The LSC Lifestand Compact offers electrical powered seated and standing mobility to users with ambulatory impairments, including people with spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc.
Device Story
LSC Lifestand Compact is a powered standup wheelchair for users with ambulatory impairments. Device features rigid treated steel chassis, adjustable tibia support, seat, backrest, and armrests. Propulsion and steering achieved via two independent rear-wheel motors controlled by standard joystick-based e-fix controller. Two 12V gel batteries provide power. Two linear actuators enable transition between seated and standing positions and adjust backrest inclination. Operated by user via joystick. Provides mobility and standing capability to assist with daily activities. Bench testing performed per ISO 7176 series and EN standards to verify stability, braking, and electrical safety.
Clinical Evidence
No clinical data. Safety and effectiveness established through non-clinical bench testing, including ISO 7176 series (stability, braking, dimensions, fatigue, obstacle climbing), EN 12184, EN 12182, and electromagnetic compatibility standards (ANSI/RESNA WC/Vol. 2-1998, EN 61000-4-2, EN 60601-4-3). Biocompatibility testing performed per EN ISO 10993-5.
Technological Characteristics
Chassis: treated steel with epoxy resin. Power: 2x12V 17Ah rechargeable electrolyte gel batteries. Motors: 24V DC e-fix motors. Actuation: two linear actuators (Linac LA 31.1, Linac 314210). Brakes: manual and electromagnetic. Dimensions: 7" front wheels, 12.5" rear wheels. Max user weight: 120kg. Connectivity: standard joystick controller. Standards: ISO 7176 series, EN 12184, EN 12182, EN 60335, EN 60601-4-3.
Indications for Use
Indicated for users with ambulatory impairments, including spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, and rheumatism, requiring electrical powered seated and standing mobility.
Regulatory Classification
Identification
A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.
{0}
K04/535 1/4
SEP 16 2004
EXHIBIT #1
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: ______.
## 1. Submitter's Identification:
Lifestand
Rond Point de Rosarge
40, rue Palverne
F-01700 Les Echets - France
Tel: +33(0)4 37 26 27 28
Fax: +33(0)4 37 26 27 29
Date Summary Prepared: June 7, 2004
## 2. Name of the Device:
LSC Lifestand Compact
## 3. Common or Usual Name:
Electrical powered standup wheelchair
## 4. Device Description:
The LSC Lifestand Compact is a powered standup wheelchair. It is propelled and steered by varying the speed of the two back wheels using independent motors. Front castors support the front of the chair and allow indirect steering through the turning back wheels. The system is controlled with a standard e-fix wheelchair controller, with direct user operation through standard wheelchair joystick. Two high quality maintenance-free watertight rechargeable electrolyte gel batteries, connected in series, supply the energy for the system. A Linear actuator drives the seat into seating or standing position. Another linear actuator drives the backrest into an individual seating-inclination. Both linear actuators are also direct user operated through the standard wheelchair joystick.
| Chassis: | Rigid, made from treated steel with epoxy resin paint |
| --- | --- |
| Tibia support: | adjustable in height and inclination |
| Seat: | Depth adjustable with cushion |
| Backrest: | electrically adjustable |
| Armrest: | Retractable, can be converted to adjustable chest support |
| Footrest: | Height can be adjusted |
Page 1 of 4
{1}
K041535
2/4
Upholstery: Fire-resistant (M4), washable fabric
Front wheels: Ø 7” x 1 3/4”
Rear wheels: Ø 12 1/2” x 2 1/4”
Brakes: manual by pushing and electromagnetic in the back wheels
User weight: max. 120kg
Idle weight: 76kg
Batteries: 2x12V, 17Ah, watertight rechargeable electrolyte gel batteries
Driving-motor: e-fix by Ulrich Alber GmbH + Co. KG – Germany, 24V DC
Linear actuator seat: Linac LA 31.1
Linear actuator back: Linac 314210
## 5. Intended Use:
The LSC Lifestand Compact offers electrical powered seated and standing mobility to users with ambulatory impairments, including people with spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc.
