design with tangential outlet stem. Silicone locking mechanism. Dimensions include depth markings up to 65 cm. Sterilization via EtO. Non-electronic, mechanical device.
Indications for Use
Indicated for adult patients requiring repeated vascular or body site access for medication, nutrition, fluid, blood product delivery, or blood sampling. Indicated for power injection of contrast media at max 5 ml/sec when used with specific 19G or 20G non-Y site LifeGuard infusion sets.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "RITA" in black letters. To the left of the word is a square with a pattern of lines and dots. The letters are bold and slightly pixelated, giving them a textured appearance. The overall image has a simple, monochrome design.
## 5.0 510(k) Summary
### Manufacturer and sponsor of the 510(k)
RITA Medical Systems, Inc. One Horizon Way Manchester, Georgia 31816 800-472-5221 Phone 706-846-5226 fax Establishment registration number: 1056436
#### Device Identification:
| Proprietary Name: | Vortex® CT Port Access System |
|------------------------|--------------------------------------------------------------------|
| Common Name: | Vascular access port |
| Classification Name: | Subcutaneous, implanted, intravascular infusion port &<br>catheter |
| Classification Number: | 21 CFR §880.5965 |
| Classification Panel: | General Hospital |
| Product Code: | LJT |
| Regulatory Class: | II |
| Proprietary Name: | LifeGuard® Safety Infusion Set |
| Common Name: | Port Access Infusion set |
| Classification Name: | Set, administration, intravascular |
| Classification Number: | 21 CFR §880.5440 |
| Classification Panel: | General Hospital |
| Product Code: | FPA |
| Regulatory Class: | II |
#### Legally marketed device to which equivalence is claimed:
C.R. Bard PowerPort 510(k) K060812
#### Primary contact for correspondence
David Smith Director of Engineering 800-472-5221 ext. 3085 voice 706-846-5226 fax dsmith@ritamed.com
#### Secondary contact
Karen Alexander Director of Regulatory Affairs 800-472-5221 ext. 3023 voice 706-846-5303 fax kalexander@ritamed.com
Page 13 of 78
Kd62414 Pg 1 of 8
- 1 2007
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "RITA" in a bold, sans-serif font. To the left of the word is a square graphic with diagonal lines. The text and graphic are black and the background is white. The image has a slightly pixelated or grainy texture.
K06 2414 Pg 2 of 8
#### Device Description
The legally marketed devices that form the basis for this 510(k) submission is Horizon Medical Products' Vortex® Access System (K010767 cleared April 10, 2001), and the LifeGuard safety infusion set K013871 cleared March 6, 2002).
The Vortex® CT Port is a Titanium port with a self sealing silicone rubber septum designed to maintain integrity after punctures with a non-coring needle. The port has a hollow area, or reservoir, under the septum through which fluid passes during infusion or aspiration.
The Vortex design features a proprietary reservoir with rounded walls giving it a toroidal shape. The outlet stem is located tangential to the reservoir wall allows fluid to pass between the reservoir and the catheter.
The Vortex® Port Access systems offer models with single lumen 7.5 French to 9.6 French catheters made from either polyurethane or silicone. The catheters all contain radiopacifiers, and depth markings.
A more detailed description of the components of the Vortex® Port Access System is as follows;
Port Body: The body of the port is crafted of Titanium. Suture sites are incorporated into the port base to facilitate anchoring to the underlying tissue.
Reservoir: The reservoir is the hollow area under the septum, into which the non-coring needle is inserted for administration or withdrawal of fluids. The toroidal wall shape and a tangential outlet stem interface to facilitate flushing of the system.
Outlet stem: The port outlet stem is made from Titanium. Connection to the stem is accomplished through barbed connectors. The stem is located tangentially to the reservoir. The catheter is placed over stem, and locked into place. This creates a fluid connection with the reservoir.
Port septum: The port septum is the component through which the non-coring needle is inserted. The septum is self sealing, designed to maintain integrity after repeated punctures and is constructed from silicone (NuSil MED 4850).
