K965209 · Arrow Intl., Inc. · DSP · Jan 23, 1998 · Cardiovascular
Device Facts
Record ID
K965209
Device Name
ACAT 1 (ARROW CARDIAC ASSIST TECHNOLOGY 1)
Applicant
Arrow Intl., Inc.
Product Code
DSP · Cardiovascular
Decision Date
Jan 23, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3535
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cardiogenic shock, Pre- shock syndrome, Post- infarction angina (Threatening extension of MI) Unstable refractory angina, or impending infraction, Ischemia related intractable ventricular dysrhythmias, Septic shock syndrome, Cardiac contusion, Support for diagnostic interventional procedures including: (Cardiac Angiography, Coronary Angioplasty (PTCA), Coronary Atherectomy, Failed Mitral Valvuloplasty), Mechanical complications due to acute myocardial infraction: (Valvular stenosis - Mitral stenosis, Mitral Valve insufficiency - Mitral regurgitation, Ventricular Septal Defect (VSD), Papillary muscle rupture), Prophylactic support in preparation for cardiac surgery or high risk cardiac patients undergoing non-cardiac surgical procedures, Post-surgical myocardial dysfunction, Cardiac support following correction of anatomical defects, Maintenance of graft patency post-coronary bypass surgery, Pulsatile flow during cardiopulmonary bypass, Mechanical bridge to other assist devices.
Device Story
ACAT 1 is an advanced multi-microprocessor-based intra-aortic balloon pump (IABP) system for in-hospital and transport use. It provides circulatory support by inflating/deflating a balloon in the aorta. Inputs include patient physiological signals for triggering; system maintains helium levels and monitors balloon pressure waveforms. Operates via a lifetime metal bellows drive system. Features include cold trap vapor removal and alarm systems. Used by clinicians to manage cardiogenic shock, post-MI complications, and surgical support. Output provides pulsatile flow and hemodynamic assistance, aiding clinical decision-making by stabilizing hemodynamics. Benefits include improved cardiac output and graft patency.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Multi-microprocessor-based IABP control system. Features a lifetime metal bellows drive system, helium-based balloon inflation, cold trap vapor removal, and continuous balloon pressure waveform monitoring. Designed for portability (smaller/lighter than predecessor).
Indications for Use
Indicated for patients experiencing cardiogenic shock, pre-shock syndrome, post-infarction angina, unstable refractory angina, impending infarction, ischemia-related intractable ventricular dysrhythmias, septic shock, cardiac contusion, mechanical complications of MI (e.g., VSD, mitral regurgitation), post-surgical myocardial dysfunction, or requiring prophylactic support for cardiac/non-cardiac surgery, diagnostic interventional procedures, or as a mechanical bridge to other assist devices.
Regulatory Classification
Identification
An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}
K965209
# SECTION 2 - 510(k) SUMMARY and CERTIFICATION
510(k) SUMMARY
JAN 23 1998
This device is substantially equivalent to the following legally marketed intra-aortic balloon pumps:
1. The KAAT II Plus IABP catalog no. IAP-00025, from Arrow International, Everett, MA, Premarket Notification K905313.
2. The System 95 & 97 IABP, from Datascope Corporation, Paramus, NJ.
3. The TransAct H-8000 IABP, from C. R. Bard, Inc. Haverhill, MA.
The ACAT 1 IABP is the next generation of the Arrow/Kontron KAAT II Plus IABP manufactured by Arrow International. The ACAT 1 system is an advanced multi microprocessor based system that provides flexibility of operation and extended patient care capabilities, designed for in hospital and transport applications.
The ACAT 1 is a completely new industrial design package from the KAAT II Plus. Several new features have been incorporated which have continued to be customer requirements or preferences. The physical design of the ACAT 1 is smaller and lighter in weight than the current KAAT II Plus design. Overall we have incorporated new updates while maintaining the current product strengths of the KAAT II Plus system which are:
- Unique lifetime metal bellows drive system
- Wide range of triggering options
- Continuous assessment/maintenance of helium
- Safe alarming system
- Calibrated, continuous usable Balloon Pressure Waveform
- Cold trap, vapor removal system
- Ease of use
The device is indicated for the following conditions:
Cardiogenic shock, Pre- shock syndrome, Post- infarction angina (Threatening extension of MI)
Unstable refractory angina, or impending infraction, Ischemia related intractable ventricular dysrhythmias, Septic shock syndrome, Cardiac contusion, Support for diagnostic interventional procedures including: (Cardiac Angiography, Coronary Angioplasty (PTCA), Coronary Atherectomy, Failed Mitral Valvuloplasty), Mechanical complications due to acute myocardial infraction: (Valvular stenosis - Mitral stenosis, Mitral Valve insufficiency - Mitral regurgitation, Ventricular Septal Defect (VSD), Papillary muscle rupture), Prophylactic support in preparation for cardiac surgery or high risk cardiac patients undergoing non-cardiac surgical procedures, Post-surgical myocardial dysfunction, Cardiac support following correction of anatomical defects, Maintenance of graft patency post-coronary bypass surgery, Pulsatile flow during cardiopulmonary bypass, Mechanical bridge to other assist devices.
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Michael Jesi
Manager, Regulatory Affairs
and Quality Assurance
Arrow International, Inc.
9 Plymouth Street
Everett, MA 02149
Re: K965209
ACAT 1 (Arrow Cardiac Assist Technology 1)
Regulatory Class: II (Two)
Product Code: DSP
Dated: October 24, 1997
Received: October 27, 1997
Dear Mr. Jesi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
JAN 23 1998
{2}
Page 2 - Mr. Michael Jesi
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{3}
870.3535 DSP III Intra-aortic Balloon and Control System
510(k) Number (if known): K965209
Indications for IABP use are as follows:
- Cardiogenic shock
- Pre-shock syndrome
- Post- Infarction angina (Threatening extension of MI)
- Unstable refractory angina, or impending infraction
- Ischemia related intractable ventricular dysrhythmias
- Septic shock syndrome
- Cardiac contusion
- Support for diagnostic interventional procedures including:
- Cardiac Angiography
- Coronary Angioplasty (PTCA)
- Coronary Atherectomy
- Failed Mitral Valvuloplasty
- Mechanical complications due to acute myocardial infarction:
- Valvular stenosis - Mitral stenosis
- Mitral Valve insufficiency - Mitral regurgitation
- Ventricular Septal Defect (VSD)
- Papillary muscle rupture
- Prophylactic support in preparation for cardiac surgery or high risk cardiac patients undergoing non cardiac surgical procedures
- Post-surgical myocardial dysfunction
- Cardiac support following correction of anatomical defects
- Maintenance of graft patency post coronary bypass surgery
- Pulsatile flow during cardiopulmonary bypass
- Mechanical bridge to other assist devices.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K965209
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.