## 6. Comparison to Predicate Devices:
The LSC Lifestand is substantially equivalent in design and intended use to the standup wheelchair LCM by LEVO, K963817
## 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:
The following tests where performed on the LSC standup wheelchair to approve its safety:
EN 12184: 1999 Electrically powered wheelchairs, scooters and their chargers - requirements and test methods
EN 12182: 1999 Technical aids for disabled persons- general requirements and test methods
ISO 7176-1: 1999-10-01 Determination of static stability
ISO 7176-2: 2001-06-15 Determination of dynamic stability of electric wheelchairs
ISO 7176-3: 1988-11-15 Determination of efficiency of brakes
ISO 7176-4: 1997-12-15 Energy consumption of electric wheelchairs and scooters for determination of theoretical distance
ISO 7176-5: 1986-03-01 Determination of overall dimensions, mass and turning space
ISO 7176-6: 2001-10-01 Determination of maximum speed, acceleration and deceleration of electric wheelchairs
ISO 7176-7: 1998-05-15 Measurement of seating and wheel dimensions
Page 2 of 4
{2}
K041535 3/4
ISO 7176-8: 1998-07-15 Static, impact and fatigue strength
ISO 7176-9: 2001-10-15 Climatic tests for electric wheelchair
ISO 7176-10: 1988-11-15 Determination of obstacle-climbing ability of electric wheelchairs
ISO 7176-14: 1997-10-15 Power and control systems for electric wheelchairs – Requirements and test methods
ISO 7176-15: 1996-11-00 Requirements for information disclosure, documentation and labeling
ISO/CD 7176-20: 2001-07-06 Determination of the performance of stand-up type wheelchairs
ANSI/RESNA WC/Vol. 2-1998 Section 21 Requirements and Test Methods for Electromagnetic Compatibility of Electric Wheelchairs and Scooters
EN 55011: 1998 Limits and methods of measurement of radio disturbance characteristics of information technology equipment
CISPR 11: 1997 Industrial, scientific and medical (ISM) radio-frequency equipment – Radio disturbance characteristics – Limits and methods of measurement, Amendment No. 1 (1999)
EN 61000-4-2: 1995 Electromagnetic Compatibility (EMC), Part 4: Testing and measuring techniques, Section 2: Electrostatic discharge immunity test
EN 60601-4-3: 1996 Electromagnetic Compatibility (EMC), Part 4: Testing and measuring techniques, Section 3: Radiated, radio-frequency electromagnetic field immunity test
EN 60335 : 1995 Specification for safety of household and similar electrical appliances
EN 50081-1: 1993 Electromagnetic compatibility. Generic emission standard. Residential, commercial and light industry
EN 50082-2: 1996 Electromagnetic compatibility. Generic immunity standard. Industrial environment
EN ISO 10993-5: 1999 Biological evaluation of medical devices. Tests for in vitro cytotoxicity
NFX 41002 Resistance to salt spray fog
NFP 92503 flammability
NFP 92505 flammability
ISO 6941:1984 Textile fabrics. Burning behavior. Measurement of flame spread properties of vertically oriented specimens
Page 3 of 4
{3}
K041535 4/4
8. **Discussion of Clinical Tests Performed:**
Clinical tests were not performed
9. **Conclusions:**
Lifestand believes that the LSC Lifestand Compact is substantially equivalent to the predicate and is safe and effective for it’s intended use.
Page 4 of 4
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
SEP 16 2004
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Lifestand
C/o Carolann Kotula
MDI Consultants, Inc.
55 Northern Boulevard, Suite 200
Great Neck, New York 11021
Re: K041535
Trade/Device Name: LSC LifeStand Compact
Regulation Number: 21 CFR 890.3900
Regulation Name: Standup wheelchair
Regulatory Class: II
Product Code: IPL
Dated: September 2, 2004
Received: September 3, 2004
Dear Ms. Kotula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}
Page 2 – Ms. Carolann Kotula
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
for Celia M. Witten, Ph.D., M.D.
Director
Division of General, Restorative
and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{6}
Exhibit #5
Page 1 of 1
510(k) Number (if known): K041535
Device Name: LSC LifeStand Compact
Indications For Use:
The LSC LifeStand Compact offers electrical powered seated and standing mobility to users with ambulatory impairments, including people with spinal cord injury, spina bifida, cerebral palsy, multiple sclerosis, muscular dystrophy, polio, rheumatism, etc.
Prescription Use ☑
(Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use ☐
(21 CFT 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number K041535
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.