Catheter: The available catheters are 7.5 or 9.6 French made from silicone (NuSil MED 4765) impregnated with a 15% BaSO4 in the cannula. The catheters include a blue stripe and tip loaded 25% BaSO4 for improved radiopacity. All catheters have depth markings to 65 cm to allow for precise placement.
Locking Mechanism: The cylindrical locking mechanism is made from silicone. The lock is advanced over the catheter until it positively locks onto the port stem, thereby compressing the catheter onto the stem and sealing the system.
Page 14 of 78
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the word "RITA" in a bold, sans-serif font. To the left of the word is a square graphic that appears to be a stylized design. The text and graphic are composed of small dots, giving the image a textured appearance. The overall impression is that of a logo or branding element.
Implantation: The Vortex® Port Access System is totally implanted in the body. The catheter is placed by percutaneous puncture technique with the distal tip at the juncture of the superior vena cava and right atrium via the jugular or subclavian vein. A subcutaneous pocket is prepared for placement of the port. The appropriate catheter length is determined, and the excess is trimmed by cutting straight across the proximal end. The catheter is marked up to 65 cm to allow more precise placement in the desired location. The catheter is attached to the port using the locking mechanism. The port body is secured by suturing through the base to the underlying fascia prior to wound closure.
The Vortex® port has significant physical and performance characteristics that make it an ideal candidate for indication for power injection. These characteristics include; All Titanium construction for maximum strength, barbed outlet stems for ease and simplicity of catheter connection, and maximum connection strength, combination locking mechanism / strain relief for simple locking, and maximum catheter protection.
The LifeGuard safety infusion set is a port access needle set with an integrated proprietary safety feature to prevent re-bound injury. The safety infusion set includes a huber needle, a winged housing, non-DEHP PVC extension legs, and a luer standard connector.
It has come to RITA's attention that ports have and are being used for power injection studies. RITA currently makes no claims, and provides no instructions for this procedure. The purpose of this 510(k), and the studies which have been conducted, is to establish a safe and effective way to conduct a power injection study through ports which are deemed to be appropriate for this type of use.
When used for power injection, the port is accessed, using qualified infusion sets, in the standard fashion, per the standard instructions. The supplemental instructions will then be followed for power injection. Power injection involves the use of a power syringe to inject contrast media in order to complete a study.
#### Intended Use / Indications
The Vortex® CT Port Access System is indicated for any adult patient requiring repeated access of the vascular system or other sclected body site, for the delivery of medications, nutritional supplementation, fluids, blood, blood products, and sampling of blood,
The Vortex® CT ports can accommodate a 5 ml/see injection rate of contrast dye. This should only be attempted using a 20 gauge or 19 gauge non-Y site LifeGuard® in Ision set model #'s LG-19-XXX, and LG-20-XXX.
Page 15 of 78
။ အိမ္း အသိန္းမွန္းမွာေရးမွာ ျပန္မားမာေတာင္းမွာ များမွန္မွာ ျပန္မာျဖစ္မွ ျပန္မာျဖစ္ေျပာင္း ေျခင္း
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the word "RITA" next to a square design. The square design is filled with a pattern of dots and lines. The word "RITA" is written in a bold, sans-serif font and is also filled with a pattern of dots.
Pre-market Notification -510(k) RITA Medical Systems, Inc. Vortex® CT Power Injectable Port Access System - August 14, 2006
## Summary of the technological characteristics of your device compared to the predicate device
The following table summarizes the technological comparison between the Vortex® CT and Bard PowerPort:
| Technological Characteristic | RITA Vortex™ CT | C.R. Bard PowerPort (K060812) |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use, power injection | Accommodates a 5 ml/sec<br>injection rate of contrast dye. This<br>should only be attempted using a<br>20 gauge or larger non-Y site<br>LifeGuard* CT infusion set. | Accommodates a 5 ml/sec<br>injection rate of contrast dye<br>injected at up to 300 psi without<br>damaging the port or catheter. |
| Design | Port system with attachable<br>catheter. | Port system with attachable<br>catheter. |
| Material | Titanium Grade 2 | Titanium |
| Shape | Round port system with tangential<br>outlet | Triangular port system with central<br>outlet |
| Included labeling | Instructions for use (General)<br>Instructions for use power<br>injection<br>Vortex™ CT poster for CT offices<br>Patient information booklet<br>Patient ID card with key fob<br>Patient chart labeling<br>Product label | Educational booklet for patients,<br>nurses, placers, and CT department<br>Wall charts for CT department and<br>placers<br>Discharge kits for patients<br>(includes Patient ID card with key<br>fob and bracelet) |
| Needles used for Access | LifeGuard non-Y site safety<br>infusion sets in 19 or 20 Ga | LiftLock non-Y site safety infusion<br>sets in 19 or 20 Ga |
ი ინტერნეტში
Page 16 of 78
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "RITA" in bold, black letters. To the left of the word is a square filled with a pattern of lines and dots, also in black. The letters are slightly pixelated, giving them a textured appearance.
