K052202 · Siemens Medical Solutions USA, Inc. · OWB · Mar 7, 2006 · Radiology
Device Facts
Record ID
K052202
Device Name
AXIOM ARTIS - MODULAR ANGIOGRAPHIC SYSTEM
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
Mar 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro-angiography, general angiography, operating room angiography, multipurpose angiography and radiographic/fluoroscopic procedures e.g. gastro-intestinal imaging, skeletal imaging etc. AXIOM Artis can also support the acquisition of position triggered imaging for spatial data synthesis.
Device Story
AXIOM Artis is a modular angiographic system for diagnostic and interventional imaging. This submission adds a wireless footswitch and voice control interface to the existing system. The wireless footswitch uses RF technology with a proprietary communication protocol and independent microcontroller hardware for safety-oriented signal transmission between the footswitch and the system receiver. The voice control interface allows clinicians to operate image processing functions via voice commands; safety-critical functions like X-ray release and movement control are excluded from voice operation. The system is used in clinical settings (e.g., OR, cath lab) by healthcare professionals. The device provides visual and audible warnings and includes automated error monitoring that blocks system functions if a fault is detected. These additions enhance workflow efficiency for the operator during complex interventional procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness are supported by system design, performance testing, and the use of established, cleared predicate technologies for the wireless and voice control components.
Technological Characteristics
System utilizes RF-based wireless communication for footswitch control and voice recognition for image processing commands. Includes independent microcontroller hardware for safety-oriented signal transmission. System is integrated into the existing AXIOM Artis angiographic platform. Safety features include visual/audible warnings and automated error-based system blocking. Operates within the existing AXIOM Artis modular architecture.
Indications for Use
Indicated for patients requiring diagnostic imaging and interventional procedures, including cardiac, neuro, general, operating room, and multipurpose angiography, as well as radiographic/fluoroscopic procedures such as gastro-intestinal and skeletal imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K052202
# MAR 7 2006 510(K) SUMMARY FOR
## SIEMENS AXIOM ARTIS
# WIRELESS FOOTSWITCH AND VOICE CONTROL
Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355
August 1, 2005
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### 1. Contact Person:
Mr. Gary Johnson Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions USA, Inc 51 Valley Stream Parkway E-50 Malvern, PA 19355-1406 Phone:(601) 448-1778 Fax: (610) 448-1787
#### 2. Device Name and Classification
| Product Name: | AXIOM Artis - Modular Angiographic System |
|-----------------------|-------------------------------------------|
| Classification Name: | Angiographic X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1600 |
| Device Class: | Class II |
| Product Code: | 90 JAA (90 IZI with flat Panel Detector) |
#### 3. Intended Use:
AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures.
Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro-angiography, general angiography, operating room angiography, multipurpose angiography and radiographic/fluoroscopic procedures e.g. gastro-intestinal imaging, skeletal imaging etc.
{1}------------------------------------------------
AXIOM Artis can also support the acquisition of position triggered imaging for spatial data synthesis.
The intended use and indications for use of the AXIOM Artis with wireless footswitch and with voice control as described in its labeling have not changed from its predicate device the AXIOM Artis Modular Angiographic System.
#### 4. Device Description:
The Wireless Footswitch uses RF technology in order to establish a short-range, low-power radio communication link between the footswitch (transmitter) and AXIOM Artis system (receiver). Additional hardware (independent of microcontroller) and Siemens own communication protocol layers will provide a safety-oriented signal transmission.
The Voice Control for the AXIOM Artis family allows for the control of selected functions by voice commands. The primary focus will be on image processing functions. Safety relevant functions such as X-ray release and movement control will not be included in the voice control functions.
Both Wireless Footswitch and Voice Control will be made available on the currently marketed Siemens AXIOM Artis family of products.
#### ട്. Substantial Equivalence:
The Wireless Footswitch and Voice Control are designed for use with the commercially available AXIOM Artis Modular Angiographic systems.
The Siemens wireless footswitch is substantially equivalent to the Stryker Wireless footswitch, 510(k) K033135 with a clearance date of August 9, 2004.
Siemens Voice Control is substantially equivalent to the voice control used on the SIOS -Siemens Integrated Operation System discussed in 510(k) K994231 "cleared on August 10, 2000 and the Philips iU22 Ultrasound System, 510(k) W0425401 with a clearance date of October 04, 2004.
These systems will be incorporated for use with SIEMENS AXIOM Artis Modular Angiographic System cleared via 510(k), K010721 on March 30, 2001 and the SIEMENS AXIOM Artis with FD, 510(k) K021021 with a clearance date of June 6, 2002. The Indication for use of the AXIOM Artis does not change with the above-mentioned additions.
{2}------------------------------------------------
### Summary of Technological Characteristics of the Principal Device as compared with 6. the Predicate Device:
A Wireless Footswitch and a Voice Control will be made available for the AXIOM Artis family. Many of the components used in AXIOM Artis with the Wireless Footswitch and Voice Control are either commercially available with current Siemens systems or include minor modifications to existing components.
#### General Safety and Effectiveness Concerns: 7.
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the AXIOM Artis Modular Angiography System is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed.
Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards. Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle or bird with its wings spread, depicted in a simple, abstract design. The image is in black and white.
Public Health Service
MAR 7 2006 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Gary Johnson Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway E-50 MALVERN PA 19355-1406
Re: K052202
Trade/Device Name: AXIOM Artis with Wireless Footswitch and Voice Control Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: II Product Code: IZI and JAA Dated: February 16, 2006 Received: February 17, 2006
Dear Mr. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device (o proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Nancy C. Bigelow
Nanev C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 3 INDICATION FOR USE
510(k) Number (if known): _KOS 2202_____________________________________________________________________________________________________________________________________________________________________
#### AXIOM Artis with Wireless Footswitch and Voice Control Device Name:
AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures.
Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro-angiography, general angiography, operating room angiography, multipurpose angiography and radiographic/fluoroscopic procedures e.g. gastro-intestinal imaging, skeletal imaging etc.
AXIOM Artis can also support the acquisition of position triggered imaging for spatial data synthesis.
The intended use and indications for use of the AXIOM Artis with wireless footswitch and with voice control as described in its labeling have not changed from its predicate device the AXIOM Artis Modular Angiographic System.
(Please do not write below this line - continue on another page if needed)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of the CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use _
Over-The-Counter Use
(Per 21 CFR §801.109)
Nancy Brogdon
(Division Sian-Off Division of Reproductive, A bdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.