K061726 · Lifestand · IPL · Dec 4, 2006 · Physical Medicine
Device Facts
Record ID
K061726
Device Name
LSC, MODEL 3
Applicant
Lifestand
Product Code
IPL · Physical Medicine
Decision Date
Dec 4, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LSC offers electrically operated seated and standing mobility to users with ambulatory impairments, including people with spinal cord injury, spina bifibia, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc.
Device Story
LSC is a powered standup wheelchair; provides seated and standing mobility. Input: user-controlled steering/speed via two back wheels; electric linear actuator system transitions seat between seated and standing positions. Output: physical mobility and postural change. Used in clinical or home settings; operated by patient. Benefits: enables standing posture for individuals with ambulatory impairments. Healthcare provider uses device to facilitate patient mobility and postural independence.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via testing to EN 12182, EN 12184, and ISO 7176 series (stability, brakes, energy consumption, dimensions, speed/acceleration, fatigue strength, obstacle climbing, power/control systems, electromagnetic compatibility). Biocompatibility testing performed per EN ISO 10993-1 and 10993-5.
Technological Characteristics
Rigid magnesium frame with epoxy paint; polyester fireproof upholstery (M4). Electric propulsion and elevation via linear actuator. Solid front (200mm x 50mm) and rear (350mm x 70mm) wheels. Automatic electro-magnetic and manual parking brakes. Dimensions: 67cm width, 87-109cm length. Weight: 94.7kg. Max user weight: 120kg. Complies with ISO 7176 series and EN 12184 standards.
Indications for Use
Indicated for users with ambulatory impairments, including spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, and rheumatism.
Regulatory Classification
Identification
A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.
{0}------------------------------------------------
K061726
Page 1 of 3
EXHIBIT #1
Image /page/0/Picture/1 description: The image shows a logo with the word "LIFESTAND" at the bottom. Above the text, there is an abstract graphic that appears to be composed of vertical lines of varying heights. The lines are arranged in a way that suggests a stylized representation of a cityscape or a series of standing figures. The overall design is simple and modern.
LifeStand "Vivre-Debout" Rond-point de Rosarge 40, rue Palverne - 01700 LES ECHETS-FRANCE Tel : +33(0)4 37 26 27 28 Fax :+33(0)4 37 26 27 29 VAT :FR 23 312 906 613 00079
BEC - 4 2006
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is:________________________________________________________________________________________________________________________________________________
- 1. Submitter's Identification:
Lifestand Rond-point de Rosarge 40, rue Palverne F 01700 LES ECHETS-FRANCE
Date Summary Prepared: June 12, 2006
- 2. Name of the Device: LSC
- 3. Common or Usual Name:
powered standup wheelchair
### 4. Device Description:
The LSC is a powered standup wheelchair. It is propelled and steered by varying the speed of the two back wheels. Front castors support the front of the caain and allow indirect steering through the turning back wheels. An electric lineanan anot system puts the seat into a seating or standing position.
| Maximum end-user weight | : | 120 kg |
|-------------------------|---|---------------------------------------------|
| Wheelchair width | : | 67 cm |
| Wheelchair length | : | 87-109 cm |
| Frame | : | Rigid, in magnesium, epoxy paint |
| Seat | : | Depth adjustable, with sore proof cushion |
| Backrest | : | Inclinable. Folds down for transport |
| Upholstery | : | Polyester fireproof material (M4), washable |
| Foot-rests | : | Height adjustable |
| Front wheels | : | Ø 200mm x 50mm, solid |
| Rear wheels | : | Ø 350mm x 70mm, solid |
| Brakes | : | automatic electro-magnetic brake and |
| | | manual standard parking brake |
| Propulsion | : | electric powered |
| Elevation | : | electric powered |
| Rear stabilization | : | Anti-tip wheels (optional). |
| Idle weight | : | 94.7kg |
### 5. Intended Use:
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with the text "LIFESTA" at the bottom. Above the text, there are three vertical lines of varying heights. The lines appear to be connected at the top, forming a stylized graphic element. The image is in black and white.