K062414 Pg 5 of 8
## Summary of the non-clinical performance data
:
.
The following non-clinical tests were performed as the basis for establishing substantial equivalence to the predicate device, as well as safety and effectiveness of the indication for use:
Image /page/4/Figure/5 description: The image shows the text "Page 17 of 78". This text likely indicates the page number within a larger document. The page number is 17, and the total number of pages in the document is 78. The text is written in a clear, legible font.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "RITA" in bold, sans-serif font. To the left of the word is a square graphic with diagonal lines. The text and graphic are black and the background is white.
# Pre-market Notification -510(k)
RITA Medical Systems, Inc.
Vortex® CT Power Injectable
Portex® CT Power Injectable
Port Access System -- August 14, 2006
| Test<br>Description | Sample<br>Size | Test Description | Required Results |
|----------------------------------------|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization<br>exposure | All | The samples shall be exposed to a 2x EtO<br>sterilization cycle. | All samples must be capable of<br>withstanding a 2X sterilization cycle |
| Physiological<br>exposure | All | Maintain samples in a 0.9% saline bath at 37°C for<br>24hrs prior to testing and between any tests.<br>During the initial 24 hour preconditioning, the<br>samples shall be flushed and locked with a<br>Vancomycin solution (1g/1000cc of normal<br>saline). | All samples must be capable of<br>withstanding physiological<br>conditioning |
| Power injection<br>Performance<br>Test | 18 | Inject 200 ml of an 11.8 cP solution through the<br>port at 5ml/sec. Evaluate system for leaks and<br>damage | Maximum Pressure: $(U-X)/s ≥ k$<br>where k = 1.96 for an AQL level of<br>0.65<br>Catheter material failure: no bursts.<br>leaks, / plastic deformations allowed |
| Simulated<br>Power Injection | 30 | Inject 150 ml of an 11.8 cP solution through the<br>port at 5ml/sec with 19 Ga and 20 Ga needles.<br>Evaluate system for leaks and damage | All samples must meet expected<br>flow rate and pressure withstand<br>requirements |
| Dynamic<br>Failure Test | 18 | Inject 200 ml an 11.8 cP solution through the port at<br>increasing flow rates until failure occurs. | Data shall be gathered to support<br>label claims and determine the<br>operational safety factor. |
| Static Burst<br>Test | 18 | Inject 100ml of a saline solution through the port<br>with the catheter occluded. Record the maximum<br>pressure | Data shall be gathered to support<br>label claims. |
| Life Cvcle<br>Power Injection<br>Test | 10 | Inject 200 ml of an 11.8 cP solution through the<br>port at 5ml/sec. Evaluate system for leaks and<br>damage. Repeat the test 9 times | Catheter material failure: no bursts,<br>leaks, / plastic deformations allowed |
| Cyclic testing | 30, 10<br>cycles per<br>port | Inject 150 ml of an 11.8 cP solution through the<br>port at 5ml/sec with 19 Ga needles. Repeat cycle<br>10 times. Evaluate system for leaks and damage. | All samples must withstand 10<br>cycles without leaking or bursting,<br>The expected maximum number of<br>power injection cycles is 5. |
| Port Patency<br>Verification | 5 | Withdraw simulated blood through port needle<br>combination to determine flow rate of fully patent<br>system. | Blood return must be easily and<br>empirically verifiable to establish<br>safety of power injection |
| Port Occlusion<br>Test | 10 | Inject 100ml of a saline solution through the port<br>with the catheter occluded. Record the maximum<br>pressure | Data shall be gathered to support<br>label claims. |
| Puncture Life | 2 | To test the durability of the septum with needles<br>expected to be used in power injection. | Establish label claim for largest<br>needle likely to be used during<br>power injection |
·
Page 18 of 78
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "RITA" in a bold, sans-serif font. To the left of the word is a square graphic with a pattern of diagonal lines. The text and graphic are composed of small dots, giving them a textured appearance.
