K053498 · Wjo, Inc. · CBF · Mar 15, 2006 · Anesthesiology
Device Facts
Record ID
K053498
Device Name
HYPEROX 101
Applicant
Wjo, Inc.
Product Code
CBF · Anesthesiology
Decision Date
Mar 15, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The expressed and intended use of the HyperOx 101 Hyperbaric Oxygen Treatment Chamber is to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class U prescriptive device, it is further intended for physician involvement in their procurement and routine use.
Device Story
HyperOx 101 is a Class A multi-place hyperbaric chamber; treats up to 11 occupants at pressures up to 32 psig. Input: compressed, dried/chilled air for pressurization; 100% medical oxygen for patient breathing via BIBBS (Built-In Breathing System). Operation: manual electropneumatic pressurization/depressurization; constant airflow ventilation. Output: pressurized environment for therapeutic oxygen absorption. Used in hospitals/clinics; operated by physicians/trained staff. Monitoring: oxygen and carbon dioxide gas analysis; wireless PA communication. Benefits: increased tissue oxygenation for wound healing, infection control, and treatment of specific ischemic/toxic conditions. Healthcare providers use chamber to deliver UHMS-approved hyperbaric oxygen therapy protocols.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on design, construction standards (ASME/NFPA), and intended use consistent with established hyperbaric oxygen therapy indications defined by the Undersea and Hyperbaric Medical Society (UHMS).
Technological Characteristics
Multi-place steel chamber; 72"W x 88"H x 16'6"L; 11,200 lbs. Construction: ASME Section VIII Div 1/2, ASME-PVHO-1-2002-2003, NFPA 99. Operating pressure 32 psig; design pressure 45 psig. Manual electropneumatic controls. Features: 12-feed oxygen manifold, 4 BIBBS, gas analysis (O2/CO2), wireless PA, optional entertainment systems. Finish: SSPC-SP-10 shotblast, primed/painted.
Indications for Use
Indicated for patients with air or gas embolism, carbon monoxide poisoning (with or without cyanide poisoning), clostridial myositis and myonecrosis, crush injury, compartment syndrome, acute traumatic ischemias, decompression sickness, problem wounds, exceptional blood loss anemia, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury, compromised skin grafts and flaps, thermal burns, and intracranial abscess.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
{0}------------------------------------------------
MAR 1 5 2006
Ko53498
# 5. Traditional 510(k) Pre-market Notification Summary (Per 21CFR807.92)
### December 19, 2005
Device Name: HyperOx 101, a Multiplace Hyperbaric Oxygen Treatment Chamber
Common Name: Multiplace Hyperbaric Chamber
Regulation No. 868.5470
Class: II
Code: CBF
- Submitted by: WJO, Inc. 424 Mill Street Canal's End Plaza Bristol, PA 19007 William J. O'Brien, D.O., President Phone: 215-826-8050 Fax 215-826-8053 Contact person: Chris Belletieri, D.O.
Establishment Registration Number: Pending
Railway Specialties Corporation (DOD Cage Code 86700) Manufacturer: 2979 State Road, P.O. Box 29 Bristol, Pennsylvania 19007-0029 Phone: 215-788-9242, Fax. 215-788-9244 Contact person: Frederic H. Calkins, Jr. - President
Establishment Registration Number: Pending
#### Predicate Device(s):
Fink Engineering Pty. Ltd Models SL8/DL8/TL20 (K031649) Pan-American Hyerbarics, Inc. PAH-S1 - K021693 Perry Sigma MP Hyperbaric Chamber -- K930748
{1}------------------------------------------------
#### SUMMARY
The Undersea and Hyperbaric Medical Society (UMHS) defines hyperbaric oxygen therapy as breathing 100% medical oxygen at pressures higher than atmospheric in a hyperbaric chamber. According to the National Fire Protection Association (NFPA), hyperbaric chambers are classified into two categories: Class A (multi-occupant) and Class B (single occupant). The HyperOx Hyperbaric Oxygen Treatmcnt Chamber is a Class A multi-place hyperbaric chamber designed to treat up to 10 patients up to 32 psig. The device will use compressed, dried/chilled air as the pressurization gas and 100% medical oxygen as the treatment gas.
The HyperOx Hyperbaric Oxygen Treatment Chamber is intended to be procured and used by physicians and hospitals to treat a variety of medical conditions that respond to hyperbaric oxygen therapy. Approved conditions responding to hyperbaric oxygen therapy are provided at the end of this summary.
The HyperOx Hyperbaric Oxygen Treatment Chamber is designed and fabricated in strict accordance with the requirements of the American Society of Mechanical Engineers (ASME) Boiler and Pressure Vessel Code Section VIII, Division 2 using Section II parts A, C and D, Section V, Section IX, ASME-PVHO-1-2002-2003 (American Society of Mechanical Engineers-Pressure Vessels for Human Occupancy) and the National Fire Protection Association (NFPA 99).
