BEAVERITE BODY FAT SCALES, MODELS 8901, 8903, 8905, 8909, 8911, 8915, 9563 AND 9568
K060538 · Beaverite International Limited · MNW · Oct 10, 2006 · Cardiovascular
Device Facts
Record ID
K060538
Device Name
BEAVERITE BODY FAT SCALES, MODELS 8901, 8903, 8905, 8909, 8911, 8915, 9563 AND 9568
Applicant
Beaverite International Limited
Product Code
MNW · Cardiovascular
Decision Date
Oct 10, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The Beaverite Body Fat Scales -- Models 8921, 8925, 8931, 8935 and 9576 -- measure body weight and impedance, and estimate percentages of body fat and total body water. They are intended for use by healthy adults. The Beaverite Body Fat Scales -- Models 8919, 8923, 8929, 9580, 9583, and 9588 -- measure body weight and impedance, and estimate percentages of body fat, total body water, and total muscle mass. They are intended for use by healthy adults.
Device Story
Beaverite Body Fat Scales are bioelectrical impedance analysis (BIA) devices used to estimate body composition. The device functions as a scale; user stands on electrodes to complete a circuit. Input: electrical impedance measured via foot-to-foot contact; body weight. Transformation: BIA technology processes impedance and weight data to calculate body fat, water, and muscle mass percentages. Output: digital display of weight and estimated body composition metrics. Used in home environment by healthy adults. Output provides users with body composition data for personal health monitoring.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Bioelectrical impedance analysis (BIA) plethysmograph. Operating frequencies: 20 kHz or 54 kHz. Electrodes: 2 or 4. Power: 9V or 3V batteries. Form factor: scale. Connectivity: standalone.
Indications for Use
Indicated for healthy adults for the measurement of body weight and impedance, and the estimation of body fat, total body water, and (for specific models) total muscle mass.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Tanita Innerscan Body Composition Monitor (K040778)
Submission Summary (Full Text)
{0}------------------------------------------------
Beaverite Body Fat Scales Traditional 510(k)
## Section 5.0 510(k) Summary
Beaverite Body Fat Scales
| Submitter Name: | Beaverite International Ltd |
|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | West Wing, 8/F<br>CNT Group Building<br>822 Lai Chi Kok Road<br>Cheung Sha Wan, Kowloon, Hong Kong |
| Contact Person: | Patsy J. Trisler, J.D., RAC<br>Regulatory Consultant - Medical Devices |
| Phone Number: | 301-652-5344 |
| Fax Number: | 301-654-6976 |
| Date: | September 25, 2006 |
| Device Trade Name: | Beaverite Body Fat Scales |
| Device Common Name: | Analyzer, Body Composition |
| Classification Name: | Impedance Plethysmograph |
| Classification Number: | 21 CFR 870.2770 |
| Product Code: | MNW |
| Predicate Device: | Tanita Innerscan Body Composition Monitor, K040778 |
| Statement of<br>Intended Use: | The Beaverite Body Fat Scales -- Models 8921, 8925, 8931, 8935<br>and 9576 -- measure body weight and impedance, and estimate<br>percentages of body fat and total body water. They are intended for<br>use by healthy adults.<br>The Beaverite Body Fat Scales -- Models 8919, 8923, 8929, 9580,<br>9583, and 9588 -- measure body weight and impedance, and<br>estimate percentages of body fat, total body water, and total muscle<br>mass. They are intended for use by healthy adults. |
| Device Description,<br>Summary of<br>Technological<br>Characteristics, and<br>Comparison to the<br>Predicate Device | The 11 models of Beaverite body fat scales submitted in this 510(k)<br>are plethysmograph body composition analyzers. The devices<br>estimate percentages of body fat, body water and muscle mass (as<br>noted in the Indications statement) based on bioelectrical impedance<br>analysis (BIA). The following table illustrates the substantial<br>equivalence in technology and intended use to the predicate device. |
OCT 1 0 2006
{1}------------------------------------------------
| Feature | Proposed Device:<br>Beaverite Body Fat Scales | Predicate Device:<br>Tanita InnerScan Body Composition<br>Monitor (Model BC-533) |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | K040778 |
| Manufacturer | Beaverite International Ltd | Tanita Corp of America |
| Classification #<br>Product Code<br>Indications for<br>Use | 21 CFR 870.2770<br>MNW<br>The Beaverite Body Fat Scales -- Models<br>8921, 8925, 8931, 8935 and 9576 --<br>measure body weight and impedance,<br>and estimate percentages of body fat<br>and total body water. They are intended<br>for use by healthy adults.<br>The Beaverite Body Fat Scales -- Models<br>8919, 8923, 8929, 9580, 9583, and 9588<br>-- measure body weight and impedance,<br>and estimate percentages of body fat,<br>total body water, and total muscle mass.<br>They are intended for use by healthy adults. | 21 CFR 870.2770<br>MNW<br>The Tanita family of InnerScan Body<br>Composition Monitors measure body<br>weight and impedance and estimate<br>percentage of body fat and body water,<br>visceral fat rating, bone mass, muscle<br>mass, physique rating, daily caloric intake<br>(DCI) and metabolic age using BIA. They<br>are intended for use by healthy children 7-<br>17 years old and healthy adults with active,<br>moderately active, to inactive lifestyles for<br>body composition assessment in the home<br>environment. |
| Device<br>description | Body composition analyzer/scale that<br>utilizes a 'foot-to-foot' bioelectrical<br>impedance (BIA) technology to determine<br>internal body composition. | Body composition analyzer/scale that<br>utilizes a 'foot-to-foot' bioelectrical<br>impedance (BIA) technology to determine<br>internal body composition. |
| Analysis<br>Method | BIA | BIA |
| Operating<br>Parameters | 20 KHz (Models 8923, 8929)<br>54 KHz (Models 8919, 8921, 8925, 8931,<br>8935, 9576, 9580, 9583, 9588) | 54 KHz |
| Power source | Replaceable 9V or 3V batteries,<br>depending on the model. | AA batteries |
| Operating keys | Range of 2 to 5, depending on the model. | 13 |
| Number of<br>electrodes | 4 (except for Model 8929 = 2) | 4 |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized eagle or bird.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## OCT 1 @ 2006
Patsy J. Trisler, J.D., R.A.C. Regulatory Consultant - Medical Devices Beaverite International Ltd. 5600 Wisconsin Avenue., Suite 509 CHEVY CHASE MD 20815
Re: K060538
> Trade/Device Name: Beaverite Body Fat Scales; Models 8919, 8921, 8923, 8925, 8929, 8931, 8935, 9576, 9580, 9583 and 9588 Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: September 25, 2006 Received: September 25, 2006
Dear Ms. Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a logo with the text "FDA Centennial 1906-2006" in a circular design. Below the logo, there is some text that reads "noting Public Health". The text is written in a decorative font. The image appears to be a scan of a document.
{3}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K060538
Device Name: __ Beaverite Body Fat Scales
Indications for Use:
The Beaverite Body Fat Scales -- Models 8921, 8925, 8931, 8935 and 9576 -measure body weight and impedance, and estimate percentages of body fat and total body water. They are intended for use by healthy adults.
The Beaverite Body Fat Scales -- Models 8919, 8923, 8929, 9580, 9583, and 9588 -- measure body weight and impedance, and estimate percentages of body fat, total body water, and total muscle mass. They are intended for use by healthy adults.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Eva
Rata Rall,
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K060538
(Posted November 13, 2003)
Section 4.0, 09/25/06
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.