The Prisma System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload and for therapeutic plasma exchange in patients with disease where removal of plasma components is indicated.
Device Story
Prisma R03.10A System is a hemodialysis delivery system for continuous renal replacement therapies (SCUF, CVVH, CVVHD, CVVHDF) and therapeutic plasma exchange (TPE). System comprises a control unit and disposable extracorporeal blood circuits (Prisma Sets). Control unit automates loading/priming of sets, pumps blood, delivers anticoagulants, controls fluid/plasma loss, and pumps replacement/dialysate fluids. Used in intensive care units by clinicians. System monitors pressures (arterial, venous, filter, effluent) and calculates Transmembrane Pressure (TMP) to trigger alarms for high pressure or clotting. Output includes real-time monitoring of therapy parameters and system alerts. Benefits include continuous management of fluid overload and solute removal in critically ill patients.
Clinical Evidence
No clinical data provided. Evidence consists of non-clinical bench testing, including unit testing, code inspections, regression testing, and human factors evaluations performed by internal and external personnel.
Technological Characteristics
Hemodialysis delivery system; consists of control unit and disposable extracorporeal blood circuits. Monitors arterial, venous, and filter pressures. Calculates TMP. No internal dialysate mixing or temperature control. Connectivity: Standalone. Software: Embedded control system.
Indications for Use
Indicated for patients with acute renal failure or fluid overload requiring continuous solute/fluid removal, and patients requiring therapeutic plasma exchange. For use in critically ill patients in intensive care units. Must be prescribed by a physician.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
K062090
Page 1 of 5
Traditional 510(k)
Prisma® 3.10 System
# 12.0 510(k) SUMMARY
OCT 1 8 2006
| Submitter's Name | Gambro Renal Products |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address | 10810 West Collins Avenue<br>Lakewood, CO 80215 |
| Establishment<br>Registration Number | 1713683 |
| Date of Summary | July 23, 2006 |
| Telephone Number | (303) 231-4094 |
| Fax Number | (303) 542-5138 |
| Contact Person | Thomas B. Dowell, Manager Regulatory Affairs |
| Name of the Device | Prisma <sup>®</sup> R03.10A System<br>Catalogue Number: 018089-507 |
| Common or Usual Name | Hemodialysis Delivery System |
| Classification Name | Classification Name: High Permeability Hemodialysis System<br>Device Class: II<br>Product Code: 78KDI<br>Regulation Number: 876.5860 |
| Indications for Use | The Prisma System is indicated for continuous solute and/or fluid<br>removal in patients with acute renal failure or fluid overload and for<br>therapeutic plasma exchange in patients with disease where removal<br>of plasma components is indicated. |
| Identification of the<br>Legally Marketed Device<br>(Predicate Device) | Prisma <sup>®</sup> 3.03 System<br>Catalogue Number: 018089-507<br>Classification Name: High Permeability Hemodialysis System<br>Device Class: II<br>Product Code: 78KDI<br>Regulation Number: 876.5860 |
{1}------------------------------------------------
K062090
Page 2 of 5
Traditional 510(k)
Prisma® 3.10 System
### 510(k) SUMMARY, continued
| Device Description | The Prisma System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload and for therapeutic plasma exchange in patients with disease where removal of plasma components is indicated. All treatment administered via the Prisma® System must be prescribed by a physician. The Prisma® System is designed for use on critically ill patients in the intensive care unit to provide the following treatments: SCUD (slow Continuous Ultrafiltrtion), CVVH (Continuous Veno-Venous Hemofiltration), CVVHD (Continuous Veno-Venous Hemodialysis), CVVHDF (Continuous Veno-venous Hemodialfiltration) and TPE (Therapeutic Plasma Exchange). The Prisma® System consists of the Prisma Control Unit and a series of disposable extracorporeal blood circuits (Prisma Sets) to allow four types of continuous renal replacements therapies as well as therapeutic plasma exchange (TPE) therapy. The blood circuit utilized will be dependent on the individual patient's therapy and needs. |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Prisma Control Unit performs the following functions: |
