The RTA 5 & RTA Model E Retinal Thickness Analyzer ("RTA 5 & RTA Model E") is a computerized slitlamp biomicroscope that is intended to provide manual and computerized tomography of the retina in vivo. The RTA 5 & RTA Model E scans successive slit images on the fundus, without the need for a contact lens, to determine the thickness and the inner structure of the retina, both by observation of the slit images and by computer analysis of these images. It is indicated for assessing the area and location of retinal thickness abnormalities, such as thickening due to macular edema and atrophy associated with degenerative diseases, and for visualizing other retinal pathologies.
Device Story
Computerized slitlamp biomicroscope for in vivo retinal tomography. System uses 532 nm green solid-state laser and conventional light sources to project thin slit beams onto fundus. Optical head with scanner and digital camera captures successive slit images; computer system performs analysis to determine retinal thickness and inner structure. Used in clinical settings by eye care professionals. Output includes visual slit images and computerized tomographic data for assessment of retinal thickness abnormalities (e.g., macular edema, atrophy). Enables non-contact visualization and quantification of retinal pathologies, aiding clinical decision-making regarding degenerative eye diseases.
Clinical Evidence
Bench testing only. Verification and validation testing performed to assess modifications to the previously cleared predicate device.
Technological Characteristics
Electro-optical system; 532 nm green solid-state laser; slitlamp biomicroscope form factor; patient-contacting materials: C-Flex Opaque (thermoplastic elastomer); non-contact imaging; computerized analysis of slit images.
Indications for Use
Indicated for assessing area and location of retinal thickness abnormalities, including macular edema and atrophy associated with degenerative diseases, and for visualizing retinal pathologies in patients requiring retinal tomography.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K061674
# AUG 2 3 2006
### 510(k) SUMMARY
## Talia Technology, Ltd.'s RTA 5 & RTA Model E Retinal Thickness Analyzer
### Contact Information:
| Submitter: | Talia Technology, Ltd.<br>2 Yodfat Lod, Israel<br>Phone: 011-972-892-090-40<br>Facsimile: 011-972-891-510-09<br>Email: eamoyal@talia.com |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. Efi Amoyal, QA Manager |
| Name of The Device: | RTA 5 & RTA Model E Retinal Thickness Analyzer |
| Common or Usual Name: | Retinal Thickness Analyzer |
| Classification Name: | Ophtalmoscope, AC-Powered (Product Code HLI) |
| Predicate Devices: | Talia Technology Ltd.'s RTA Retinal Thickness Analyzer |
## Intended Use:
The RTA 5 & RTA Model E Retinal Thickness Analyzer ("RTA 5 & RTA Model E") is a computerized slitlamp biomicroscope that is intended to provide manual and computerized tomography of the retina in vivo. The RTA 5 & RTA Model E scans successive slit images on the fundus, without the need for a contact lens, to determine the thickness and the inner structure of the retina, both by observation of the slit images and by computer analysis of these images. It is indicated for assessing the area and location of retinal thickness abnormalities, such as thickening due to macular edema and atrophy associated with degenerative diseases, and for visualizing other retinal pathologies.
# Device Description, Principles of Operation, and Technological Characteristics:
The RTA 5 & RTA Model E is a computerized electro-optical system comprised of two primary components, namely the optical head and the computer system. The main elements of the optical head include laser and conventional light sources, optics, a scanner, and a digital camera.
{1}------------------------------------------------
The RTA 5 & RTA Model E is a computerized slitlamp biomicroscope that provides manual and computerized tomography of the retina in vivo. The RTA 5 & RTA Model E scans successive slit images of the fundus to determine the thickness and the inner structure of the retina, both by observation of the slit images and by computer analysis of these images. The RTA 5 & RTA Model E uses a solid-state laser source that emits green light at a wavelength of 532 nm. The beam is focused into a thin slit and, by means of a beam-splitter, is directed toward the eye. The scanner and optics then detect the image of the illuminated portion of the retina and transmit the image to the digital camera. The digital camera then captures the image, where it can then be stored and analyzed by the computer system.
### Substantial Equivalence:
The RTA 5 & RTA Model E is a modification to the previously cleared RTA Model D Retinal Thickness Analyzer. The only differences between the previously cleared RTA and the modified RTA 5 & RTA Model E are:
- 1. An additional scanning procedure was added with the capability to scan 24 sequential positions with overall coverage of 6x3 mm (HxV) on the retina. The basic scan of 16 sequential positions with overall coverage of 3x3 mm (HxV) on the retina was preserved and has not been changed.
- 2. Stereo Angle can be set at 9.8° or 5.4°, compared to 12.2° and 7.4°, respectively. in Model D.
- 3. Filament Light Illumination the light illumination intensity on the retina was reduced for patient convenience to allow better penetration through a small pupil and an undilated pupil.
- 4. Target Mechanism new target mechanism allows grabbing of fundus image without dark pattern artifact on the fundus image.
- 5. Human Engineering and External Product Design
- 6. Material change for patient contacting materials from C-Flex R70-091 to C-Flex Opaque, both of which are primarily composed of thermoplastic elastomer (TPE).
- 7. Software Change
Through design control assessment, including verification and validation testing, Talia has demonstrated that the modifications to the cleared RTA do not raise any new questions of safety or effectiveness. Accordingly, the RTA 5 & RTA Model E is substantially equivalent.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle with its wings spread. The eagle is facing to the right and has three lines representing its feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 3 2006
Talia Technology, Ltd. c/o Jonathan S. Kahan Hohan & Hartson L.L.P. 555 Thirteenth St. N.W. Washington, DC 20004-1109
K061674 Re:
Trade/Device Name: RTA 5 & RTA Model E Retinal Thickness Analyzer Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope, AC-Powered Regulatory Class: II Product Code: HLI Dated: August 14, 2006 Received: August 14, 2006
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device wf nave roviewed your we determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for ass based of to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dicrerer, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r toase of actived that I Drimination that your device complies with other requirements of the Act that i Drederal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (21 CFR Part 820); and if applicable, the electronic forth in the quarty by ovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 – Jonathan S. Kahan
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Eychler SimMD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### ATTACHMENT 8
### Indications For Use Statement
510(K) Number (if known):
Device Name:
RTA 5 & RTA Model E Retinal Thickness Analyzer
Indications for Use:
The RTA 5 & RTA Model E Retinal Thickness Analyzer ("RTA 5 & RTA Model E") is a computerized slitlamp biomicroscope that is intended to provide manual and computerized tomography of the retina in vivo. The RTA 5 & RTA Model E scans successive slit images on the fundus, without the need for a contact lens, to determine the thickness and the inner structure of the retina, both by observation of the slit images and by computer analysis of these images. It is indicated for assessing the area and location of retinal thickness abnormalities, such as thickening due to macular edema and atrophy associated with degenerative diseases, and for visualizing other retinal pathologies.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Denis L. Mc Carthy
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
510(k) Number K061674
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.