K060924 · Szabocsik and Associates · LPN · Jun 28, 2006 · Ophthalmic
Device Facts
Record ID
K060924
Device Name
JSZ MULTIACTION GP CONTACT LENS SOLUTION
Applicant
Szabocsik and Associates
Product Code
LPN · Ophthalmic
Decision Date
Jun 28, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
The JSZ MultiAction GP Contact Lens Solution is indicated for use in cleaning, rinsing, chemical (not heat) disinfection, storage and conditioning of rigid gas permeable contact lenses, including fluorosilicone acrylate and silicone acrylate lenses, as recommended by your eye care practitioner. JSZ Wetting/Rewetting Eyedrop is indicated to rewet soft (hydrophilic) and to cushion and wet rigid gas permeable contact lenses, including fluoro-silicone acrylate and silicone acrylate lenses, before application and during lens wear. The product may be used with daily wear or extended wear lenses and with disposable lenses or lenses prescribed for frequent replacement.
Device Story
JSZ MultiAction GP Contact Lens Solution and JSZ Wetting/Rewetting Eyedrop are sterile, isotonic ophthalmic solutions. Solutions contain poloxamer, sodium phosphate buffer, sodium chloride, and disodium edetate; preserved with polyquaternium-1 (0.0001%). Products are used by contact lens wearers to clean, disinfect, store, or rewet lenses. No mechanical or electronic components; no software or algorithms. Products are identical in formulation to previously cleared device (K050517).
Clinical Evidence
No clinical data required; substantial equivalence established based on formulation identity to previously cleared devices.
Technological Characteristics
Sterile, isotonic aqueous solution. Ingredients: poloxamer, sodium phosphate buffer, sodium chloride, disodium edetate. Preservative: polyquaternium-1 (0.0001%). No chlorhexidine, thimerosal, or mercury-containing products. Packaging: 2oz/4oz bottles (solution) and 0.5oz/1oz bottles (eyedrop).
Indications for Use
Indicated for users of rigid gas permeable (fluoro-silicone acrylate and silicone acrylate) contact lenses for cleaning, rinsing, disinfection, storage, and conditioning. Also indicated for users of soft (hydrophilic) and rigid gas permeable contact lenses (including daily, extended, disposable, and frequent replacement wear) for rewetting and cushioning before application and during wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
Optimum Cleaning, Disinfecting and Storage Solution
{0}------------------------------------------------
# JUN 2 8 2006
FOI Summary K060924
Júne 6, 2006
## 510(k) SUMMARY
## K060924 JSZ MultiAction GP Contact Lens Solution JSZ Wetting/Rewetting Eyedrop
This summary uses the format provided in 21 CFR 807.92:
(a)(1). Submitter:
John M. Szabocsik, PhD President SZABOCSIK AND ASSOCIATES INC 203 N WABASH AVE STE 1200 CHICAGO IL 60601
Phone 312-553-0828 FAX 312-553-0611
June 6, 2006
Summary prepared:
(a)(2) Device Trade Name:
Device Common Name:
Rigid Gas Permeable contact lens solution Contact lens Wetting/Rewetting Eyedrop
JSZ Wetting/Rewetting Eyedrop
JSZ MultiAction GP Contact Lens Solution
### Device Classification Name:
## JSZ MultiAction GP Contact Lens Solution:
Products, Contact Lens Care, Rigid Gas Permeable (MRC)
# JSZ Wetting/Rewetting Eyedrop:
Accessories Soft Lens Products (LPN) and Products, Contact Lens Care, Rigid Gas Permeable (MRC)
### (a)(3) Identification of Predicate Device:
The JSZ MultiAction GP Contact Lens Solution product is substantially equivalent to other currently marketed rigid gas permeable (GP) contact lens care multipurpose solutions, such as Optimum Cleaning, Disinfecting and Storage Solution, Boston Simplus Multiaction Solution, and Sauflon Delta Plus Multiaction Solution.
The JSZ Wetting/Rewetting Eyedrop is substantially equivalent to other currently marketed wetting/rewetting drops for use with contact lenses, such as Complete Blink-n-Clean Lens Drops.
{1}------------------------------------------------
Both the JSZ MultiAction GP Contact Lens Solution and the JSZ Wetting/Rewetting Eyedrop are identical in composition to the JSZ-Multipurpose Solution previously cleared for market in K050517, which is also identical to Sauflon Delta Plus Multiaction Solution.