| | 1. Loads and primes the Prisma Set automatically. |
| | 2. Pumps blood through the blood flowpath of the set. |
| | 3. Delivers anti-coagulant solution into blood flowpath. |
| | 4. Controls fluid removal/plasma loss from the patient. |
| | 5. Pumps sterile replacement solution/fluid and/or sterile dialysate. Pumps effluent. |
| | 6. Monitors the system and alerts the operator to abnormal situations through alarms. |
{2}------------------------------------------------
K062090
Page 3 of 5
Traditional 510(k)
Prisma® 3.10 System
:
# 510(k) SUMMARY, continued
## Comparison Table
| | PREDICATE<br>Prisma ® System V 3.03 | MODIFIED DEVICE<br>Prisma ® System R03.10A |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Prisma System is indicated for<br>continuous solute and/or fluid removal<br>in patients with acute renal failure or<br>fluid overload and for therapeutic<br>plasma exchange in patients with<br>disease where removal of plasma<br>components is indicated | The Prisma System is indicated for<br>continuous solute and/or fluid removal<br>in patients with acute renal failure or<br>fluid overload and for therapeutic<br>plasma exchange in patients with<br>disease where removal of plasma<br>components is indicated |
| Dedicated<br>Disposable Sets | M60/M100 (K032431), 66D pre and<br>post dilution sets, PF2000N (PMA<br>P830063) HF1000 | M60/M100 (K032431), 66D pre and<br>post dilution sets, PF2000N (PMA<br>P830063) HF1000 |
| Replacement<br>Solutions Used | Sterile commercial fluid labeled for<br>intravenous injection or solution<br>prepared in the hospital pharmacy.<br>Prescribed by physician. | Sterile commercial fluid labeled for<br>intravenous injection or solution<br>prepared in the hospital pharmacy.<br>Prescribed by physician. |
| Dialysate Solutions<br>Used | AAMI RD-5 Standard dialysate<br>solution, composition prescribed by<br>physician and Prismasate (K013448).<br>None used for TPE. | AAMI RD-5 Standard dialysate<br>solution, composition prescribed by<br>physician and Prismasate (K013448).<br>None used for TPE. |
| Anticoagulantion | Delivered continuously or in bolus into<br>blood path at a point before blood enters<br>the dialyzer. | Delivered continuously or in bolus into<br>blood path at a point before blood<br>enters the dialyzer. |
| Fluid Removal<br>Rate from Patient | SCUF: up to 2 L/hr.<br>CVVH: up to 1 L/hr.<br>CVVHD: up to 1 L/hr.<br>CVVHDF: up to 1 L/hr.<br>TPE: 0 - 1000 ml/hr. | SCUF: up to 2 L/hr.<br>CVVH: up to 1 L/hr.<br>CVVHD: up to 1 L/hr.<br>CVVHDF: up to 1 L/hr.<br>TPE: 0 - 1000 ml/hr. |
| Blood Flow Rate | Up to 180 ml/min. | Up to 180 ml/min. |
| Fluid Replacement<br>Rate | SCUF: 0 L/hr.<br>CVVH: up to 4.5 L/hr.<br>CVVHD: 0 L/hr.<br>CVVHDF: up to 2 L/hr.<br>TPE: up to 2 L/hr. | SCUF: 0 L/hr.<br>CVVH: 0, or 0.1 to 4.5 L/hr.<br>CVVHD: 0 L/hr.<br>CVVHDF: 0, or 0.1 to 2 L/hr.<br>TPE: up to 2 L/hr. |
| Effluent Flow Rate | SCUF: up to 2 L/hr.<br>CVVH: up to 5.5 L/hr.<br>CVVHD: up to 3.5 L/hr.<br>CVVHDF: 10-5500 ml/hr<br>TPE: up to 3 L/hr. | SCUF: up to 2 L/hr.<br>CVVH: up to 5.5 L/hr.<br>CVVHD: up to 3.5 L/hr.<br>CVVHDF: 10-5500 ml/hr<br>TPE: up to 3 L/hr. |
| Primary Solute<br>Removal<br>Mechanism | SCUF: Convection<br>CVVHD: Convection<br>CVVHD: Diffusion<br>CVVHDF: Convection & Diffusion<br>TPE: Convection | SCUF: Convection<br>CVVHD: Convection<br>CVVHD: Diffusion<br>CVVHDF: Convection & Diffusion<br>TPE: Convection |
| | PREDICATE<br>Prisma® System V 3.03 | MODIFIED DEVICE<br>Prisma® System R03.10A |
| Transmembrane<br>Pressure | See table below | See table below |
| Ultrafiltration Rate | Off, 10 - 5500 ml/hr | Off, 10 - 5500 ml/hr |
| Dialysate<br>Temperature | *N/A: See explanation below | *N/A: See explanation below |
| Dialysate<br>Conductivity | **N/A: See explanation below | **N/A: See explanation below |
| Arterial and<br>Venous Pressure | Operating Range -250 to +50 mmHg<br>Accuracy ±10% of reading or ±8<br>mmHg, whichever is greater | Operating Range -250 to +50 mmHg<br>Accuracy ±10% of reading or ±8<br>mmHg, whichever is greater |
:
{3}------------------------------------------------
K062090
Page 4 of 5
Traditional 510(k)
Prisma® 3.10 System
#### 510(k) SUMMARY, continued
## TMP Alarm Limits in Prisma (same for 3_03 & 3_10)
| Alarm | Default | Option |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| "TMP Too High"<br>Advisory Limit | +350 mmHg | +70 to +350 mmHg<br>Increment: 10 mmHg |
