The JSZ Multipurpose Solution is indicated for use in the daily cleaning, rinsing, chemical (not heat) disinfection, removal of proteins and storage of soft (hydrophilic) contact lenses as recommended by the eye care practitioner.
Device Story
JSZ Multipurpose Solution is a sterile, isotonic chemical solution for soft (hydrophilic) contact lens maintenance. It contains poloxamer, sodium phosphate buffer, sodium chloride, disodium edetate, and polyhexanide (0.0001%) preservative; it is free of chlorhexidine and thimerosal. The solution is packaged in plastic bottles with controlled dropper tips. It is used by contact lens wearers at home to clean, rinse, disinfect, and store lenses. The solution acts as a chemical disinfectant and protein remover. By maintaining lens hygiene, it supports comfortable lens wear and reduces the risk of ocular complications associated with contaminated lenses.
Clinical Evidence
Clinical trial of 6 months duration involving 246 subjects wearing representative soft (hydrophilic) contact lenses. The study compared the JSZ-Multipurpose Solution against control groups using currently marketed care products. Results demonstrated the device is substantially equivalent to existing solutions in terms of safety, efficacy, and comfort.
Technological Characteristics
Sterile, isotonic solution. Ingredients: poloxamer, sodium phosphate buffer, sodium chloride, disodium edetate, and polyhexanide 0.0001%. Packaging: plastic bottles (high density polyethylene) with controlled dropper tips. No chlorhexidine or thimerosal. Chemical disinfection principle (non-heat).
Indications for Use
Indicated for daily cleaning, rinsing, chemical disinfection, protein removal, and storage of soft (hydrophilic) contact lenses for patients using such lenses as recommended by an eye care practitioner.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of a bird.
Food and Druq Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 8 2005
John M. Szabocsik, Ph.D. SZABOCSIK AND ASSOCIATES Suite 1200 203 North Wabash Avenue Chicago, IL 60601
Re: K050517 Trade/Device Name: JSZ Multipurpose Solution Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LPN Dated (Date on orig SE 1tr): February 18, 2005 Received (Date on orig SE Itr): March 1, 2005
Dear Dr. Szabocsik:
This letter corrects our substantially equivalent letter of April 13, 2005 regarding the trade name section of the reference block.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - John M. Szabocsik, Ph.D.
CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
David M. Whipple
David M. Whipple Acting Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) NUMBER (IF KNOWN) K 05 05 05 17
Device Name: İSZ Multipurpose Solution
- ﻧ
Indications for Use The JSZ Multipurpose Solution is indicated for use in the daily cleaning, rinsing, chemical (not heat) disinfection, removal of proteins and storage of soft (hydrophilic) contact lenses as recommended by the eye care practitioner.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Ophthalmic Ear,
Nose and Throat Devises
| 510(k) Number | K050517 |
|--------------------------|---------------------------------------|
| Prescription Use | OR Over-The-Counter-Use <span></span> |
| (Per 21 CFR 801.109) | |
| (Optional Format 1-2-96) | |
ﺔ ﺍﻟﻤﺘ
{3}------------------------------------------------
APR 1 3 2005
# 1050517
JSZ-Multipurpose Solution 510(k) FOI Summary
# 510(k) SUMMARY
JSZ-MULTIPURPOSE SOLUTION FOR SOFT (HYDROPHILIC) CONTACT LENSES
This summary uses the format provided in 21 CFR 807.92:
| (a)(1). Submitter: | John M. Szabocsik, PhD<br>President<br>SZABOCSIK AND ASSOCIATES INC<br>203 N WABASH AVE STE 1200<br>CHICAGO IL 60601<br>Phone 312-553-0828<br>FAX 312-553-0611 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary prepared: | February 18, 2005 |
| (a)(2) Device Trade Name: | JSZ-Multipurpose Solution |
| Device Common Name: | Soft (hydrophilic) contact lens solution |
Device Classification Name:
Accessories to Contact Lens Solution (86LPN)
# (a)(3) Identification of Predicate Device:
This product is substantially equivalent to other currently marketed contact lens care multipurpose solutions.
# (a)(4) Device Description:
The JSZ-Multipurpose Solution is a sterile, isotonic solution that contains poloxamer, sodium phosphate buffer, sodium chloride, and disodium edetate; preserved with polyhexanide 0.0001%. Contains no chlorhexidine or thimerosal.
This product is a clear, colorless solution packaged in plastic bottles with controlled dropper tips.
# (a)(5) Intended Use (Indications for Use):
The JSZ Multipurpose Solution is indicated for use in the daily cleaning, rinsing, chemical (not heat) disinfection, removal of proteins and storage of soft (hydrophilic) contact lenses as recommended by the eye care practitioner.