Specifications of the HyperOx 101 Hyperbaric Oxygen Treatment Chamber are:
| Design Code | ASME Section VIII, Div 1 and ASME-PVHO- |
|-----------------------|-----------------------------------------|
| Operating Pressure | 32 psig |
| Operating Temperature | 65 to 95 degrees F |
| Design Pressure | 45 psig |
| Design Temperature | 65 to 95 degrees F |
| Design Life | 70,000 cycles (approx. 30 years) |
| Hydrostatic Pressure | 42.0 psig |
| Weight | 11,200 lbs. (5.6 tons) |
| Dimensions | 72" W X 88" H X 16' 6" L |
| Doorway Opening Size | 30" X 60" clear opening |
{2}------------------------------------------------
| Lighting | Four (4) external dimmable lights<br>(1) 12" dia. clear view portlight in door<br>(8) 9" dia. clear view portlights in shell |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Capacity | Up to 11 occupants |
| Fire Suppression | IAW NFPA 99 |
| Finish | SSPC-SP-10 shotblast, primed & painted |
| Controls | Manual electropneumatic pressurization<br>Manual electropneumatic depressurization |
| Ventilation | Constant airflow |
| BIBBS with dump | Four |
| Oxygen Flow Manifold | Twelve feeds to O2 mask |
| Gas Analysis | Oxygen and carbon dioxide |
| Communications | Wireless PA |
| Entertainment<br>(optional) | Wireless internet connection<br>External TV system w/remote & DVD player<br>AM/FM tuner with CD player |
WJO, Inc. and Railway Specialties Corporation suggest that the general design approach, method of pressurization and the intended use of the HyperOx Hyperbaric Oxygen Treatment Chamber is substantially equivalent to the Fink Engineering Pty. Ltd models SL8/DL8/TL20 (K031649), Pan-American Hyperbarics, Inc PAH-S1 Chamber (K021693), and Perry Sigma MP Hyperbaric Chamber (K930748) and is proposing them as predicate devices.
## INTENDED USE:
Hyperbaric Oxygen Therapy (HBOT) is a painless procedure in which a person is exposed to increased pressure, thus allowing greater absorption of oxygen throughout body tissues. This increased pressure allows more oxygen to reach the cells within the body therefore contributing to the many healing and therapeutic benefits.
{3}------------------------------------------------
When utilizing HBOT, oxygen is forced into the tissues, organs, brain and fluids throughout the body through the pressurization of the hyperbaric chamber.
- Oxygen floods arcas that are oxygen starved to stimulate cell growth and . regencration.
- Oxygen displaces toxins and other impurities to assist in detoxification of the 。 patients system.
- Hyperbaric oxygen acts as an anti-viral and anti-bacterial, as bacteria and # viruses typically cannot tolerate oxygen.
- Hyperbaric oxygen is an immune modulator, supporting the immune system 。 to bring T and B cells within normal levels.
- Oxygen reduces tumor growth in cancer patients. �
- Hyperbaric oxygen increases neural brain function due to oxygen saturation. で
- Hyperbaric oxygen provides many other condition specific benefits. .
The expressed and intended use of the HyperOx 101 Hyperbaric Oxygen Treatment Chamber is to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class U prescriptive device, it is further intended for physician involvement in their procurement and routine use.
The UHMS is the professional medical organization chartered with setting the standards of care defining the appropriate use of hyperbaric oxygen. More specifically, the UIIMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection.
The conditions listed as appropriate for the use of Hyperbaric Oxygen Treatment are:
- Air or gas embolism .
- Carbon monoxide poisoning and carbon monoxide poisoning complicated by . cyanide poisoning
- Clostridial myositis and myonecrosis .
- Crush injury, compartment syndrome and other acute traumatic ischemias .
- Decompression sickness .
- Enhanced healing of selected problem wounds .
- . Exceptional blood loss anemia
- Necrotizing soft tissue infections ●
- Osteomyclitis (refractory) .
- Delayed radiation injury (soft tissue and bony necrosis) .
- Skin grafts and flaps (compromised) .
- Thermal burns .
- Intercranial abscess .
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three lines forming its wings and head. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 5 2006
Mr. Chris Belletieri WJO. Incorporated 424 Mill Street at Canal's End Plaza Bristol, Pennsylvania 19007
Re: K053498
Trade/Device Name: Hyperbaric Oxygen Treatment Chamber - Model
101 Regulation Number: 868.5470 Regulation Name: Hyperbaric chamber Regulatory Class: II Product Code: CBF Dated: December 16, 2005 Received: December 16, 2005
Dear Mr. Belletieri :
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Chris Belletieri
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Snyte y. Michael Oms
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): k053498
Device Name: HyperOx Hyperbaric Oxygen Treatment Chamber - Model 101
Indications For Use:
- . Air or gas embolism
- Carbon monoxide poisoning and carbon monoxide poisoning . complicated by cyanide poisoning
- Clostridial myositis and myonecrosis .
- Crush injury, compartment syndrome, and other acute traumatic . ischemias
- Decompression sickness .
- Enhanced of selected problem wounds .
- Exceptional blood loss anemia .
- Necrotizing soft tissue infections .
- . Osteomvelitis (refractory)
- Delayed radiation injury (soft tissue and bony necrosis) .
- . Skin grafts and flaps (compromised)
- Thermal burns •
- . Intercranial abscess
Prescription Use Only (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Mh. Katerl
Singlogy, General Hospitzl. Of Dental Davices
K053498
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.