| "Filter Is Clotting"<br>Advisory Limits | Advisory alarm occurs | One or more limits are reached. |
| a) Filter pressure dropb (ΔP filter) variation | 100 mmHg | a) User settable: 10 to 100 mmHg<br>greater than initial ΔP.<br>Increment: 10 mmHg |
| b) TMP increase | Default: 150 mmHg | Service settable: 50 to 200 mmHg<br>greater than initial TMP.<br>Increment: 5 mmHg |
| Filter Clotted" Warning<br>Limit | Filter pressure minus return<br>pressure is ≥ 250 mmHg OR One<br>or both of the "Filter is Clotting"<br>Advisory Limits are reached and<br>TMP is ≥ 450 mmHg. | N/A |
| "TMP Excessive" Caution<br>Limit | TMP ≥ 450 mmHg | N/A |
TMP = (Filter Pressure + Return Pressure)/2 - Effluent Pressure
{4}------------------------------------------------
Koc 2090
Page 5 of 5
Traditional 510(k) Prisma® 3.10 System
#### 510(k) SUMMARY, continued
* Dialysate conductivity is not applicable because the Prisma does not mix water and concentrates to produce dialysate. There is also no temperature control or monitoring included in Prisma. Accessory warmers for Prisma include appropriate protective measures. There are no conductivity or temperature controls or monitoring.
** The term Access Pressure in Prisma is equivalent to Arterial Pressure; the term Return Pressure in Prisma is equivalent to Venous Pressure. Prisma includes an Effluent Flow Rate in therapies where dialysate is not used, the effluent rate is the ultrafiltration rate (everything pulled in to the effluent bag is coming across the dialyzer membrane). In therapies where there is also a dialysate flow (CVVHD, CVVHDF), the ultrafiltration rate is the difference between effluent rate and the dialysate rate.
TMP (CRRT) and TMPa (TPE) are values calculated from measured pressures. There is no control to TMP (TMPa) in the system, but there are various alarms that are triggered by the calculated TMP.
Description and Nonclinical Testing: The nonclinical testing included unit testing, code inspections, testing Conclusion of Testing targeted to the changes implemented in R03.10A, regression testing and human factors evaluations and testing that was performed by internal and external independent personnel with the appropriate skills.
#### Conclusion:
The successful non-clinical testing demonstrates the safety and effectiveness of the Prisma® R03.10A System when used for the defined indications for use and demonstrates that the device for which the 510(k) is submitted performs as well as or better than the legally marketed device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized depiction of an eagle or bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the emblem in a circular fashion. The text is in a simple, sans-serif font.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
# OCT 1 8 2006
Mr. Thomas B. Dowell Regulatory Project Manager Gambro Renal Products 10810 W. Collins Avenue LAKEWOOD CO 80215
Re: K062090
Trade/Device Name: Prisma® R03.10A System Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: July 23, 2006 Received: July 24, 2006
#### Dear Mr. Dowell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce arior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that prove to the first and in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket 11 your ab, it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/5/Picture/10 description: The image is a circular logo with the text "FDA Centennial" and the years "1906-2006". The letters "FDA" are in a bold, stylized font in the center of the logo. Below the letters, the word "Centennial" is written in a cursive font. There are three stars below the word "Centennial". The logo is surrounded by a dotted circle.
*Protecting and Promoting Public Health*
{6}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Traditional 510(k) Prisma® 3.10 System
Indications for Use
510(k) Number (if known): K062080
Device Name: Prisma® R03.10A System
Indications for Use:
"The Prisma® System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload and for therapeutic plasma exchange in patients with disease where removal of plasma components is indicated."
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David h. Ingram
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Devic
510(k) Number.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.