The LSC offers electrically operated seated and standing mobility to users with ambulatory impairments, including people with spinal cord injury, spina bifibia, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc.
### 6. Comparison to Predicate Devices:
The LSC is substantially equivalent to its forerunner model LSC Compact by Lifestand, K041535 and the standup wheelchair LCM by LEVO, K963817
- Discussion of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence are as follows:
To approve the performance of the LSC, tests according to current applicable standards where performed at test-laboratories of European notified bodies:
| EN 12182: 1999 | Technical aids for disabled persons. General requirements<br>and test methods |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| EN 12184: 1999 | Electrically powered wheelchairs, scooters and their chargers<br>-requirements and test methods |
| ISO 7176-1: 1999 | Wheelchairs. Determination of static stability |
| ISO 7176-2: 2001 | Wheelchairs. Determination of dynamic stability of electric<br>wheelchairs |
| ISO 7176-3: 1988 | Wheelchairs. Determination of effectiveness of brakes |
| ISO 7176-4: 1997 | Wheelchairs. Energy consumption of electric wheelchairs and<br>scooters for determination of theoretical distance |
| ISO 7176-5: 1986 | Wheelchair tests. Methods for determination of overall<br>dimensions, mass and turning space |
| ISO 7176-6: 2001 | Wheelchairs. Determination of maximum speed, acceleration<br>and deceleration of electric wheelchairs |
| ISO 7176-7: 1998 | Wheelchairs. Measurement of seating and wheel dimension |
| ISO 7176-8: 1998 | Wheelchairs. Requirements and test methods for static,<br>impact and fatigue strength |
| ISO 7176-9: 2001 | Wheelchairs. Climatic test for electric wheelchairs |
| ISO 7176-10: 1988 | Wheelchairs. Determination of obstacle-climbing ability of<br>electric wheelchairs |
| ISO 7176-14: 1997 | Power and Control systems for electric wheelchairs –<br>Requirements and test methods |
| ISO 7176-15: 1996 | Wheelchairs. Requirements for information disclosure,<br>documentation and labeling |
| ISO 7176-16: 1997 | Wheelchairs - Part 16: Resistance to ignition of upholstered<br>parts - Requirements and test methods |
| ISO 7176-20: 2001 | Wheelchairs. Determination of the performance of stand-up<br>type wheelchairs |
| ISO 7176-21: 2003 | Wheelchairs. Requirements and test methods for<br>electromagnetic compatibility of electrically powered<br>wheelchairs and motorized scooters |
| EN ISO 10993-1: 2003 | Biological evaluation of medical devices. Evaluation and<br>testing |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the word "LIFESTAND" at the bottom. Above the text, there is an abstract graphic that appears to be three vertical lines of varying heights. The lines are not solid but seem to be made up of smaller elements, giving them a textured appearance.
K061726
Page 3 of 3
Biological evaluation of medical devices. Tests for in vitro EN ISO 10993-5: 1999 cytotoxicity
- Discussion of Clinical Tests Performed: 8. Clinical tests were not performed
- 9. Conclusions:
Lifestand believes that the LSC is substantially equivalent to the predicate and is safe and effective for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 4 2006
LifeStand "Vivre-Debout" % Ms. Stefanie D. Bankston Rond-point de Rosarge 40, rue Palverne F 01700 Les Echets- France
Re: K061726
Trade/Device Name: LSC Regulation Number: 21 CFR 890.3900 Regulation Name: Standup wheelchair Regulatory Class: Class II Product Code: IPL Dated: November 20, 2006 Received: November 28, 2006
Dear Ms. Bankston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Ms. Stefanie D. Bankston
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours
Cahmz
Mark N. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Exhibit #3
Page ﻠﺴﻴ of 1
K061726 510(k) Number (if known):
Device Name: LSC
Indications For Use:
The LSC offers electrically powered seated and standing mobility to users with ambulatory impairments, including people with spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc.
Prescription Use _ × (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use (21 CFT 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K061726
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number. K061726
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.