The AutoCAT Intra-Aortic Balloon Pump Series is clinically indicated for the following conditions: a. Acute Coronary Syndrome b. Cardiac and Non-Cardiac Surgery c. Complications of Heart Failure
Device Story
The AutoCAT Intra-Aortic Balloon Pump (IABP) provides cardiac assist therapy via counterpulsation to support patients with impaired left ventricular function; it increases coronary/systemic perfusion while decreasing afterload and preload. The system takes physiological patient data as input to control IAB inflation and deflation timing. It features two operational modes: 'Autopilot' (automatic selection/control of functions) and 'Operator' (manual control). Used in clinical settings, the device assists physicians in managing hemodynamics. The system processes signals to optimize timing, handle arrhythmias, and manage noise, providing automated support to improve patient cardiac function.
Clinical Evidence
Bench testing only. The device was evaluated against FDA guidance documents for IABP consoles and software validation, as well as international standards including IEC 60601-1-4 (programmable electrical medical systems), EN 55024 (immunity), and EN 55022 (radio disturbance). No clinical trial data was required for this 510(k) submission.
Technological Characteristics
The system is a programmable electrical medical device utilizing computer-controlled pneumatic actuation for IAB inflation/deflation. It features software-driven timing and triggering based on physiological inputs. Connectivity includes RS232 output and modem functionality. Standards met include IEC 60601-1-4, EN 55024, and EN 55022.
Indications for Use
Indicated for patients with acute coronary syndrome, those undergoing cardiac or non-cardiac surgery, or patients experiencing complications of heart failure.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 17, 2022
Arrow Intl., Inc. Karen Provencher Senior Regulatory Affairs Specialist 9 Plymouth St. Everett, Massachusetts 02149
Re: K060309
Trade/Device Name: Autocat Intra-Aortic Balloon Pump Series Regulation Number: 21 CFR 870.3535 Regulation Name: Intra-aortic balloon and control system Regulatory Class: Class II Product Code: DSP
Dear Karen Provencher:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 6, 2005. Specifically, FDA is updating this SE Letter to reflect an administrative correction corresponding to the reclassification of intra-aortic balloon and control system (1ABP) devices when indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure, a preamendments class III device, into class II (special controls), as detailed in the final order published on December 20, 2019 (see here for more information: https://www.federalregister.gov/documents/2013/12/30/2013-31218/cardiovascular-devicesreclassification-of-intra-aortic-balloon-and-control-systems-foracute#:~:text=The%20Food%20Drug%20Administration%20(FDA)%20is%20issuing%20a%20fina 1,(special%20controls)%2C%20and%20to). In addition, IABP devices indicated for septic shock or pulsatile flow generation will remain Class III devices and would not be appropriate for the premarket notification pathway(510(k)), instead requiring a premarket approval (PMA).
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Alejandra Cambonchi, OHT2: Office of Cardiovascular Devices, 301-796-0552, Alejandra.Cambonchi(@fda.hhs.gov.
> Sincerely, Nicole M. Gillette -S
Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
6 2006 APR
6 2006
Arrow International, Inc. c/o Ms. Karen Provencher Senior Regulatory Affairs Specialist 9 Plymouth Street Everett, MA 02149
Re: K060309
K000507
Arrow AutoCAT Intra-Aortic Balloon Pump (IABP) Series Regulation Number: 21 CFR 870.3535 Regulation Name: Balloon, Intra-Aortic and Control System Regulatory Class: Class III Product Code: DSP Dated: March 17, 2006 Reccived: March 23, 2006
Dear Ms. Provencher:
We have reviewed your Section 510(k) premarket notification of intent to market the device . we nave reviewed your bection >10(tr) premation is substantially equivalent (for the indications ferenced above and nave determined ly marketed predicate devices marketed in interstate for use stated in the encreative to regard) the enactment date of the Medical Device Amendments, or to commerce prior to May 20, 7776, the encordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmette Herefore, market the device, subject to the general controls provisions of the Act. The r our may, therefore, market the act include requirements for annual registration, listing of general controls provisions of uactice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can may ve subject to back to begulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
# Page 2 - Ms. Karen Provencher
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please oe advised that 1 Dri 3 issuation of a vith other requirements of the Act that TDA has made a determination that your in the Federal agencies. You must or any I ederal statutes and regulations and united to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFK Fart 807), ademig (21 CFR Part 820), and if applicable, the electronic forth in the quality systems (QD) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (2001) on the vice as described in your Section 510(k) This letter with anow you to obgin maneting your and equivalence of your device to a legally premarket notification. The PDA miding of Saccanada for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acries ion 708 101 - 101 - 120 - 120. Also, please note the regulation entitled, Connact the Office of Compulline and (21) (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general information on your responser Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Befremdender for
Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K060309
Device Name
AutoCAT Intra-Aortic Balloon Pump Series
Indications for Use (Describe)
The AutoCAT Intra-Aortic Balloon Pump Series is clinically indicated for the following conditions:
a. Acute Coronary Syndrome
b. Cardiac and Non-Cardiac Surgery
c. Complications of Heart Failure
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Teleflex" in a stylized, sans-serif font. The word is a dark blue color, and the "R" in "Teleflex" has a circle around it, indicating that it is a registered trademark. The font is bold and the letters are close together.