{4}------------------------------------------------
### (a)(6) Comparison of Technological Characteristics:
No changes have been made to the product formulation subject to this application.
### Discussion of Nonclinical: b)(1)
- I. Chemistry
A. Solution compatibility
The compatibility of the JSZ-Multipurpose Solution was demonstrated by cycling lenses through 30 cycles of simulated use, using the JSZ-Multipurpose Solution for cleaning, rinsing, disinfecting and storing. Parameters of lenses were measured before and after the 30 cycles, and no differences were found.
### B. Cleaning effectiveness
The efficacy of the JSZ-Multipurpose Solution as a daily cleaner was shown by determining the critical micelle concentration.
#### II. Toxicology
The toxicological testing is summarized below, and reports are attached. The solution was shown to be non-toxic in all tests. Additional toxicity testing (cytotoxicity, systemic toxicity and ocular irritation) was done to verify the safety of the solution in the contract manufacturer's bottle.
Agar Overlay Cytotoxicity: A.
Representative lenses from all four groups of soft (hydrophilic) lens types were exposed to the JSZ-Multipurpose Solution for 24 hours, then tested in a direct contact cytotoxicity assay. All test lens types were noncytotoxic.
Systemic Toxicity: B.
The JSZ-Multipurpose Solution was evaluated for systemic toxicity by intraperitoneal (ip) injection in healthy mice, 50ml/kg body weight. The animals were observed over a 72 hour period, and showed no difference from control animals injected with saline. The solution passed the test requirements, that there be no difference between the response of test and control animals.
{5}------------------------------------------------
C. Acute Oral Toxicity
The JSZ-Multipurpose Solution was evaluated for acute oral toxicity by intubation in healthy rats, 5ml/kg body weight. The animals were observed immediately after intubation, after 2 and 4 hours, then daily for fourteen days. The animals were weighed prior to intubation, at 7 days, and at 14 days. All animals showed no clinical signs of toxicity from test initiation to Day 14, therefore the JSZ-Multipurpose Solution passed the test requirements of no acute oral toxicity.
#### Acute Ocular Irritation: D.
The JSZ-Multipurpose Solution was instilled directly into one eye of each of three rabbits, the other eye receiving sterile water as a control. Examinations over 72 hours showed no differences between test and control eyes, with no evidence of ocular irritation with either the test or control solutions. The JSZ-Multipurpose Solution therefore meets the requirements of the acute ocular irritation test, that it does not cause ocular irritation.
- Full USP Class VI Testing of JSZ-Multipurpose Solution in bottles E. manufactured of PETROTHENE® LR 7320-01 high density polyethylene.
- Cytotoxicity Test: 1.
Under the conditions of this study, the SC test extract showed no evidence of causing cell lysis or toxicity. The negative controls, reagent controls, and the positive controls performed as anticipated. The SC test extract was not cytotoxic and passed this ISO study.
Systemic Toxicity Test: 2. Under the conditions of this study, there was no mortality or evidence of significant systemic toxicity from the extracts. Each test article extract met with
USP requirements.
### Ocular Irritation Study: 3.
Under the conditions of this study, there was no evidence of significant irritation in the test or control eye of any rabbit. The SC and CSO test article extracts would not be considered irritants to the ocular tissue of the rabbits.
#### Microbiology III.
Sterility A.
Sterility data is included from the contract manufacturer.
{6}------------------------------------------------
Preservative efficacy B.
The JSZ-Multipurpose Solution solution passed the requirements of the preservative efficacy test.
..
Disinfection Efficacy C.
The JSZ-Multipurpose Solution passed the requirements of the stand-alone of The JSZ-Multipurpose Solution passed the requirements for the multi-item testing.
> Stability D.
Stability data is included from the contract manufacturer.
# (b)(2) Discussion of Clinical Data:
A clinical trial of 6 months usage of the JSZ-Multipurpose Solution by A clinical trial of 6 monuls usage of the 382 mount of soft (hydrophilic)
246 subjects, wearing representative lenses from all marked care 246 subjects, wearing representative rouses using currently marketed care
contact lenses, compared to control groups using currently marketed care contact lenses, compared to control groups asing our unstally
products, showed that the product is substantially equivalent to those current solutions.
(b)(3) Conclusions Drawn from Data Supporting Equivalence Determination:
Conclusions Drawn Barns of States of divised studies supporting the application. sions Drawn from Data Supporting Equites supporting the application, the In view of the in Vitto and cinical starter starter oppose Solution is substantially
safety, efficacy and comfort of the JSZ-Multipurpose Solution is substions safety, efficacy and comfort of the JSZ-Multipurpose Bountillers of any and
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.