Kob 24/14 pg 7 of 8
## Summary of the clinical performance data
No clinical tests were performed to determine substantial equivalence.
:
Page 19 of 78
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "RITA" in a bold, sans-serif font. To the left of the word is a square graphic with a pattern of diagonal lines. The word and the graphic are both black and white. The image appears to be a logo or branding element.
K062414 pg 8of8
## Conclusions drawn from the non-clinical performance data
The non-clinical tests demonstrate that the device is as safe, as effective for the modified intended use.
Page 20 of 78
.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David Smith Director, of Vascular Products Engineering RITA Medical Systems, Incorporated One Horizon Way Manchester, Georgia 31816
FEB - 1 2007
Re: K062414
Trade/Device Name: Vortex® CT Port Access Systems and LifeGuard® Safety Infusion Set Regulation Number: 880.5965 Regulation Name: Subcutaneous, Implanted Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT, FPA Dated: January 24, 2007 Received: January 25, 2007
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
Page 2 - Mr. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clars
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "RITA" in a bold, sans-serif font. The letters are black and appear to be made up of small dots or pixels, giving them a slightly textured or grainy appearance. To the left of the word, there is a geometric shape, possibly a triangle or arrow, also composed of similar dots or pixels.
Pre-market Notification response -- 510(k) # K062414 RITA Medical Systems, Inc. Vortex® CT Power Injectable Port Access System
The LifeGuard® Safety Administration Set intended use has been reproduced below in the FDA's required format.
## Indications for Use
510(k) Number (if known): K062414
LifeGuard® Safety Infusion Set Device Name:
Indications for Use:
The LifeGuard® Safety Infusion Sct is an administration set with a non-coring right angle needle and manually activated needle-stick prevention safety mechanism. The device is used to access surgically implanted vascular ports.
The LifeGuard® Safety Infusion Set is indicated for use in the administration of fluids and drugs, as well as blood sampling through surgically implanted vascular ports.
When used with the Vortex® CT Port Access System, the LifeGuard® Safety Infusion Set is also indicated for power injection of contrast media into the central venous system. For power injection of contrast media, only models LG-19-75, LG-19-100, LG-20-75, LG-20-100, and LG-20-150 may be used at a maximum infusion rate of Sml/sec.
Prescription Use AND/OR X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807
Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony D. Watson
K462414
Page 4 of 34
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "RITA" in a stylized font. To the left of the word is a design that looks like a field or a pattern of lines. The letters of the word "RITA" are textured, giving them a slightly rough or grainy appearance. The overall impression is of a logo or branding element.
#### Indications for Use Statement 4.0
The Vortex® CT Port Access System intended use has been reproduced below in the FDA`s required format.
## Indications for Use
510(k) Number (if known): K062414
Vortex® CT Port Access System Device Name:
Indications for Use:
The Vortex® CT Port Access System is indicated for any patient requiring repeated access of the vascular system or other selected body site, for the delivery of medications, nutritional supplementation, fluids, blood, blood products, and the sampling of blood.
When used with non Y site LifeGuard Safety infusion sets in 20 Ga or 19Ga sizes, the Vortex® CT Port Access System is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec.
Prescription Use X X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpari C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
itz Lont
1462414
and and the britain and the comments of the first of the count
Page 11 of 78
이 부분했다. 이 지역 1915년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975년 1975
이 이용 (1) 이름을 하고 있다. 이
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.