### (a)(4) Device Description:
Both products are sterile, isotonic solutions that contain poloxamer, sodium phosphate buffer, sodium chloride, and disodium edetate; preserved with polydum ide 0.0001%. Contain no chlorhexidine, no thimerosal, nor any other mercury containo products.
The products are packaged in sizes appropriate to their intended uses: JSZ MultiAction GP Contact Lens Solution in 2oz (60ml) and 4oz (120ml) bottles, JSZ Wetting/Rewetting Eyedrop in 0.5oz (15ml) and 1oz (30ml) bottles.
# (a)(5) Intended Use (Indications for Use):
The JSZ MultiAction GP Contact Lens Solution is indicated for use in cleaning, rinsing, chemical (not heat) disinfection, storage and conditioning of rigid gas permeable contact lenses, including fluoro-silicone acrylate and silicone acrylate lenses, as recommended by your eye care practitioner.
JSZ Wetting/Rewetting Eyedrop is indicated to rewet soft (hydrophilic) and condition and wet rigid gas permeable contact lenses, including fluoro-silicone acrylate and silicone acrylate lenses, before application and during lens wear. The product may be used with daily wear or extended wear lenses and with disposable lenses or lenses prescribed for frequent replacement.
# (a)(6) Comparison of Technological Characteristics:
No changes have been made to the product formulation subject to this application. All technical information is contained in K050517.
#### b)(1) Discussion of Nonclinical:
No new data have been submitted in this application. All information is contained in K050517.
### (b)(2) Discussion of Clinical Data:
No new data have been submitted in this application. Based on the identity of the solution to the previously cleared solutions, no clinical data was required.
# (b)(3) Conclusions Drawn from Data Supporting Equivalence Determination:
This application presents new labeling for the product. There were no changes in
{2}------------------------------------------------
FOI Summary K060924
June 6, 2006
the product formulation and no new data were presented. The products are substantially equivalent to the predicate devices.
:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the words "Public Health Service" in a simple, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall impression is clean and straightforward, suggesting an official or informational context.
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle's head facing to the right, with three stylized lines representing the eagle's feathers.
JUN 2 8 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Szabocsik and Associates c/o Mr. John M. Szabocsik 203 North Wabash Ave., Suite 1200 Chicago, IL 60601
Re: K060924
Trade/Device Name: JSZ MultiAction GP Contact Lens Solution Regulation Number: 21 CFR 886.5918 Regulation Name: Rigid Gas Permeable Contact Lens Care Products Regulatory Class: II Product Code: MRC
Trade/Device Name: JSZ Wetting/Rewetting Eyedrop Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products Regulatory Class: II Product Code: LPN; MRC Dated: June 7, 2006 Received: June 8, 2006
Dear Mr. Szabocsik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act (11.1 The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misburanti, and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. IDA max publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not me . . ct
{4}------------------------------------------------
Page 2 - Mr. John M. Szabocsik
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
J. Soveta
Marvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
FOI Summary K060924
June 6, 2006
Page 1 of 2
# INDICATIONS FOR USE STATEMENT
# 510(k) NUMBER (IF KNOWN) K060924
# Device Name: JSZ MultiAction GP Contact Lens Solution
Indications for Use
The JSZ MultiAction GP Contact Lens Solution is indicated for use in cleaning, rinsing, chemical (not heat) disinfection, storage and conditioning of rigid gas permeable contact lenses, including fluorosilicone acrylate and silicone acrylate lenses, as recommended by your eye care practitioner.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mig. Vinn Shiz
Division Sign-Off
Division of Ophthalmic Ear, Nose and Throat Devises
510(k) Number
Prescription Use OR (Per 21 CFR 801.109) (Optional Format 1-2-96)
Over-The-Counter-Use
ટ વ
{6}------------------------------------------------
### FOI Summary K060924
June 6, 2006
Page 2 of 2
# INDICATIONS FOR USE STATEMENT
# 510(k) NUMBER (IF KNOWN) K060924
# Device Name: JSZ Wetting/Rewetting Eyedrop
Indications for Use
JSZ Wetting/Rewetting Eyedrop is indicated to rewet soft (hydrophilic) and to cushion and wet rigid gas permeable contact lenses, including fluoro-silicone acrylate and silicone acrylate lenses, before application and during lens wear. The product may be used with daily wear or extended wear lenses and with disposable lenses or lenses prescribed for frequent replacement.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mei-Ann Shih i
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
Prescription Use ાર (Per 21 CFR 801.109) (Optional Format 1-2-96)
Over-The-Counter-Use
ર્દેન
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.