PAN-AMERICA HYPERBARICS HYPERBARIC CHAMBER SYSTEM, MODEL PAH-S1
K021693 · Pan-America Hyperbarics, Inc. · CBF · Oct 24, 2002 · Anesthesiology
Device Facts
Record ID
K021693
Device Name
PAN-AMERICA HYPERBARICS HYPERBARIC CHAMBER SYSTEM, MODEL PAH-S1
Applicant
Pan-America Hyperbarics, Inc.
Product Code
CBF · Anesthesiology
Decision Date
Oct 24, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is the expressed, intended use of the Pan-America Hyperbarics, Inc., PAH-S1 Hyperbaric Chamber System to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in its procurement and routine use. The UHMS is the professional medical organization chartered with setting the standards of care defining the appropriate use of hyperbaric oxygen. More specifically, the UHMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit. The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows: - Air or gas embolism 1. - Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide 2. poisoning - Clostridial myositis and myonecrosis 3. - Crush injury, compartment syndrome, and other acute traumatic ischemias 4. - Decompression sickness 5. - Enhanced healing of selected problem wounds 6. - Exceptional blood loss anemia 7. - Necrotizing soft tissue infections 8. - Osteomyelitis (refractory) 9. - Delayed radiation injury (soft tissue and bony necrosis) 10. - Skin grafts and flaps (compromised) 11. - Thermal burns 12. - Intracranial abscess 13. The Pan-America Hyperbarics, Inc., PAH-S1 Hyperbaric Chamber System is designed to be installed and operated in medical facilities as defined by the NFPA 99, Health Care Facilities, Chapter 20, Hyperbaric Facilities. Further, this system is intended to be operated only by medical personnel specifically trained in the appropriate use of HBO and the safe operation of all related equipment such as the hyperbaric chamber.
Device Story
PAH-S1 is a Class B monoplace hyperbaric chamber; treats one patient at up to 3 ATA (29.4 psig). Uses 100% oxygen for both pressurization and patient breathing. System includes operator control panel with manual pneumatic pressure controls; fire-retardant bunk on retractable gurney for patient loading. Low-voltage intercom system allows communication between patient and operator. Installed in medical facilities per NFPA 99; operated by trained medical personnel. Provides hyperbaric oxygen therapy to treat specific medical conditions; physician involvement required for procurement and use. Benefits include therapeutic application of oxygen for validated clinical indications.
Clinical Evidence
Bench testing only. Device design and fabrication verified against ANSI/ASME Boiler and Pressure Vessel Code, ANSI/ASME-PVHO-1, and NFPA 99 standards. No clinical data presented.
Technological Characteristics
Monoplace hyperbaric chamber; 2200mm length, 813mm internal diameter. Construction: half-steel, half-acrylic tube. Operating pressure: max 3 ATA (29.4 psig). Pressurization gas: 100% oxygen. Manual pneumatic pressure control. Low-voltage patient intercommunication system (max 250 mW at 1.4 volts). Compliant with ANSI/ASME Section VIII Div 1, ANSI/ASME-PVHO-1, and NFPA 99.
Indications for Use
Indicated for patients with conditions responsive to hyperbaric oxygen therapy, including air/gas embolism, carbon monoxide poisoning, clostridial myositis/myonecrosis, crush injury/compartment syndrome/acute traumatic ischemias, decompression sickness, problem wounds, exceptional blood loss anemia, necrotizing soft tissue infections, refractory osteomyelitis, delayed radiation injury, compromised skin grafts/flaps, thermal burns, and intracranial abscess.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
{0}------------------------------------------------
K021693
# OCT 2 4 2002
# 510(k) NOTIFICATION SUMMARY (Per 21 CFR 807.92)
Prepared: May 12, 2002
TRADE NAME: Pan-America Hyperbarics, Inc., PAH-S1 Hyperbaric Chamber System
COMMON NAME OF DEVICE: Monoplace Hyperbaric Chamber
CLASSIFICATION: 73 CBF, 21 CFR 868.5470
## ESTABLISHMENT REGISTRATION NUMBER: Pending
## CLAIMED PREDICATE DEVICE(S):
Sechrist 2500 (K934164) Sechrist 2800 (K950386) Sechrist 3200 (K950386) Perry Sigma 1 (K832127) Perry Sigma 34 (K990927) ETC BaraMed (K993010) Tampa Hyperbarics "T.H.E." Chamber (K981938) HyperTec 3200 (K002795)
#### ADDRESS OF MANUFACTURER:
Pan-America Hyperbarics, Inc. No. 9, Lane 12, Guang Feng 1st Street Ba-de City, Tao Yuan County Taiwan, R.O.C. (886)-3-3676676
CONTACT PERSON: Cheng, Kuo-Chung, General Manager
## EXECUTIVE SUMMARY
The PAH-S1 Hyperbaric Chamber System is a Class B monoplace hyperbaric chamber designed to treat one patient at up to a maximum operating pressure of 3 Atmospheres Absolute (ATA) or 29.4 pounds per square inch gauge (psig). The chamber uses 100% oxygen as the pressurization gas and patient breathes the oxygen contained inside the chamber as the hyperbaric treatment das.