### Arrow International, Inc
9 Plymouth Street Everett, MA 02149 USA Phone: 617-389-6400 617-387-2157 Fax:
www.teleflex.com
# 510(k) Summary K060309
AutoCAT Series IABP Series Date Prepared: February 3, 2006 Date Summary Updated: July 28, 2015
# A. Submitter's Name:
Arrow International, Inc. 9 Plymouth Street, Everett, MA 02149
### Updated Correspondent Address:
Fusun Tufan Senior Requlatory Affairs Manager Arrow International, Inc. 16 Elizabeth Drive, Chelmsford, MA 01824 Phone (978)250-5100 Fax (978)250-5105
### B. Company Contact
Michael Malis Director RA/QA Arrow International, Inc. 16 Elizabeth Drive, Chelmsford, MA 01824 Phone (978)250-5100 Fax (978)250-5105
### C. Device Name
Trade Name: AutoCAT Intra-Aortic Balloon Pump Series Common Name: Intra-Aortic Balloon Pump (IABP) Classification Name: Intra-Aortic Balloon and Control System
#### Predicate Devices D.
Arrow International AutoCAT Series IABP System represents the integration of four Arrow IABP technologies which is substantially equivalent in Intended Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution:
- · AutoCAT, cleared in K983866, (1/6/00)
- · ACAT 1, cleared in K965209, (1/23/98)
- · ACAT 2, cleared in K002256, (5/3/01)
- · FOS / FOMS IAB Black Box, cleared in K021462, (6/6/02)
#### Description of Device ய்
The Intra-Aortic Balloon Pump (IABP) provides cardiac assist therapy. The IABP provides temporary support to patients with impaired left ventricular function through the therapeutic method referred to as counterpulsation increases coronary and systemic profusion, decreasing after load (myocardial work) and decreasing preload.
The AutoCAT Series IABP System utilizes computer technology to select and maintain precise
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Teleflex" in a stylized, sans-serif font. The word is in a dark blue color, and there is a registered trademark symbol to the right of the "x". The logo appears to be for a company called Teleflex.
IAB inflation and deflation timing and triggering based on current physiological data from the patient. The system offers two modes of operation, the Autopilot mode, where functions are automatically selected and controlled by the IABP and the Operator mode where the user has control over settings and selections.
#### F. Indications for Use
The AutoCAT Intra-Aortic Balloon Pump Series is clinically indicated for the following conditions:
- a. Acute Coronary Syndrome
- Cardiac and Non-Cardiac Surgerv ﻘ
- Complications of Heart Failure C.
#### G. Comparison of Technological Characteristics
Software V2.22 features the modifications to the following- Clarification I The modified Modification to Pre-set Start-up Settings, General Improvements in AutoPilot Timing, Modifications to AP Zero and Calibration Function, Improvements in AutoPilot Signal Switching & Selection, General Improvements in Triggering, Improved Purge Cycle & Helium Refill Function, Recognition of IAB Connector Volumes, Improved Noise Recognition and Handling, Modifications to Arrhythmia Timing Function, Alarm Handling Improvements, Additions to Help Text. Activation of Modem Function. Reqular Output of RS232
#### Summary Performance Data H.
The AutoCAT IABP Series with modified software meets all the same performance standards of the unmodified device including the following:
- 1. FDA Guidance: "Guidance for the Preparation and Content of Applications to the Food and Drug Administration for Determining the Equivalence of Intra-Aortic Balloon Catheters and Consoles under the 510(k) Requlations - Preliminary Draft. December 8.1993
- FDA Guidance: "General Principles of Software Validation -Final Guidance' January 11, 2002 2.
- 3. FDA Guidance: "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Final Guidance' May 11, 2005
- 4. IEC 60601-1-4:1997: Medical Electrical Equipment- General Requirements for Safety-Collateral Standard: Programmable Electrical Medical Systems
- 5. EN 55024:1998: Information Technology Equipment-Immunity Characteristics
- 6. EN 55022:1998: Information Technology Equipment- Radio Disturbance Characteristics Limits and Methods of Measurement
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.