The PAH-S1 Hyperbaric Oxygen System is designed and fabricated in accordance with the requirements of the ANSI/ASME Boiler and Pressure Vessel Code, Section VIII, Division 1, Pressure Vessels; ANSI/ASME-PVHO-1 (American Society of Mechanical Engineers-Pressure Vessels for Human Occupancy); and, NFPA 99, Health Care Facilities, Chapter 20, Hyperbaric Facilities, 2002 Edition. The overall external length of the chamber is 2200mm (approx, 7.22 ft). Its internal diameter is 813mm (approx. 32 inches). The chamber is constructed by half-steel and half-acrylic tube. A low-voltage patient intercommunication system designed and installed in accordance with NFPA 99, Chapter 19 and provides communications between the patients in the
{1}------------------------------------------------
chamber and the outside chamber operator. Maximum power to the speaker is limited by the intercom circuitry to 250 mW at 1.4 volts (NFPA 99, Chapter 20 limit is 500 mW and 28 volts).
The system consists of an operator control panel that contains all of the controls and connection points. Single operator chamber pressure control is achieved via a simple manual pneumatic control. Spare penetrators are provided to allow user supplied medical monitoring leads. etc., to be used as required. Patient is loaded and unloaded by a retractable qurnev. When loading. patient lies down on the fire retardant foamed bunk, aligned to the rails of the chamber and fixes four wheels of the qurney, then push the bunk into the chamber. Unloading the patient with an opposite procedure. The chamber is also equipped with for pressurization. There is no gas supply for pressurization before the chamber's door is closed and secured thoroughly.
#### Intended Use:
It is the expressed, intended use of the Pan-America Hyperbarics, Inc., PAH-S1 Hyperbaric Chamber System to provide therapy to those patients with selected medical conditions that have been determined to respond to the application of hyperbaric oxygen. As a Class II prescriptive device, it is further intended for physician involvement in its procurement and routine use.
The UHMS is the professional medical organization chartered with setting the standards of care defining the appropriate use of hyperbaric oxygen. More specifically, the UHMS publishes a listing of medical conditions that have been clearly established as appropriate primary or adjunctive use of hyperbaric oxygen (HBO). The disorders on the list have been scientifically validated and verified through extensive data collection. It should be noted that the list is dynamic. Based on the strength of the scientific data, disorders are both added and removed from the list, depending on the outcomes of scientific pursuit.
The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- Air or gas embolism 1.
- Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide 2. poisoning
- Clostridial myositis and myonecrosis 3.
- Crush injury, compartment syndrome, and other acute traumatic ischemias 4.
- Decompression sickness 5.
- Enhanced healing of selected problem wounds 6.
- Exceptional blood loss anemia 7.
- Necrotizing soft tissue infections 8.
- ഗ് Osteomyelitis (refractory)
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
- 12. Thermal burns
- 13. Intracranial abscess
The Pan-America Hyperbarics, Inc., PAH-S1 Hyperbaric Chamber System is designed to be installed and operated in medical facilities as defined by the NFPA 99, Health Care Facilities, Chapter 20, Hyperbaric Facilities. Further, this system is intended to be operated only by medical personnel specifically trained in the appropriate use of HBO and the safe operation of all related equipment such as the hyperbaric chamber.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract design of three human profiles facing right, stacked on top of each other. The profiles are rendered in black against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the design.
# OCT 2 4 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Pan-America Hyperbarics, Incorporated C/O Mr. W. T. Workman Workman Hyperbaric Services, Incorporated 18111 Copper Ridge Drive San Antonio, Texas 78259-3612
Re: K021693
Trade/Device Name: Hyperbaric Chamber System, Model PAH-S1 Regulation Number: 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: II Product Code: CBF Dated: September 5, 2002 Received: September 6, 2002
Dear Mr. Workman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
## Page 2 - Mr. Workman
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 510(k) Number (if known): KO21643
Device Name: Pan-America Hyperbarics, Inc., PAH-S1 Hyperbaric Chamber System
# Indications for Use:
The conditions listed as appropriate for the use of HBO in the current edition of the Hyperbaric Oxygen Therapy Committee Report (1999) are as follows:
- 1. Air or gas embolism
- 2. Carbon monoxide poisoning and carbon monoxide poisoning complicated by cyanide poisoning
- 3. Clostridial myositis and myonecrosis
- 4. Crush injury, compartment syndrome, and other acute traumatic ischemias
- 5. Decompression sickness
- Enhanced healing of selected problem wounds 6.
- 7. Exceptional blood loss anemia
- 8. Necrotizing soft tissue infections
- 9. Osteomyelitis (refractory)
- 10. Delayed radiation injury (soft tissue and bony necrosis)
- 11. Skin grafts and flaps (compromised)
- 12. Thermal burns
- 13. Intracranial abscess
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
fothin
Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K032693
(Optional Format 3-10-98)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.